(266 days)
TransFix/EDTA Vacuum Blood Collection Tubes (TVTs) are intended for collection and storage of human whole blood specimens for immunophenotyping of white blood cells by flow cytometry. Recovery of lymphocyte subset markers of the HIV panel can be accomplished over a 14-day period following collection. TVTs are in-vitro diagnostic medical devices.
TransFix®/EDTA Vacuum Blood Collection Tubes (TVTs) are intended for collection and storage of human whole blood specimens for immunophenotyping of white blood cells (WBC, leukocytes) by flow cytometry up to 14-days following collection; the TVTs preserve the qualitative and quantitative leukocyte subset characteristics. Flow cytometry is used to identify subsets of leukocytes that can be distinguished on the basis of cell surface antigens using fluorescent antibodies.
The TransFix®/EDTA Vacuum Blood Collection Tubes (TVTs) consist of purple-capped polyethylene terephthalate blood collection tubes containing the paraformaldehyde based preservative, TransFix, and K2EDTA at the correct volume to simultaneously stabilize and anticoagulate human whole blood at the time of collection. TVTs hold 3mL (final draw volume) with a tube dimension of 13 x 75mm. TVTs are sterilized by gamma radiation. Further, the vacuum contained within the TVT ensures that the TransFix/EDTA reagent is administered at the correct ratio of 1 part TransFix/EDTA to 5 parts whole blood.
The provided text describes the performance of a medical device, the TransFix®/EDTA Vacuum Blood Collection Tubes (TVTs), and presents data to support its substantial equivalence to a predicate device. However, it does not explicitly state acceptance criteria or a specific study proving the device meets those criteria in the typical sense of a diagnostic test performance study (e.g., sensitivity, specificity thresholds).
Instead, the documentation focuses on reproducibility (precision) and equivalence (method comparison) to a reference method and a predicate device. The "acceptance criteria" are implied by statements regarding "sufficiently precise" and the comparison tables to accepted methods.
Here's an attempt to extract and present the information given available:
Acceptance Criteria and Device Performance
Implied Acceptance Criteria:
Based on the text, the device aims to demonstrate:
- Reproducibility (Precision): Individual and total %CVs (Coefficient of Variation) are "less than 10% and 15%, respectively" for absolute cell counts of specific lymphocyte biomarkers (CD3, CD8, CD45, CD16/CD56, CD19). This is the key performance metric measured.
- Equivalence: The absolute counts of lymphocyte subsets (CD3, CD4, CD8, CD16/56, CD19, and CD45) in TVT tubes should be comparable to those obtained from the predicate device (Cyto-Chex® BCT) and a gold standard (BD EDTA tubes) over a 14-day storage period. This is assessed via Passing-Bablok regression with slopes and intercepts within an acceptable range, and high R-values indicating strong correlation. The exact thresholds for acceptable slope/intercept/R are not explicitly stated as "acceptance criteria" but are implied by the presentation of the data for a substantial equivalence claim.
- Non-Interference: The performance of the TVTs should not be significantly affected by common interfering substances (bilirubin, hemoglobin, triglycerides).
Reported Device Performance:
Reproducibility Studies:
The tables in the document (Tables 1-12) present %CVs for various biomarkers under different conditions (Between-Day, Between-Lot, Between-Tube, Between-Replicate, Between-Site) and indicate that the TVT lots were "sufficiently precise" against the implied criteria of individual and total %CVs being less than 10% and 15% respectively. For example, in Table 1 (CD3+ biomarker), all total %CVs are well below 15% (ranging from 4.7% to 6.2%). Similar observations are made for all other biomarkers across both reproducibility studies.
Equivalence Study (Method Comparisons):
Tables 12 and 14 (for HIV positive subjects and healthy subjects, respectively) show Passing-Bablok regression results comparing TVTs and BCTs against BD EDTA tubes.
- Slopes: Most slopes for TVT (and BCT) are close to 1.0, indicating good agreement with the BD EDTA control. For HIV-positive subjects, TVT slopes are generally between 0.95 and 1.12. For healthy subjects, TVT slopes are generally between 0.90 and 1.10.
- Intercepts: Intercepts are generally small for both TVT and BCT, indicating minimal systematic bias.
- R-values: All R-values are high (0.95-1.00), demonstrating a strong correlation between the TVTs (and BCTs) and the BD EDTA control for the listed markers.
- The text explicitly states: "The TVT tubes were found to be safe and effective for the intended use."
Interference Study:
The study observed "no significant effects from the potential interferents listed" when comparing results from aliquots with added interfering substances to a control aliquot. This indicates the device meets the non-interference aspect of its performance.
Study Details:
-
A table of acceptance criteria and the reported device performance:
(Combined above based on implicit criteria in the text). -
Sample size used for the test set and the data provenance:
-
Reproducibility Study 1 (Intra-lab variability):
- Sample Size: 5 healthy subjects. For each subject, blood was collected into six TVTs (two tubes from each of three lots). Testing was performed at three time points (0, 12, 15 days post venipuncture) with triplicate determinations on each of the duplicate TVTs from each of the three lots. (N=54 mentioned for each subject for each biomarker calculation, indicating total determinations).
- Data Provenance: Not explicitly stated, but the context of submitting to FDA suggests a regulated environment. Likely prospective, as it's a specific study design.
-
Reproducibility Study 2 (Inter-site variability):
- Sample Size: 5 healthy subjects. Blood collected into six TVTs (from the same lot) from each subject. Tested at three clinical flow cytometry laboratories. Testing performed with triplicate determinations on each of the duplicate TVTs. (N=18 mentioned for each donor for each biomarker calculation).
- Data Provenance: UK (from "three clinical flow cytometry laboratories in the UK"). Likely prospective.
-
Equivalence Study (Method Comparison):
- Sample Size: 30 HIV positive donors and 12 normal donors.
- Data Provenance: Not explicitly stated, but the context implies data collected for this registration. Likely prospective.
-
Interference Study:
- Sample Size: 5 healthy subjects and 5 HIV positive subjects.
- Data Provenance: Not explicitly stated. Likely prospective.
-
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This document describes a device for sample collection and preservation rather than a diagnostic algorithm that interprets images or patient data. Therefore, the concept of "experts establishing ground truth" in the way it applies to AI/radiology devices is not directly applicable here. The ground truth for lymphocyte counts is established by flow cytometry using a reference method (BD EDTA tubes) and accepted laboratory procedures. The "experts" would be the trained laboratory personnel performing the flow cytometry and data analysis. Their specific qualifications are not detailed in this summary.
