(382 days)
BA400: The BA400 analyser is used to determine analyte concentrations by in vitro biochemical and turbidimetric measurements of human samples of serum, urine, plasma, cerebrospinal fluid or total blood. This device is intended to replace manual analytical procedures by performing automatically various steps such as pipetting, heating, and measuring color intensity.
ALBUMIN: Reagent for the measurement of albumin concentration in human serum or plasma. The obtained values are useful as an aid in the evaluation of protein synthesis of the liver in the chronic liver diseases and for the nutritional status. This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory.
ALKALINE PHOSPHATASE (ALP) - AMP: Reagent for the measurement of alkaline phosphatase (ALP)-AMP concentration in human serum or plasma. The obtained values are useful as an aid in the diagnosis and treatment of hepatobiliary and bone diseases with impaired osteoblastic activity diseases. This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory.
GLUCOSE-HEXOKINASE: Reagent for the measurement of glucose concentration in human serum, plasma, urine or cerebrospinal fluid. The obtained values are useful as an aid in the diagnosis and monitoring of the diabetes mellitus. This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory.
Not Found
This document is a 510(k) premarket notification decision letter from the FDA for several in vitro diagnostic reagents and an analyzer. It only provides information about the intended use of the devices and their regulatory classification. It does not contain any data, studies, acceptance criteria, or performance results for the devices mentioned (ALBUMIN, ALKALINE PHOSPHATASE (ALP)-AMP, GLUCOSE-HEXOKINASE reagents, or the BA400 analyzer).
Therefore, I cannot provide the requested information in the requested format because the input document does not contain it.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
April 13, 2018
BioSystems S.A. Teresa Maria Cortes Quality Assurance Manager Costa Brava, 30. 08030 Barcelona Spain
Re: K170901
Trade/Device Name: ALBUMIN, ALKALINE PHOSPHATASE (ALP)-AMP, GLUCOSE-HEXOKINASE, BA400 Regulation Number: 21 CFR 862.1035 Regulation Name: Albumin test system Regulatory Class: Class II Product Code: CIX, CJE, CFR, JJE Dated: March 9, 2018 Received: March 12, 2018
Dear Teresa Maria Cortes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Kellie B. Kelm -S
for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K170901
Device Name
BA400, ALBUMIN, ALKALINE-PHOSPHATASE (ALP) - AMP, GLUCOSE-HEXOKINASE
Indications for Use (Describe)
BA400
The BA400 analyser is used to determine analyte concentrations by in vitro biochemical and turbidimetric measurements of human samples of serum, urine, plasma, cerebrospinal fluid or total blood.
This device is intended to replace manual analytical procedures by performing automatically various steps such as pipetting, heating, and measuring color intensity.
ALBUMIN
Reagent for the measurement of albumin concentration in human serum or plasma. The obtained values are useful as an aid in the evaluation of protein synthesis of the liver in the chronic liver diseases and for the nutritional status. This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory.
ALKALINE PHOSPHATASE (ALP) - AMP
Reagent for the measurement of alkaline phosphatase (ALP)-AMP concentration in human serum or plasma. The obtained values are useful as an aid in the diagnosis and treatment of hepatobiliary and bone diseases with impaired osteoblastic activity diseases. This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory. GLUCOSE-HEXOKINASE
Reagent for the measurement of glucose concentration in human serum, plasma, urine or cerebrospinal fluid. The obtained values are useful as an aid in the diagnosis and monitoring of the diabetes mellitus.
This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
§ 862.1035 Albumin test system.
(a)
Identification. An albumin test system is a device intended to measure the albumin concentration in serum and plasma. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.(b)
Classification. Class II.