K Number
K162982

Validate with FDA (Live)

Date Cleared
2017-02-27

(124 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Description

The Medline Nitrile Powder-Free Examination Glove with Colloidal Oatmeal USP (Tested for Use with Chemotherapy Drugs) are non-sterile single use only, disposable gloves intended for medical purposes to be worn on the hands of examiners to prevent contamination between the patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, white colored nitrile gloves featuring an inner coating of colloidal oatmeal USP. The gloves are offered in five sizes, extra-small, small, medium, large and extra-large.

The gloves are designed and manufactured per ASTM D6319-10 standard and tested for use with chemotherapy drugs per ASTM D6978-05 (reapproved 2013).

AI/ML Overview

The provided text describes the regulatory submission for the "Medline Nitrile Powder-Free Examination Glove with Colloidal Oatmeal USP (Tested for Use with Chemotherapy Drugs)" and the testing conducted to support its claims, specifically its resistance to chemotherapy drugs.

Here's the breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance:

The primary acceptance criteria for the chemotherapy drug resistance are the breakthrough times for various chemotherapy drugs, as tested per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. The reported device performance is the minimum breakthrough time for each drug.

Chemotherapy Drug TestedConcentrationAcceptance Criteria (Implied by standard and results)Reported Device Performance (Minimum Breakthrough Time)
Blenoxane15 mg/mlMinimum acceptable time depends on risk assessment>240 minutes
Bortezomib1 mg/ml>240 minutes
Busulfan6 mg/ml>240 minutes
Carmustine (BCNU)3.3 mg/ml11.9 minutes
Cetuximab (Erbitux)2 mg/ml>240 minutes
Cisplatin1.0 mg/ml>240 minutes
Cyclophosphamide (Cytoxan)20.0 mg/ml>240 minutes
Cytarabine100 mg/ml>240 minutes
Dacarbazine (DTIC)10.0 mg/ml>240 minutes
Daunorubicin5 mg/ml>240 minutes
Docetaxel10.0 mg/ml>240 minutes
Doxorubicin Hydrochloride2.0 mg/ml>240 minutes
Ellence2 mg/ml>240 minutes
Etoposide (Toposar)20.0 mg/ml>240 minutes
Fludarabine25 mg/ml>240 minutes
Fluorouracil50.0 mg/ml>240 minutes
Gemcitabine (Gemzar)38 mg/ml>240 minutes
Idarubicin1 mg/ml>240 minutes
Ifosfamide50.0 mg/ml>240 minutes
Irinotecan20.0 mg/ml>240 minutes
Mechlorethamine HCI1.0 mg/ml>240 minutes
Melphalan5 mg/ml>240 minutes
Methotrexate25 mg/ml>240 minutes
Methotrexate50 mg/ml>240 minutes
Mitomycin C0.5 mg/ml>240 minutes
Mitoxantrone2.0 mg/ml>240 minutes
Oxaliplatin5 mg/ml>240 minutes
Paclitaxel (Taxol)6.0 mg/ml>240 minutes
Paraplatin10 mg/ml>240 minutes
Pemetrexed Disodium25 mg/ml>240 minutes
Raltitrexed0.5 mg/ml>240 minutes
Rituximab10 mg/ml>240 minutes
Thiotepa10.0 mg/ml18.6 minutes
Trisonex0.1 mg/ml>240 minutes
Vidaza (5-Azacytidine)25 mg/ml>240 minutes
Vinblastine1 mg/ml>240 minutes
Vinorelbine10 mg/ml>240 minutes
Vincristine Sulfate1.0 mg/ml>240 minutes

Other acceptance criteria are derived from ASTM D6319-10 Standard Specification for Nitrile Examination Gloves for Medical Application, including:

