K Number
K152379
Date Cleared
2015-09-23

(33 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

iHealth View BP7S Wireless Blood Pressure Wrist Monitor is for use by medical professionals or at home and is a noninvasive blood pressure measurement system intended to measure the diastolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 13.5cm-22cm.

Device Description

iHealth View BP7S Wireless Blood Pressure Monitors are designed and manufactured according to IEC 80601-2-30. The operational principle is based on oscillometric and silicon integrates pressure sensor technology. it can calculate the systolic and diastolic blood pressure, the measurements results can also be classified by the function of blood pressure classification indicator. If any irregular heartbeat is detected, it can be shown to the user. The new devices achieves its function by an iOS or Andriod devices.

AI/ML Overview

The provided document is a 510(k) summary for a medical device called the "iHealth View Wireless Blood Pressure Wrist Monitor (BP7S)". It primarily focuses on demonstrating substantial equivalence to a predicate device, rather than detailing a specific study proving the device meets acceptance criteria. As such, information on specific acceptance criteria and a study to prove them is limited.

However, I can extract the available information related to performance and testing:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present acceptance criteria in a tabular format alongside device performance for a specific study. Instead, it refers to performance standards that the device conforms to. The closest information is under "Accuracy" in the technological characteristics comparison table.

Performance MetricAcceptance Criteria (Standard Reference)Reported Device Performance
Pressure AccuracyIEC 80601-2-30 (implied by "Same as predicate" for accuracy)$\pm$ 3mmHg
Pulse AccuracyIEC 80601-2-30 (implied by "Same as predicate" for accuracy)$\pm$ 5%

2. Sample Size and Data Provenance for Test Set

The document does not specify a sample size for a test set or the data provenance (country of origin, retrospective/prospective). The non-clinical tests mentioned are adherence to electrical safety and electromagnetic compatibility standards, and performance characteristics according to a blood pressure monitor standard. These are general compliance tests for the device, not a specific clinical performance study with a test set of patient data.

3. Number and Qualifications of Experts for Ground Truth

Not applicable. The document does not describe a study involving expert-established ground truth for a test set. The tests mentioned are compliance tests against established engineering and medical device standards.

4. Adjudication Method for Test Set

Not applicable. There is no mention of a test set requiring adjudication in the provided text.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. The document makes no mention of an MRMC comparative effectiveness study, nor does it discuss human reader improvement with or without AI assistance. This device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool for human readers.

6. Standalone Performance Study

Yes, implicitly. The "Non-clinical Tests" and "Performance summary" sections indicate that the device's performance was evaluated against recognized standards (IEC 80601-2-30 for "Safety and performance characteristics"). While not a dedicated "standalone performance study" in the context of AI algorithms, these tests assess the device's functionality and accuracy independently.

7. Type of Ground Truth Used

The ground truth for the device's accuracy claims (pressure $\pm$ 3mmHg, pulse $\pm$ 5%) would typically be based on reference measurements from a validated, calibrated standard or method, as outlined in the IEC 80601-2-30 standard for automated sphygmomanometers. This standard specifies clinical investigation requirements for blood pressure device accuracy. The document states the device "conforms to" this standard, implying that the ground truth methodology prescribed by the standard was followed for these accuracy claims.

8. Sample Size for Training Set

Not applicable. This device is a traditional medical device (blood pressure monitor) governed by engineering and performance standards, not an AI/machine learning algorithm requiring a "training set" in the computational sense.

9. How Ground Truth for Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Public Health Service

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 23, 2015

Andon Health Co., Ltd Ms. Liu Yi President No 3, Jinping Street Ya' An Rd Nankai District Tianjin, 300190 CN

Re: K152379

Trade/Device Name: iHealth View Wireless Blood Pressure Wrist Monitor (BP7S) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: August 19, 2015 Received: August 26, 2015

Dear Ms. Liu Yi,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

Page 2 - Ms. Liu Yi

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

M.A. Hilleman

for

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152379

Device Name

iHealth View Wireless Blood Pressure Wrist Monitor (BP7S)

Indications for Use (Describe)

iHealth View BP7S Wireless Blood Pressure Wrist Monitor is for use by medical professionals or at home and is a noninvasive blood pressure measurement system intended to measure the diastolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 13.5cm-22cm.

Type of Use (Select one or both, as applicable)

For use in identifying a U.S. Vehicle Safety Standard
To correct the information in a U.S. Vehicle Safety Standard

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.