-
Adjudication method for the test set:
- Not applicable in this context. This is a quantative measurement of cell counts, not an interpretation requiring adjudication.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC study was not done. This device is a blood collection tube, not an AI diagnostic algorithm for human interpretation.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is not an algorithm.
-
The type of ground truth used:
- For the equivalence study, the ground truth was established by immediate (Day 0, within 6 hours) analysis of blood samples collected in BD EDTA tubes, considered the reference condition for immunophenotyping. This represents a "reference method" ground truth.
-
The sample size for the training set:
- Not applicable as this is a physical device (blood collection tube) and not an algorithm requiring a training set in the machine learning sense. The "training" of the tubes is inherent in their manufacturing process and formulation.
-
How the ground truth for the training set was established:
- Not applicable (as above).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping each other to create a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 21, 2017
Caltag Medsystems Ltd. Erika B Ammirati 575 Shirlynn Court Los Altos, CA 94022
Re: K162723
Trade/Device Name: TransFix®/EDTA Vacuum Blood Collection Tubes Regulation Number: 21 CFR 862.1675 Regulation Name: Blood specimen collection device Regulatory Class: II Product Code: GIM Dated: May 24, 2017 Received: May 25, 2017
Dear Ms. Ammirati:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kelly Oliner -S
FOR
Leonthena R. Carrington, MS, MBA, MT(ASCP) Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K162723
Device Name TransFix/EDTA Vacuum Blood Collection Tubes (TVTs)
Indications for Use (Describe)
TransFix/EDTA Vacuum Blood Collection Tubes (TVTs) are intended for collection and storage of human whole blood specimens for immunophenotyping of white blood cells by flow cytometry. Recovery of lymphocyte subset markers of the HIV panel can be accomplished over a 14-day period following collection.
TVTs are in-vitro diagnostic medical devices.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is K162723.
807.92 (a)(1): Name: Caltag Medsystems Ltd.
Address: Whiteleaf Business Centre 11. Litle Balmer Buckingham MK18 1TF UK
- Phone: 011 44 1280 827 460 FAX: 011 44 1280 827 466
Contact: Mr. Daniel Harrison Quality Manager
807.92 (a)(2): Device name- trade name and common name, and classification
Trade name:
TransFix®/EDTA Vacuum Blood Collection Tubes
Common Name: Blood specimen collection device- tubes, vacuum sample, with anticoagulant
Classification: 21 CFR § 862.1675- blood specimen collection device Class: Class II Panel: Clinical Chemistry and Toxicology Product Code: GIM
807.92 (a)(3): Identification of the legally marketed predicate devices
Cyto-Chex® BCT (Blood Collection Tubes, Streck Laboratories, Inc., Omaha, NE), cleared under K080552
807.92 (a)(4): Device Description
TransFix®/EDTA Vacuum Blood Collection Tubes (TVTs) are intended for collection and storage of human whole blood specimens for immunophenotyping of white blood cells (WBC, leukocytes) by flow cytometry up to 14-days following collection; the TVTs preserve the qualitative and quantitative leukocyte subset characteristics. Flow cytometry is used to identify subsets of leukocytes that can be distinguished on the basis of cell surface antigens using fluorescent antibodies.
{4}------------------------------------------------
The TransFix®/EDTA Vacuum Blood Collection Tubes (TVTs) consist of purple-capped polyethylene terephthalate blood collection tubes containing the paraformaldehyde based preservative, TransFix, and K2EDTA at the correct volume to simultaneously stabilize and anticoagulate human whole blood at the time of collection. TVTs hold 3mL (final draw volume) with a tube dimension of 13 x 75mm. TVTs are sterilized by gamma radiation. Further, the vacuum contained within the TVT ensures that the TransFix/EDTA reagent is administered at the correct ratio of 1 part TransFix/EDTA to 5 parts whole blood.
807.92 (a)(5): Intended Use
TransFix/EDTA Vacuum Blood Collection Tubes (TVTs) are intended for collection and storage of human whole blood specimens for immunophenotyping of white blood cells by flow cytometry. Recovery of lymphocyte subset markers of the HIV panel can be accomplished over a 14-day period following collection. TVTs are in-vitro diagnostic medical devices.
The TransFix®/EDTA Vacuum Blood Collection Tubes are substantially equivalent to the Cyto-
| Parameter | TransFix® / EDTA VacuumBlood Collection Tube | Cyto-Chex® BCT(K080552) |
|---|---|---|
| Intended use | For the collection and storage ofblood specimens forimmunophenotyping of WBCby flow cytometry. | Same |
| Tube size | 13 x 75mm | Same |
| Preservation of HIV markers | 14 days | Same |
| Tube type | Polyethylene terephthalate | Glass |
| Contents | Liquid reagent: K3EDTA andTransFix preservative | Liquid reagent: K3EDTA andpreservative |
| Sample volume | 2.5mL | 5 mL |
| Liquid reagent | 0.5mL | 75.8μl |
807.92 (a)(6): Technological Similarities and Differences to the Predicate
Chex® BCT (BCT), Streck Laboratories, Inc., La Vista, NE), (K080552).
| Liquid reagent | 0.5mL | |
807.92 (b)(1,2): Brief Description of Nonclinical and Clinical Data
Reproducibility- Repro Studies 1 and 2 Repro Study 1:
The objective of the study (along with Reproducibility Study 2) was to assess the variability of the 3mL TransFix®/EDTA Vacuum Blood Collection Tube (TVT). This was achieved by collecting blood into six TVTs (two tubes from each of three lots) from five healthy subjects whose blood samples spanned the absolute cell count range for lymphocytes and their subsets. The study compared the lymphocyte absolute cell counts recorded from each TVT sample for three test time points (0, 12 and 15 days post venipuncture). At each time point, testing was performed with triplicate determinations on each of the duplicate TVTs from each of the three lots.
{5}------------------------------------------------
For each sample, all 6 biomarkers were tested (CD3, CD8, CD45, CD16/CD56, CD19). Thus, the components of variation for analysis included:
-
- Between-Day
-
- Between- Tube Lot
-
- Between-Tube
-
- Between-Replicate
-
- Total precision (the combined precision of factors 1-4)
The data are presented in Tables 1-6.