CharacteristicAcceptance Criteria (ASTM D6319-10 minimum)Reported Device Performance
Length230mm min.Complies (230mm min.)
WidthX-small – 70±10mm, Small – 80±10mm, Medium – 95±10mm, Large – 110±10mm, X-large – 120±10mmComplies
ThicknessPalm – 0.05mm min., Finger – 0.05mm min.Complies
Tensile StrengthBefore Aging ≥14 MPa, min.; After Aging ≥14 MPa, min.Complies
ElongationBefore Aging 500%, min.; After Aging 400%, min.Complies
Freedom from HolesG-1, AQL 1.5 (per ASTM D6319-10 and ASTM D5151-06)Complies
Residual Powder<2mg per glove (per ASTM D6319-10)Complies
BiocompatibilityNot a skin irritant, Not a skin sensitizer (per AAMI/ANSI/ISO 10993-10)Complies

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: The document does not explicitly state the sample size used for each individual chemotherapy drug permeation test. However, the testing was conducted per ASTM D6978-05 (Reapproved 2013). This standard outlines procedures for testing, and typically specifies minimum sample sizes or replicates required to ensure statistically significant results. Without access to the specific test report, the exact number of gloves tested for each drug cannot be determined from this summary.
  • Data Provenance: The testing was conducted in accordance with recognized ASTM (American Society for Testing and Materials) standards. The document does not specify the country of origin of the data or whether the tests were retrospective or prospective, but as it's a 510(k) submission, it would be prospective testing performed specifically for this submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not applicable to this type of device and testing. The "ground truth" for the device's performance regarding chemotherapy drug permeation is established through a standardized, objective laboratory test (ASTM D6978-05). It does not involve human expert interpretation of data in the way medical imaging or diagnostic algorithms would. The ASTM standard itself defines the methodology and interpretation of results.

4. Adjudication Method for the Test Set:

This is not applicable. The testing involves a direct measurement of breakthrough time using analytical methods, not subjective assessment requiring adjudication among multiple human observers.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable. This device is a physical examination glove, not an AI-powered diagnostic system. Therefore, MRMC studies are not relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This is not applicable. The device is a physical glove, not an algorithm or AI system. The performance testing described is standalone for the glove itself, meaning its physical properties are tested without human interaction as part of the measurement process (e.g., a person wearing the glove and interacting with chemotherapy drugs). The test determines the intrinsic resistance of the material.

7. The Type of Ground Truth Used:

The ground truth used is based on standardized laboratory measurements and physical property tests. Specifically:

  • Permeation by Chemotherapy Drugs: Determined by the "breakthrough time" measured according to ASTM D6978-05 (Reapproved 2013). This is an objective, quantitative measurement using analytical instrumentation.
  • Physical Properties (e.g., tensile strength, elongation, freedom from holes, dimensions, residual powder): Determined by standardized test methods outlined in ASTM D6319-10 and ASTM D5151-06. These are also objective, quantitative measurements.
  • Biocompatibility: Determined by standardized tests for primary skin irritation and dermal sensitization according to ISO 10993-10. These are biological tests that produce objective results, often interpreted by qualified toxicologists/biocompatibility specialists.

8. The Sample Size for the Training Set:

This is not applicable. The device is a physical product (a glove), not a machine learning model. Therefore, there is no "training set." The performance is inherent to the manufactured product's material and design, tested against established standards.

9. How the Ground Truth for the Training Set Was Established:

This is not applicable as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 27, 2017

Medline Industries, Inc. Jennifer Mason Senior Regulatory Affairs Specialist One Medline Place Mundelein, Illinois 60060

Re: K162982

Trade/Device Name: Medline Nitrile Powder-free Examination Glove With Colloidal Oatmeal USP (tested For Use With Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: January 31, 2017 Received: January 31, 2017

Dear Jennifer Mason:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Susan Runne DDS, MA

For Tina Kiang, Ph.D. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162982

Device Name

Medline Nitrile Powder-Free Examination Glove with Colloidal Oatmeal USP (Tested for Use with Chemotherapy Drugs)

