K Number
K162821

Validate with FDA (Live)

Device Name
Babyleo TN500
Date Cleared
2017-06-23

(259 days)

Product Code
Regulation Number
880.5400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Babyleo TN500 is intended for use with premature babies and can be used as both an incubator and a radiant warmer. When the product is switched between incubator and radiant warmer operation, patients continue to be kept warm during the transition. The device provides a thermally regulated environment for patients with a body weight of up to 5 kg (11 lbs) and a height of up to 55 cm (22 in). The device can be operated as either a closed care unit or an open care unit. As a closed care unit, Babyleo TN500 is an incubator.Neonates are kept warm in the patient with humidifiable air, which can be enriched with oxygen (option). As an open care unit, Babyleo TN500 is a radiant warmer. Babyleo TN500 provides controlled ambient conditions for premature babies and neonates. The following parameters are regulated, according to the intended use: - Temperature - Humidity - Oxygen (option)

Device Description

The device Babyleo TN500 is a medical device used to maintain environmental conditions suitable for preterm babies and neonates. The important features are temperature and humidity. The operational principle is a combination of a neonatal incubator and an open warmer.

Babyleo TN500 can be operated in closed care therapy as an incubator according to IEC60601-2-19 or in open care therapy as a warmer according to IEC60601-2-21. Additionally, an optional x-ray translucent heated mattress is intended to provide sensible heat to the preterm babies and neonates.

The device consists of the following components: Incubator (=bassinet) with a convection heater and closed humidification system for the patient, trolley, radiant warmer, display, mattress, x-ray and height adjustment. Optional components are scale, heated mattress, drawer, oxygen regulation with cylinder holder and Auto Thermo package and Developmental Care package.

AI/ML Overview

This document is a 510(k) Premarket Notification for the Babyleo TN500, a neonatal incubator and radiant warmer. It aims to demonstrate substantial equivalence to legally marketed predicate devices. The information provided heavily focuses on design verification and validation testing, and therefore, does not contain the details on acceptance criteria and study proving performance in the context of an AI/ML medical device that would typically involve clinical studies, multi-reader multi-case studies, and explicit ground truth establishment with expert consensus.

Given the document talks about a physical medical device (neonatal incubator/radiant warmer) and not an AI/ML diagnostic or prognostic tool, the questions posed about "acceptance criteria and the study that proves the device meets the acceptance criteria" in terms of clinical performance metrics like sensitivity, specificity, or AUC, multi-reader multi-case studies, expert consensus for ground truth, effect size for human reader improvement with AI assistance, and training/test set sample sizes and provenance for AI/ML models are not applicable to this submission.

The "acceptance criteria" and "study that proves the device meets the acceptance criteria" for this conventional medical device primarily revolve around safety and performance standards, electrical safety, EMC, software verification, biocompatibility, human factors, and reprocessing validation.

Here's how to address the prompt based on the provided document:

1. A table of acceptance criteria and the reported device performance:

The document lists compliance with various standards and provides specifications that serve as performance criteria.