{6}------------------------------------------------
| CD3+ | Between-Day | Between-Lot | Between-Tube | Between-Replicate | Total | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Sample # | N | Mean | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) |
| HC A | 54 | 1470.3 | 49.2 (41.3-60.7) | 3.3 (2.8-4.1) | 52.2 (43.8-64.4) | 3.5 (3.0-4.4) | 46.0 (38.7-56.8) | 3.1 (2.6-3.9) | 14.6 (12.3-18.1) | 1.0 (0.8-1.2) | 89.9 (75.6-111.0) | 6.1 (5.1-7.5) |
| HC B | 54 | 1183.6 | 46.0 (38.7-56.8) | 3.9 (3.3-4.8) | 41.4 (34.8-51.1) | 3.5 (2.9-4.3) | 28.5 (24.0-35.2) | 2.4 (2.0-3.0) | 16.4 (13.7-20.2) | 1.4 (1.2-1.7) | 72.8 (61.2-89.9) | 6.2 (5.2-7.6) |
| HC C | 54 | 404.7 | 7.1 (6.0-8.8) | 1.8 (1.5-2.2) | 10.4 (8.7-12.8) | 2.6 (2.2-3.2) | 2.7 (2.3-3.4) | 0.7 (0.6-0.8) | 6.1 (5.1-7.5) | 1.5 (1.3-1.9) | 19.4 (16.3-23.9) | 4.8 (4.0-5.9) |
| HC D | 54 | 1334.9 | 5.9 (4.9-7.2) | 0.4 (0.4-0.5) | 35.8 (30.1-44.3) | 2.7 (2.3-3.3) | 32.0 (26.9-39.5) | 2.4 (2.0-3.0) | 7.3 (6.1-9.0) | 0.5 (0.5-0.7) | 62.4 (52.5-77.0) | 4.7 (3.9-5.8) |
| HC E | 54 | 1458.9 | 25.7 (21.6-31.7) | 1.8 (1.5-2.2) | 62.9 (52.9-77.6) | 4.3 (3.6-5.3) | 22.0 (18.5-27.1) | 1.5 (1.3-1.9) | 16.4 (13.7-20.2) | 1.1 (0.9-1.4) | 82.5 (69.4-101.8) | 5.7 (4.8-7.0) |
Table 1 demonstrates that the 3mL TVT lots were soft the CD3 biomarker (individual and total %CVs were less than 10% and 15%, respectively).
| CD8+ | Between-Day | Between-Lot | Between-Tube | Between-Replicate | Total | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Sample # | N | Mean | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) |
| HC A | 54 | 517.7 | 44.7 (37.5-55.1) | 8.6 (7.2-10.6) | 29.2 (24.6-36.1) | 5.6 (4.7-7.0) | 12.5 (10.5-15.5) | 2.4 (2.0-3.0) | 1.4 (1.1-1.7) | 0.3 (0.2-0.3) | 56.0 (47.1-69.1) | 10.8 (9.1-13.4) |
| HC B | 54 | 451.2 | 25.4 (21.3-31.3) | 5.6 (4.7-6.9) | 19.6 (16.5-24.3) | 4.4 (3.7-5.4) | 8.1 (6.8-10.0) | 1.8 (1.5-2.2) | 5.2 (4.4-6.4) | 1.1 (1.0-1.4) | 34.1 (28.7-42.1) | 7.6 (6.4-9.3) |
| HC C | 54 | 167.7 | 8.6 (7.3-10.7) | 5.2 (4.3-6.4) | 11.3 (9.5-13.9) | 6.7 (5.6-8.3) | 3.6 (3.1-4.5) | 2.2 (1.8-2.7) | 3.1 (2.6-3.8) | 1.8 (1.5-2.3) | 16.6 (14.0-20.5) | 9.9 (8.3-12.2) |
| HC D | 54 | 758.4 | 11.2 (9.4-13.8) | 1.5 (1.2-1.8) | 20.2 (17.0-24.9) | 2.7 (2.2-3.3) | 7.4 (6.2-9.1) | 1.0 (0.8-1.2) | 4.5 (3.8-5.6) | 0.6 (0.5-0.7) | 46.8 (39.3-57.8) | 6.2 (5.2-7.6) |
| HC E | 54 | 572.1 | 3.4 (2.9-4.2) | 0.6 (0.5-0.7) | 27.1 (22.8-33.5) | 4.7 (4.0-5.9) | 8.8 (7.4-10.9) | 1.5 (1.3-1.9) | 11.0 (9.2-13.6) | 1.9 (1.6-2.4) | 40.2 (33.8-49.6) | 7.0 (5.9-8.7) |
Table 2 demonstrates that the 3mL TVT lots were sufficiently precise for the CD8 biomarker (individual and total %CVs respectively).
| CD4+ | Between-Day | Between-Lot | Between-Tube | Between-Replicate | Total | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Sample # | N | Mean | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) |
| HC A | 54 | 767.5 | 29.9 (25.2-37.0) | 3.9 (3.3-4.8) | 29.5 (24.8-36.4) | 3.8 (3.2-4.7) | 21.6 (18.1-26.6) | 2.8 (2.4-3.5) | 12.5 (10.5-15.4) | 1.6 (1.4-2.0) | 59.0 (49.6-72.8) | 7.7 (6.5-9.5) |
| HC B | 54 | 685.0 | 25.4 (21.4-31.4) | 3.7 (3.1-4.6) | 7.9 (6.6-9.8) | 1.2 (1.0-1.4) | 11.6 (9.7-14.3) | 1.7 (1.4-2.1) | 9.3 (7.9-11.5) | 1.4 (1.1-1.7) | 44.5 (37.4-54.9) | 6.5 (5.5-8.0) |
| HC C | 54 | 197.9 | 7.2 (6.1-8.9) | 3.7 (3.1-4.5) | 0.9 (0.8-1.1) | 0.5 (0.4-0.6) | 0.8 (0.7-1.0) | 0.4 (0.3-0.5) | 2.3 (1.9-2.8) | 1.2 (1.0-1.4) | 16.0 (13.4-19.8) | 8.1 (6.8-10.0) |
| HC D | 54 | 496.3 | 11.9 (10.0-14.7) | 2.4 (2.0-3.0) | 11.4 (9.6-14.1) | 2.3 (1.9-2.8) | 8.0 (6.8-9.9) | 1.6 (1.4-2.0) | 4.9 (4.1-6.1) | 1.0 (0.8-1.2) | 35.0 (29.4-43.2) | 7.1 (5.9-8.7) |
| HC E | 54 | 828.2 | 6.9 (5.8-8.5) | 0.8 (0.7-1.0) | 24.8 (20.9-30.6) | 3.0 (2.5-3.7) | 7.1 (5.9-8.7) | 0.9 (0.7-1.1) | 23.7 (19.9-29.2) | 2.9 (2.4-3.5) | 52.4 (44.0-64.7) | 6.3 (5.3-7.8) |
Table 3 demonstrates that the 3mL TVT lots were sufficiently precise (individual and total %CVs were less than 10% and 15%, respectively).
