K Number
K171198
Date Cleared
2017-12-07

(227 days)

Product Code
Regulation Number
892.1750
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NeuViz 16 Essence Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Device Description

The NeuViz 16 Essence Multi-Slice CT Scanner System is composed of a gantry, a patient couch, an operator console and includes image acquisition hardware and software, and associated accessories. It is designed to be a head and whole body X-ray computed tomography scanner which features a continuously rotating tube-detector system and functions according to the fan beam principle.

The system provides the filter back-projection (FBP) and iterative reconstruction algorithm(ClearView cleared in K133373) to reconstruct images. The end user can choose to apply either ClearView or the FBP to the acquired raw data.

The system software is an interactive program used for X-ray scan control, image reconstruction, and image archive/evaluation. It provides the following digital image processing and visualization tools:

  • Support following scan speed: 0.5s、0.6s、0.75s、1.0s、1.5s、2.0s.
  • · Surview scan
  • Dual surview
  • · Spiral scan
  • Axial scan
  • Image reconstruction
  • Plan scan
  • Patient information management
  • Patient information registration
  • Protocol selection
  • O-Dose
  • . Bolus tracking
  • SAS
  • Home
  • Film
  • Report
  • 2D
  • MPR
  • 3D
  • VE(Virtual Endoscopy)
  • Vessel Analysis
  • Dicom Viewer
  • · Bar code Reader
  • Dual Monitor
  • Continuous CT
  • ClearView
  • Dental Analysis
  • Virtual Colonoscopy
  • Brain Perfusion
  • Body Perfusion
  • Lung Nodule Analysis
AI/ML Overview

This document describes the Neusoft Medical Systems NeuViz 16 Essence Multi-Slice CT Scanner System (K171198). The information provided focuses on the substantial equivalence claim to its predicate devices, K092742 (NeuViz 16 Multi-Slice CT Scanner System) and K151383 (NeuViz 128 Multi-Slice CT Scanner System).

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not provide a formal table of quantitative acceptance criteria for image quality metrics or a direct comparison of measured performance against specific thresholds. Instead, it states that "The clinical validation was executed as planned and acceptance criteria met for each requirement." It also mentions that "The Results indicated that the images were of diagnostic quality."

For image quality, the document states:
"Some of the bench tests included the sag and the speed of patient support table, the gantry wobble, CT image quality metrics testing including CT number accuracy and uniformity, MTF, noise, slice sensitivity profiles and CTDI." It also states, "Sample images were provided to show the performance of the system in presence of implants."

The conclusions state: "The results of these tests demonstrate that the subject device performs as intended." and "The result of all conducted testing was found acceptable to support the claim of substantial equivalence."

2. Sample Size Used for the Test Set and Data Provenance:

The document indicates that a clinical validation was performed, which included "image evaluation...to evaluate images of the brain, chest, abdomen and spine/extremities of the body area." However, it does not specify the sample size (number of patients or images) used for this test set nor the country of origin of the data. It also does not explicitly state whether the data was retrospective or prospective.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

The document mentions that "An image evaluation was performed...Images were scored using a 5 point Likert scale by a qualified radiologist." This indicates that one qualified radiologist was used for ground truth assessment. Their specific qualifications (e.g., years of experience, subspecialty) are not detailed beyond "qualified radiologist."

4. Adjudication Method for the Test Set:

Based on the information provided, it appears that no specific adjudication method (like 2+1 or 3+1) was used, as only one radiologist is mentioned for scoring the images.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study, nor does it provide an effect size for human reader improvement with or without AI assistance. The study focuses on demonstrating the diagnostic quality of the images produced by the device.

6. Standalone Performance Study:

The information provided describes the performance of the "NeuViz 16 Essence Multi-Slice CT Scanner System" as a standalone medical device capable of image acquisition and reconstruction. The clinical validation assessed whether the images produced by the device were of diagnostic quality. Therefore, a standalone performance assessment (device only, without human interpretation for improving diagnostic accuracy) was implicitly performed in that the output of the device (the images) was evaluated for its diagnostic utility. The study doesn't contrast with a human-in-the-loop scenario.