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Blenoxane (Bleomycin) 15 mg/ml >240 minutes Bortezomib 1 mg/ml >240 minutes Busulfan 6 mg/ml >240 minutes Carboplatin 10.0 mg/ml >240 minutes Carmustine (BCNU) 3.3 mg/ml 11.9 minutes Cetuximab (Erbitux) 2 mg/ml >240 minutes Cisplatin 1.0 mg.ml >240 minutes Cyclophosphamide (Cytoxan) 20.0 mg/ml >240 minutes Cytarabine 100 mg/ml >240 minutes Dacarbazine (DTIC)10.0 mg/ml >240 minutes Daunorubicin 5 mg/ml >240 minutes Docetaxel 10.0 mg/ml >240 minutes Doxorubicin Hydrochloride 2.0 mg/ml >240 minutes Ellence 2 mg/ml >240 minutes Etoposide (Toposar) 20.0 mg/ml >240 minutes Fludarabine 25 mg/ml >240 minutes Fluorouracil 50.0 mg/ml >240 minutes Gemcitabine (Gemzar) 38 mg/ml >240 minutes Idarubicin 1 mg/ml >240 minutes Ifosfamide 50.0 mg/ml >240 minutes Irinotecan 20.0 mg/ml >240 minutes Mechlorethamine HCI 1.0 mg/ml >240 minutes Melphalan 5 mg/ml >240 minutes Methotrexate 25 mg/ml >240 minutes Methotrexate 50 mg/ml >240 minutes Mitomycin C 0.5 mg/ml >240 minutes Mitoxantrone 2.0 mg/ml >240 minutes Oxaliplatin 5 mg/ml >240 minutes Paclitaxel (Taxol) 6.0 mg/ml >240 minutes Paraplatin 10 mg/ml >240 minutes Pemetrexed Disodium 25 mg/ml >240 minutes Raltitrexed 0.5 mg/ml >240 minutes Rituximab 10 mg/ml >240 minutes Thiotepa 10.0 mg/ml 18.6 minutes Trisonex 0.1 mg/ml >240 minutes Vidaza (5-Azacytidine) 25 mg/ml >240 minutes Vinblastine 1 mg/ml >240 minutes

{3}------------------------------------------------

Vinorelbine 10 mg/ml >240 minutes Vincristine Sulfate 1.0 mg/ml >240 minutes

Please note that the following drugs have extremely low permeation times: Carmustine (BCNU) (3.3 mg/ml) 11.9 minutes Thiotepa (10.0 mg/ml) 18.6 minutes

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image features the Medline logo, set against a blue background. The logo consists of the word "MEDLINE" in a bold, sans-serif font, positioned to the left of a white, stylized starburst symbol. The starburst has four points, with the top and bottom points being longer than the side points. A gray bar is visible on the right side of the image.

K162982 510(k) SUMMARY [AS REQUIRED BY 21CFR807.92(c)]

Submitter / 510(k) Sponsor

Medline Industries, Inc. 1 Medline Place Mundelein, IL 60060

Registration Number: 1417592

Contact Person

Jennifer Mason Senior Regulatory Affairs Specialist Phone: 847-643-3652 Email: jamason@medline.com

Summary Preparation Date February 27, 2017

Type of 510(k) Submission Traditional

Device Name / Classification

Name of Device: Patient Examination Glove (Tested for Use with Chemotherapy Drugs) Proprietary Name: Medline Nitrile Powder-Free Examination Glove with Colloidal Oatmeal USP (Tested for Use with Chemotherapy Drugs) Common Name: Patient Examination Glove Classification Name: Patient Examination Glove Product Code: LZA, LZC Classification Panel: General Hospital Regulation #: 21 CFR 880.6250

Predicate Device

Nitrile Powder Free Examination Glove with Colloidal Oatmeal USP Tested for Use with Chemotherapy Drugs K160562

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image features the logo for Medline, a medical supply company. The logo consists of the word "MEDLINE" in bold, white sans-serif font, positioned to the left of a stylized white cross. The cross is oriented diagonally, with its top point extending upwards and to the right. The background is a solid dark blue color, providing a strong contrast to the white text and cross. A small gray rectangle is visible in the upper right corner of the image.