Acceptance Criteria (Standard Compliance / Performance Spec)Reported Device Performance (Compliance / Value)
Electrical Safety: IEC 60601-1, IEC 60601-2-19, IEC 60601-2-21, IEC 60601-2-35Device complies with all listed standards.
Electromagnetic Compatibility (EMC): IEC 60601-1-2Device complies with the standard.
Software Verification and Validation: FDA Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." (2005)Software verification and validation testing conducted; documentation provided; considered "major" level of concern.
Biocompatibility: ISO 10993, ISO 10993-1, ISO 10993-17, ISO 10993-18Evaluation conducted in accordance with FDA guidance; battery of testing included: Integral Test for volatile organic compounds, Photogenic bacteria test, Emission of particles, Material characterization, Toxicological Evaluation, Cytotoxicity, Irritation, Sensitization.
Human Factors: FDA Guidance: Applying Human Factors and Usability Engineering to Medical Devices Guidance for Industry and Food and Drug Administration Staff, 2016Usability evaluation conducted in accordance with FDA guidance.
Reprocessing: FDA Guidance: Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling; Guidance for Industry and Food and Drug Administration Staff, 2015Reprocessing validation conducted in accordance with FDA guidance.
Operating Noise Volume in Patient CompartmentTypically 40 dB(A) (Measured without oxygen application) - (Compared to predicate <49 dB(A), Caleo 47 ± 2 dB(A))
Alarm Volume50 to 70 dB(A) (according to IEC 60601-2-19 and IEC 60601-2-21) - (Compared to predicate "according to IEC 60601-2-19")
Power Failure AlarmYes (65 dB(A), 10 min) - (Compared to predicate "Yes")
Maximum Irradiance of Warmer (at 100% power)32 mW/cm2 - (Compared to Babytherm 30 mW/cm2)
Warm-up Time15 to 20 min - (Compared to predicate <50 min, Caleo 20 min)
Flow Velocity over Mattress Surface10 cm/s (3.94 in/s) (at 36°C) - (Compared to predicate <10 cm/s, Caleo <8 cm/s)
Maximum Carbon Dioxide (CO2) concentration in incubator<0.5 Vol% - (Compared to predicate 0.2 %, Caleo <0.5 Vol%, Babytherm max 0.5 Vol%)
Air Temperature Measurement Range13 to 45 °C (55.4 to 113 °F) - (Compared to Caleo 13 to 42 °C)
Air Temperature Measurement Uncertainty± 0.8°C (1.44 °F) - (Compared to Caleo ± 0.8°C)
Skin Temperature Measurement Uncertainty (Entire System)± 0.3°C (0.54 °F) - (Matches predicate and reference devices)
Oxygen Measurement Range (Normal)18 to 65 Vol% - (Matches Caleo)
Oxygen Measurement Uncertainty± 2.5 Vol% (+ 2.5 % of measured value) - (Compared to predicate 5%, Caleo ± 3 Vol%)
Humidity Measurement Range (Normal)30 to 99 % r. H. - (Matches Caleo)
Humidity Measurement Uncertainty±10 % - (Matches predicate and Caleo)
Heated Mattress Setting RangeOff, 35°C to 39°C, AUTO - (Compared to Babytherm Off, 30°C to 38.5°C)
Scale Range200 g to 10 kg - (Compared to predicate 300 g to 8 kg, Caleo 250 g to 10 kg, Babytherm 250 g to 8 kg)
Scale Resolution5 g (OIML) or 1 g (standard version) - (Compared to predicate 10 g or 5 g, Caleo 1 g or 5 g, Babytherm 1 g or 10 g)

2. Sample sized used for the test set and the data provenance:

  • Test Set Sample Size: Not applicable. This is a conventional medical device; performance testing is often conducted on a representative number of units or components, but the "sample size" concept as used in AI/ML clinical studies for a test dataset does not directly apply. The testing mentioned (electrical safety, EMC, software verification, biocompatibility, human factors, reprocessing) would be performed on the device itself or its materials.
  • Data Provenance: Not applicable in the sense of patient data. The provenance for the testing data would be from internal lab testing conducted by the manufacturer (Drägerwerk AG & Co. KGaA, Germany) or their designated testing facilities. It's not retrospective or prospective patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. Ground truth in the context of an AI/ML clinical study (e.g., diagnosis of disease) is not relevant for this device. The "ground truth" for this device's performance is established by objective measurements against engineering specifications and recognized international standards. Experts involved would be engineers, quality assurance personnel, and regulatory specialists.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable. Adjudication methods are typically for resolving discrepancies in expert labeling or diagnoses in AI/ML studies. For a physical device, performance is measured against established limits derived from standards or engineering requirements.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, not done. This is a physical medical device, not an AI-assisted diagnostic tool. MRMC studies are not relevant for demonstrating the effectiveness of an incubator/warmer.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. There is no "algorithm only" performance for a neonatal incubator/radiant warmer as it is not an AI/ML-driven diagnostic or treatment algorithm. Its performance is measured by its ability to maintain specified environmental conditions.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • Objective Measurement against Standards and Specifications: The "ground truth" for this device is the adherence to internationally recognized safety and performance standards (e.g., IEC 60601 series, ISO 10993) and the manufacturer's own detailed engineering specifications. For example, the "truth" of the noise level is the scientifically measured sound pressure level, not an expert's subjective assessment.

8. The sample size for the training set:

  • Not applicable. There is no "training set" in the context of an AI/ML model for this device. Design and development of the device would involve engineering principles, materials science, and iterative testing, not machine learning model training.

