K Number
K142049
Date Cleared
2014-12-18

(143 days)

Product Code
Regulation Number
892.1680
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. Exposures may be taken with the patient sitting, standing, or in the prone position.

The Multix Fusion system is not ment for mammography. The Multix Fusion uses a mobile (wired), or fixed (integrated) or wreless digital detector for generating diagnostic images signals. The Multix Fusion is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.

Device Description

Siemens Medical Solutions USA, Inc. intends to market a Radiography X-ray system the Multix Fusion VA30 and VA40. This 510(k) submission describes several modifications to the previously cleared predicate devices the Multix Fusion K121513 and the YSIO Max Radiography X-ray system K133259. The following Modifications are made to the cleared predicate devices and developed into the Subject Device the Multix Fusion VA 30 and VA40:

1). A new 43x43cm fixed detector, Pixium 4343RG manufactured by Trixell; 2). A new 43x36cm wireless detector, Paxscan 4336W manufactured by Varian. 3). A new 43x43cm fixed detector, Pixium 4343RC manufactured by Trixell. 4) Slight change in the Indication for Use Statement as compared to the secondary predicate Ysio Max K133259). 5). Bucky Wall Stand (BWS) changed from manual to motor driven movement. 6). Radiographic Grid change from movable 50-line grid to stationary 80-line grid. 7). A remote control console to control BWS motorized movement, center align, collimator size and collimator light and 8). An optional Multix DR-Upgrade with wireless detector (Varian Paxscan 4336W) for the predicate the Multix Fusion, K121513 is also available within Multix Fusion VA40.

The Multix Fusion VA30 and VA40 Radiography X-ray system is designed as a modular system of x-ray components (ceiling suspension with X-ray tube, Bucky wall stand, Bucky table, X-ray generator, portable wireless and integrated detectors) which is the same as the predicate the Multix Fusion K121513.

AI/ML Overview

This FDA document is a 510(k) summary for the Siemens Multix Fusion VA30 and VA40 X-ray system. It focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than establishing acceptance criteria or reporting performance as one would for a novel AI device with specific performance metrics.

Therefore, the requested information cannot be fully provided from the given document as it does not contain the level of detail regarding acceptance criteria, specific device performance numbers, sample sizes for test sets, expert qualifications, or details about comparative effectiveness studies typically found in submissions for AI/ML devices.

However, based on the provided text, here's what can be extracted and inferred:

1. A table of acceptance criteria and the reported device performance

The document does not provide a table with specific quantitative acceptance criteria or numerical performance metrics (e.g., sensitivity, specificity, AUC) for the Multix Fusion VA30 and VA40. The evaluation is based on demonstrating that modifications do not affect the intended use and that the device conforms to various performance standards and safety regulations.

The "acceptance criteria" are implied by adherence to:

  • Performance standards: IEC 60601-1 (various editions), IEC 62366, ISO 14971, IEC 62304, PS 3.1 - 3.20 (DICOM), IEC 60601-1-6, ISO 10993-1.
  • Safety and Effectiveness: Demonstrated through verification and validation testing, risk analysis, and conformance to EMC/electrical safety standards.
  • Substantial Equivalence: The core acceptance is that the device is "substantially equivalent" to predicate devices (Multix Fusion K121513 and YSIO Max K133259) in terms of intended use, design, material, functionality, technology, and energy source, and that it does not introduce new safety risks.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document primarily discusses non-clinical performance testing (verification and validation testing, integration and functional tests, risk analysis). It explicitly states: "Clinical testing is not applicable due to the fact that no new clinical applications were introduced to the system." Therefore, there is no mention of a test set with human patient data or its provenance for evaluating the device's diagnostic performance for the new modifications.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable, as clinical testing with a human-read test set for diagnostic performance was not performed or deemed necessary for this 510(k) submission.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as clinical testing with a human-read test set for diagnostic performance was not performed or deemed necessary.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No such study was conducted or reported. This device is a stationary X-ray system, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a physical X-ray system, not a standalone algorithm. Its performance is evaluated based on its technical specifications, conformance to standards, and safety.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the non-clinical and verification/validation testing mentioned, the "ground truth" would be established by engineering specifications, calibration standards, and adherence to recognized industry and regulatory standards (e.g., verifying radiation output against expected values, confirming software functionality as per design documents, ensuring electrical safety within limits). There is no patient-level ground truth in the context of diagnostic accuracy.

8. The sample size for the training set

Not applicable. This device is an X-ray system. The modifications are related to physical components (detectors, Bucky stand, grid, etc.) and associated software for controlling the system, not a machine learning algorithm trained on a dataset.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or movement.

