K Number
K170293
Date Cleared
2017-10-25

(267 days)

Product Code
Regulation Number
862.3250
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Emit® II Plus Cocaine Metabolite Assay is a homogeneous enzyme immunoassay with a 150 ng/mL or 300 ng/mL cutoff. The assay is intended for use in the qualitative and semiquantitative analyses of benzoylecgonine (cocaine metabolite) in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers.

The Emit® II Plus Cocaine Metabolite Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

Device Description

The Emit II Plus Cocaine Metabolite assay is a homogeneous enzyme immunoassay that qualitatively and semiquantitatively measures benzoylecgonine. The assay has cutoffs of 150 ng/mL and 300 ng/mL benzoylecgonine.

The assay is based on competition between drug in the specimen and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for antibody binding sites. Enzyme activity decreases upon binding to the antibody, so the drug concentration in the specimen can be measured in terms of enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that is measured spectrophotometrically at 340 nm.

The Emit II Plus Cocaine Metabolite Assay consists of two ready-to-use reagents:

Antibody/Substrate Reagent 1
Sheep polyclonal antibodies to benzoylecgonine (2.2 µg/mL), bovine serum albumin, G6P (15 mM), NAD (12 mM), preservatives, and stabilizers

Enzyme Reagent 2
Benzoylecgonine labeled with bacterial G6PDH (0.46 U/mL), HEPES buffer, bovine serum albumin, preservatives, and stabilizers

AI/ML Overview

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" for the method comparison studies. However, the reported "Result" for agreement percentages can be interpreted as the performance achieved. Similarly, for precision studies, "Results" indicate the number of negative/positive determinations. For linearity, it's % recovery.

Table 1: Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance Criteria (Implied)Reported Device PerformanceComments
Method Comparison (Qualitative & Semi-Quantitative)
150 ng/mL CutoffHigh agreement with GC/MS91% agreementThe document does not specify a numerical threshold for "high agreement."
300 ng/mL CutoffHigh agreement with GC/MS88% agreementThe document does not specify a numerical threshold for "high agreement."
Precision (Qualitative & Semi-Quantitative)
150 ng/mL Cutoff
0 ng/mL100% Negative80 NegativeAchieved
38 ng/mL (-75%)100% Negative80 NegativeAchieved
75 ng/mL (-50%)100% Negative80 NegativeAchieved
113 ng/mL (-25%)100% Negative80 NegativeAchieved
150 ng/mL (Cutoff)Consistent classification (~50% Positive/Negative)9 Negative / 71 PositiveAt the cutoff, some variability in classification is expected. The exact split for "acceptance" isn't defined but this is a common observation.
188 ng/mL (+25%)100% Positive80 PositiveAchieved
225 ng/mL (+50%)100% Positive80 PositiveAchieved
263 ng/mL (+75%)100% Positive80 PositiveAchieved
300 ng/mL (+100%)100% Positive80 PositiveAchieved
300 ng/mL Cutoff
0 ng/mL100% Negative80 NegativeAchieved
75 ng/mL (-75%)100% Negative80 NegativeAchieved
150 ng/mL (-50%)100% Negative80 NegativeAchieved
225 ng/mL (-25%)100% Negative80 NegativeAchieved
300 ng/mL (Cutoff)Consistent classification (~50% Positive/Negative)54 Negative / 26 PositiveAt the cutoff, some variability in classification is expected.
375 ng/mL (+25%)100% Positive80 PositiveAchieved
450 ng/mL (+50%)100% Positive80 Negative (This appears to be an error in the document, should likely be positive)Note: The document states 80 Negative for +50% of 300 ng/mL cutoff (450 ng/mL) which is contradictory to expectation for a positive result above cutoff. This might be a typo in the provided text. Based on other results, it should be 80 Positive.
525 ng/mL (+75%)100% Positive80 PositiveAchieved
600 ng/mL (+100%)100% Positive80 PositiveAchieved
Recovery/Linearity (Semiquantitative)
% RecoveryNot explicitly stated (e.g., ±X%)Ranges from -0.7% to 8.6%The document does not specify an acceptance range for % Recovery (e.g., 90-110%).
Specificity - Structurally Related Compounds
Cross-ReactivityLow cross-reactivity for non-benzoylecgonine compoundsEcgonine: 2-3%, Cocaine: 0.5%, others <0.01%The document lists values but does not define an explicit acceptance criteria (e.g., <5% for non-target compounds).

