K Number
K162911
Manufacturer
Date Cleared
2017-02-22

(128 days)

Product Code
Regulation Number
866.3480
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sofia RSV FIA employs immunofluorescence for detection of respiratory syncytial virus (RSV) nucleoprotein antigen in nasopharyngeal swab and nasopharyngeal aspirate/wash specimens taken directly from symptomatic patients. This qualitative test is intended for use as an aid in the rapid diagnosis of acute RSV infections in pediatric patients. Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or for other management decisions. A negative result is presumptive, and it is recommended these results be confirmed by viral culture or an FDA-cleared RSV molecular assay.

The Sofia RSV FIA may be used with the Sofia or Sofia 2.

Device Description

The Sofia RSV FIA test employs immunofluorescence technology that is used with Sofia for the rapid detection of RSV antigens. The Sofia RSV FIA test involves the disruption of RSV viral antigens. The patient specimen is placed in the Reagent Tube, during which time the virus particles in the specimen are disrupted, exposing internal viral nucleoproteins. After disruption, the specimen is dispensed into the Cassette sample well. From the sample well, the specimen migrates through a test strip containing various unique chemical environments. If RSV viral antigens are present, they will be trapped in a specific location.

Note: Depending upon the user's choice, the cassette is either placed inside of Sofia or Sofia 2 for automatically timed development (Walk Away Mode) or placed on the counter or bench top for a manually timed development and then placed into Sofia 2 to be scanned (Read Now Mode).

Sofia or Sofia 2 will scan the test strip and measure the fluorescent signal by processing the results using method-specific algorithms. Test results will be displayed (Positive, or Invalid) on the screen. The results can also be automatically printed on an integrated printer if this option is selected.

Sofia 2 is a microprocessor-controlled device about the size of a desk top telephone and weighs less than 3 pounds. Sofia 2 uses a fluorescent tag that is illuminated by an Ultraviolet (UV) light source to generate specific results.

AI/ML Overview

The document describes the Sofia RSV FIA device and its performance, but it does not contain a detailed table of acceptance criteria and reported device performance with specific metrics like sensitivity and specificity, nor does it detail the specific clinical study design, sample sizes, and ground truth establishment methods for a clinical evaluation.

However, based on the provided text, here's what can be extracted and inferred:

1. Table of Acceptance Criteria and Reported Device Performance

The submission states that "Numerous studies were undertaken to document the performance characteristics of Sofia 2 and the Sofia RSV assay, as well as to compare the performance between Sofia and Sofia 2." The conclusion is that "These studies demonstrated equivalent performance of the Sofia RSV FIA on the Sofia and Sofia 2 analyzer."
While specific numerical acceptance criteria (e.g., "sensitivity must be >90%") are not explicitly stated within this document, the studies listed (Limit of Detection, Precision, Assay Development Time, Early Read, Method Comparison, Reproducibility) confirm the equivalence of the Sofia RSV FIA on Sofia 2 to the predicate device (Sofia RSV FIA on Sofia). This implies that the acceptance criteria for performance are met if the performance on Sofia 2 is comparable to the already cleared Sofia device.

Study TypeReported Device PerformanceAcceptance Criteria (Inferred)
Limit of Detection (LoD)LoD for Sofia RSV FIA on Sofia 2 is equivalent to Sofia.LoD on Sofia 2 must be equivalent to LoD on predicate Sofia device.
PrecisionEquivalent qualitative results between Sofia and Sofia 2.Sofia 2 must demonstrate equivalent precision to the predicate Sofia device.
Assay Development Time8 to 30 minutes acceptable in Read Now mode.Acceptable range for assay development time.
Early ReadPositive samples detectable as early as 3 minutes in Walk Away mode.Early detection capability demonstrated.
Method Comparison (Clinical Samples)Sofia and Sofia 2 have comparable performance.Performance of Sofia 2 must be comparable to the predicate Sofia device using clinical samples.
ReproducibilityIntra- and inter-operator, and interlaboratory reproducibility demonstrated with comparable performance between Sofia and Sofia 2.Sofia 2 must demonstrate equivalent reproducibility to the predicate Sofia device.

