K Number
K162971
Date Cleared
2016-11-22

(28 days)

Product Code
Regulation Number
892.1680
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Multix Fusion Max system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion Max enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatic, adult and bariatic patients. Exposures may be taken with the patient sitting, standing, or in the prone position. The Multix Fusion Max system is not meant for mammography.

Multix Fusion Max uses a mobile(wired), or a fixed (integrated) or wireless digital detector for generating diagnostic images by converting x-rays into image signals. The Multix Fusion Max is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.

Device Description

The Multix Fusion Max Radiography X-ray system is a modular system of x-ray components (ceiling suspension with X-ray tube, Bucky wall stand, Bucky table, X-ray generator, portable wireless and integrated detectors) same as the predicate the Multix Fusion. This 510(k) submission describes modifications to the predicate device the Multix Fusion cleared via K142049.

AI/ML Overview

The provided text is a 510(k) summary for the Siemens Multix Fusion Max X-ray system. This document is a premarket notification to the FDA to demonstrate that the new device is substantially equivalent to a legally marketed predicate device. As such, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" refer to demonstrating substantial equivalence, not typically a clinical performance study with specific metrics like sensitivity or specificity for an AI-powered diagnostic device.

This document describes the device as a radiographic system with various hardware and software modifications compared to its predicate. The "acceptance criteria" here are primarily about demonstrating that the new modifications do not introduce new safety concerns and that the device performs functionally similarly to its predicates.

Here's an analysis of the provided information, framed within your request, but acknowledging the nature of a 510(k) for a general radiography system rather than an AI diagnostic algorithm:

1. Table of Acceptance Criteria (as implied by a 510(k) for an X-ray system) and Reported Device Performance

For an X-ray system like the Multix Fusion Max, "acceptance criteria" are defined by its substantial equivalence to predicate devices, focusing on technical specifications, safety, and operational performance. The "performance" is the demonstration that these are met or that differences are not significant and do not raise new safety/effectiveness questions.

Acceptance Criteria (Implied by 510(k) Process for X-ray System)Reported Device Performance
Similar Indications for Use"Same" as predicate (Multix Fusion K142049) and within same classification regulation.
Similar Technological Characteristics (Hardware)Many components are "same" (e.g., tube, generator, ceiling-mounted support, imaging system). New components (detectors, Bucky wall stand, table, grid) have comparable specifications (e.g., DQE, MTF values are "Difference not significant" or "Same"). Wireless Bucky wall stand is "Only motorized height adjustment" (compared to predicate's motor/manual option) which is a functional difference, but not affecting safety or fundamental performance.
Similar Technological Characteristics (Software)Operating system upgraded from Windows XP to Windows 7; software version VE21. Compliance with software guidance documents is stated. Tested for "continued conformance with special controls for medical devices containing software."
Safety and Effectiveness Not CompromisedNon-clinical tests (verification and validation) were conducted. Conformance to various IEC and ISO standards (e.g., IEC 60601-1:2012, ISO 14971:2007 for risk management). Risk analysis completed and risk control implemented. Software testing demonstrated all specifications met acceptance criteria. Visual and audible warnings are incorporated, and error messages are displayed if an error occurs.
Image Quality (for new detectors)Comparative tables show "Difference not significant" or "Same" for key image quality metrics (DQE, MTF, pixel size, resolution) between new and predicate detectors. Functional tests included exposure workflow, detector sharing, and image resend. Image quality tests included flat field uniformity.
No New Safety Risks Introduced"Siemens is of the opinion that the Multix Fusion Max does not introduce any new potential safety risk and is substantially equivalent to and performs as well as the predicate device."

2. Sample Size Used for the Test Set and the Data Provenance

This is an X-ray system submission, not a study of an AI-powered diagnostic device. Therefore, there isn't a "test set" in the sense of a dataset of patient images for an AI algorithm. The testing described is primarily non-clinical, focusing on system function, image quality a phantom/test environment, and safety standards compliance.

  • Sample Size: Not applicable in the context of clinical "test set" for an AI algorithm. Non-clinical tests were conducted at the product development stage, but specific "sample sizes" of components or tests are not detailed.
  • Data Provenance: Not applicable. The "data" are internal testing results, not patient data from a specific country or collected retrospectively/prospectively.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable. There is no mention of expert readers or ground truth establishment in this type of 510(k) submission for a general X-ray system. The performance demonstration is based on technical specifications and engineering tests.

