K Number
K163276
Date Cleared
2017-05-11

(171 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

KD-721 and KD-723 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm-25cm.

Device Description

KD-721 and KD-723 Fully Automatic Electronic Blood Pressure Monitor is designed and manufactured according to IEC 80601-2-30.

The operational principle is based on oscillometric and silicon integrates pressure sensor technology. It can calculate the systolic and diastolic blood pressure, and display the result on the LCD. The measurements results can also be classified by the function of blood pressure classification indicator. If any irregular heartbeat is detected, it can be shown to the user.

AI/ML Overview

The provided text describes the acceptance criteria and the study for the KD-721 and KD-723 Fully Automatic Electronic Blood Pressure Monitor.

1. Table of Acceptance Criteria and Reported Device Performance:

The document refers to compliance with particular standards, which implicitly define the acceptance criteria for certain performance aspects. The device characteristics are compared to those of a predicate device (iHealth View BP7S Wireless Blood Pressure Wrist Monitor, K152379) and a previous device (KD-972, K121470).

Feature / Acceptance CriteriaReported Device Performance (KD-721/723)
Pulse Rate Range40 - 180 times/min
Pulse Rate AccuracyWithin ±5%
Systolic Range60 - 260 mmHg
Diastolic Range40 - 199 mmHg
Pressure AccuracyWithin ±3 mmHg
Cuff Pressure Range0 - 300 mmHg
Overpressure Limit300 mmHg
Electromagnetic CompatibilityComplies with IEC 60601-1-2:2014
Electrical SafetyComplies with IEC 60601-1:2005/(R)2012 And A1:2012,C1:2009/(R)2012 And A2:2010/(R)2012
Safety & Performance CharacteristicsComplies with IEC 80601-2-30:2009 & A1:2013

2. Sample size used for the test set and the data provenance:

The document states that the clinical test report of KD-972 (cleared in 2012 as K121470) was used for evaluating KD-721/723. This implies retrospective use of clinical data.
The document states that KD-972 conformed to ANSI/AAMI/ISO 81060-2:2009 & 2013. These standards typically outline the requirements for clinical validation studies for automated sphygmomanometers, which involve a specific number of subjects. However, the exact sample size for the test set is not explicitly stated in the provided text. The data provenance is not specified beyond being from a previous device's clinical trial conforming to international standards (ANSI/AAMI/ISO 81060-2), which generally requires multi-center data from diverse populations.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

The document does not specify the number or qualifications of experts used to establish the ground truth for the test set. Clinical validation studies for blood pressure monitors generally involve trained observers (often medical professionals) to perform reference measurements.

4. Adjudication method for the test set:

The document does not specify the adjudication method used for the test set.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This device is a fully automatic electronic blood pressure monitor, not an AI-assisted diagnostic tool that involves human readers interpreting output. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not done.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Yes, a standalone performance evaluation was done through the reference to the KD-972 clinical test report, which conformed to ANSI/AAMI/ISO 81060-2:2009 & 2013. This standard specifically outlines the requirements for evaluating the accuracy of automated sphygmomanometers, which inherently tests the algorithm's performance in measuring blood pressure values.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The ground truth for blood pressure monitors in clinical validation studies conforming to ANSI/AAMI/ISO 81060-2 is typically established through auscultatory measurements performed by trained observers using a mercury sphygmomanometer or an equivalent reference device, often in a double-blinded protocol to minimize bias. This is a form of expert reference measurement.

8. The sample size for the training set:

The document states that the algorithm version (BPM-WAU V2.0-201109) is the same for KD-721/723 and the previously cleared KD-972. This implies that the algorithm was trained prior to the KD-972 clearance (K121470 in 2012). The sample size for the training set is not provided in this document, as the focus is on the clinical validation of the device using an existing algorithm.

9. How the ground truth for the training set was established:

Given that the algorithm is based on "oscillometric and silicon integrates pressure sensor technology" and is a "fully automatic" device, the ground truth for its original training would have been established through a combination of simulated data and clinical data where reference blood pressure measurements (e.g., auscultatory method by trained experts) were used to optimize the algorithm's performance in detecting systolic, diastolic, and pulse rate. The specific details of the training data and ground truth establishment are not provided in this document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it, overlaid with three human profiles facing right. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 11, 2017

Andon Health Co.,ltd Liu Yi President No 3, Jin Ping St, Ya An Rd Nankai District Tianjin, 300190 CN

Re: K163276

Trade/Device Name: KD-721 and KD-723 Fully Automatic Electronic Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: April 6, 2017 Received: April 10, 2017

Dear Liu Yi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

M.A. Hillemann

for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K163276

Device Name

KD-721 and KD-723 Fully Automatic Electronic Blood Pressure Monitor

Indications for Use (Describe)

KD-721 and KD-723 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm-25cm.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D) Over-the-Counter Use
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

ver-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.