{7}------------------------------------------------
| Lymph CD45+ | Between-Day | Between-Lot | Between-Tube | Between-Replicate | Total | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Sample # | N | Mean | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) |
| HC A | 54 | 2306.7 | 70.3 (59.1-86.8) | 3.0 (2.6-3.8) | 67.7 (56.9-83.6) | 2.9 (2.5-3.6) | 72.0 (60.6-88.9) | 3.1 (2.6-3.9) | 18.3 (15.4-22.6) | 0.8 (0.7-1.0) | 127.1 (106.8-156.9) | 5.5 (4.6-6.8) |
| HC B | 54 | 1730.2 | 49.3 (41.4-60.9) | 2.8 (2.4-3.5) | 63.5 (53.4-78.4) | 3.7 (3.1-4.5) | 29.7 (25.0-36.6) | 1.7 (1.4-2.1) | 26.1 (21.9-32.2) | 1.5 (1.3-1.9) | 97.1 (81.6-119.9) | 5.6 (4.7-6.9) |
| HC C | 54 | 720.5 | 9.9 (8.3-12.2) | 1.4 (1.2-1.7) | 27.7 (23.2-34.1) | 3.8 (3.2-4.7) | 9.6 (8.1-11.9) | 1.3 (1.1-1.6) | 11.3 (9.5-14.0) | 1.6 (1.3-1.9) | 37.5 (31.5-46.3) | 5.2 (4.4-6.4) |
| HC D | 54 | 1683.0 | 6.7 (5.6-8.2) | 0.4 (0.3-0.5) | 34.2 (28.8-42.3) | 2.0 (1.7-2.5) | 34.3 (28.8-42.3) | 2.0 (1.7-2.5) | 10.4 (8.7-12.8) | 0.6 (0.5-0.8) | 75.8 (63.7-93.6) | 4.5 (3.8-5.6) |
| HC E | 54 | 1952.0 | 19.3 (16.2-23.8) | 1.0 (0.8-1.2) | 88.5 (74.4-109.2) | 4.5 (3.8-5.6) | 34.4 (28.9-42.4) | 1.8 (1.5-2.2) | 20.2 (17.0-24.9) | 1.0 (0.9-1.3) | 107.8 (90.6-133.1) | 5.5 (4.6-6.8) |
Table 4 demonstrates that the 3mL TVT lots were sufficiently precise for the CD45 biomarker (individual and total %CVs respectively).
| CD16+CD56+ | Between-Day | Between-Lot | Between-Tube | Between-Replicate | Total | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Sample # | N | Mean | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) |
| HC A | 54 | 377.1 | 12.3 (10.4-15.2) | 3.3 (2.8-4.0) | 0.1 (0.1-0.1) | 0.0 (0.0-0.0) | 11.6 (9.7-14.3) | 3.1 (2.6-3.8) | 2.3 (2.0-2.9) | 0.6 (0.5-0.8) | 24.9 (20.9-30.7) | 6.6 (5.6-8.2) |
| HC B | 54 | 366.7 | 15.9 (13.4-1.9.6) | 4.3 (3.6-5.4) | 21.0 (17.7-26.0) | 5.7 (4.8-7.1) | 4.3 (3.6-5.4) | 1.2 (1.0-1.5) | 8.0 (6.7-9.9) | 2.2 (1.8-2.7) | 31.1 (26.1-38.4) | 8.5 (7.1-10.5) |
| HC C | 54 | 209.5 | 6.4 (5.4-7.9) | 3.1 (2.6-3.8) | 15.4 (13.0-19.0) | 7.4 (6.2-9.1) | 5.2 (4.3-6.4) | 2.5 (2.1-3.0) | 3.4 (2.8-4.2) | 1.6 (1.4-2.0) | 17.2 (14.5-21.2) | 8.2 (6.9-10.1) |
| HC D | 54 | 217.7 | 6.4 (5.4-7.9) | 3.0 (2.5-3.6) | 5.1 (4.3-6.3) | 2.3 (2.0-2.9) | 4.2 (3.6-5.2) | 1.9 (1.6-2.4) | 3.9 (3.3-4.8) | 1.8 (1.5-2.2) | 15.5 (13.0-19.1) | 7.1 (6.0-8.8) |
| HC E | 54 | 296.9 | 11.5 (9.7-14.2) | 3.9 (3.3-4.8) | 15.1 (12.7-18.7) | 5.1 (4.3-6.3) | 9.6 (8.1-11.8) | 3.2 (2.7-4.0) | 4.2 (3.5-5.2) | 1.4 (1.2-1.8) | 26.6 (22.4-32.8) | 9.0 (7.5-11.1) |
Table 5 demonstrates that the 3mL TVT lots were sufficiently precise for the CD16/56 biomarker (individual and total 15%, respectively).
| CD19+ | Between-Day | Between-Lot | Between-Tube | Between-Replicate | Total | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Sample # | N | Mean | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) |
| HC A | 54 | 352.7 | 9.0 (7.6-11.2) | 2.6 (2.2-3.2) | 14.0 (11.8-17.3) | 4.0 (3.3-4.9) | 4.7 (3.9-5.8) | 1.3 (1.1-1.6) | 6.5 (5.5-8.1) | 1.9 (1.6-2.3) | 26.2 (22.0-32.3) | 7.4 (6.2-9.2) |
| HC B | 54 | 112.9 | 2.0 (1.6-2.4) | 1.7 (1.5-2.1) | 1.4 (1.1-1.7) | 1.2 (1.1-1.5) | 0.1 (0.1-0.2) | 0.1 (0.1-0.1) | 4.4 (3.7-5.4) | 3.9 (3.3-4.8) | 8.2 (6.9-10.1) | 7.3 (6.1-9.0) |
| HC C | 54 | 53.7 | 0.5 (0.4-0.6) | 0.8 (0.7-1.0) | 0.7 (0.6-0.9) | 1.3 (1.1-1.6) | 0.8 (0.7-1.0) | 1.5 (1.2-1.8) | 0.7 (0.6-0.9) | 1.4 (1.2-1.7) | 4.4 (3.7-5.4) | 8.2 (6.9-10.1) |
| HC D | 54 | 106.5 | 1.6 (1.4-2.0) | 1.5 (1.3-1.9) | 2.9 (2.4-3.5) | 2.7 (2.3-3.3) | 2.9 (2.5-3.6) | 2.7 (2.3-3.4) | 3.3 (2.7-4.0) | 3.1 (2.6-3.8) | 10.5 (8.8-13.0) | 9.9 (8.3-12.2) |
| HC E | 54 | 151.2 | 3.6 (3.0-4.4) | 2.4 (2.0-2.9) | 4.6 (3.8-5.6) | 3.0 (2.5-3.7) | 2.4 (2.0-2.9) | 1.6 (1.3-2.0) | 2.6 (2.2-3.2) | 1.7 (1.4-2.1) | 10.0 (8.4-12.3) | 6.6 (5.6-8.2) |
Table 6 demostrates that the 3mL TVT lots were sufficiently precise for the CD19 biomarker (individual and total WCVs were less than 10% and 15%, respectively).