2. Sample Size Used for the Test Set and Data Provenance

The document mentions "a panel of clinical samples" for the Method Comparison study and "a panel of test samples at various RSV concentrations" for the Reproducibility study. However, the specific sample sizes for these test sets are NOT provided. The data provenance (e.g., country of origin, retrospective or prospective) is also not explicitly stated in this document.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not provided in the document. For an in vitro diagnostic device like Sofia RSV FIA, the "ground truth" for clinical samples is typically established by a reference method, such as viral culture or an FDA-cleared molecular assay for RSV, as mentioned in the "Indications for Use" for confirming negative results. Expertise would be in performing and interpreting these reference methods.

4. Adjudication Method for the Test Set

This information is not provided in the document, as it likely pertains to expert review of ambiguous cases, which is not described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study involving human readers with and without AI assistance is not described in this document. This device is an in vitro diagnostic (IVD) test, which is a standalone algorithm-based test. The "readers" would be the Sofia/Sofia 2 instruments themselves, not human interpreters of AI output.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the studies described are standalone performance evaluations of the Sofia RSV FIA device (algorithm only), specifically when performed on the Sofia 2 analyzer, and comparing its performance to the Sofia analyzer. The device automatically scans the test strip and processes results using "method-specific algorithms." Humans load the cassette and read the displayed result.

7. The Type of Ground Truth Used

The "Indications for Use" section states: "A negative result is presumptive, and it is recommended these results be confirmed by viral culture or an FDA-cleared RSV molecular assay." This indicates that the ground truth for clinical samples, especially for confirmation, would be established by these reference methods. For laboratory studies like LoD and Reproducibility, the ground truth would be based on known concentrations of RSV.

8. The Sample Size for the Training Set

The document does not mention a training set sample size. For an IVD device like this, the "algorithm" is likely developed and refined during the product development phase (which would involve internal testing and optimization data, akin to a training set), but specific details on that are not provided in this regulatory submission summary for substantial equivalence. The studies described are performance validation studies.

9. How the Ground Truth for the Training Set Was Established

Since a training set is not explicitly mentioned, the method for establishing its ground truth is also not provided. If an algorithm was trained, the ground truth would typically be established through highly accurate reference methods (like viral culture or PCR) on a diverse set of samples.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping to create a sense of depth.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Quidel Corporation Edward C. Brehm, PhD Regulatory Affairs Manager 12544 High Bluff Drive, Suite 200 San Diego, CA 92130

February 22, 2017

Re: K162911

Trade/Device Name: Sofia® RSV FIA Regulation Number: 21 CFR 866.3480 Regulation Name: Respiratory syncytial virus serological reagents Regulatory Class: I Product Code: GOG Dated: October 14, 2016 Received: October 24, 2016

Dear Dr. Brehm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Uwe Scherf -S

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162911

Device Name Sofia® RSV FIA

Indications for Use (Describe)

The Sofia RSV FIA employs immunofluorescence for detection of respiratory syncytial virus (RSV) nucleoprotein antigen in nasopharyngeal swab and nasopharyngeal aspirate/wash specimens taken directly from symptomatic patients. This qualitative test is intended for use as an aid in the rapid diagnosis of acute RSV infections in pediatric patients. Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or for other management decisions. A negative result is presumptive, and it is recommended these results be confirmed by viral culture or an FDA-cleared RSV molecular assay.

The Sofia RSV FIA may be used with the Sofia or Sofia 2.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a square with rounded corners, divided into two sections. The top section is a gradient of colors, including red, orange, yellow, green, blue, and purple. The bottom section is a smaller square that is blue and purple. Below the square is the word "QUIDEL" in a sans-serif font.

5. 510(K) SUMMARY

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for QuidelOrtho Corporation. The logo features a square with a smaller square cut out of the bottom right corner. The larger square is filled with a gradient of colors, including red, orange, yellow, green, blue, and purple. The smaller square is filled with a gradient of blue and purple. Below the square is the word "QUIDEL" in a sans-serif font.

5.1. Submitter

Quidel Corporation 12544 High Bluff Drive, Suite 200 San Diego, California 92130 Telephone: 858-552-1100, X-21015 Fax: 858-646-8045

5.2. Submission Contact

Edward C. Brehm, Ph.D.

5.3. Date Prepared

February 21, 2017

5.4. Proprietary and Established Names

Sofia® RSV FIA performed on Sofia 2

5.5. Common Name

Same as above

Regulatory Information 5.6.