e Medline Place

Reference Device

Powder Free Nitrile Patient Examination Glove, White Colored, Non-Sterile. Tested for Use with Chemotherapy Drugs K151750

Device Description

The Medline Nitrile Powder-Free Examination Glove with Colloidal Oatmeal USP (Tested for Use with Chemotherapy Drugs) are non-sterile single use only, disposable gloves intended for medical purposes to be worn on the hands of examiners to prevent contamination between the patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, white colored nitrile gloves featuring an inner coating of colloidal oatmeal USP. The gloves are offered in five sizes, extra-small, small, medium, large and extra-large.

The gloves are designed and manufactured per ASTM D6319-10 standard and tested for use with chemotherapy drugs per ASTM D6978-05 (reapproved 2013).

Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy Drug TestedBreakthrough Time(in minutes)
Blenoxane 15 mg/ml>240
Bortezomib 1 mg/ml>240
Busulfan 6 mg/ml>240
Carmustine (BCNU) 3.3 mg/ml11.9
Cetuximab (Eributux) 2 mg/ml>240
Cisplatin 1.0 mg.ml>240
Cyclophosphamide (Cytoxan) 20.0 mg/ml>240
Cytarabine 100 mg/ml>240
Dacarbazine (DTIC)10.0 mg/ml>240
Daunorubicin 5 mg/ml>240
Docetaxel 10.0 mg/ml>240
Doxorubicin Hydrochloride 2.0 mg/ml>240
Ellence 2 mg/ml>240
Etoposide (Toposar) 20.0 mg/ml>240
Fludarabine 25 mg/ml>240

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in white, set against a blue background. To the right of the word is a white star-like symbol with four points. The background is a solid blue color, providing a stark contrast to the white text and symbol.

Fluorouracil 50.0 mg/ml>240
Gemcitabine (Gemzar) 38 mg/ml>240
Idarubicin 1 mg/ml>240
Ifosfamide 50.0 mg/ml>240
Irinotecan 20.0 mg/ml>240
Mechlorethamine HCI 1.0 mg/ml>240
Melphalan 5 mg/ml>240
Methotrexate 25 mg/ml>240
Methotrexate 50 mg/ml>240
Mitomycin C 0.5 mg/ml>240
Mitoxantrone 2.0 mg/ml>240
Oxaliplatin 5 mg/ml>240
Paclitaxel (Taxol) 6.0 mg/ml>240
Paraplatin 10 mg/ml>240
Pemetrexed Disodium 25 mg/ml>240
Raltitrexed 0.5 mg/ml>240
Rituximab 10 mg/ml>240
Thiotepa 10.0 mg/ml18.6
Trisonex 0.1 mg/ml>240
Vidaza (5-Azacytidine) 25 mg/ml>240
Vinblastine 1 mg/ml>240
Vinorelbine 10 mg/ml>240
Vincristine Sulfate 1.0 mg/ml>240

Please note that the following drugs have extremely low permeation times:

Carmustine (BCNU) (3.3 mg/ml) 11.9 minutes

Thiotepa (10.0 mg/ml) 18.6 minutes

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Medline. The logo consists of the word "MEDLINE" in a sans-serif font, with the letters stacked on top of each other. To the right of the word is a white starburst symbol. The background of the logo is a solid blue color.