{8}------------------------------------------------
Repro Study 2:
The objective of this study was to build on the data set provided in Reproducibility Study 1, and add the effect of site-to-site testing. This was achieved by collecting blood into six TVTs (from the same lot) from five healthy subjects whose blood samples spanned the absolute cell count range for lymphocytes and their subsets. The study compared the lymphocyte absolute cell counts recorded from each TVT sample after they were tested at three clinical flow cytometry laboratories in the UK in order to analyze the following components of variation:
-
- Between-Site (TVT sample results from different clinical laboratories).
-
- Between-Tube (TVT sample results from a different tube, but of the same lot).
-
- Between-Replicate (TVT sample results from the same tube).
-
- Total precision (the combined precision of factors 1-3).
At each site, testing was performed with triplicate determinations on each of the duplicate TVTs using one lot of tubes. For each sample, all six biomarkers were tested (CD3, CD8, CD45, CD16/CD56, CD19). Thus, the components of variation for analysis included.
The data are presented in Tables 7-12.
{9}------------------------------------------------
| CD3 | Between-Site | Between-Tube | Between-Replicate | Total | ||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Donor | N | Mean | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) |
| A | 18 | 1441.1 | 85.5 (64.2-128.2) | 5.9 (4.5-8.9) | 17.4 (13.1-26.1) | 1.2 (0.9-1.8) | 24.2 (18.2-36.3) | 1.7 (1.3-2.5) | 97.0 (72.8-145.4) | 6.7 (5.1-10.1) |
| B | 18 | 1435.7 | 107.4 (80.6-161.0) | 7.5 (5.6-11.2) | 13.2 (9.9-19.8) | 0.9 (0.7-1.4) | 28.7 (21.5-43.0) | 2.0 (1.5-3.0) | 103.9 (78.0-155.8) | 7.2 (5.4-10.8) |
| C | 18 | 372.3 | 12.4 (9.3-18.5) | 3.3 (2.5-5.0) | 9.5 (7.1-14.2) | 2.5 (1.9-3.8) | 20.7 (15.6-31.1) | 5.6 (4.2-8.4) | 28.4 (21.3-42.6) | 7.6 (5.7-11.4) |
| D | 18 | 1416.6 | 90.1 (67.6-135.1) | 6.4 (4.8-9.5) | 46.8 (35.2-70.2) | 3.3 (2.5-5.0) | 78.3 (58.7-117.4) | 5.5 (4.1-8.3) | 122.2 (91.7-183.2) | 8.6 (6.5-12.9) |
| E | 18 | 1195.1 | 52.0 (39.0-78.0) | 4.4 (3.3-6.5) | 49.9 (37.5-74.9) | 4.2 (3.1-6.3) | 67.1 (50.3-100.6) | 5.6 (4.2-8.4) | 86.0 (64.5-128.9) | 7.2 (5.4-10.8) |
Table 7 demonstrates that the 3mL TVT lots were sufficiently precise for the CD3 biomarker (individual and total %CVs respectively).
| CD8 | Between-Site | Between-Tube | Between-Replicate | Total | ||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Donor | N | Mean | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) |
| A | 18 | 592.5 | 48.3 (36.2-72.4) | 8.1 (6.1-12.2) | 26.7 (20.0-40.0) | 4.5 (3.4-6.8) | 8.8 (6.6-13.1) | 1.5 (1.1-2.2) | 54.4 (40.8-81.6) | 9.2 (6.9-13.8) |
| B | 18 | 599.2 | 54.0 (40.5-80.9) | 9.0 (6.8-13.5) | 8.5 (6.4-12.8) | 1.4 (1.1-2.1) | 15.0 (11.3-22.5) | 2.5 (1.9-3.8) | 52.4 (39.3-78.6) | 8.7 (6.6-13.1) |
| C | 18 | 187.3 | 13.0 (9.8-19.5) | 6.9 (5.2-10.4) | 6.5 (4.9-9.7) | 3.5 (2.6-5.2) | 12.2 (9.1-18.2) | 6.5 (4.9-9.7) | 21.0 (15.8-31.5) | 11.2 (8.4-16.8) |
| D | 18 | 868.9 | 61.2 (46.0-91.8) | 7.0 (5.3-10.6) | 23.0 (17.2-34.4) | 2.6 (2.0-4.0) | 47.7 (35.8-71.4) | 5.5 (4.1-8.2) | 89.6 (67.2-134.3) | 10.3 (7.7-15.5) |
| E | 18 | 527.1 | 28.0 (21.0-42.0) | 5.3 (4.0-8.0) | 34.9 (26.2-52.4) | 6.6 (5.0-9.9) | 30.1 (22.6-45.1) | 5.7 (4.3-8.6) | 51.0 (38.3-76.5) | 9.7 (7.3-14.5) |
Table 8 demonstrates that the 3mL TVT lots were sufficiently precise (individual and total %CVs were less than 10% and 15%, respectively).