4. Adjudication Method for the Test Set

Not applicable.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

Not applicable. This is not an AI-powered diagnostic device, and no MRMC study is mentioned or required for this type of submission.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Not applicable. This is an X-ray imaging system, not an AI algorithm.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

Not applicable in the context of clinical ground truth. The "ground truth" for this device's performance demonstration relies on engineering specifications, adherence to international standards (e.g., IEC, ISO), and performance metrics like DQE and MTF measured in a controlled environment, not clinical outcomes or expert labels.

8. The Sample Size for the Training Set

Not applicable. This is an X-ray system, not an AI algorithm requiring a "training set."

9. How the Ground Truth for the Training Set was Established

Not applicable.

In summary, the provided document
describes a 510(k) premarket notification for a conventional X-ray system, not an AI-powered diagnostic device. Therefore, the questions related to AI algorithm development, clinical datasets, expert ground truth, and reader studies are not addressed and are not applicable to this type of device submission.
The acceptance criteria are based on demonstrating substantial equivalence to predicate devices through non-clinical performance testing and compliance with recognized standards.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: a circular text element and a symbol. The text element reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The symbol is a stylized representation of three human profiles facing to the right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 22, 2016

Siemens Medical Solutions USA, Inc. % Ms. Denise Adams Regulatory Affairs Specialist 40 Liberty Boulevard, 65-1A MALVERN PA 19355

Re: K162971

Trade/Device Name: Multix Fusion Max Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR Dated: October 21, 2016 Received: October 25, 2016

Dear Ms. Adams:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert Oolo

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K162971

Device Name Multix Fusion Max

Indications for Use (Describe)

Multix Fusion Max system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion Max enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatic, adult and bariatic patients. Exposures may be taken with the patient sitting, standing, or in the prone position. The Multix Fusion Max system is not meant for mammography.

Multix Fusion Max uses a mobile(wired), or a fixed (integrated) or wireless digital detector for generating diagnostic images by converting x-rays into image signals. The Multix Fusion Max is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The letters are a teal color. The background is white.

Healthcare

510(k) Summary: Multix Fusion Max

Company: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355

Date Prepared: November 21, 2016

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.

1. General Information

Importer / Distributor

Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Establishment Registration Number 2240869

Location of Manufacturing Site

Siemens Shanghai Medical Equipment Ltd. 278 Zhou Zhu Road, Shanghai 201318, China Establishment Registration Number 3003202425

Siemens Healthcare GmbH, Business Unit XP Siemensstrasse 1 Forchheim, Germany 91301 Establishment Registration Number: 3004977335

2. Contact Person

Denise Adams, RAC Regulatory Affairs Specialist Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 610-448-6139 adams.denise@siemens.com

40 Liberty Boulevard Malvern, PA 19355-9998 USA

{4}------------------------------------------------

3. Device Name and Classification

Trade Name:Multix Fusion Max
Classification Name:Stationary X-Ray System
Classification Panel:Radiology
Classification Regulation:21 CFR §892.1680
Device Class:Class II
Product Code:KPR

4. Legally Marketed Predicate Devices

Primary Predicate
Trade Name:Multix Fusion
510(k) #:K142049
Clearance Date:December 18, 2014
Classification Name:Stationary X-Ray System
Classification Panel:Radiology
CFR Section:21 CFR §892.1680
Device Class:Class II
Product Code:KPR

Reference Predicate

|

Trade Name:YSIO Max
510(k) #:K133259
Clearance Date:January 24, 2014
Classification Name:Stationary X-Ray System
Classification Panel:Radiology
CFR Section:21 CFR §892.1680
Device Class:Class II
Product Code:KPR

5. Device Description

The Multix Fusion Max Radiography X-ray system is a modular system of x-ray components (ceiling suspension with X-ray tube, Bucky wall stand, Bucky table, X-ray generator, portable wireless and integrated detectors) same as the predicate the Multix Fusion. This 510(k) submission describes modifications to the predicate device the Multix Fusion cleared via K142049.

The following modifications have been made to the cleared predicate device and the system will be rebranded the Multix Fusion Max:

{5}------------------------------------------------

Hardware

The following components are being integrated into the system: the Pixium 3543 EZh wireless detector, the Pixium 2430 EZ wireless detector; the Pixium 4343RCE fixed detector, a new wireless Bucky wall stand with motorized height adjustment, a new table that has a charging tray and a manual Bucky tray, and a new 92-line grid with a 13:1 ratio.