1.0 submitter's information

Name:Andon Health Co., Ltd.
Address:No 3, Jinping Street Ya An Road, Nankai District, Tianjin, P.R. China
Phone number:86-22-6052 6161
Fax number:86-22-6052 6162
Contact:Liu Yi
Date of Preparation:11/15/2016

2.0 Device information

Tradename:KD-721andKD-723FullyAutomatic
Electronic Blood Pressure Monitor
Common name:Noninvasive blood pressure measurement system

Classification name: Noninvasive blood pressure measurement system

3.0 Classification

Production code: DXN- Noninvasive blood pressure measurement system. Regulation number: 870.1130 Classification: II Panel: Cardiovascular

4.0 Predicate device information

Manufacturer: Andon Health Co., Ltd. Device: iHealth View BP7S Wireless Blood Pressure Wrist Monitor 510(k) number: K152379

{4}------------------------------------------------

5.0 Intended use

KD-721 and KD-723 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm-25cm.

The intended use and the indication for use of KD-721 and KD-723, as described in its labeling are the same as the predicate device BP7S (K152379).

6.0 Device description

KD-721 and KD-723 Fully Automatic Electronic Blood Pressure Monitor is designed and manufactured according to IEC 80601-2-30.

The operational principle is based on oscillometric and silicon integrates pressure sensor technology. It can calculate the systolic and diastolic blood pressure, and display the result on the LCD. The measurements results can also be classified by the function of blood pressure classification indicator. If any irregular heartbeat is detected, it can be shown to the user.

7.0 Summary comparing technological characteristics with predicate device

Technological CharacteristicsComparison result
Design principleIdentical
AppearanceSimilar
Patients contact MaterialsSimilar
PerformanceSimilar
BiocompatibilityIdentical
Mechanical safetyIdentical
Standards metIdentical
Electrical safetyIdentical
EMCIdentical
FunctionSimilar

{5}------------------------------------------------

8.0 Discussion of non-clinical and clinical test performed

Non-clinical Tests have been done as follows:

a. Electromagnetic compatibility test according to IEC 60601-1-2;

b. Electrical safety according test to IEC 60601-1;

c. Safety and performance characteristics of the test according to IEC 80601-2-30

None of the test demonstrates that KD-721 and KD-723 Fully Automatic Electronic Blood Pressure Monitor bring new questions of safety and effectiveness.

Clinical Tests discussion

KD-721/723 and KD-972(cleared in 2012 as K121470) are blood pressure monitors based on oscillometric and silicon integrates pressure sensor technology, when inflating the cuff, systolic and diastolic signal will be collected, using the algorithm to get final value and record. The algorithm version of both the new device and KD-972 is: BPM-WAU V2.0-201109, and the cuff size used is exactly the same, so the clinical test report of KD-972(that conformed to ANSI/AAMI/ISO 81060-2:2009 & 2013) can be used for evaluating KD-721/723.

9.0 Performance summary

KD-721 and KD-723 Fully Automatic Electronic Blood Pressure Monitor conforms to the following standards:

  • IEC 60601-1:2005/(R)2012 And A1:2012,C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD)
  • · IEC 60601-1-2:2014.Medical Electrical Equipment -- Part 1-2: General Requirements For Basic Safety And Essential Performance -- Collateral Standard: Electromagnetic Disturbances -- Requirements And Tests
  • IEC 80601-2-30:2009 & A1:2013,Medical Electrical Equipment Part 2-30: Particular Requirements For The Basic Safety And Essential Performance Of Automated Noninvasive Sphygmomanometers

10.0 Comparison to the predicate device and the conclusion

{6}------------------------------------------------

Our device KD-721 and KD-723 Fully Automatic Electronic are substantially equivalent to iHealth BP7S Wireless Blood Pressure Wrist Monitor whose 510(k) number is K152379.