9. How the ground truth for the training set was established:

  • Not applicable. As there is no training set for an AI/ML model, this question is not relevant. The "ground truth" for the device's design and manufacturing is derived from scientific and engineering principles, regulatory requirements, and established medical device standards.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 23, 2017

Draegerwerk AG & Co. KGaA % Ms. Beth Zis Director, Regulatory Affairs Draeger Medical Systems, Inc. 6 Tech Drive Andover, Massachusetts 01810

Re: K162821

Trade/Device Name: Babyleo TN500 Regulation Number: 21 CFR 880.5400 Regulation Name: Neonatal Incubator Regulatory Class: Class II Product Code: FMZ, FMT Dated: May 24, 2017 Received: May 30, 2017

Dear Ms. Beth Zis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

James P. Bertram -S

for

Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162821

Device Name Babyleo TN500

Babyleo TN500 is intended for use with premature babies and can be used as both an incubator and a radiant warmer. When the product is switched between incubator and radiant warmer operation, patients continue to be kept warm during the transition. The device provides a thermally regulated environment for patients with a body weight of up to 5 kg (11 lbs) and a height of up to 55 cm (22 in). The device can be operated as either a closed care unit or an open care unit. As a closed care unit, Babyleo TN500 is an incubator.Neonates are kept warm in the patient with humidifiable air, which can be enriched with oxygen (option). As an open care unit, Babyleo TN500 is a radiant warmer. Babyleo TN500 provides controlled ambient conditions for premature babies and neonates. The following parameters are regulated, according to the intended use: - Temperature - Humidity - Oxygen (option)

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue and the background is white. The two dots above the "a" are also blue.

510(k) Summary

K162821

Applicants / Manufacturer Name and Address:

Drägerwerk AG & Co. KGaA Moislinger Allee 53-55, 23558 Luebeck, Germany

Contact Person:

Dr. Bettina Möbius Senior Manager Regulatory Affairs

Tel. No.: 011 49 (451) 882-4100 Fax No.: 011 49 (451) 882-3018

Applicants US Contact Person

Beth Zis Director, Regulatory Affairs Draeger Medical Systems, Inc. 6 Tech Drive, Andover, MA 01810 USA

Tel: +1 (978) 379-8265 Fax: +1 (978) 379-8335

Date submission was prepared:

06/05/2017

Device Name:

Device Trade Name: Common / Usual Name: Classification Name Regulatory Class: Product Code:

Babyleo TN500 Incubator and Warmer Incubator, neonatal 21 CFR 880.5400 = FMZ, FMT

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "Dräger" in blue font. The font is bold and sans-serif. The two dots above the "a" are also blue. The background is white.

Legally Marketed Devices to which Substantial Equivalence is claimed:

Legally Marketed Predicate Device: K010222, Ohmeda Medical Giraffe Incubator Reference Devices: K003067, CALEO K971198, Babytherm 8004/8010

Device Description:

The device Babyleo TN500 is a medical device used to maintain environmental conditions suitable for preterm babies and neonates. The important features are temperature and humidity. The operational principle is a combination of a neonatal incubator and an open warmer.

Babyleo TN500 can be operated in closed care therapy as an incubator according to IEC60601-2-19 or in open care therapy as a warmer according to IEC60601-2-21. Additionally, an optional x-ray translucent heated mattress is intended to provide sensible heat to the preterm babies and neonates.

The device consists of the following components: Incubator (=bassinet) with a convection heater and closed humidification system for the patient, trolley, radiant warmer, display, mattress, x-ray and height adjustment. Optional components are scale, heated mattress, drawer, oxygen regulation with cylinder holder and Auto Thermo package and Developmental Care package.

Indications for Use:

Babyleo TN500 is intended for use with premature babies and can be used as both an incubator and a radiant warmer. When the product is switched between incubator and radiant warmer operation, patients continue to be kept warm during the transition. The device provides a thermally regulated environment for patients with a body weight of up to 5 kg (11 lbs) and a height of up to 55 cm (22 in). The device can be operated as either a closed care unit or an open care unit. As a closed care unit, Babyleo TN500 is an incubator. Neonates are kept warm in the patient compartment with humidifiable air, which can be enriched with oxygen (option). As an open care unit, Babyleo TN500 is a radiant warmer. Babyleo TN500 provides controlled ambient conditions for premature babies and neonates. The following parameters are regulated, according to the intended use: Temperature; Humidity; Oxygen (option)

Substantial Equivalence Comparison:

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the word "Dräger" in a bold, blue font. The "a" in Dräger has an umlaut, which is a diacritical mark consisting of two dots placed over the letter. The logo is simple and modern, and the blue color gives it a professional look.