Product CodeClassificationRegulatory SectionPanel
GQGI21 CFR 866.3480Microbiology
KHOI21 CFR 862.2560Clinical Chemistry per regulation; Microbiology because it is used with the Sofia RSV FIA

5.7. Predicate Device

Sofia RSV FIA performed on Sofia

5.8. Device Description

The Sofia RSV FIA test employs immunofluorescence technology that is used with Sofia for the rapid detection of RSV antigens. The Sofia RSV FIA test involves the disruption of RSV viral antigens. The patient specimen is placed in the Reagent Tube, during which time the virus particles in the specimen are disrupted, exposing internal viral nucleoproteins. After disruption, the specimen is dispensed into the Cassette sample well. From the sample well, the specimen

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image features the logo of Quidel. The logo consists of a square with rounded corners, filled with a gradient of colors resembling a rainbow. Inside the square is a smaller, white square with rounded corners, creating a frame-like effect. The word "QUIDEL" is written in a simple, sans-serif font below the square logo.

migrates through a test strip containing various unique chemical environments. If RSV viral antigens are present, they will be trapped in a specific location.

Note: Depending upon the user's choice, the cassette is either placed inside of Sofia or Sofia 2 for automatically timed development (Walk Away Mode) or placed on the counter or bench top for a manually timed development and then placed into Sofia 2 to be scanned (Read Now Mode).

Sofia or Sofia 2 will scan the test strip and measure the fluorescent signal by processing the results using method-specific algorithms. Test results will be displayed (Positive, or Invalid) on the screen. The results can also be automatically printed on an integrated printer if this option is selected.

Sofia 2 is a microprocessor-controlled device about the size of a desk top telephone and weighs less than 3 pounds. Sofia 2 uses a fluorescent tag that is illuminated by an Ultraviolet (UV) light source to generate specific results.

5.9. Intended Use

The Sofia RSV FIA employs immunofluorescence for detection of respiratory syncytial virus (RSV) nucleoprotein antigen in nasopharyngeal swab and nasopharyngeal aspirate/wash specimens taken directly from symptomatic patients. This qualitative test is intended for use as an aid in the rapid diagnosis of acute RSV infections in pediatric patients. Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or for other management decisions. A negative result is presumptive, and it is recommended these results be confirmed by viral culture or an FDA-cleared RSV molecular assay.

The Sofia RSV FIA may be used with Sofia or Sofia 2.

5.10. Substantial Equivalence Information:

  • Predicate Device Name: Sofia RSV FIA performed on Sofia 1.
    1. Predicate 510(k) Numbers: K130398
    1. Comparison with Predicate:

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image is a logo for Quidel. The logo features a square with rounded corners, filled with a gradient of colors from yellow to red to blue. Inside the square is a smaller, white square that is offset to the top left corner. Below the square is the word "QUIDEL" in a sans-serif font.