7. Type of Ground Truth Used:

The ground truth for the clinical validation was established through expert consensus/reader evaluation. Specifically, "Images were scored using a 5 point Likert scale by a qualified radiologist," who determined if the images were of "diagnostic quality."

8. Sample Size for the Training Set:

The document does not provide any information regarding the sample size of a training set. The descriptions focus on the validation of the device and its algorithms (FBP, ClearView), not on the training data for an AI model. "ClearView" is mentioned as an iterative reconstruction algorithm cleared in K133373, but no details about its training data are given in this K171198 submission.

9. How the Ground Truth for the Training Set Was Established:

Since no information is provided about a training set for an AI model, there is no information on how its ground truth would have been established. The "ClearView" algorithm, where potentially AI/machine learning could be involved, references a previous 510(k) (K133373), and the current document does not detail its development or training.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

NEUSOFT MEDICAL SYSTEMS CO., LTD.

December 7, 2017

% Shao Xiancheng R&A No. 16, Shiji Road Hunnan Industrial Area Shenyang, Liaoning 110179 P R CHINA

Re: K171198

Trade/Device Name: NeuViz 16 Essence Multi-Slice CT Scanner System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: November 6, 2017 Received: November 13, 2017

Dear Shao Xiancheng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

Page 2 - Shao Xiancheng

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D.'Hara
For

Robert Ochs. Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K171198

Device Name

NeuViz 16 Essence Multi-Slice CT Scanner System

Indications for Use (Describe)

The NeuViz 16 Essence Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other spect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

510(K) Summary

This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92

General Information:

Manufacturer:Neusoft Medical Systems Co., Ltd.No.177-1 Chuangxin Road, Hunnan District,Shenyang, Liaoning, China, 110167.
Submitter:Contact : Shao XianchengTitle : RA EngineerTel : 86-24-83660645Fax : 86-24-83660563E-Mail : shaoxc@neusoft.com
510(k) Summary Dateof Preparation:November 6, 2017

Device Name and Classification:

Trade Name:NeuViz 16 Essence Multi-Slice CT Scanner System
Common Name:CT Scanner
Regulation:21 CFR 892.1750
Classification Name:Computed tomography x-ray system
Product Code:JAK
Classification:Class II
Performance Standard:21 CFR Subchapter J, Federal Diagnostic X-ray EquipmentStandard

Primary Predicate device:

Trade Name:NeuViz 16 Multi-Slice CT Scanner System
510(k) number:K092742
Clearance Date:09/22/2009
Regulation:21 CFR 892.1750
Classification Name:Computed tomography x-ray system
Product Code:JAK
Classification:Class II
Manufacturer:Neusoft Medical Systems Co., Ltd.

Reference Device:

Trade Name:NeuViz 128 Multi-Slice CT Scanner System
510(k) number:K151383
Clearance Date:11/04/2015
Regulation:21 CFR 892.1750
Classification Name:Computed tomography x-ray system
Product Code:JAK
Classification:Class II
Manufacturer:Neusoft Medical Systems Co., Ltd.

{4}------------------------------------------------

Safety and Effectiveness information

Indications for use:

The NeuViz 16 Essence Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Device Description:

The NeuViz 16 Essence Multi-Slice CT Scanner System is composed of a gantry, a patient couch, an operator console and includes image acquisition hardware and software, and associated accessories. It is designed to be a head and whole body X-ray computed tomography scanner which features a continuously rotating tube-detector system and functions according to the fan beam principle.

The system provides the filter back-projection (FBP) and iterative reconstruction algorithm(ClearView cleared in K133373) to reconstruct images. The end user can choose to apply either ClearView or the FBP to the acquired raw data.