Summary of Technological Characteristics

TABLE 1: Comparison of Proposed and Predicate Devices

DeviceCharacteristicProposedDevicePredicateDeviceReferenceDeviceComparisonAnalysis
Product NameMedline NitrilePowder-FreeExamination Glovewith ColloidalOatmeal USP (Testedfor Use withChemotherapy Drugs)Nitrile Powder FreeExamination Glovewith ColloidalOatmeal USP Testedfor Use withChemotherapy DrugsPowder Free NitrilePatient ExaminationGlove, WhiteColored, Non Sterile.Tested for Use withChemotherapy DrugsN/A
510(k) ReferenceK162982K160562K151750N/A
Product OwnerMedline Industries,Inc.HartalegaKossanDifferent
Product CodeLZA, LZCLZA., LZCLZA, LZCSame
Intended UseA patient examinationglove is a disposabledevice intended formedical purposes thatis worn on theexaminer's hand toprevent contaminationbetween patient andexaminer.These gloves weretested for use withchemotherapy drugsas per ASTM D6978-05 (Reapproved 2013)Standard Practice forAssessment ofMedical Gloves toPermeation byChemotherapy DrugsA non-steriledisposable deviceintended for medicalpurpose that is wornon the examiner'shand to preventcontaminationbetween patient andexaminer. Thesegloves were testedfor use withChemotherapy Drugsin accordance withASTM D6978-05,Standard Practice forAssessment ofResistance ofMedical Gloves toPermeation byChemotherapy DrugsA patientexamination glove isa disposable deviceintended for medicalpurposes that is wornon the examiner'shand to preventcontaminationbetween patient andexaminer.These gloves weretested for use withchemotherapy drugsas per ASTMD6978-05(Reapproved 2013)Standard Practice forAssessment ofMedical Gloves toPermeation byChemotherapy DrugsSame
RegulationNumber21 CFR 880.625021 CFR 880.625021 CFR 880.6250Same
MaterialsNitrileColloidal oatmealNitrileColloidal oatmealNitrileSame as K160562
ColorWhiteWhiteWhiteSame
Sizesx-small, small, medium, large, x-largex-small, small, medium, large, x-largex- small, small, medium, large, x-large, xx-largeSame
Dimensions – LengthComplies with ASTM D6319-10230mm min.Complies with ASTM D6319-10240mm min.Complies with ASTM D6319-10230mm min.Same as K151750 and similar to K160562 which is 10mm longer than the subject device.
Dimensions – WidthComplies with ASTM D6319-10X-small – 70±10mmSmall – 80±10mmMedium – 95±10mmLarge – 110±10mmX-large – 120±10mmComplies with ASTM D6319-10Medium – 95mmComplies with ASTM D6319-10X-small – 70-80mmSmall – 80-90mmMedium – 90-100mmLarge – 101-11mmX-large – 111-121mmXX-large – 121-131mmSame as K160562 and similar to K151750.
Dimensions – ThicknessComplies with ASTM D6319-10Palm – 0.05mm min.Finger – 0.05mm min.Complies with ASTM D6319-10Palm – 0.07mm min.Finger – 0.07mm min.Complies with ASTM D6319-10Palm – 0.05mm min.Finger – 0.05mm min.Same as K151750 and similar to K160562
Physical PropertiesComplies with ASTM D6319-10 minimum:Complies with ASTM D6319-10 minimum:Complies with ASTM D6319-10 minimum:Same
Tensile Strength:Before Aging ≥14 MPa, min.After Aging ≥14 MPa, min.Elongation:Before Aging 500%, min.After Aging 400%, min.Tensile Strength:Before Aging ≥14 MPa, min.After Aging ≥14 MPa, min.Elongation:Before Aging 500%, min.After Aging 400%, min.Tensile Strength:Before Aging ≥14 MPa, min.After Aging ≥14 MPa, min.Elongation:Before Aging 500%, min.After Aging 400%, min.