| CD4 | Between-Site | Between-Tube | Between-Replicate | Total | ||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Donor | N | Mean | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) |
| A | 18 | 782.1 | 26.7 (20.0-40.0) | 3.4 (2.6-5.1) | 4.3 (3.2-6.5) | 0.6 (0.4-0.8) | 17.4 (13.0-26.1) | 2.2 (1.7-3.3) | 48.8 (36.6-73.2) | 6.2 (4.7-9.4) |
| B | 18 | 828.2 | 46.0 (34.5-68.9) | 5.6 (4.2-8.3) | 24.7 (18.5-37.0) | 3.0 (2.2-4.5) | 12.9 (9.7-19.4) | 1.6 (1.2-2.3) | 51.6 (38.7-77.4) | 6.2 (4.7-9.3) |
| C | 18 | 165.9 | 5.0 (3.8-7.5) | 3.0 (2.3-4.5) | 0.6 (0.5-1.0) | 0.4 (0.3-0.6) | 7.6 (5.7-11.4) | 4.6 (3.4-6.9) | 16.3 (12.2-24.4) | 9.8 (7.4-14.7) |
| D | 18 | 488.2 | 23.8 (17.9-35.7) | 4.9 (3.7-7.3) | 10.9 (8.2-16.4) | 2.2 (1.7-3.4) | 19.5 (14.7-29.3) | 4.0 (3.0-6.0) | 42.7 (32.0-64.0) | 8.7 (6.6-13.1) |
| E | 18 | 646.1 | 13.7 (10.3-20.5) | 2.1 (1.6-3.2) | 36.4 (27.3-54.6) | 5.6 (4.2-8.5) | 28.0 (21.0-42.0) | 4.3 (3.3-6.5) | 48.5 (36.4-72.7) | 7.5 (5.6-11.3) |
Table 9 demonstrates that the 3mL TVT lots were sufficiently precise (individual and total %CVs were less than 10% and 15%, respectively).
{10}------------------------------------------------
| CD45 | Between-Site | Between-Tube | Between-Replicate | Total | ||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Donor | N | Mean | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) |
| A | 18 | 2221.6 | 124.3 (93.3-186.4) | 5.6 (4.2-8.4) | 27.1 (20.4-40.7) | 1.2 (0.9-1.8) | 29.1 (21.9-43.7) | 1.3 (1.0-2.0) | 146.2 (109.7-219.2) | 6.6 (4.9-9.9) |
| B | 18 | 2122.5 | 167.2 (125.5-250.7) | 7.9 (5.9-11.8) | 35.0 (26.3-52.5) | 1.6 (1.2-2.5) | 47.4 (35.6-71.1) | 2.2 (1.7-3.4) | 162.6 (122.0-243.8) | 7.7 (5.7-11.5) |
| C | 18 | 613.3 | 22.1 (16.6-33.1) | 3.6 (2.7-5.4) | 19.0 (14.3-28.5) | 3.1 (2.3-4.6) | 38.8 (29.1-58.1) | 6.3 (4.7-9.5) | 53.1 (39.8-79.6) | 8.7 (6.5-13.0) |
| D | 18 | 1806.2 | 111.6 (83.7-167.2) | 6.2 (4.6-9.3) | 60.1 (45.1-90.1) | 3.3 (2.5-5.0) | 102.8 (77.2-154.2) | 5.7 (4.3-8.5) | 154.2 (115.7-231.2) | 8.5 (6.4-12.8) |
| E | 18 | 1743.8 | 80.2 (60.2-120.2) | 4.6 (3.5-6.9) | 77.4 (58.1-116.0) | 4.4 (3.3-6.7) | 100.6 (75.5-150.9) | 5.8 (4.3-8.7) | 131.6 (98.8-197.3) | 7.5 (5.7-11.3) |
Table 10 demonstrates that the 3mL TVT lots were sufficiently precise for the CD45 bionariser (individual and total MCVs respectively).
| CD16/CD56 | Between-Site | Between-Tube | Between-Replicate | Total | ||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Donor | N | Mean | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) |
| A | 18 | 358.1 | 22.9 (17.2-34.3) | 6.4 (4.8-9.6) | 11.2 (8.4-16.8) | 3.1 (2.4-4.7) | 21.9 (16.5-32.9) | 6.1 (4.6-9.2) | 33.5 (25.1-50.2) | 9.4 (7.0-14.0) |
| B | 17 | 484.1 | 36.3 (27.0-55.2) | 7.5 (5.6-11.4) | 37.4 (27.9-57.0) | 7.7 (5.7-11.7) | 28.1 (21.0-42.8) | 5.8 (4.3-8.8) | 57.6 (42.9-87.7) | 11.9 (8.8-18.1) |
| C | 18 | 153.5 | 5.2 (3.9-7.8) | 3.4 (2.6-5.1) | 7.6 (5.7-11.5) | 5.0 (3.7-7.5) | 10.0 (7.5-15.0) | 6.5 (4.9-9.8) | 14.7 (11.0-22.0) | 9.6 (7.2-14.4) |
| D | 18 | 275.1 | 12.3 (9.2-18.4) | 4.5 (3.3-6.7) | 14.2 (10.7-21.3) | 5.2 (3.9-7.7) | 19.1 (14.3-28.6) | 6.9 (5.2-10.4) | 29.1 (21.8-43.6) | 10.6 (7.9-15.9) |
| E | 18 | 402.0 | 26.4 (19.8-39.6) | 6.6 (4.9-9.9) | 16.1 (12.1-24.1) | 4.0 (3.0-6.0) | 20.3 (15.2-30.4) | 5.0 (3.8-7.6) | 34.7 (26.0-52.0) | 8.6 (6.5-12.9) |
Table 11 demonstrates that the 3mL TVT lots were sufficiently precise for the CVs were less than 10% and 15%, respectively).
| CD19 | Between-Site | Between-Tube | Between-Replicate | Total | ||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Donor | N | Mean | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) | SD (95% CI) | % CV (95% CI) |
| A | 18 | 351.7 | 25.5 (19.1-38.2) | 7.2 (5.4-10.8) | 1.9 (1.4-2.8) | 0.5 (0.4-0.8) | 5.2 (3.9-7.8) | 1.5 (1.1-2.2) | 31.2 (23.4-46.8) | 8.9 (6.7-13.3) |
| B | 17 | 141.9 | 7.5 (5.6-11.5) | 5.4 (4.0-8.1) | 6.2 (4.6-9.4) | 4.4 (3.3-6.6) | 8.8 (6.5-13.4) | 6.2 (4.6-9.5) | 14.2 (10.6-21.6) | 10.1 (7.5-15.3) |
| C | 18 | 53.7 | 0.5 (0.3-0.7) | 0.8 (0.6-1.3) | 0.9 (0.7-1.3) | 1.7 (1.3-2.5) | 3.2 (2.4-4.8) | 6.0 (4.5-9.0) | 4.0 (3.0-6.0) | 7.4 (5.6-11.2) |
| D | 18 | 90.4 | 6.2 (4.7-9.3) | 6.9 (5.2-10.3) | 5.9 (4.4-8.8) | 6.5 (4.9-9.8) | 5.9 (4.4-8.9) | 6.6 (4.9-9.8) | 10.8 (8.1-16.2) | 11.9 (9.0-17.9) |
| E | 18 | 97.7 | 6.2 (4.6-9.2) | 6.3 (4.7-9.4) | 0.1 (0.1-0.2) | 0.1 (0.1-0.2) | 8.9 (6.7-13.3) | 9.1 (6.8-13.6) | 10.9 (8.2-16.3) | 11.2 (8.4-16.7) |
Table 12 demonstrates that the 3mLTVT lots were sufficiently precise for the CD19 biomarker (individual and otal %CVs were less than 10% and 15%, respectively).