The system software is an interactive program used for X-ray scan control, image reconstruction, and image archive/evaluation. It provides the following digital image processing and visualization tools:

  • Support following scan speed: 0.5s、0.6s、0.75s、1.0s、1.5s、2.0s.
  • · Surview scan
  • Dual surview
  • · Spiral scan
  • Axial scan
  • Image reconstruction
  • Plan scan
  • Patient information management
  • Patient information registration
  • Protocol selection
  • O-Dose
  • . Bolus tracking
  • SAS
  • Home
  • Film
  • Report
  • 2D
  • MPR
  • 3D
  • VE(Virtual Endoscopy)
  • Vessel Analysis
  • Dicom Viewer
  • · Bar code Reader
  • Dual Monitor
  • Continuous CT
  • ClearView
  • Dental Analysis
  • Virtual Colonoscopy
  • Brain Perfusion
  • Body Perfusion
  • Lung Nodule Analysis

{5}------------------------------------------------

  • Lung Density Analysis
  • CTDSA
  • Tumor Assessment

Statement of Substantial Equivalence:

The NeuViz 16 Essence Multi-Slice CT Scanner System has the same indications for use as the predicate device, NeuViz 16 Multi-Slice CT Scanner System (K092742). The NeuViz 16 Essence is a computed Tomography X-Ray System intended to produce images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. The subject device does not have significant changes in technological characteristics when compared to the predicate device. The subject device and predicate device are the same in regards to:

  • Detector/Tube system that can rotate continuously. The Tube emits X-ray beam ● of fan shape, detector system with multiple channels;
  • Low voltage slip ring and On-Board HV Generator;
  • Parallel data acquisition, simultaneous scan and image reconstruction;
  • Gantry that may be tilted forward or backward;
  • Patient Couch that can be elevated or lowered;
  • Computerized control of scan and other operations;
  • Computerized image reconstruction and post processing;
  • Digitarized display, storage and output of data/image.

The subject device and the reference device, NeuViz 128 Multi-Slice CT Scanner System (K151383) are the same in regard to most of application features.

A complete comparison table is included in this submission. The following table shows a brief comparison of the application features, including both similarities and differences among the subject, the predicate and the reference devices.

NeuViz 16 EssenceNeuViz 16NeuViz 128
O-Dose×
Bolus tracking
SAS
Home×
Film
Report
2D×
MPR
3D
VE(Virtual Endoscopy)
Vessel Analysis
Dicom Viewer×
Bar code Reader
Dual Monitor×

{6}------------------------------------------------

Continuous CT
ClearView×
Dental Analysis
Virtual Colonoscopy×
Brain Perfusion
Body Perfusion×
Lung Nodule Analysis×
Lung Density Analysis×
CTDSA×
Tumor Assessment×
Comparison“√” symbol means the comparable type and substantiallyequivalent to the contrasted devices.“x” symbol means the new clinical functionality.Including O-Dose, ClearView, Virtual Colonoscopy, BodyPerfusion, Lung Nodule Analysis, Lung Density Analysis, CTDSAand Tumor Assessment, these advanced image-processingsoftware features were drawn from the NeuViz 128 without anychange.

According to the comparison based on the requirements of 21 CFR 807.87, we stated that these devices are substantially equivalent.

Nonclinical Testing:

The safety and effectiveness of the NeuViz 16 Essence was assured by adherence to Good Manufacturing Practices(GMP) 21CFR 820 and to International Standards ISO 13485:2003. This device is in conformance with the applicable parts of the following standards:

  • ANSI/AAMI ES60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R)2012 and ● A2:2010/(R)2012 medical electrical equipment -- part 1: general requirements for basic safety and essential performance
  • IEC 60601-1-2: 2007, Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
  • IEC 60601-1-3: 2013, Medical Electrical Equipment - Part 1-3: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Radiation Protection In Diagnostic X-Ray Equipment
  • IEC 60601-1-6: 2013, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
  • IEC 60601-2-28: 2010, Medical electrical equipment Part 2-28: Particular ● requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis
  • IEC 60601-2-44: 2012. Medical Electrical Equipment - Part 2-44: Particular Requirements For The Basic Safety And Essential Performance Of X-Ray Equipment For Computed Tomography
  • IEC 60825-1:2007, Safety of laser products - part 1: equipment classification, and requirements
  • IEC 61223-3-5: 2004, Evaluation and routine testing in medical imaging departments - Part 3-5: Acceptance tests - Imaging performance of computed tomography X-ray equipment
  • AAMI ANSI IEC 62304:2006, Medical device software - Software life-cycle processes
  • IEC 62366:2014,Medical Devices - Application Of Usability Engineering To