Software

The operating system is being upgraded from Windows XP to Windows 7 and the software version will now be VE21.

6. Indication for Use

Multix Fusion Max system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion Max enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. Exposures may be taken with the patient sitting, standing, or in the prone position. The Multix Fusion Max system is not meant for mammography.

Multix Fusion Max uses a mobile (wired), or a fixed (integrated) or wireless digital detector for generating diagnostic images by converting x-rays into image signals. The Multix Fusion Max is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.

7. Substantial Equivalence

The Multix Fusion Max is a modification to the primary predicate device the Multix Fusion cleared via K142049. The Multix Fusion Max utilizes the same wireless detectors as the reference device the Ysio Max cleared via K133259. It is within the same classification regulation for the same indications for use and the same mechanical design as the predicate devices. The Multix Fusion Max is substantially equivalent to the predicate devices and documentation is provided to support a claim of substantial equivalence.

Multix Fusion Max(subject)Multix Fusion K142049(primary predicate)ComparisonResults
Indicationsfor UseThe Multix Fusion Maxsystem is a radiographicsystem used in hospitals,clinics, and medicalpractices. Multix Fusion Maxenables radiographicexposures of the whole bodyincluding: skull, chest,abdomen, and extremitiesand may be used onpediatric, adult and bariatricpatients. Exposures may betaken with the patient sitting.The Multix Fusion systemis a radiographic systemused in hospitals, clinics,and medical practices.Multix Fusion enablesradiographic exposures ofthe whole body including:skull, chest, abdomen, andextremities and may beused on pediatric, adult andbariatric patients.Exposures may be takenwith the patient sitting.Same

Comparison of Indications for Use

{6}------------------------------------------------

standing, or in the proneposition. The Multix FusionMax system is not meant formammography.standing, or in the proneposition. The MultixFusion system is not meantfor mammography.
The Multix Fusion uses amobile (wired), or fixed(integrated) or wirelessdigital detector forgenerating diagnostic imagesby converting x-rays intoimage signals.The Multix Fusion uses amobile (wired), or fixed(integrated) or wirelessdigital detector forgenerating diagnosticimages by converting x-rays into image signals.
The Multix Fusion is alsodesigned to be used withconventional film/ screen orComputed Radiography(CR) cassettes.The Multix Fusion is alsodesigned to be used withconventional film/ screenor Computed Radiography(CR) cassettes.

Comparison of Subject Device to the Primary Predicate Device

AttributeMultix Fusion Max (subject)Multix Fusion K142049 (primary predicate)Comparison Results
Ceiling mounted support (unchanged)Semi-automatedSemi-automatedSame
X-ray tube assembly (unchanged)80 kWTwo-focus80 kWTwo-focusSame
Collimator (unchanged)Standard collimator(ACSS)Standard collimator(ACSS)Same
Patient TableRadiographic table motorized with floating table top (wireless detector)Radiographic table motorized with floating table top (wireless detector)Charging tray
Radiographic table motorized with floating table top (integrated detector)Radiographic table motorized with floating table top (integrated detector)Same
Bucky Wall StandBWS for integrated detector motorized height adjustment (vertical movement)BWS for integrated detector motorized height adjustment (vertical movement)Same
BWS for wireless detectormotorized height adjustment(vertical movement)BWS for wireless detectormotorized or manualheight adjustment (verticalmovement)Only motorizedheightadjustment
TouchInterface(unchanged)Graphical user interfaceGraphical user interfaceSame
X-raygenerator(unchanged)55kW, 65 kW or 80 kW55kW, 65 kW or 80 kWSame
IntegrateddetectorTRIXELLPixium 4343RCE43 cm x 43 cmPixium 4343RG43 cm x 43 cmPixium 4343RC43 cm x 43 cmCompared tothe predicatePixium4343RC-theactive area issmaller but theperformance(e.g. DQE,MTF) is similar
PortableWirelessDetector(wi-D)TRIXELLPixium 3543EZh43 cm x 35 cmPixium 2430EZ30 cm x 24 cmPaxScan 4336W43cm x 36 cmSametechnology withWIFI feature
Operatingmodes(unchanged)RAD Single ExposureRAD Single ExposureSame
ImagingSystem(unchanged)Fluorospot Compact PCbased,high res digitalFluorospot Compact PCbased,high res digitalSame
Display 19"(unchanged)Black & Whiteor ColorBlack & Whiteor ColorSame
DICOM 3Functions(unchanged)Send, StC, Print,Query/Retrieve, GetWorklist, MPPSSend, StC, Print,Query/Retrieve, GetWorklist, MPPSSame
RadiographicGrid92-line grid(ratio 13:1)85-line grid(ratio 5:1) (optional)80-line grid (ratio10:1)50-line grid (ratio10:1)More lines,different ratio
Accessory(unchanged)Remote ControlRemote ControlSame