{11}------------------------------------------------
Results showed that after evaluating the TVT investigational product for Between-Site, Between-Tube, and Between-Replicate variability across all biomarkers, it was sufficiently precise.
Equivalence- Method Comparisons against Becton Dickinson EDTA (BD EDTA Day 0, control) and predicate
A clinical study validated the stability of lymphocyte subsets (CD3, CD4, CD8, CD16/56, CD19 and CD45) in TVT tubes. Samples were collected 30 HIV positive donors and 12 normal donors into BD EDTA tubes (reference condition). TVT tubes and BCT tubes. All tube types underwent testing for all markers on Day 0 (within 6 hours of collection), and TVT and BCT tubes were assayed on Days 11 and 15 post collection. All samples at all time points were analyzed by flow cytometry using a single-platform method. Passing -Bablok regression was performed to assess the relationship between the measurement for each marker from each time point and the control BD EDTA tube.
{12}------------------------------------------------
Summary for HIV positive subjects as compared to control (n=30) (Absolute counts) This analysis is BCT and TVT vs BD for each marker.
| Time | Passing-Bablok Regression | |||||
|---|---|---|---|---|---|---|
| Marker | Tube | Interval | Intercept | Slope | 95% CI for Slope | R |
| CD3 | BCT | 6 hour | -3.4 | 0.99 | 0.9478 to 1.039 | 1.00 |
| TVT | 6 hour | -18.1 | 1.05 | 0.9773 to 1.142 | 0.99 | |
| BCT | 11 day | 28.9 | 1.01 | 0.9071 to 1.074 | 0.99 | |
| TVT | 11 day | -31.7 | 1.05 | 0.9493 to 1.207 | 0.98 | |
| BCT | 15 day | 52.2 | 0.95 | 0.8882 to 1.034 | 0.99 | |
| TVT | 15 day | -13.8 | 1.04 | 0.9425 to 1.154 | 0.97 | |
| CD4 | BCT | 6 hour | 5.3 | 0.96 | 0.9325 to 1.013 | 1.00 |
| TVT | 6 hour | 0.6 | 1.02 | 0.9636 to 1.089 | 0.99 | |
| BCT | 11 day | 1.9 | 0.96 | 0.9076 to 1.024 | 0.99 | |
| TVT | 11 day | 11.5 | 0.97 | 0.9159 to 1.027 | 0.99 | |
| BCT | 15 day | 2.4 | 0.99 | 0.9222 to 1.027 | 0.99 | |
| TVT | 15 day | 2.1 | 0.98 | 0.8558 to 1.067 | 0.98 | |
| CD8 | BCT | 6 hour | -12.2 | 1.02 | 0.9467 to 1.074 | 1.00 |
| TVT | 6 hour | -60.5 | 1.12 | 1.0481 to 1.196 | 1.00 | |
| BCT | 11 day | -0.3 | 1.04 | 0.9398 to 1.142 | 0.99 | |
| TVT | 11 day | -26.3 | 1.08 | 0.9823 to 1.191 | 0.98 | |
| BCT | 15 day | 47.9 | 0.97 | 0.8793 to 1.055 | 0.99 | |
| TVT | 15 day | 4.1 | 1.03 | 0.8790 to 1.179 | 0.97 | |
| CD16/56 | BCT | 6 hour | 3.8 | 1.08 | 0.9451 to 1.189 | 0.98 |
| TVT | 6 hour | 1.5 | 1.05 | 0.9278 to 1.172 | 0.96 | |
| BCT | 11 day | 4.1 | 1.10 | 0.9681 to 1.205 | 0.97 | |
| TVT | 11 day | 10.3 | 0.96 | 0.7929 to 1.118 | 0.95 | |
| BCT | 15 day | 20.3 | 0.96 | 0.8554 to 1.090 | 0.96 | |
| TVT | 15 day | 15.7 | 0.95 | 0.8079 to 1.139 | 0.96 | |
| CD19 | BCT | 6 hour | 0.4 | 1.01 | 0.9735 to 1.072 | 1.00 |
| TVT | 6 hour | 3.2 | 1.02 | 0.9361 to 1.078 | 0.99 | |
| BCT | 11 day | -5.8 | 1.05 | 0.9779 to 1.109 | 0.99 | |
| TVT | 11 day | -4.8 | 1.07 | 0.9437 to 1.149 | 0.98 | |
| BCT | 15 day | -2.2 | 0.96 | 0.8950 to 1.031 | 0.99 | |
| TVT | 15 day | -5.8 | 1.05 | 0.9489 to 1.123 | 0.99 | |
| CD45 | BCT | 6 hour | 13.6 | 1.00 | 0.9501 to 1.049 | 1.00 |
| TVT | 6 hour | -10.5 | 1.07 | 0.9978 to 1.153 | 0.99 | |
| BCT | 11 day | 19.3 | 1.02 | 0.9435 to 1.101 | 0.99 | |
| TVT | 11 day | -32.5 | 1.05 | 0.9718 to 1.167 | 0.98 | |
| BCT | 15 day | 9.4 | 1.00 | 0.9384 to 1.076 | 0.99 | |
| TVT | 15 day | 5.3 | 1.05 | 0.9232 to 1.137 | 0.97 |
{13}------------------------------------------------
| Marker | Tube | Time Interval | Mean | Min | Max |
|---|---|---|---|---|---|
| CD3 | BD | 6 hour | 1415.9 | 226.3 | 3741.1 |
| CD4 | BD | 6 hour | 520.3 | 12.5 | 1142.4 |
| CD8 | BD | 6 hour | 878.0 | 217.4 | 3017.8 |
| CD16/56 | BD | 6 hour | 227.6 | 33.2 | 656.6 |
| CD19 | BD | 6 hour | 199.7 | 0.2 | 520.7 |
| CD45 | BD | 6 hour | 1853.7 | 298.6 | 4556.0 |
Dynamic range summary for HIV positive subjects (n=30) (Absolute counts):
{14}------------------------------------------------
| Marker | Tube | Time Interval | Intercept | Slope | 95% CI for Slope | R |
|---|---|---|---|---|---|---|
| CD3 | BCT | 6 hour | 75.9 | 0.92 | 0.7386 to 1.048 | 0.99 |
| TVT | 6 hour | -14.3 | 1.03 | 0.9415 to 1.226 | 0.99 | |
