(251 days)
The C3 Wave System is indicated for use in the positioning of Peripherally Inserted Central Catheters (PICC). The C3 Wave provides real-time catheter tip location information by displaying changes in the patient's cardiac electrical activity. The C3 Wave is indicated for use as an alternative method to chest X-ray or fluoroscopy confirmation of PICC tip placement in adult patients.
Note: Limiting, but not contraindicated, situations for this technique are patients where cardiac rhythms may change presentation of the P-Wave:
-Atrial fibrillation
-Atrial flutter
- Severe tachycardia
-Pacemaker-Driven Rhythm
-Chronic obstructive pulmonary disease (COPD)
Such patients are easily identified prior to PICC insertion. Use of additional method is necessary to confirm catheter tip location.
C3 Wave is designed to provide a continuous display of electrocardiograph [ECG] waveform to be used as a quide in placement of peripherally-inserted central catheters [PICC] in the lower third of the Superior Vena Cava [SVC] of a patient. The principle for operation of this system uses three ECG leads placed on the patient's chest and generates a third ECG lead by switching from RA to PICC stylet. The ECG waveform is wirelessly transmitted to a tablet which allows the operator to view and record changes to the ECG waveform as the tip of the catheter approaches the heart. As the PICC catheter approaches the atrium of the heart, the P wave in the ECG waveform shows substantial changes. This system is designed to aid the visualization of changes in P wave amplitude.
Here's an analysis of the acceptance criteria and study information for the C3 Wave System, based on the provided text:
Acceptance Criteria and Device Performance
The document doesn't explicitly list "acceptance criteria" as a separate table with specific numerical targets. However, the outcome of the clinical study serves as the primary performance metric for the expanded indications for use.
Table 1: Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Successful demonstration of clinical acceptance between the MedComp C3 Wave System ECG Tip Confirmation System positioning results and traditional portable chest x-ray in determining proper distal tip location during a PICC procedure, to support its use as an alternative method to X-ray or fluoroscopy. | 98% success rate when the Bundle Protocol Parameters were met. |
No critical task failures during human factors testing for PICC tip confirmation. | No critical task failures were observed. |
Study Details
Here's a breakdown of the study information as requested:
2. Sample size used for the test set and the data provenance:
- Sample Size: 303 PICCs were placed. Each PICC placement likely represents a single case in the test set.
- Data Provenance: Retrospective, as indicated by the completion of "Clinical Surveys" and data collection from "two healthcare facilities." The study design involved obtaining an X-ray after the C3 Wave System was used, implying a comparison against established practice. Given it's a 510(k) submission, the data is typically from a clinical trial conducted for regulatory purposes. The specific country of origin is not mentioned, but "two healthcare facilities" implies clinical settings.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts: Not explicitly stated how many individual experts independently reviewed each X-ray for ground truth. However, the process refers to "established hospital protocol (e.g., Chest X-Ray, Fluoroscopy)," which implies that the interpretation of the chest X-rays to determine "acceptable tip location" was performed by qualified medical professionals as part of standard clinical practice. This would typically involve radiologists or trained clinicians responsible for interpreting such images.
- Qualifications of Experts: Not specifically detailed. It can be inferred that the chest X-rays were interpreted by clinicians or radiologists according to "clinical judgement and established hospital protocol."
4. Adjudication method for the test set:
- Adjudication Method: Not explicitly detailed. The study mentions that a "subsequent portable chest x-ray was to be obtained... to ascertain acceptable tip location of PICC." This implies that the chest X-ray served as the gold standard against which the C3 Wave System's positioning results were compared. There's no mention of a formal adjudicated consensus process if multiple readers were involved in interpreting the X-rays; it likely followed standard clinical practice for X-ray interpretation.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done in the context of human readers improving with AI vs without AI assistance.
- Nature of Study: This study evaluated the C3 Wave System as a standalone alternative to chest X-ray or fluoroscopy, not as an AI-assistance tool for human readers. The "success rate" refers to the device's ability to correctly identify tip placement compared to the X-ray, not an improvement in human reader performance.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Standalone Performance: Yes, the clinical study primarily evaluated the C3 Wave System for its standalone ability to determine PICC tip location. The system provides real-time electrical activity display, and the "clinician was to complete an electronic survey document to capture specific data points within each procedure including insertion related assessments" which implies the system's output (ECG display) was used by the clinician to guide placement, and then the system's output was judged against the X-ray. The 98% success rate reflects the system's performance in guiding tip placement without requiring an X-ray for initial confirmation.
7. The type of ground truth used:
- Ground Truth: Expert Consensus (implied by clinical practice) / Imaging (Chest X-ray). The "acceptable tip location of PICC" was ascertained by a "subsequent portable chest x-ray," which is then interpreted by medical professionals according to "clinical judgement and established hospital protocol." The X-ray image and its interpretation serve as the ground truth.
8. The sample size for the training set:
- Training Set Sample Size: The document does not provide information about the sample size used for training the C3 Wave System's algorithms. As it's an ECG-based system, its "training" might involve developing and refining the algorithms that interpret ECG changes to infer catheter tip position, which could be based on prior physiological models or data, but specific data for algorithmic training is not present in this summary.
9. How the ground truth for the training set was established:
- Training Set Ground Truth Establishment: This information is not provided in the document. Without details on the training set, how its ground truth was established cannot be determined from this summary.
§ 880.5970 Percutaneous, implanted, long-term intravascular catheter.
(a)
Identification. A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.(b)
Classification. Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”