Similarities and Differences
ItemSofia RSV FIA on SofiaK130398Sofia RSV FIA on Sofia 2K162911
510(k) NumberK130398K162911
Intended UseThe Sofia RSV FIA employsimmunofluorescence for detection ofrespiratory syncytial virus (RSV)nucleoprotein antigen in nasopharyngealswab and nasopharyngeal aspirate/washspecimens taken directly fromsymptomatic patients. This qualitativetest is intended for use as an aid in therapid diagnosis of acute RSV infectionsin pediatric patients. Negative results donot preclude RSV infection and shouldnot be used as the sole basis fortreatment or for other managementdecisions. A negative result ispresumptive, and it is recommendedthese results be confirmed by viralculture or an FDA-cleared RSVmolecular assay.The Sofia RSV FIA employsimmunofluorescence for detection ofrespiratory syncytial virus (RSV)nucleoprotein antigen in nasopharyngealswab and nasopharyngeal aspirate/washspecimens taken directly fromsymptomatic patients. This qualitativetest is intended for use as an aid in therapid diagnosis of acute RSV infectionsin pediatric patients. Negative results donot preclude RSV infection and shouldnot be used as the sole basis fortreatment or for other managementdecisions. A negative result ispresumptive, and it is recommendedthese results be confirmed by viralculture or an FDA-cleared RSVmolecular assay.The Sofia RSV FIA may be used withSofia or Sofia 2.
Calibration CheckCalibration Check required every 30days or less, as set by the supervisor. Aspecial Calibration Cassette is providedwith the Installation Pack.Same and uses the same CalibrationCassette
DevelopmentModesTwo basic assay development modes:• Walk-Away: User can walk awayduring the assay cassettedevelopment period• Read Now: User manually times theassay cassette development periodoutside of Sofia, then places cassettein Sofia to image and provide testresult.Same
Development Time15 minutes for Sofia RSV FIAPotential for early read in Walk-AwayMode. Sofia 2 will image cassette at 3,5, 8, 10, and 15 minutes until a positiveresult is received.
SystemComponents
User interface3.5 inch diagonal color LCD display andnumeric keypad with function specificbuttons4 inch color LCD touchscreen display
Similarities and Differences
ItemSofia RSV FIA on SofiaSofia RSV FIA on Sofia 2
510(k) NumberK130398K162911
User TypesHas 2 distinct security levels; user and supervisor plus a Quidel only service level.Same
Barcodescanner(sample)External hand held barcode scannerIntegrated barcode scanner but same functionality
Barcodescanner(cassette)Integrated barcode scannerSame using custom integrated 0.3 MP camera
Assay / instrument interfaceDrawer (electro-mechanical)Same (manual)
Determine test typeInstrument scans barcode on cassetteSame
Power Supply100 – 240 VAC, self-switching, or with 4 AA batteries100 – 240 VAC, self-switching, or with rechargeable lithium polymer battery
PrinterIntegrated printerExternal printer connected via USB port (DYMO LabelWriter 450 Printer supported), optional network printer.
Dimensions24 cm deep x 16 cm wide x 10 cm high19.7 cm deep x 11.4 cm wide x 12.7 cm high
Weight3 lbs~2.5 lbs

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image features the logo for Quidel. The logo consists of a square divided into four quadrants, each filled with a different color. The top left quadrant is a gradient of warm colors, including red, orange, and yellow. The top right quadrant is a gradient of green and blue. The bottom left quadrant is a gradient of purple and pink. The bottom right quadrant is a gradient of blue and purple. Below the square is the word "QUIDEL" in a sans-serif font.

5.11. Performance Data

Numerous studies were undertaken to document the performance characteristics of Sofia 2 and the Sofia RSV assay, as well as to compare the performance between Sofia and Sofia 2. The studies included the following:

  • a. Limit of Detection (LoD)
    This study confirmed that the LoD generated for the Sofia RSV FIA on Sofia 2 is equivalent to the LoD generated on Sofia.

  • b. Precision
    This study confirmed that Sofia and Sofia 2 generated equivalent qualitative results when used by multiple operators to test negative and positive concentrations that are close to the positivity threshold, on multiple device lots, operated over multiple days, and two calibration cycles.

  • c. Assay development time
    This study confirmed that when running Sofia 2 in Read Now mode, a development time of eight (8) to thirty (30) minutes is acceptable.

  • d. Early Read
    This study confirmed that when running Sofia 2 in Walk Away mode, positive samples (depending on the viral load) can be interpreted as positive as early as 3 minutes.

  • Method Comparison e.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image features the logo of QuidelOrtho Corporation. The logo consists of a square divided into four quadrants, each filled with a different color gradient. The top-left quadrant transitions from yellow to green, the top-right from orange to red, the bottom-left from light blue to purple, and the bottom-right from dark blue to purple. Below the square is the word "QUIDEL" in a simple, sans-serif font.

This study demonstrated that Sofia and Sofia 2 have comparable performance when using a panel of clinical samples.

  • f. Reproducibility
    This study demonstrated intra- and inter-operator reproducibility and interlaboratory reproducibility with a panel of test samples at various RSV concentrations. This study also demonstrated comparable performance between Sofia and Sofia 2.

5.12. Conclusion

These studies demonstrated equivalent performance of the Sofia RSV FIA on the Sofia and Sofia 2 analyzer.

§ 866.3480 Respiratory syncytial virus serological reagents.

(a)
Identification. Respiratory syncytial virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to respiratory syncytial virus in serum. Additionally, some of these reagents consist of respiratory syncytial virus antisera conjugated with a fluorescent dye (immunofluorescent reagents) and used to identify respiratory syncytial viruses from clinical specimens or from tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of respiratory syncytial virus infections and provides epidemiological information on diseases caused by these viruses. Respiratory syncytial viruses cause a number of respiratory tract infections, including the common cold, pharyngitis, and infantile bronchopneumonia.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.