{7}------------------------------------------------

{8}------------------------------------------------

consoleconsole
Fixed Subject Detector Compared to the Fixed Primary Predicate Detector
TechnicalSpecificationsTrixell Pixium4343RCE detector(fixed) (subject)Trixell Pixium4343RC detector(fixed) (predicate)ComparisonResults
Dimensions423.3 mm x 425.4 mm424.6 mm x 425.5 mmDifference notsignificant
Resolution2860 x 2874 pixels2869 x 2874 pixelsDifference notsignificant
Pixel size148 um148 umSame
SemiconductorMaterialAmorphous silicon, a-SiAmorphous silicon, a-SiSame
ScintillatorCesium iodide (CsI)Cesium iodide (CsI)Same
Acquisition depth16 bit16 bitSame
DQE (DetectiveDuantinDOE @ 0.05 lp/mm (2μGy), 67%DOE @ 0.05 lp/mm (2uGy), 65%Difference notsignificant
(Modulationstransfer function)MTF @ 1 lp/mm, 62%MTF @ 1 lp/mm, 63%Difference notsignificant

|

Wireless Subject Detector Compared to the Wireless Primary Predicate Detector

TechnicalSpecificationsTrixell Pixium 3543EZh detector(wireless) (subject)Varian PaxScan4336W detector(wireless)(predicate)ComparisonResults
Dimensions349 mm x 425 mm356 mm x 427 mmSmaller
Resolution2356 x 2872 pixels3032 x 2520 pixelsDifference notsignificant
Pixel size148 139 Larger
SemiconductorMaterialAmorphous silicon, a-SiAmorphous silicon, a-SiSame
ScintillatorCesium iodide (CsI)Cesium iodide (CsI)Same
Acquisition depth16 bit16 bitSame
DQE (DetectiveQuantumEfficiency)DQE @ 1 lp/mm (2 DQE @ 1 lp/mm (2 Difference notsignificant
MTF(Modulationstransfer function)MTF @ 1 lp/mm, 61%MTF @ 1 lp/mm, 57%Difference notsignificant

Wireless Subject Detector Compared to the Wireless Reference Predicate Detector

TechnicalSpecificationsTrixell Pixium 3543EZh detector(wireless) (subject)Trixell Pixium 3543EZh detector(wireless)(reference predicate)ComparisonResults
Dimensions349 mm x 425 mm349 mm x 425 mmSame

{9}------------------------------------------------

Resolution2356 x 2872 pixels2356 x 2872 pixelsSame
Pixel size148 μm148 μmSame
SemiconductorMaterialAmorphous silicon, a-SiAmorphous silicon, a-SiSame
ScintillatorCesium iodide (CsI)Cesium iodide (CsI)Same
Acquisition depth16 bit16 bitSame
DQE (DetectiveQuantumEfficiency)DQE @ 1 lp/mm (2 μGy), 50%DQE @ 1 lp/mm (2 μGy), 50%Same
MTF(Modulationstransfer function)MTF @ 1 lp/mm, 61%MTF @ 1 lp/mm, 61%Same

Wireless Subject Detector Compared to the Wireless Primary Predicate Detector

TechnicalSpecificationsTrixell Pixium2430EZ detector(wireless) (subject)Varian PaxScan4336W detector(wireless) (primary predicate)ComparisonResults
Dimensions228 mm x 284 mm356 mm x 427 mmSmaller
Resolution1560 x 1920 pixels3032 x 2520 pixelsDifference notsignificant
Pixel size148 μm139 μmLarger
SemiconductorMaterialAmorphous silicon, a-SiAmorphous silicon, a-SiSame
ScintillatorCesium iodide (CsI)Cesium iodide (CsI)Same
Acquisition depth16 bit16 bitSame
DQE (DetectiveQuantumEfficiency)DQE @ 1 lp/mm (2μGy), 50%DQE @ 1 lp/mm (2μGy), 54%Difference notsignificant
MTF(Modulationstransfer function)MTF @ 1 lp/mm,61%MTF @ 1 lp/mm,57%Difference notsignificant