2. Sample Size Used for the Test Set and Data Provenance

  • Method Comparison:
    • 150 ng/mL Cutoff: n = 102 native patient urine samples.
    • 300 ng/mL Cutoff: n = 95 native patient urine samples.
    • Data Provenance: Retrospective, as indicated by "native patient urine samples." The country of origin is not specified.
  • Precision (Repeatability and Within-Lab Precision):
    • Sample Size: 9 sample levels for each cutoff (150 ng/mL and 300 ng/mL). Each level was determined 80 times (20 days, 2 runs/day, 2 replicates/run). This implies 9 * 80 = 720 individual determinations per cutoff, across the various concentrations.
    • Data Provenance: The samples were "prepared by spiking," implying laboratory-controlled samples, not native patient samples. Data provenance (country of origin) is not specified.
  • Recovery/Linearity:
    • Sample Size: Two aliquots of negative urine pool spiked to create samples. Five replicates were tested at each of the 10 concentration levels. This implies 5 * 10 = 50 individual determinations.
    • Data Provenance: Laboratory-prepared samples, not native patient samples. Data provenance (country of origin) is not specified.
  • Specificity - Structurally Related Compounds:
    • Sample Size: Not explicitly stated as a "sample size" for a study, but rather concentrations at which various compounds were tested for cross-reactivity.
    • Data Provenance: Laboratory-controlled testing, not native patient samples. Data provenance (country of origin) is not specified.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

The ground truth for the test set (Method Comparison) was established by Gas Chromatography/Mass Spectrometry (GC/MS). GC/MS is an analytical chemistry technique, not a human expert. Therefore, the concept of "number of experts" and their "qualifications" is not applicable in this context.

4. Adjudication Method for the Test Set

Not applicable. The ground truth was established by an objective analytical method (GC/MS), not through human interpretation or adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This device is an automated in vitro diagnostic assay (an enzyme immunoassay) designed to detect a cocaine metabolite in urine. It does not involve human readers for interpretation, nor does it incorporate AI in a way that would assist human interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the performance studies described (Method Comparison, Precision, Recovery/Linearity, Specificity) represent standalone performance of the device (assay on the analyzer). The results are generated directly by the Emit® II Plus Cocaine Metabolite Assay on the Beckman Coulter DxC 700 AU Clinical Chemistry Analyzer, without human interpretive input. The output is a qualitative (positive/negative) or semi-quantitative result.

7. The Type of Ground Truth Used

The primary ground truth used for the method comparison study was Gas Chromatography/Mass Spectrometry (GC/MS). For precision, linearity, and specificity studies, the ground truth was based on known concentrations of the analyte in prepared samples.

8. The Sample Size for the Training Set

The document does not describe a "training set" in the context of machine learning or AI. This is a traditional in vitro diagnostic device, not an AI/ML-based device.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no "training set" described for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 25, 2017

SIEMENS HEALTHCARE DIAGNOSTICS INC. ALAN HALEY REGULATORY CLINICAL AFFAIRS SPECIALIST 500 GBC Drive, M/S 514 NEWARK, DE. 19702

Re: K170293

Trade/Device Name: Emit II Plus Cocaine Metabolite Assay Regulation Number: 21 CFR 862.3250 Regulation Name: Cocaine and cocaine metabolite test system Regulatory Class: II Product Code: DIO Dated: October 16, 2017 Received: October 17, 2017

Dear Alan Haley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Courtney H. Lias -S

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170293

Device Name Emit® II Plus Cocaine Metabolite Assay

Indications for Use (Describe)

The Emit® II Plus Cocaine Metabolite Assay is a homogeneous enzyme immunoassay with a 150 ng/mL or 300 ng/mL cutoff. The assay is intended for use in the qualitative and semiquantitative analyses of benzoylecgonine (cocaine metabolite) in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers.

The Emit® II Plus Cocaine Metabolite Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SIEMEN?

510(k) Summary

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of Safe Medical Device Act of 1990 and 21 CFR 807.92.

The assigned 510(k) number is K170293

1. Submitter

Company Siemens Healthcare Diagnostics Inc.

500 GBC Drive, M/S 514 Newark, DE 19702 Address

Contact Alan Haley

302.631.9883 Telephone

302.631.6299 Fax

Date of Preparation 2.

October 24, 2017

3. Device Information

Trade NameEmit® II Plus Cocaine Metabolite Assay
Common NameCocaine Metabolite Assay
Classification NameEnzyme Immunoassay, Cocaine and Cocaine Metabolites
Regulation21 CFR 862.3250
Device ClassClass II
Product CodeDIO

4. Indications for Use

The Emit® II Plus Cocaine Metabolite Assay is a homogeneous enzyme immunoassay with a 150 ng/mL or 300 ng/mL cutoff. The assay is intended for use in the qualitative and semiguantitative analyses of benzoylecgonine (cocaine metabolite) in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers.