{7}------------------------------------------------

Medical Devices

  • ISO 10993-1: 2009, Biological evaluation of medical devices - part 1: evaluation and testing within a risk management process.
  • ISO 14971:2007, Medical devices - Application of risk management to medical devices
  • NEMA PS 3.1 - 3.20 (2016), Digital Imaging And Communications In Medicine (DICOM) Set
  • . NEMA XR 25: 2010. Computed tomography dose check
  • NEMA XR 28: 2013, Supplemental Requirements for User Information and System Function Related to Dose in CT

Additionally, this device complies with all applicable requirements of the radiation safety performance standards, as outlined in 21 CFR §1010 and §1020.

Risk manaqement is ensured via a hazard analysis which is used to identify potential hazards. These potential hazards are controlled during product development, and verification and validation testing. To minimize electrical, and radiation hazards, Neusoft adheres to recognized and established industry practice and standards.

Software safety is assured by the company procedures that conform to accepted practices. Quality assurance procedures are adhered to, and meeting the specifications and functional requirements if demonstrated via testing. Testing included functional, smoke and regression tests. The vast majority of tests passed our testing criteria. Any defects found or reported were either fixed or logged in the Unresolved Anomalies report included with this submission and annotated as to any impact on safety or effectiveness including applicable workarounds. The traceability between design requirements, design specifications, testing requirements, and relevant hazards with the implementation and testing of the mitigations are described in the Traceability Matrix. The Traceability Matrix also shows the overall test results per requirement and per hazard mitigation.

Cybersecurity requirements is assured by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. Cybersecurity information in accordance with quidance document "Content of Premarket Submissions for Management of Cybersecurity Medical Devices issues on October 2, 2014" is included within this submission.

Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.

Verification and validation activities(including performance testing, safety testing and simulated use testing) was performed. Some of the bench tests included the sag and the speed of patient support table, the gantry wobble, CT image quality metrics testing including CT number accuracy and uniformity, MTF, noise, slice sensitivity profiles and CTDI. Sample images were provided to show the performance of the system in presence of implants.

The results of these tests demonstrate that the subject device performs as intended. The result of all conducted testing was found acceptable to support the claim of substantial equivalence.

Clinical Testing:

{8}------------------------------------------------

Neusoft

The clinical validation plan identified the tasks, deliverables, methodology, requirements and the resources for validation of the intended use and meets customer needs. The clinical validation covered requirements related to clinical workflows and features. The validation was executed as planned and acceptance criteria met for each requirement. All validation tests demonstrate the safety and effectiveness of NeuViz 16 Essence.

The results of the validation are available in the Clinical Validation Summary. Two CT clinical specialists of Neusoft validated the system under simulative clinical circumstances. An image evaluation was performed to evaluate images of the brain, chest, abdomen and spine/extremities of the body area. Images were scored using a 5 point Likert scale by a qualified radiologist. The Results indicated that the images were of diagnostic quality. The proposed NeuViz 16 Essence can be used as defined in its clinical workflow and intended use.

Conclusions:

The NeuViz 16 Essence performs in a manner similar to and is intended for the same use as the predicate device, as indicated in product labeling. Based upon this information, conformance to standards, successful completion of software validation, application of risk management and design controls and the performance data presented in this submission it is concluded that the subject device is substantially equivalent in safety and effectiveness to the predicate device. The comparison of technological characteristics, non-clinical performance data, software validation and clinical images demonstrates that the subject device is as safe and effective when compared to the predicate device that is currently marketed for the same intended use.

According to the comparison based on the requirements of 21.CFR 807.87, Neusoft believes that the subject device and the predicate device are substantially equivalent.

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.