K Number
K170314
Device Name
ACL AcuStar
Date Cleared
2017-03-03

(30 days)

Product Code
Regulation Number
864.5425
Panel
HE
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Automated immunoassay analyzer designed specifically for in vitro diagnostic use in a clinical laboratory. The assay analysis is based on chemiluminescent technology. The system provides results for both direct measurements and calculated parameters.

Device Description

The AcuStar is an automated, bench-top system for lab use that measures the analyte amount in samples by: Subjecting the sample to reagents that cause a reaction with an antigen or antibody in the sample. Placing the cuvettes in a controlled environment to let the reactants bind into a complex. Separating out the complex from unused reactants. Treating this complex with a chemical that produces light in proportion to the analyte concentration. Measuring the light output to determine the amount of antibodies or antigens that were in the sample.

AI/ML Overview

This document is a 510(k) premarket notification for a software update to an existing medical device, the ACL AcuStar™. The submission qualifies as a "Special 510(k)" because the modification (updating the operating system from Windows XP to Windows 7) does not change the indications for use, operating principle, labeled performance claims, hardware, data reduction software, fluidic design, test parameters, calibration, quality control, consumables, or reagents. Therefore, the information provided focuses on demonstrating that the updated device is substantially equivalent to the predicate device and does not involve a new clinical study to establish acceptance criteria or device performance in the same way a de novo submission would.

Given the nature of this Special 510(k), the prompt's questions regarding acceptance criteria, device performance, sample sizes, expert ground truth, adjudication methods, MRMC studies, and standalone performance are not directly applicable in the traditional sense for a new device evaluation. Instead, the "acceptance criteria" here relate to demonstrating that the change did not negatively impact the device's original validated performance.

Here's how the information aligns with the prompt's request, interpreted in the context of a Special 510(k) for a software update:

1. A table of acceptance criteria and the reported device performance

The document does not present a table of specific performance acceptance criteria (e.g., sensitivity, specificity, accuracy) for the device itself because it's a software update to an already cleared device. Instead, the implicit acceptance criterion is that the device's performance remains unchanged despite the operating system upgrade.

The "reported device performance" in this context is that the updated device is substantially equivalent to the predicate device, meaning its performance characteristics relevant to its intended use have not been altered.

Acceptance Criteria (Implicit for Special 510(k) Software Update)Reported Device Performance (Conclusion)
No change in Indications for UseSame
No change in Operating PrincipleSame
No change to Labeled Performance ClaimsSame (implies original performance claims are met)
No change to HardwareSame
No change to Data Reduction SoftwareSame
No change to Fluidic DesignSame
No change to Test ParametersSame
No change to CalibrationSame
No change to Quality ControlSame
No change to ConsumablesSame
No change to ReagentsSame
Overall substantial equivalence to predicate deviceThe ACL AcuStar™ with software v3.0.1 running on Windows 7 is substantially equivalent to the cleared and currently marketed predicate device (K083518).

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document does not detail a specific "test set" or its sample size in the context of clinical performance evaluation for this Special 510(k). The focus is on demonstrating that the software change itself does not affect the existing device's validated performance. For a software update of this nature, testing would typically involve verification and validation (V&V) activities against the original design specifications and potentially comparing outputs from the Windows XP and Windows 7 versions using a set of representative samples, but these details are not provided in this summary. There is no mention of country of origin or retrospective/prospective data for a new performance study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

This is not applicable to this Special 510(k). The regulatory submission for a software operating system update does not involve establishing new diagnostic ground truth with medical experts. The ground truth for the underlying assays was established during the initial clearance of the predicate device (K083518).

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

This is not applicable to this Special 510(k) as no new clinical test set requiring expert adjudication is described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The ACL AcuStar™ is an automated immunoassay analyzer, not a device involving human interpretation of images or data that would benefit from AI assistance in the way described for an MRMC study.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

The ACL AcuStar™ is an automated system that performs tests in a standalone manner (without continuous human intervention for each test analysis beyond loading samples and reagents). The original predicate device's performance was established as a standalone automated system. The current submission simply states that the software update does not change the operating principle, performance claims, or data reduction software, implying the standalone performance remains identical to the predicate.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth for the diagnostic assays performed by the ACL AcuStar™ would have been established during the original clearance of the predicate device (K083518), likely through comparison to reference methods, clinical diagnosis, or other accepted gold standards for in vitro diagnostic tests. This Special 510(k) does not provide details on the ground truth for the original assays, as it is assumed to be established with the predicate.

8. The sample size for the training set

This is not applicable. The device is not an AI/ML algorithm that requires a training set in the common understanding of an "AI device." It's an automated immunoassay analyzer with a software operating system update.

9. How the ground truth for the training set was established

This is not applicable for the same reason as point 8.

§ 864.5425 Multipurpose system for in vitro coagulation studies.

(a)
Identification. A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.(b)
Classification. Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.