Image /page/5/Picture/2 description: The image shows the word "Dräger" in a bold, blue font. The two dots above the "a" are also blue. The word is centered in the image and is the only element present.

BabyleoTN500Omnibed(Giraffe)CaleoBabytherm8004/8010Comments
Statusnew devicepredicate devicereference devicereference device
ManufacturerDrägerwerk AG & Co.KGaAOhmeda Medical,a Division ofDatex-Ohmeda,Inc. A GECompanyDrägerwerk AG &Co. KGaADrägerwerk AG &Co. KGaA
Classification Name /Product codeIncubator, neonatal /FMZIncubator, neonatal/ FMZIncubator, neonatal/ FMZInfant RadiantWarmer / FMTSame
510(k)K162821K010222;K003067K971198
Regulation Number:21 CFR 880.540021 CFR 880.540021 CFR 880.540021 CFR 880.5130Same
Class:IIIIIIIISame
Standards60601-1: 60601-1-2:60601-2-19: 60601-2-2160601-1: 60601-1-2: 60601-2-19:60601-2-2160601-1: 60601-1-2: 60601-2-19:60601-1; 60601-1-2;60601-2-21Same
Indications for useThe Babyleo TN500 isintended for use withpremature babies andneonates and can beused as both anincubator and a radiantwarmer. When theproduct is switchedbetween incubator andradiant warmeroperation, patientscontinue to be kept warmduring the transition. Thedevice provides athermally regulatedenvironment for patientswith a body weight of upThe OmniBed is acombination of aninfant incubator(
BabyleoTN500Omnibed(Giraffe)CaleoBabytherm8004/8010Comments
to 5 kg (11 lbs) and aheight of up to 55 cm (22in). The device can beoperated as either aclosed care unit or anopen care unit. As aclosed care unit, BabyleoTN500 is an incubator.Neonates are kept warmin the patientcompartment withhumidifiable air, whichcan be enriched withoxygen (option). As anopen care unit, BabyleoTN500 is a radiantwarmer. Babyleo TN500provides controlledambient conditions forpremature babies andneonates. The followingparameters areregulated, according tothe intended use:Temperature; Humidity;Oxygen (option)temperaturecontrolledenvironment andwarmers provideinfrared heat in anopen environment.They may also beused for shortperiods of time tofacilitate theneonate'stransition from theuterus to theexternalenvironment.treating twins islimited to 5 kg (11lbs). Caleo isintended for use inclinicalenvironmentswhere prematurebabies or infantsare treated whorequire a controlledclimaticenvironment.neonatal andpediatric intensivecare units.
Operating noise
Operating noise volume inpatient compartmentTypically 40 dB(A)(Measured withoutoxygen application)<49 dB(A)47 $\pm$ 2 dB(A)no closed patientcompartmentSimilar
BabyleoTN500Omnibed(Giraffe)CaleoBabytherm8004/8010Comments
Alarm Volume50 to 70 dB(A)(according to IEC 60601-2-19 and IEC 60601-2-21)according to IEC60601-2-19according to IEC60601-2-19Not providedSimilar
Power failure alarmYes (65 dB(A), 10 min)YesYesYesSame
GeneralRadiant Warmer andNeonatal IncubatorRadiant Warmerand NeonatalIncubatorNeonatal IncubatorRadiant WarmerSame
WarmerRadiant WarmerRadiant WarmerNA (Incubator only)Radiant WarmerSame
Maximum Irradiance ofWarmer32 mW/cm2 (at 100%power); 18 mW/cm2 (at60% power) 10 mW/cm2(at 30% Power)Not providedNA (Incubator only)30 mW/cm2 (heatlevel 10); 10mW/cm2(heat level 3)Same
HeaterConvective heaterConvective heaterConvective heaterNA (Radiant Warmeronly)Same
Warm up time15 to 20 min< 50 min20 minNA (warmer only)Similar
