510(k) Data Aggregation

    K Number
    K180567
    Device Name
    C3 Wave System
    Date Cleared
    2018-04-12

    (38 days)

    Product Code
    Regulation Number
    880.5970
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The C3 Wave System is indicated for use in the positioning of Peripherally Inserted Central catheters (PICC). The C3 Wave provides real-time catheter tip location by displaying changes in the patient's cardiac electrical activity. The C3 Wave is indicated for use as an alternative method to chest X-ray or fluoroscopy confirmation of PICC tip placement in adult patients.

    Note: Limiting, but not contrindicated, situations for this technique are patients where cardiac rhythms may change presentation of the P-Wave: Atrial fibrillation Atrial flutter Severe tachycardia Pacemaker-Driven Rhythm Chronic obstructive pulmonary disease (COPD)

    Such patients are easily identified prior to PICC insertion. Use of additional method is necessary to confirm catheter tip location.

    Device Description

    C3 Wave is designed to provide a continuous display of electrocardiograph [ECG] waveform to be used as a guide in placement of peripherally-inserted central catheters [PICC] in the lower third of the Superior Vena Cava [SVC] of a patient. The principle for operation of this system uses three ECG leads placed on the patient's chest and generates a third ECG lead by switching from RA to PICC stylet. The ECG waveform is wirelessly transmitted to a tablet which allows the operator to view and record changes to the ECG waveform as the tip of the catheter approaches the heart. As the PICC catheter approaches the atrium of the heart, the P wave in the ECG waveform shows substantial changes. This system is designed to aid the visualization of changes in P wave amplitude.

    The C3 Wave system must only be operated by a skilled nurse, physician, or trained medical professional who has been qualified in placement of PICC's and trained in the proper use of this device.

    AI/ML Overview

    This document details the substantial equivalence review for the Medcomp C3 Wave System (K180567), which is an update to a previously cleared device (K170934). The current submission focuses on expanding accessory pack offerings, specifically adding a larger drape and revising the packaging system. Therefore, the "study" described relates to the testing performed to ensure the safety and effectiveness of these modifications, rather than a clinical efficacy study of the core device functionality.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Related to Modifications)Reported Device Performance (Related to Modifications)
    Biocompatibility (ISO 10993-1, -5, -10)Biocompatibility was performed and met for skin-contacting surface, limited exposure.
    Cytotoxicity (ISO 10993-5:2009)Testing performed. (Implied successful if SE is granted)
    Sensitization (ISO 10993-10)Testing performed. (Implied successful if SE is granted)
    Irritation (ISO 10993-10)Testing performed. (Implied successful if SE is granted)
    Sterilization Cycle Development (ANSI/AAMI/ISO 11135-1:2014, Annex B)Developed using Overkill Approach.
    Microbiological Performance Qualification (ANSI/AAMI/ISO 11135-1:2014, Section 9.4.2)Performed in accordance with standard.
    Physical Performance Qualification (ANSI/AAMI/ISO 11135-1:2014, Section 9.4.3)Performed in accordance with standard.
    Ethylene Oxide Sterilization Residuals (AAMI/ANSI/ISO 10993-7:2008)Residuals for C3 ECG Cable Accessory Pack (MRC3RD004) were in accordance with the standard, with a maximum level of 4/mg/device.
    Packaging Integrity (ASTM F 1929-15)Validated with a dye test.
    Packaging Seal Strength (ASTM F1140/F1140M-13)Validated with a compressed air burst test.
    Shipping/Transit TestingISTA 2A performed. (This is listed under "Testing (Accessory Packs)" in Table 6.1 and section "Non-Clinical Testing")

    Note: The document states that "All testing was completed and is the subject of this submission" and concludes that the device raises "no new questions of safety or effectiveness compared to the predicate device." This implies that all acceptance criteria for the non-clinical testing performed on the modified accessory pack were met. Specific numerical performance metrics for each test (beyond the EO residuals) are not detailed in this summary.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text describes non-clinical laboratory testing performed on the modified accessory pack. It does not mention a "test set" in the context of patient data or clinical performance. Therefore, information on sample size for a test set and data provenance (country of origin, retrospective/prospective) is not applicable or not provided for this specific submission, as it focuses on accessory modifications rather than the primary device's clinical performance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable or not provided as the submission focuses on non-clinical testing of device accessories (drape and packaging), not on clinical performance requiring expert ground truth for interpretation of patient data.

    4. Adjudication Method for the Test Set

    This information is not applicable or not provided for the same reasons as point 3.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not performed or described in this document. The C3 Wave System is an ECG-based positioning aid for PICC lines, not an AI-assisted diagnostic tool that would typically involve human readers interpreting images. This submission specifically addresses modifications to accessory packs.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    A standalone performance study of an algorithm is not discussed or applicable to this submission. The C3 Wave System is a device that provides real-time data to a human operator (skilled nurse, physician, or trained medical professional) to aid in PICC placement. The submission pertains to modifications to non-electronic accessories.

    7. The Type of Ground Truth Used

    For the non-clinical testing of accessories, the "ground truth" would be established by standardized laboratory testing procedures and the specifications of relevant international and national consensus standards (e.g., ISO 10993 for biocompatibility, ANSI/AAMI/ISO 11135 for sterilization, ASTM F1929 and F1140 for packaging). This is not expert consensus, pathology, or outcomes data.

    8. The Sample Size for the Training Set

    This information is not applicable or not provided. Since the submission refers to modifications of accessories and non-clinical testing, there is no "training set" in the context of machine learning or AI algorithms.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable or not provided as there is no "training set" in the context of this submission.