The Emit® II Plus Cocaine Metabolite Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred

confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

5. Device Description

The Emit II Plus Cocaine Metabolite assay is a homogeneous enzyme immunoassay that qualitatively and semiquantitatively measures benzoylecgonine. The assay has cutoffs of 150 ng/mL and 300 ng/mL benzoylecgonine.

The assay is based on competition between drug in the specimen and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for antibody binding sites. Enzyme activity decreases upon binding to the antibody, so the drug concentration in the specimen can

{4}------------------------------------------------

SIEMEN?

be measured in terms of enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that is measured spectrophotometrically at 340 nm.

The Emit II Plus Cocaine Metabolite Assay consists of two ready-to-use reagents:

Antibody/Substrate Reagent 1

Sheep polyclonal antibodies to benzoylecgonine (2.2 µg/mL), bovine serum albumin, G6P (15 mM), NAD (12 mM), preservatives, and stabilizers

Enzvme Reagent 2

Benzoylecgonine labeled with bacterial G6PDH (0.46 U/mL), HEPES buffer, bovine serum albumin, preservatives, and stabilizers

Medical Device to Which Equivalence is Claimed 6.

The Siemens Healthcare Diagnostics Inc. Emit® II Plus Cocaine Metabolite Assay has the same intended use and indications for use as the Emit II Plus Cocaine Metabolite Assay previously cleared under 510(k) K993988 (January 27, 2000). Siemens Healthcare Diagnostics provides instructions for using this assay on a number of chemistry analyzers.

The predicate and proposed device were considered side to examine the similarities and differences between the devices.

AttributeEmit II Plus CocaineMetabolite Assay(K993988)Emit II Plus CocaineMetabolite Assay(Proposed)
Intended UseThe Emit® II Plus CocaineMetabolite Assay is ahomogeneous enzymeimmunoassay with a 150 ng/mL(SAMHSA initial test cutoff level)or 300 ng/mL cutoff. The assayis intended for use in thequalitative and semiquantitativeanalyses of benzoylecgonine(cocaine metabolite) in humanurine. Emit® II Plus assays aredesigned for use with a numberof chemistry analyzers.The Emit® II Plus CocaineMetabolite Assay provides onlya preliminary analytical testresult. A more specificalternative chemical methodmust be used to obtain aconfirmed analytical result. Gaschromatography/massspectrometry (GC/MS) is theThe Emit® II Plus CocaineMetabolite Assay is ahomogeneous enzymeimmunoassay with a 150 ng/mLor 300 ng/mL cutoff. The assayis intended for use in thequalitative and semiquantitativeanalyses of benzoylecgonine(cocaine metabolite) in humanurine. Emit® II Plus assays aredesigned for use with a numberof chemistry analyzers.The Emit® II Plus CocaineMetabolite Assay provides onlya preliminary analytical testresult. A more specificalternative chemical methodmust be used to obtain aconfirmed analytical result. Gaschromatography/massspectrometry (GC/MS) is thepreferred confirmatory method.

7. Comparison of Technological Characteristics with the Predicate Device

{5}------------------------------------------------

SIER

AttributeEmit II Plus CocaineMetabolite Assay(K993988)Emit II Plus CocaineMetabolite Assay(Proposed)
preferred confirmatory method.Other chemical confirmationmethods are available. Clinicalconsideration and professionaljudgment should be applied toany drug-of-abuse test result,particularly when preliminarypositive results are used.Other chemical confirmationmethods are available. Clinicalconsideration and professionaljudgment should be applied toany drug-of-abuse test result,particularly when preliminarypositive results are used.
MeasurandBenzoylecgonineSame
Type of TestQualitative and semi-quantitativehomogeneous enzymeimmunoassaySame
AntibodySheep polyclonalSame
Reagent FormLiquid, ready to useSame
ReagentCompositionAntibody/Substrate Reagent 1Sheep polyclonal antibodies tobenzoylecgonine (2.2 µg/mL),bovine serum albumin, G6P(15 mM), NAD (12 mM),preservatives, and stabilizersEnzyme Reagent 2Benzoylecgonine labeled withbacterial G6PDH (0.46 U/mL),HEPES buffer, bovine serumalbumin, preservatives, andstabilizersSame
Cutoffs/Controls150 ng/mL (±25%)300ng/mL (±25%)Same
Sample MatrixHuman UrineSame

8. Performance Data

The following tests were conducted to assess the critical performance parameters of the Emit® II Plus Cocaine Metabolite assay on the Beckman Coulter DxC 700 AU Clinical Chemistry Analyzer.