Humidification principleBoiling of Aquadest(destilled water)DrawoverEvaporatorBoiling of AquadestNA (warmer only)Similar
Flow velocity over matresssurface10 cm/s (3.94 in/s) (at36°C)<10 cm/s< 8 cm/sNA (warmer only)Same
Maximum Carbon Dioxide(CO2) concentration in theincubator measured inaccordance with IEC60601-2-19<0.5 Vol%0.2 %<0.5 Vol%max 0.5 Vol% (withbed canopy)Same
BabyleoTN500Omnibed(Giraffe)CaleoBabytherm8004/8010Comments
Electrical Description
Power supply100 V to 240 V, 50/60 Hz100V, 115V, 220V,230V, 240V, 50/60 Hz100V, 110V, 120 V,127V, 220V to240V; 50 Hz / 60 Hz100V, 110 - 127 V;220 - 240 VSame
Maximum current / powerconsumption10A / 1000Wnot provided8.7A / 732W15A / 1750 WSimilar
Electrical safetyaccording to IEC 60601-1according to IEC 60601-1according to IEC 60601-1according to IEC 60601-1Same
Gas supply (option)
O2 operating pressure270 to 600 kPanot provided300 and 600 kPa(43.5 and 87 psi).NA (warmer only)Similar
Measured value display
Radiant Warmer
Setting rangeOff, 10 % to 100 %0 to100 % in 5%INCNA (Incubator only)not providedSame
Prewarm ModeYesYesNA (Incubator only)YesSame
Air temperatureregulation
Measurement regulation13 to 45 °C (55.4 to 113 °F)Not provided13 to 42 °C (55.4 °F to 107.6 °F)NA (warmer only)Similar
Measurement uncertainty$\pm$ 0.8°C (1.44 °F)Not provided$\pm$ 0.8°C (1.44 °F)NA (warmer only)Same
Air Mode temperaturesetting range20 to 39 °C (68 to 102.2 °F) in 0.1 ° inc.20 to 39 °C in 0.1 ° inc.20 to 39 °C (68 to 102.2 °F)NA (warmer only)Same
Air Mode override<28 °C (82.4 °F), afterconfirmation>37 °C (98.6 °F), afterconfirmationNot provided<28 °C (82.4 °F),after confirmation>37 °C (98.6 °F),after confirmationNA (warmer only)Same
BabyleoTN500Omnibed(Giraffe)CaleoBabytherm8004/8010Comments
Skin Temperatureregulation
Measurement range13 to 43 °C (55.4 to109.4 °F)Not providedNA (Incubator only)15 °C to 42 °CSimilar
Temperature Sensor forSkinThermo Trace (Dräger)Not providedThermo Trace(Dräger)Thermo Trace(Dräger)Same
Measurement uncertaintysensors± 0.1ºC (0.18 °F)Not provided± 0.1ºC (0.18 °F)± 0.1ºC (0.18 °F)Same
Measurement uncertaintyentire system± 0.3°C (0.54 °F)± 0.3°C (0.54 °F)± 0.3°C (0.54 °F)± 0.3°C (0.54 °F)Same
Skin temperature Modesetting range34 to 38 °C (93.2 to100.4 °F) in 0.1 °C35 to 37.5 °C in 0.1°C34 to 38 °C (93.2to 100.4 °F)not providedSame
Skin temperature Modeoverride>37 °C (98.6 °F), afterconfirmationNot provided>37 °C (98.6 °F),after confirmationnot providedSame
Heated mattress (Option)
Setting rangeOff, 35°C to 39°C, AUTONot providedNAOff, 30°C to 38.5°CSimilar
Oxygen regulation(Option)
measurement rangeNormal range: 18 to 65Vol %Extended range: 66 to99 Vol%Not providedNormal range: 18to 65 Vol %Extended range:66 to 99 Vol%NA (warmer only)Same
Measurement uncertainty± 2.5 Vol% (+ 2.5 % ofmeasured value)5%± 3 Vol%NA (warmer only)Same
Setting range21 to 65 Vol%(>40 Vol%, afterconfirmation)21 to 65%21 to 75 Vol%NA (warmer only)Same
02 rise time from 21 to 65Vol%<10 min10 minutes from21% to 5% belowset pointNot providedNA (warmer only)Same
Oxygen Cylinder HolderYesYesYesNA (warmer only)Same
BabyleoTN500Omnibed(Giraffe)CaleoBabytherm8004/8010Comments
Humidity SystemDescription
Measurement rangeNormal range: 30 to 99% r. H.Extended range: 10 to29 % r. H.up to 85%set point range: 30to 99 % r. H.measure range: 10to 99 % r. H.NA (warmer only)Same