December 18, 2014

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Siemens Medical Solutions USA, Inc. % Ms. Patricia Jones Regulatory Affairs Specialist 51 Valley Stream Parkway, D-02 MALVERN PA 19355

Re: K142049

Trade/Device Name: Multix Fusion VA30 & VA40 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR Dated: November 25, 2014 Received: November 26, 2014

Dear Ms. Jones:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert A Ochs

Robert A. Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K142049

Device Name Multix Fusion

Indications for Use (Describe)

The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. Exposures may be taken with the patient sitting, standing, or in the prone position.

The Multix Fusion system is not ment for mammography. The Multix Fusion uses a mobile (wired), or fixed (integrated) or wreless digital detector for generating diagnostic images signals. The Multix Fusion is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SIEMENS

510(k) Summary: Multix Fusion X-ray System

Siemens Medical Systems, Inc. Company: 51 Valley Stream Parkway Malvern, PA 19355

November 25, 2014 Date Prepared:

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.

1. General Information:

Importer / Distributor:

Siemens Medical Solutions, Inc. 51 Valley Stream Parkway, E-50 Malvern, PA 19355 Establishment Registration Number: 2240869

Location of Manufacturing Site:

Siemens Shanghai Medical Equipment Ltd. 278 Zhou Zhu Road, Shanghai 201318, China Establishment Registration Number: 3003202425

Establishment Registration Number: 3004977335

2. Contact Person:

Ms. Patricia D Jones Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway D-02 Malvern, PA 19355 Phone: (610) 448 -3536 Fax: (610) 448-1787 Email: patricia.d.jones@siemens.com

3. Device Name and Classification:

Trade Name: Classification Name: Classification Panel: Classification Requlation: Device Class: Product Code:

Multix Fusion Stationary X-Ray System Radiology 21 CFR §892.1680 Class II 90 KPR

    1. Legally Marketed Predicate Devices Trade Name: Multix Fusion 510(k) #: K121513

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the word "SIEMENS" in a large, sans-serif font. The letters are a light blue color, and there is a black line underneath the word. The word is the main focus of the image.

Clearance Date:August 10, 2012
Classification Name:Stationary X-ray System
Classification Panel:Radiology
CFR Section:21 CFR §892.1680
Device Class:Class II
Product Code:90 MQB
Trade Name:YSIO Max
510(k)#:K133259
Clearance Date:January 24, 2014
Classification Name:Stationary X-ray System
Classification Panel:Radiology
CFR Section:21 CFR§892.1680
Device Class:Class II
Product Code:90 KPR

5. Device Description:

Siemens Medical Solutions USA, Inc. intends to market a Radiography X-ray system the Multix Fusion VA30 and VA40. This 510(k) submission describes several modifications to the previously cleared predicate devices the Multix Fusion K121513 and the YSIO Max Radiography X-ray system K133259. The following Modifications are made to the cleared predicate devices and developed into the Subject Device the Multix Fusion VA 30 and VA40:

1). A new 43x43cm fixed detector, Pixium 4343RG manufactured by Trixell; 2). A new 43x36cm wireless detector, Paxscan 4336W manufactured by Varian. 3). A new 43x43cm fixed detector, Pixium 4343RC manufactured by Trixell. 4) Slight change in the Indication for Use Statement as compared to the secondary predicate Ysio Max K133259). 5). Bucky Wall Stand (BWS) changed from manual to motor driven movement. 6). Radiographic Grid change from movable 50-line grid to stationary 80-line grid. 7). A remote control console to control BWS motorized movement, center align, collimator size and collimator light and 8). An optional Multix DR-Upgrade with wireless detector (Varian Paxscan 4336W) for the predicate the Multix Fusion, K121513 is also available within Multix Fusion VA40.

The Multix Fusion VA30 and VA40 Radiography X-ray system is designed as a modular system of x-ray components (ceiling suspension with X-ray tube, Bucky wall stand, Bucky table, X-ray generator, portable wireless and integrated detectors) which is the same as the predicate the Multix Fusion K121513.

6. Indication for Use:

The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. Exposures may be taken with the patient

{5}------------------------------------------------

sitting, standing, or in the prone position.

The Multix Fusion system is not meant for mammography. The Multix Fusion uses a mobile (wired), or fixed (integrated) or wireless digital detector for generating diagnostic images by converting x-rays into image signals. The Multix Fusion is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.

7. Substantial Equivalence:

The Multix Fusion VA 30 and VA40 is substantially equivalent to the commercially available Siemens Multix Fusion cleared 8/10/2012 in K121513; and the YSIO Max, cleared on 01/24/2014 in K133259. See Table 1.