8.1 Method Comparison

The benzoylecgonine values of native patient urine samples were measured using the Emit® II Plus Cocaine Metabolite assay on the Beckman Coulter DxC 700 AU Clinical Chemistry Analyzer (proposed) and were compared to the values measured using GC/MS.

{6}------------------------------------------------

GC/MS
Negative(<75 ng/mL)Negative(75 - 149 ng/mL)Positive(150 - 225 ng/mL)Positive(> 225 ng/mL)
Qualitative
DxCPositive271051
700 AUNegative28400
Semi-Quantitative
DxCPositive271051
700 AUNegative28400

Table 8.1(a) Method Comparison Results for the 150 ng/mL Cutoff (n = 102)

Result: 91% agreement

Table 8.1(b) Summary of Discordant Results, 150 ng/mL cutoff, Qualitative

Emit II PlusCocaine MetaboliteAssay (pos/neg)GC/MS(ng/mL)Major Drug Present byGC/MS
Positive124Benzoylecgonine
Positive34Benzoylecgonine
Positive73Benzoylecgonine
Positive83Benzoylecgonine
Positive102Benzoylecgonine
Positive80Benzoylecgonine
Positive105Benzoylecgonine
Positive101Benzoylecgonine
Positive75Benzoylecgonine

Table 8.1(c) Summary of Discordant Results, 150 ng/mL cutoff, Semi-Quantitative

Emit II PlusCocaine MetaboliteAssay (pos/neg)GC/MS(ng/mL)Major Drug Present byGC/MS
Positive124Benzoylecgonine
Positive34Benzoylecgonine
Positive73Benzoylecgonine
Positive83Benzoylecgonine
Positive102Benzoylecgonine
Positive80Benzoylecgonine
Positive105Benzoylecgonine
Positive101Benzoylecgonine
Positive75Benzoylecgonine

Table 8.1(d) Method Comparison Results for the 300 ng/mL Cutoff (n = 95)

GC/MS
Negative(<150 ng/ml)Negative(150 - 299 ng/mL)Positive(300 - 450 ng/mL)Positive(> 450 ng/mL)
Qualitative
DxCPositive1101825
700 AUNegative40100
Semi-Quantitative
DxCPositive1101825
700 AUNegative40100

Result: 88% agreement

{7}------------------------------------------------

Emit II PlusCocaine MetaboliteAssay (pos/neg)GC/MS(ng/mL)Major Drug Present byGC/MS
Positive217Benzoylecgonine
Positive240Benzoylecgonine
Positive173Benzoylecgonine
Positive207Benzoylecgonine
Positive233Benzoylecgonine
Positive143Benzoylecgonine
Positive217Benzoylecgonine
Positive173Benzoylecgonine
Positive232Benzoylecgonine
Positive157Benzoylecgonine
Positive234Benzoylecgonine

Table 8.1(e) Summary of Discordant Results, 300 ng/mL cutoff, Qualitative

Table 8.1(f) Summary of Discordant Results, 300 ng/mL cutoff, Semi-Quantitative

Emit II PlusCocaine MetaboliteAssay (pos/neg)GC/MS(ng/mL)Major Drug Present byGC/MS
Positive217Benzoylecgonine
Positive240Benzoylecgonine
Positive173Benzoylecgonine
Positive207Benzoylecgonine
Positive233Benzoylecgonine
Positive143Benzoylecgonine
Positive217Benzoylecgonine
Positive173Benzoylecgonine
Positive232Benzoylecgonine
Positive157Benzoylecgonine
Positive234Benzoylecgonine

8.2 Repeatability and Within-Lab Precision

A precision study was performed for 20 days, 2 runs per day, 2 replicates per run per day (N=80). Nine sample levels were tested for each cutoff : equal to the cutoff, ± 25% of the cutoff, ± 50% of the cutoff, ± 75% of the cutoff, and ±100% of the cutoff. Data were analyzed using a nested, two factor (days and runs nested within days) ANOVA model. Testing was performed on a Beckman Coulter DxC 700 AU Clinical Chemistry Analyzer.