Measurement uncertainty±10 %±10 %±10 %NA (warmer only)Same
Setting RangeOff, 30 to 99 % r. H. insteps of 1 %, AUTO30 to 95% in 5%incrementsOff, 30 to 99 % r.H. in steps of 1 %,AUTONA (warmer only)Same
Noise and lightmeasurement(Developmental carepackage Option)YesNoNoNoDifferent
Noise Measurement range40 to 90 dB(A)not availablenot availablenot availableDifferent
Light Measurement range3 to 999 Lux (1 to 93 fc)not availablenot availablenot availableDifferent
AutoThermo package(Option)YesNoNoNoDifferent
Data exchange andinterfacesSerial Port (COM,RS232), USB (USB 1.1or USB 2.0); Service port(RJ45 / LAN)ThermaLink Serialdata interface(RS232);Serial Port (COM,RS232), Serviceport (RJ45 / LAN)Serial Port (COM,RS232), Service port(RJ45 / LAN)Same
Physical Description
Weight (without optionsand accessories)<140 kg138 kg130 kg (max. 230kg)133 kgSimilar
Width & Depth690 mm x 1154 mm(27.17 in x 45.43 in)690 mm x 1120mm687 mm x 1167mm750 mm x 1315 mmSimilar
Height (with hood closed)1850 mm to 2250 mm(72.83 in to 88.58 in)(height adjustmentrange: 400 mm)1470 mm to 1770mm (heightadjustment range:300 mm)1220 mm to 1520mm1896 mm to 2210mmSimilar
BabyleoTN500Omnibed(Giraffe)CaleoBabytherm8004/8010Comments
Height of the mattresssurface700 to 1100 mm (27.56in to 43.31 in)Not provided800 mm to 1100mm885 mm to 1180 mmSimilar
Mattress450 mm x 690 mm(17.72 in x 27.17 in)Not provided500 mm x 645 mm490 mm x 750 mmSimilar
Material (no implants)
Material used for indirectpatient contactMetal (e.g. Aluminium)Synthetic material (e.g.TPE)UnknownMetal (e.g. Aluminium)Synthetic material(e.g. TPE)Metal (e.g. Aluminium)Synthetic material(e.g. TPE)Similar
Material used for directpatient contactTextile (VowalonMedilind)UnknownTextile(Polyurethane)Textile(Polyurethane)Similar
Biocompatibilityaccording to ISO 10993according to ISO10993according to ISO10993according to ISO10993Same
SterilizationN/A; non sterileN/A; non sterileN/A; non sterileN/A; non sterileSame
Scale Description
Range200 g to 10 kg300 g to 8 kg250 g to 10 kg250 g to 8 kgSimilar
Resolution5 g (OIML)) or 1 g(standard version)10 g or 5 g1 g or 5 g (OIML)1 g or 10 g (OIML)Similar
Integrated X-ray cassette/ trayYesYesYesYesSame
Drawer (Option)YesYesYesYesSame
Bed-tilt mechanism(Trendelenburg)Continuous up to 13°12° mechanical tiltUp to 13° (± tiltangle)15 to 20°Similar
EnviromentalDescription duringoperation
Ambient operatingtemperature20 to 35 °C (68 to 95°F)20 to 30 °C20 to 25 °C15 to 35 °CSimilar
Operating barometricrange620 to 1100 hPa (9.0 to16.0 psi)Not provided600 to 1060 hPa900 hPa to 1060 hPaSimilar
BabyleoTN500Omnibed(Giraffe)CaleoBabytherm8004/8010Comments
Humidity operating range20 to 95 %, non-condensing10 to 95% noncondensating10 to 95%, non.-condensating0 to 75 %, nocondensationSimilar
EnviromentalDescription duringstorageTemperature - 20 to 60°C (to 4 to 140 °F)Ambient pressure 500 to1100 hPa (7.3 to 16.0psi)Relative humidity 10 to95 %, non-condensingTemperature - 25to 60°CHumidity 0 to 95%Non-condensingrelative humidityPressure 50 to 106kPaTemperature - 20to 60 °C (- 4 to 140°F)Ambient pressure210 - 1060 hPaRelative humidity10 - 95 %, non-condensingTemperature - 20 to60 °C (- 4 to 140 °F)Ambient pressure700 - 1060 hPaRelative humidity 0 -90 %, non-condensingSimilar