| BCT | 11 day | 189.6 | 0.83 | 0.6452 to 0.965 | 0.99 | |
| TVT | 11 day | -178.6 | 1.10 | 1.0392 to 1.445 | 0.99 | |
| BCT | 15 day | 38.9 | 0.95 | 0.7709 to 1.107 | 0.99 | |
| TVT | 15 day | -69.5 | 1.05 | 0.9630 to 1.221 | 0.99 | |
| CD4 | BCT | 6 hour | 33.4 | 0.94 | 0.8592 to 1.131 | 1.00 |
| TVT | 6 hour | 19.0 | 0.97 | 0.8749 to 1.209 | 0.99 | |
| BCT | 11 day | 123.9 | 0.82 | 0.7653 to 0.925 | 0.99 | |
| TVT | 11 day | -76.7 | 1.04 | 0.9407 to 1.212 | 0.99 | |
| BCT | 15 day | 66.0 | 0.88 | 0.8084 to 1.180 | 0.99 | |
| TVT | 15 day | -28.4 | 1.05 | 0.9592 to 1.303 | 0.99 | |
| CD8 | BCT | 6 hour | 66.0 | 0.86 | 0.6987 to 0.976 | 0.99 |
| TVT | 6 hour | 20.5 | 0.99 | 0.5831 to 1.176 | 0.97 | |
| BCT | 11 day | 35.3 | 0.91 | 0.7903 to 1.021 | 0.99 | |
| TVT | 11 day | 32.8 | 0.90 | 0.4914 to 1.015 | 0.97 | |
| BCT | 15 day | 37.2 | 0.98 | 0.8106 to 1.090 | 0.98 | |
| TVT | 15 day | 48.1 | 0.90 | 0.4299 to 1.032 | 0.98 | |
| CD16/56 | BCT | 6 hour | 23.4 | 0.95 | 0.5042 to 1.117 | 0.93 |
| TVT | 6 hour | -23.0 | 1.10 | 0.6485 to 1.403 | 0.92 | |
| BCT | 11 day | -24.4 | 1.23 | 0.6715 to 1.528 | 0.90 | |
| TVT | 11 day | 6.0 | 0.96 | 0.6636 to 1.311 | 0.89 | |
| BCT | 15 day | -30.2 | 1.19 | 0.8838 to 1.363 | 0.92 | |
| TVT | 15 day | -11.7 | 1.00 | 0.7030 to 1.377 | 0.90 | |
| CD19 | BCT | 6 hour | -4.1 | 1.06 | 0.8942 to 1.233 | 0.99 |
| TVT | 6 hour | 14.4 | 1.00 | 0.5285 to 1.238 | 0.98 | |
| BCT | 11 day | -1.6 | 0.97 | 0.8060 to 1.451 | 0.95 | |
| TVT | 11 day | 2.7 | 0.99 | 0.7014 to 1.248 | 0.98 | |
| BCT | 15 day | -0.3 | 0.94 | 0.7719 to 1.258 | 0.97 | |
| TVT | 15 day | -5.0 | 1.08 | 0.8658 to 1.369 | 0.99 | |
| CD45 | BCT | 6 hour | 131.5 | 0.92 | 0.5996 to 1.018 | 0.98 |
| TVT | 6 hour | -78.1 | 1.05 | 0.8239 to 1.354 | 0.98 | |
| BCT | 11 day | 295.4 | 0.82 | 0.5367 to 1.248 | 0.96 | |
| TVT | 11 day | -322.8 | 1.14 | 0.9225 to 1.427 | 0.97 | |
| BCT | 15 day | 137.9 | 0.91 | 0.7036 to 1.095 | 0.97 | |
| TVT | 15 day | 42.4 | 1.04 | 0.7490 to 1.293 | 0.98 |
Summary for Healthy Subjects (n=12) (Absolute counts)-This analysis is BCT and TVT vs BD for each marker.
{15}------------------------------------------------
| Marker | Tube | Time Interval | Mean | Min | Max |
|---|---|---|---|---|---|
| CD3 | BD | 6 hour | 1493.1 | 772.2 | 2819.4 |
| CD4 | BD | 6 hour | 953.6 | 493.8 | 2068.0 |
| CD8 | BD | 6 hour | 522.1 | 220.5 | 990.6 |
| CD16/56 | BD | 6 hour | 279.3 | 113.7 | 497.3 |
| CD19 | BD | 6 hour | 211.3 | 122.0 | 428.0 |
| CD45 | BD | 6 hour | 2009.0 | 1453.2 | 3452.4 |
Dynamic range summary for Healthy Subjects (n=12) (Absolute counts):
Interference
One site recruited five healthy subjects (individuals with no known disease process and ≥21 vears of age) and five HIV positive subjects. From each subject, two peripheral blood samples were collected in 4mL K3 EDTA Vacutainers. Immediately after collection, the 2x 4mL blood samples were split into seven 1mL aliquots.
- Aliquot 1) was treated as a control (no interfering substance). The other aliquots were manipulated as follows:
- Aliquot 2) had 2 µg/ml (0.2 mg/dL) bilirubin added to it. ●
- Aliquot 3) had 200 ug/ml (20 mg/dL) bilirubin added to it.
- Aliquot 4) had 0.025 g/ml (2.5 g/dL) hemoglobin added to it.
- Aliquot 5) had 0.035 g/ml (3.5 g/dL) hemoglobin added to it. ●
- Aliquot 6) had 1.5 mg/ml (150 mg/dL) triglycerides added to it. ●
- Aliquot 7) had 5 mg/ml (500 mg/dL) triglycerides added to it. ●
The aliquots were then stabilized using the appropriate amount of TransFix following the manufacturer's instructions, i.e., 0.2 mL TransFix administered to the 1mL aliquot, and were then stored at 2-8°C between test days. Aliquots were analyzed for a full lymphocyte count on Days 0, 11, and 15. The data indicated that there were no significant effects from the potential interferents listed above, as the results from testing Aliquots 2-7 were similar to results from the testing of Aliquot 1 (control).
807.92 (b)(3): Conclusions from Nonclinical and Clinical Testing
The TVT tubes were found to be safe and effective for the intended use.
§ 862.1675 Blood specimen collection device.
(a)
Identification. A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.(b)
Classification. Class II.