BenzoylecgonineConcentration(ng/mL)% ofcutoff# ofdeterminationsRepeatabilityResultWithin-LabResult
0-100%8080 Negative80 Negative
38-75%8080 Negative80 Negative
75-50%8080 Negative80 Negative
113-25%8080 Negative80 Negative
150Cutoff809 Negative71 Positive9 Negative71 Positive
188+25%8080 Positive80 Positive
225+50%8080 Positive80 Positive
263+75%8080 Positive80 Positive
300100%8080 Positive80 Positive

Table 8.2(a) Precision, Qualitative, 150 ng/mL Cutoff

{8}------------------------------------------------

SIEMENS

BenzoylecgonineConcentration(ng/mL)% ofcutoff# ofdeterminationsRepeatabilityResultWithin-LabResult
0-100%8080 Negative80 Negative
38-75%8080 Negative80 Negative
75-50%8080 Negative80 Negative
113-25%8080 Negative80 Negative
150Cutoff809 Negative71 Positive9 Negative71 Positive
188+25%8080 Positive80 Positive
225+50%8080 Positive80 Positive
263+75%8080 Positive80 Positive
300100%8080 Positive80 Positive

Table 8.2(b) Precision. Semi-Quantitative. 150 ng/mL Cutoff

Table 8.2(c) Precision, Qualitative, 300 ng/mL Cutoff

BenzoylecgonineConcentration(ng/mL)% ofcutoff# ofdeterminationsRepeatabilityResultWithin-LabResult
0-100%8080 Negative80 Negative
75-75%8080 Negative80 Negative
150-50%8080 Negative80 Negative
225-25%8080 Negative80 Negative
300Cutoff8054 Negative26 Positive54 Negative26 Positive
375+25%8080 Positive80 Positive
450+50%8080 Negative80 Negative
525+75%8080 Positive80 Positive
600+100%8080 Positive80 Positive

Table 8.2(d) Precision, Semi-Quantitative, 300 ng/mL Cutoff

BenzoylecgonineConcentration(ng/mL)% ofcutoff# ofdeterminationsRepeatabilityResultWithin-LabResult
0-100%8080 Negative80 Negative
75-75%8080 Negative80 Negative
150-50%8080 Negative80 Negative
225-25%8080 Negative80 Negative
300Cutoff8054 Negative26 Positive54 Negative26 Positive
375+25%8080 Positive80 Positive
450+50%8080 Negative80 Negative
525+75%8080 Positive80 Positive
600+100%8080 Positive80 Positive

{9}------------------------------------------------

SIEME

8.3 Recovery/Linearity

Samples were prepared by spiking two aliguots of negative urine pool with known concentrations of benzoylecqonine and sequentially mixing them to create samples across the analytical range. Five replicates were tested at each level. Testing was performed on a Beckman Coulter DxC 700 AU Clinical Chemistry Analyzer.

NominalBenzoylecgonine (ng/mL)Mean MeasuredBenzoylecgonine (ng/mL)% Recovery
50536.1%
1001033.1%
1501500.2%
2252354.3%
300298-0.7%
3753801.4%
5005438.6%
7507925.6%
9009060.6%
100010292.9%

Table 8.3(a) Recovery/Linearity. Semiquantitative

8.4 Specificity - Structurally Related Compounds

The Emit® II Plus Cocaine Metabolite Assay detects benzoylecgonine, the major metabolite of cocaine, in human urine. Tables 8.4(a) and 8.4(b) list the cross-reactivity for structurally related compounds. Data presented are representative of typical performance of this assay.

Table 8.4(a) Structurally Related Compounds, 150 ng/mL cutoff

CompoundConcentration Tested(ng/mL)% Cross-Reactivity
Ecgonine*5,0003%
Cocaine*29,0000.5%
Norcocaine100,000<0.01%
Cocaethylene100,000<0.01%
Ecgonine Methyl Ester100,000<0.01%

Table 8.4(b) Structurally Related Compounds, 300 ng/mL cutoff

CompoundConcentration Tested(ng/mL)% Cross-Reactivity
Ecgonine*15,0002%
Cocaine*61,0000.5%
Norcocaine100,000<0.01%
Cocaethylene100,000<0.01%
Ecgonine Methyl Ester100,000<0.01%

*Ecgonine and cocaine tested at the concentrations above produced a result approximately equivalent to the cutoff.

9. Conclusion

The information presented in this premarket notification demonstrates that the Emit® II Plus Cocaine Metabolite Assay is substantially equivalent to the legally marketed predicate device identified above.

§ 862.3250 Cocaine and cocaine metabolite test system.

(a)
Identification. A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.(b)
Classification. Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).