ItemPredicate DeviceSubject DeviceDeviceComparison
Name andmodeBP7S WirelessBloodPressure WristMonitorKD-721 WirelessBloodPressure WristMonitorKD-723 WirelessBloodPressure WristMonitorThe name is differentbecause the model isdifferent.
Intended UseMonitor anddisplay diastolic,systolic bloodpressure and pulserateMonitor anddisplay diastolic,systolic bloodpressure and pulserateMonitor anddisplay diastolic,systolic bloodpressure andpulse rateSame
Rx or OTCOTCOTCOTCSame
PopulationAdultAdultAdultSame
Cuff LocationWristWristWristSame
Physical Attributes
WeightAbout 114g(include cuff)About 67g(exclude batteries)About 69g(excludebatteries)The weight and isdifferent, because theappearance is changed.
Dimensions72mmX74mmX17.6mm78mm×60mm×28mm80mm×60mm×22mmBut the test conform thatthis does not impactsafety and effectiveness.
MemoryNA2×60 times2×60 timesAlthough the memoryfunction is changed, thereis no impact on safety andeffectiveness for thischange.
DisplayedCalculatedParametersSYSDIAPulseIHBSYSDIAPulseIHBSYSDIAPulseIHBSame
Displaycomponent1 LCD2 Display on SmartPhone1 LCD2 Display on SmartPhone1 LCD2 Display onSmart PhoneSame
AveragefunctionNo averagefunctionaverage value ofall the resultsmeasured from 5o'clock to 9 o'clockin last 7 daysAveraging all theresultsAlthough the averagefunction is changed , thiswill not impact on safetyand effectiveness.
OtherDisplayedDateTimeDateTimeDateTimeSame
Information
MemoryBattery usageBluetooth symbolcharge symbolsync symbol(Displayed ondevice)MemoryBattery usageBluetooth symbolBlood pressureclassification(Displayed ondevice)MemoryBattery usageBluetooth symbolBlood pressureclassification(Displayed ondevice)
OtherFeaturesTransmit data tosmart phone viaBluetoothWHO classificationTransmit data tosmart phone viaBluetoothJNC classificationTransmit data tosmart phone viaBluetoothWHOclassificationSame data transmissionThe classification ofKD-721 is different, butthis change will not affectsafety and effectiveness.
Apps on smart phone
Function onappStorage test result/Read memories inApp/Deletememories in app/Draw table andgraph/Set alarm inapp/Set thepersonalinformation in app/Share history datain appStorage anddisplay test resultsonlyStorage anddisplay testresults onlyThe new device can onlystore test results anddisplay, but this changewill not affect safety andeffectiveness
Electrical Power
Battery1*3.7V Li-ion400mAh2x1.5V SIZE AAA2x1.5V SIZE AAAThe battery is changed, tbut this will not affectsafety and effectiveness.
Environmental Operation
Temperature10~40°C10~40°C10~40°CSame
Humidity≤85%≤85%≤85%Same
Environmental Storage
Temperature-20~55°C-20~50°C-20~50°CAlthough the storagetemperature of KD-721and KD-723 is changed,this will not affect safetyand effectiveness
Humidity≤85%≤85%≤85%Same
PerformanceNIBP
Pulse Rate40 ---40 ---40 ---Same
Range180times/min180times/min180times/minSame
Pulse RateAccuracyWithin ±5%Within ±5%Within ±5%Same

5- 4

{7}------------------------------------------------

K163276

{8}------------------------------------------------

K163276

KD-721 and 723 Fully Automatic Electronic Blood Pressure Monitor FDA 510(k) Files
-----------------------------------------------------------------------------------
Technique/MethodOscillometricOscillometricOscillometricSame
Measure processMeasure during inflatingMeasure during inflatingMeasure during inflatingSame
Systolic Range60-260mmHg60-260mmHg60-260mmHgSame
Diastolic Range40-199mmHg40-199mmHg40-199mmHgSame
Pressure AccuracyWithin ±3mmHgWithin ±3mmHgWithin ±3mmHgSame
Cuff Pressure Range0-300mmHg0-300mmHg0-300mmHgSame
Overpressure Limit300mmHg300mmHg300mmHgSame
Microprocess orMB9AFB44NцPD78F0485цPD78F0485Although the MCU is different, this change will not affect safety and effectiveness
AlgorithmAmplitudeAmplitudeAmplitudeSame

KD-721 and KD-723 is very similar with its predicate device in the intended use, the design principle, the data transmission mode, the material, the performance and the applicable standards. Only their appearance, the memory capacity, the average function, the power and the MCU are different.

However, the test in this submission provides demonstrates that these small differences do not raise any new questions of safety and effectiveness.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).