Wireless Subject Detector Compared to the Wireless Reference Predicate Detector

TechnicalSpecificationsTrixell Pixium2430EZ detector(wireless) (subject)Trixell Pixium2430EZ detector(wireless) (referencepredicate)ComparisonResults
Dimensions228 mm x 284 mm228 mm x 284 mmSame
Resolution1560 x 1920 pixels1560 x 1920 pixelsSame
Pixel size148 µm148 µmSame
SemiconductorMaterialAmorphous silicon, a-SiAmorphous silicon, a-SiSame
ScintillatorCesium iodide (CsI)Cesium iodide (CsI)Same
Acquisition depth16 bit16 bitSame
DQE (DetectiveQuantumEfficiency)DQE @ 1 lp/mm (2 µGy), 50%DQE @ 1 lp/mm (2 µGy), 50%Same
MTF (Modulationstransfer function)MTF @ 1 lp/mm,61%MTF @ 1 lp/mm,61%Same

{10}------------------------------------------------

8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device

The Multix Fusion Max is comparable in indications for use, design, material, functionality, technology, energy source and is substantially equivalent to the commercially available predicates. It uses the same or similar components cleared in the Multix Fusion (e.g. tube, generator, ceiling-mounted tube support, table, Bucky wall stand and imaging system) and the Ysio Max (wireless detectors).

The components of the subject device have many of the same technological characteristics as the ones from the predicate devices. There are some technological characteristics that differ slightly as shown in the comparison tables.

Testing and validation have been successfully completed and test results show that the subject device Multix Fusion Max with all of its components are comparable to the predicate devices and therefore is substantially equivalent to the predicate devices.

The modifications made to the subject device Multix Fusion Max do not affect the intended use of the device nor do they alter its fundamental scientific technology from the 510(k) cleared predicate devices.

9. Nonclinical Performance Testing

Non-clinical tests were conducted for the Multix Fusion Max during product development. The modifications described in this Premarket Notification are supported with verification and validation testing.

Multix Fusion Max conforms to the following standards: IEC 60601-1:2012, ed.3.1; IEC 60601-1-2 3rd 2007; IEC 60601-1-3:2008+A1:2013; IEC 62366: 2014; ISO 14971:2007; IEC 62304 Ed. 1.1 2015; IEC 60601-2-28 Edition 2.0 (2010); IEC 60601-2-54 Edition 1.1 2015; PS 3.1 - 3.20 (2011); IEC 60601-1-6: 2013 and ISO 10993-1:2009.

Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. Non-clinical tests (integration and functional) were conducted on the Multix Fusion Max during product development.

Detector integration testing consisted of functional and image quality tests. Functional tests included exposure workflow, detector sharing function and image resend functions. Image quality tests included flat field uniformity of the detector.

The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria. Testing for verification and validation for the device was found acceptable to support the claims of substantial equivalence.

10. General Safety and Effectiveness Concerns

Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner. Several safety features including visual and audible warnings are incorporated into the system design. In addition, the

{11}------------------------------------------------

Multix Fusion Max Radiography X-ray system is continually monitored, and if an error occurs, the system functions will be blocked and an error message will be displayed. Risk management is ensured via a hazard analysis which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. Furthermore the operators are health care professionals familiar with and responsible for the X-ray examinations to be performed.

11. Conclusion as to Substantial Equivalence

The Multix Fusion Max has the same indications for use as the predicate Multix Fusion. The operating environment and mechanical design are similar. It is Siemens opinion, that the Multix Fusion Max is substantially equivalent to the Multix Fusion, cleared 12/18/2014 in K142049.

The predicate device was cleared based on non-clinical supportive information. Therefore, the subject device non-clinical data supports the safety of software with verification and validation testing. Verification and validation testing demonstrates that the Multix Fusion Max should perform as intended. The non-clinical test data demonstrate that the Multix Fusion Max device performance is comparable to the predicate device that is currently marketed for the same intended use.

In summary, Siemens is of the opinion that the Multix Fusion Max does not introduce any new potential safety risk and is substantially equivalent to and performs as well as the predicate device.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.