Freedom from HolesComplies with ASTM D6319-10 and ASTM D5151-06(reapproved 2011) G-1, AQL 1.5Complies with ASTM D6319-10 and ASTM D5151-06 (reapproved 2011) AQL 1.5Complies with ASTM D6319-10 and ASTM D5151-06 (reapproved 2011) G-1, AQL 1.5Same
Powder or Powder-FreePowder-FreePowder-FreePowder-FreeSame
Residual PowderComplies with ASTM D6319-10 <2mg per gloveComplies with ASTM D6319-10 <2mg per gloveComplies with ASTM D6319-10 <2mg per gloveSame
ContactDurationsLimited ≥24 hoursLimited ≥24 hoursLimited ≥24 hoursSame
BiocompatibilityAAMI/ANSI/ISO10993-10:Not a skin irritantNot a skin sensitizerAAMI/ANSI/ISO10993-10:Not a skin irritantNot a skin sensitizerAAMI/ANSI/ISO10993-10:Not a skin irritantNot a skin sensitizerSame
Sterile vs. Non-SterileNon-sterileNon-sterileNon-sterileSame
ChemotherapyDrugs Tested withMinimumBreakthroughDetection Time astested per ASTMD6978Blenoxane 15 mg/ml>240 min.Different
Bortezomib 1 mg/ml>240 min.Different
Busulfan 6 mg/ml>240 min.Different
Carmustine (BCNU)3.3 mg/ml11.9 min.Carmustine3.3 mg/ml10.1 min.Carmustine (BCNU)3.3 mg/ml10.1 min.Same
Cetuximab (Erbitux)2 mg/ml>240 min.Different
Cisplatin 1.0 mg.ml>240 min.Cisplatin 1.0 mg.ml>240 min.Cisplatin 1.0 mg.ml>240 min.Same
Cyclophosphamide(Cytoxan) 20.0 mg/ml>240 min.Cyclophosphamide(Cytoxan) 20 mg/ml>240 min.Cyclophosphamide(Cytoxan) 20 mg/ml>240 min.Same
Cytarabine 100 mg/ml>240 min.Cytarabine 100mg/ml>240 min.Different
Dacarbazine(DTIC)10.0 mg/ml>240 min.Dacarbazine(DTIC)10.0 mg/ml>240 min.Dacarbazine(DTIC)10.0 mg/ml>240 min.Same
Daunorubicin 5mg/ml>240 min.Different
Docetaxel 10.0 mg/ml>240 min.Different
DoxorubicinHydrochloride 2.0mg/ml>240 min.DoxorubicinHydrochloride 2.0mg/ml>240 min.DoxorubicinHydrochloride 2.0mg/ml>240 min.Same
Ellence 2 mg/ml>240 min.Different
Etoposide (Toposar)20.0 mg/ml>240 min.Etoposide (Toposar)20.0 mg/ml>240 min.Etoposide (Toposar)20.0 mg/ml>240 min.Same
Fludarabine 25 mg/ml>240 min.Different
Fluorouracil 50.0mg/ml>240 min.Fluorouracil 50.0mg/ml>240 min.Fluorouracil 50.0mg/ml>240 min.Same
Gemcitabine(Gemzar) 38 mg/ml>240 min.Different
Idarubicin 1 mg/ml>240 min.Different
Ifosfamide 50.0mg/ml>240 min.Ifosfamide (IFEX)50 mg/ml>240 min.Different
Irinotecan 20.0 mg/ml>240 min.Different
MechlorethamineHCI 1.0 mg/ml>240 min.Different
Melphalan 5 mg/ml>240 min.Different
Methotrexate 25mg/ml>240 min.Methotrexate 25mg/ml>240 minMethotrexate 25mg/ml>240 min.Same
Methotrexate 50mg/ml>240 min.Different
Mitomycin C 0.5mg/ml>240 min.Mitomycin C 0.5mg/ml>240 min.Mitomycin C 0.5mg/ml>240 min.Same
Mitoxantrone 2.0mg/ml>240 min.Mitoxantrone 2.0mg/ml>240 min.Different
Oxaliplatin 5 mg/ml>240 min.Different
Paclitaxel (Taxol) 6.0mg/ml>240 min.Paclitaxel (Taxol)6.0 mg/ml>240 min.Paclitaxel (Taxol)6.0 mg/ml>240 min.Same
Paraplatin 10 mg/ml>240 min.Different
Pemetrexed Disodium25 mg/ml>240 min.Different
Raltitrexed 0.5 mg/ml>240 min.Different
Rituximab 10 mg/ml>240 min.Different
Thiotepa 10.0 mg/ml18.6 min.Thiotepa 10.0 mg/ml30.4 min.Thiotepa 10.0 mg/ml10.4 min.Same
Trisonex 0.1 mg/ml>240 minDifferent
Vidaza (5-Azacytidine) 25mg/ml>240 min.Different
Vinblastine 1 mg/ml>240 min.Different
Vinorelbine 10 mg/ml>240 min.Different
Vincristine Sulfate1.0 mg/ml>240 min.Vincristine Sulfate1.0 mg/ml>240 min.Vincristine Sulfate(Oncovin) 1.0 mg/ml>240 min.Same