Predicate DevicesName & Manufacturer510(k) NumberClearanceDateComparable Properties
Multix FusionSiemens AGK1215138/10/2012Indication for useX-ray technologyImage processingMechanical design
Ysio MAXSiemens AGK13325901/24/2014Indication for useX-ray technologyImage processingMechanical design

Table 1: Predicate Device Comparable Properties

Summary of Technological Characteristics of the Subject Device as 8. Compared with the Predicate Device:

The Multix Fusion VA30 and VA40 Radiography X-ray system is designed as a modular system of x-ray components (ceiling suspension with X-ray tube, Bucky wall stand, Bucky table, X-ray generator, portable wireless and integrated detectors) same as the predicate Multix Fusion. The Multix Fusion VA30 and VA40 is comparable in indication for use, design, material, functionality, technology, energy source and is considered substantially equivalent to the commercially available Siemens' Ysio Max and the Multix Fusion. It uses the same or similar components cleared in the Multix Fusion and Ysio Max (e.g. tube, generator, ceiling-mounted tube support, table, Bucky wall stand and imaging system).

The components of the subject device have many of the same technological characteristics as the ones from the predicate devices. There are several technological characteristics that differ slightly as shown in the comparison tables (indicated as "similar").

Testing and validation is completed. Test results shows that the subject device the Multix Fusion VA30 and VA40 with all its components are comparable to the predicate devices and therefore is substantial equivalent to its predicate devices.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The word is written in a light blue color. There is a black line underneath the word.

The subject device the Multix Fusion VA30 and VA40 does not affect the intended use of the device nor does it alter its fundamental scientific technology from the 510(k) cleared predicate devices the Multix Fusion and the Ysio Max.

9. Nonclinical Performance Testing

Non- clinical tests were conducted for the Multix Fusion VA30 and VA40 during product development. The modifications described in this Premarket Notification were supported with verification and validation testing.

a signed statement of conformance to the following eleven (11) performance standards: 60601-1 3rd; 60601-1-2 3rd 2007; 60601-1-3: 2nd 2008; 62366: 2007; 14971:2010; 62304 Ed. 1.0 2006; 60601-2-28 Edition 2.0 (2010); 60601-2-54 2009; PS 3.1 - 3.20 (2011) 60601-1-6 and 10993-1:2009.

Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "quidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. Nonclinical tests (integration and functional) were conducted on the Multix Fusion VA30 and VA40 during product development.

The Risk analysis was completed and risk control implemented to mitigate The testing results supports that all the software identified hazards. specifications have met the acceptance criteria. Testing for verification and validation for the device was found acceptable to support the claims of substantial equivalence.

EMC/electrical safety was evaluated according to the IEC Standards. Siemens certify conformance to Voluntary Standards covering Electrical hazards was mitigated and is substantially equivalent to the predicate device in terms of safety and effectiveness. All testing and validation has been completed.

Clinical testing is not applicable due to the fact that no new clinical applications were introduced to the system.

10. General Safety and Effectiveness Concerns:

Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner. Several safety features including visual and audible warnings are incorporated into the system design. In addition the Multix Fusion VA30 and VA40 Radiography X-ray system is continually monitored, and if an error occurs, the system functions will be blocked and an error message will be displayed.

Risk management is ensured via a hazard analysis which is used to identify potential hazards. These potential hazards are controlled via software

{7}------------------------------------------------

SIEMENS

development, verification and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. Furthermore the operators are health care professionals familiar with and responsible for the X-ray examinations to be performed.

11. Conclusion as to Substantial Equivalence:

The Multix Fusion VA30 and VA40 is intended for a similar indication for use as the predicate Multix Fusion and the predicate YSIO Max. The operating environment and mechanical design is similar. It is Siemens opinion, that the Multix Fusion VA30 and VA40 is substantially equivalent to the Multix Fusion, cleared 8/10/2012 in K121513; and the YSIO Max, cleared 01/24/2014 in K133259 x-ray system.

The predicate device was cleared based on non-clinical supportive information. Therefore, the Subject Device non-clinical data supports the safety of software with verification and validation testing. Verification and validation testing demonstrates that the Multix Fusion VA30 and VA40 should perform as intended. The non-clinical test data demonstrate that the Multix Fusion VA30 and VA40 device performance is comparable to the predicate device that is currently marketed for the same intended use.

In summary, Siemens is of the opinion that the Multix Fusion VA30 and VA40 do not introduce any new potential safety risk and is substantially equivalent to and performs as well as the predicate devices.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.