1.0 submitter's information

Name:Andon Health Co., Ltd.
Address:No 3, Jinping Street Ya An Road, Nankai District, TianjinP.R. China
Phone number:86-22-6052 6161
Fax number:86-22-6052 6162
Contact:Liu Yi
Date of Preparation:08/21/2015

2.0 Device information

Trade name: iHealth View Wireless Blood Pressure Wrist Monitor (BP7S)

Device name: Wireless Blood Pressure Monitor

Classification name: Noninvasive blood pressure measurement system

3.0 Classification_

Production code: DXN- Noninvasive blood pressure measurement system. Regulation number: 870.1130 Classification: II Panel: Cardiovascular

4.0 Predicate device information

Manufacturer: Andon Health Co., Ltd. KD-972 Fully Automatic Wireless Blood Pressure Wrist Monitor Device: 510(k) number: K121470

{4}------------------------------------------------

5.0 Intended use

iHealth View BP7S Wireless Blood Pressure Wrist Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 13.5cm-22cm.

The intended use and the indication for use of iHealth BP7S View Wireless Blood Pressure Wrist Monitor, as described in its labeling are the same as the predicate device KD-972(K121470).

6.0 Device description

iHealth View BP7S Wireless Blood Pressure Monitors are designed and manufactured according to IEC 80601-2-30.

The operational principle is based on oscillometric and silicon integrates pressure sensor technology. it can calculate the systolic and diastolic blood pressure, the measurements results can also be classified by the function of blood pressure classification indicator. If any irregular heartbeat is detected, it can be shown to the user. The new devices achieves its function by an iOS or Andriod devices. .

ItemPredicate deviceBP7 (K121470)Subject deviceBP7S(K152379)
Intended use andindication for useFor use by medical professionals or at homeand is a non-invasive blood pressuremeasurement system intended to measurethe diastolic and systolic blood pressures andpulse rate of an adult individual by using anon-invasive technique in which an inflatablecuff is wrapped around the wrist. The cuffcircumference is limited to 13.5cm-22cm.Same as predicate
Method ofmeasurementOscillometricSame as predicate
Unit WeightAbout 106gAbout 114g
Average functionYesNo

7.0 Summary comparing technological characteristics with predicate device

{5}------------------------------------------------

Memory functionYesNo
Power Supply1*3.7V Li-ion 400mAhSame as predicate
AccuracyPressure $\pm$ 3mmHgPulse $\pm$ 5%Same as predicate
Range ofmeasurementCuff pressure Range0 ~ 300 mmHgDetermination Range40 ~ 260 mmHgCuff pressure Range0 ~ 300 mmHgDetermination Range40 ~ 260 mmHg
StorageEnvironment-20~55°C<95%RH-20~55°C≤85%RH
OperatingEnvironment5~40°C<90%RH10℃~40℃≤85%RH
Data transmissionBluetoothSame as predicate
MeasurementdisplayiOS device1. iOS device orAndroid device,2. LED display onblood pressuremeter,
Software platformiOSiOS and Android
PerformanceStandardISO 80601-2-30Same as predicate

8.0 Discussion of non-clinical and clinical test performed

Non-clinical Tests have been done as follows:

  • a. Electromagnetic compatibility test according to IEC 60601-1-2;
  • b. Electrical safety according test to IEC 60601-1;

c. Safety and performance characteristics of the test according to IEC 80601-2-30

None of the test demonstrates that BP7S Wireless Blood Pressure Wrist Monitor bring new questions of safety and effectiveness.

9.0 Performance summary

iHealth View BP7S Wireless Blood Pressure Wrist Monitor conforms to the following standards:

  • AAMI ANSI ES 60601-1:2005/(R) 2012 And C1:2009(R)2012 And A2:2010 (Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests)
  • · IEC60601-1-2 Edition 3: 2007-03, (Medical electrical equipment -- Part 1-2:

{6}------------------------------------------------

General requirements for basic safety and essential performance -Collateral standard: Electromagnetic compatibility - Requirements and tests)

  • IEC 80601-2-30:2009+Cor.2010/EN 80601-2-30:2010(Medical electrical equipment -Part 2-30: Particular requirements for the basic safety and essential performance of of of automated sphygmomanometers)AAMI IEC80601-2-30:2002, Manual, electronic or automated sphygmomanometers.

10.0 Comparison to the predicate device and the conclusion

iHealth View BP7S Wireless Blood Pressure Wrist Monitor is substantially equivalent to KD-972 blood pressure monitor whose 510(K) number is K121470

The new device BP7S is very similar with its predicate device in the intended use, the design principle, the material, the performance and the applicable standards. Only their data transmission function, the MCU, the memory function and the appearance are different, and a new software platform has been added.

However, the test in this submission provides demonstrates that these small differences do not raise any new questions of safety and effectiveness.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).