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image features the Medline logo, which is set against a blue background. The logo consists of the word "MEDLINE" in a sans-serif font, positioned to the left of a white, stylized starburst or cross-like symbol. The symbol is composed of four pointed shapes converging at a central point, creating a visual effect of radiating light or energy. The overall design is clean and professional, reflecting the company's presence in the medical or healthcare industry.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image features the Medline logo, which is set against a blue square background. The logo consists of the word "MEDLINE" in a sans-serif font, positioned to the left of a stylized white starburst symbol. The starburst is composed of four pointed shapes converging at a central point, creating a dynamic and modern design.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in white, bold, sans-serif font, placed horizontally. A white, stylized cross is behind the text. The background is a solid dark blue.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image features the Medline logo, which is set against a blue square background. The word "MEDLINE" is prominently displayed in white, with a stylized white cross extending vertically through the text. The cross's upper and lower points extend beyond the word, adding a distinctive visual element to the logo. A small gray rectangle is visible in the upper right corner of the image.

Summary of Non-Clinical Testing

The biocompatibility evaluation for the Medline Nitrile Powder-Free Examination Glove with Colloidal Oatmeal USP (Tested for Use with Chemotherapy Drugs) was conducted in accordance with ANSI/AAMI/ISO 10993-1:2009 Biological Evaluation of Medical Devices—Part 1: Evaluation and Testing Within a Risk Management Process, as recognized by FDA.

The following tests were performed to evaluate the biocompatibility of the Medline Nitrile Powder-Free Examination Glove with Colloidal Oatmeal USP (Tested for Use with Chemotherapy Drugs):

  • . ISO 10993-10: Primary Skin Irritation
  • . ISO 10993-10: Dermal Sensitization
  • ASTM D 6319-10 Standard Specification for Nitrile Examination Gloves for Medical Application ●
  • ASTM D 6124-06 (Reaffirmation 2011) Standard Test Method for Residual Powder on Medical ● Gloves
  • ASTM D 5151-06 (Reapproved 2011) Standard Test Method for Detection of Holes in Medical ● Gloves
  • ASTM D 6978-05 (Reapproved 2013) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs

Performance Testing (Bench)

Physical performance qualities were evaluated per ASTM D6319- 10, Standard Specification for Nitrile Examination Gloves for Medical Application.

Permeation testing was conducted to support the addition of the labeling claim: Tested for use with chemotherapy drugs. The gloves were tested according to ASTM D6978- 05 (Reapproved 2013), Standard

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the Medline logo. The logo is set against a blue square background. The word "MEDLINE" is in white, sans-serif font, stacked on top of a white, stylized star shape. The star has four points, with the top and bottom points being longer than the side points.

Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. Minimum breakthrough times were determined for a wide range of chemotherapy drugs.

Summary of Clinical Testing

Not applicable.

Conclusion

In accordance with 21 CFR Part 807, and based on the information provided in this premarket notification, Medline Industries, Inc. concludes that the Medline Nitrile Powder-Free Examination Glove with Colloidal Oatmeal USP (Tested for Use with Chemotherapy Drugs) is as safe and as effective for their intended use as the predicate device, the Nitrile Powder Free Examination Glove with Colloidal Oatmeal USP Tested for Use with Chemotherapy Drugs, K160562.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.