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the word "Dräger" in blue font. The font is bold and sans-serif. The two dots above the "a" are also blue.

Image /page/6/Picture/2 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue. The two dots above the "a" are also blue.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the word "Dräger" in a bold, blue font. The two dots above the "a" are also blue. The word is slightly slanted to the right.

Image /page/7/Picture/2 description: The image shows the word "Dräger" in a bold, blue font. The "ä" in Dräger has two dots above it. The logo is simple and clean, with a focus on the company name.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue. The "a" in Dräger has an umlaut above it.

Image /page/8/Picture/2 description: The image shows the word "Dräger" in a bold, blue font. The "ä" in the word has two dots above it. The word is centered in the image and is the only element present.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the word "Dräger" in a bold, blue font. The "ä" in the word has two dots above it. The word is the logo for the Dräger company, which is a German company that makes medical and safety technology.

Image /page/9/Picture/2 description: The image shows the word "Dräger" in a bold, blue font. The two dots above the "a" are also blue. The word is centered in the image and is the only element present.

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the word "Dräger" in a bold, blue font. The "a" in Dräger has an umlaut, which is two dots above the letter. A black line is located underneath the word.

Image /page/10/Picture/2 description: The image shows the word "Dräger" in a bold, blue font. The two dots above the "a" are also blue. The word is centered in the image and is the only element present.

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the word "Dräger" in a bold, blue font. The two dots above the "a" are also blue. The word is slightly slanted to the right.

Image /page/11/Picture/2 description: The image shows the word "Dräger" in a bold, blue font. The "ä" in Dräger has two dots above it. The logo is simple and clean, with a focus on the company name.

{12}------------------------------------------------

Image /page/12/Picture/1 description: The image shows the word "Dräger" in blue font. The font is bold and sans-serif. The two dots above the "a" are also blue. The background is white.

Image /page/12/Picture/2 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue and the background is white. The two dots above the "a" are also blue.

The Omnibed (Giraffe) is the predicate device for Babyleo TN500. Both devices are operated as an infant radiant warmer.

The Caleo is a reference device for Babyleo TN500. Both devices are operated as a neonatal incubator.

The Babytherm 8004/8010 is a reference device for Babyleo TN500. Both devices are operated as an infant radiant warmer.

Differences:

The noise and light measurement and audio stimulation are new optional features to enable further developmental care activities in the nicu.

The Auto Thermo Package is a new optional feature cooling, warm up and weaning to ease processes in the NICU. The indications for use of the Babyleo TN500 and the predicate device are equivalent.

The differences do not raise questions about the safety or effectiveness for the subject device.

{13}------------------------------------------------

Image /page/13/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue. The two dots above the "a" are also blue.

Summary of Testing

The following performance data were provided:

Electrical safety and electromagnetic compatibility (EMC), incl. performance testing

Electrical safety and EMC testing were conducted on the Babyleo TN500. The device complies with the IEC 60601-1, IEC 60601-2-19, IEC 60601-2-21 and IEC 60601-2-35 standards for safety and performance and the IEC 60601-1-2 standard for EMC.

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."2005. The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.

Biocompatibility testing

The biocompatibility evaluation for the Babyleo TN500 was conducted in accordance with the FDA guidance on Biocompatibility on the International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,', and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:

Integral Test for volatile organic compounds Photogenic bacteria test Emission of particles Material characterization according to ISO 10993-18 Toxicological Evaluation according to ISO 10993-17 Cytotoxicity Irritation Sensitization

Human Factor

The usability evaluation for the Babyleo TN500 device was conducted in accordance with the FDA guidance: Applying Human Factors and Usability Engineering to Medical Devices Guidance for Industry and Food and Drug Administration Staff, 2016.

Reprocessing

The reprocessing validation for the Babyleo TN500 device was conducted in accordance with the FDA guidance: Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling; Guidance for Industry and Food and Drug Administration Staff, 2015

Animal Study

No animal study have been conducted.

Clinical Studies

No Clinical testing has been done.

Conclusion:

Based on device comparison information and non-clinical bench testing data, the proposed device is substantially equivalent to legally marketed predicate device K010222.

§ 880.5400 Neonatal incubator.

(a)
Identification. A neonatal incubator is a device consisting of a rigid boxlike enclosure in which an infant may be kept in a controlled environment for medical care. The device may include an AC-powered heater, a fan to circulate the warmed air, a container for water to add humidity, a control valve through which oxygen may be added, and access ports for nursing care.(b)
Classification. Class II (performance standards).