K Number
K080552
Device Name
CYTO-CHEX BCT
Manufacturer
Date Cleared
2008-07-31

(154 days)

Product Code
Regulation Number
862.1675
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Cyto-Chex® BCT™ is intended for collection and storage of blood specimens for immunophenotyping of WBC by flowcytometry. Recovery of lymphocyte subset cell markers of the HIV panel can be accomplished over a 14-day period following collection.

Device Description

Cyto-Chex BCT consists of a standard 13 x 75mm glass blood collection tube containing 75.8ul of sterile KJEDTA anti-coagulant and WBC preservative. It is manufactured with a vacuum to draw 5ml of blood by venipuncture.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that demonstrates the device meets these criteria, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
Clinically important HIV markers (CD3, CD4, and CD8) are stable in samples collected in Cyto-Chex BCT.Confirmed. CD3, CD4, and CD8 showed agreement with reference values (fresh K2EDTA samples within 6 hours).
Agreement of values with fresh samples: trendline slope better than 0.90 (preferably 0.95), R^2 values of 0.85 or better.Values for CD3, CD4, and CD8 from Cyto-Chex BCT stabilized samples confirmed through calculated correlation statistics meeting acceptance criteria. (Specific slope and R^2 values are not explicitly given, but the text states they meet the criteria).
Recovery of CD3, CD4 and CD8 markers well within the acceptance criteria for absolute cell count.Confirmed in both HIV-positive and healthy donors.
Preservation of HIV markers for a 14-day period following collection.Confirmed for CD3, CD4, and CD8 in both HIV-positive and healthy donors.
CD19 stability.Showed "some deterioration at 14 days" but the report suggests "additional patient sampling would probably improve the results." (This indicates it may not have fully met the criteria or was a borderline finding that Streck believes could be improved with more data, but was deemed acceptable for the intended use of CD3/CD4/CD8).

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: The text states "multiple healthy donors" and "samples from HIV positive patients." Specific numerical sample sizes for either group are not provided.
  • Data Provenance: Not explicitly stated, but it implies a prospective clinical study where fresh blood samples were collected and then split into different tubes and analyzed over time. The mention of "resource-poor countries" in the conclusion suggests a global relevance, but doesn't specifically state the origin of the study participants. The study tested both healthy and HIV-positive individuals.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • Number of Experts: Not mentioned.
  • Qualifications of Experts: Not mentioned. The ground truth was established by laboratory analysis using flow cytometry, a standardized method. The expertise would lie in the technicians performing and interpreting these tests, but their specific qualifications are not detailed.

4. Adjudication Method for the Test Set:

  • Adjudication Method: Not applicable. This device is a blood collection tube for sample preservation, not an AI diagnostic tool requiring expert adjudication of image interpretations or similar subjective assessments. The "ground truth" was derived from direct laboratory measurements (flow cytometry results of fresh samples).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

  • MRMC Study: No. This is not an imaging or diagnostic device that involves human interpretation of results. Therefore, an MRMC study is not relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

  • Standalone Performance: No. This is a medical device (blood collection tube) that is designed to preserve biological samples. Its performance is assessed by how well it maintains the integrity of the sample, not as an algorithm performing a task autonomously.

7. The Type of Ground Truth Used:

  • Type of Ground Truth: The ground truth was established by direct laboratory measurement from fresh samples (specifically, flow cytometric data for lymphocyte subset cell-surface markers from K2EDTA blood collection tubes analyzed within 6 hours of draw). This serves as the "reference value" against which the Cyto-Chex BCT preserved samples were compared.

8. The Sample Size for the Training Set:

  • Sample Size for Training Set: Not applicable. This is not an AI/ML device that requires training. The device's performance is based on its physical properties and chemical composition to preserve blood samples.

9. How the Ground Truth for the Training Set Was Established:

  • Ground Truth for Training Set: Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Konoss2

1 2008

510(k) Summary of Safety and Effectiveness

510(k) Submitter:Streck7002 South 109th StreetOmaha, NE 68128
Official Correspondent:Carol Thompson, Quality Assurance Manager(402) 537-5313
Date Prepared:July 14, 2008
Names of Device:Trade Name:Common Name:Classification Name:Cyto-Chex® BCTBlood specimen collection deviceBlood specimen collection device, 21CFR862.1675
Predicate Device:Cyto-Chex® BCT, K040107

Description:

Cyto-Chex BCT consists of a standard 13 x 75mm glass blood collection tube containing 75.8ul of sterile KJEDTA anti-coagulant and WBC preservative. It is manufactured with a vacuum to draw 5ml of blood by venipuncture.

Intended Use:

Cyto-Chex BCT is intended for the collection and storage of blood specimens for immunophenotyping of WBC by flow-cytometry. Recovery of lymphocyte subset cell markers of the HIV panel can be accomplished over a 14-day period following collection.

Comparison with Predicate Device:

The New Cyto-Chex BCT is identical with Cyto-Chex BCT with one difference. The volume of reagent in each tube is 75.8ul instead of 57ul. The concentration of the active agent in the collected blood sample is the same in both tubes. It is used in exactly the same manner as the original Cyto-Chex BCT. Preservation of HIV markers has been established to 14 days from collection instead of 7 days in the original Cyto-Chex BCT.

Testing Performed:

Flow cytometric data for lymphocyte subset cell-surface markers was obtained by analysis of peripheral blood samples collected from multiple healthy donors. A separate clinical study was set up in which samples from HIV positive patients were collected and analyzed. Samples were collected in both KyEDTA blood collection tubes and Cyto-Chex BCT tubes. Testing was performed over a period of 14 days using both Becton-Dickinson FACSCalibur and Beckman Coulter EPICS XL flow cytometers. Results were compared to those from fresh samples (6 hours after draw in KzEDTA). Testing was also performed to establish stability of Cyto-Chex BCT reagent and to verify that under filling the tube would not compromise results.

Conclusions Drawn from the Tests:

Clinically important HIV markers (CD3, CD4 and CD8) are shown to be stable in samples collected in Cyto-Chex BCT. The results obtained in the EDTA tube within 6 hours represent the reference values for all comparisons. The agreement of these values with those recovered from the Cyto-Chex BCT stabilized samples is

confirmed through the calculated correlation statistics. Acceptable stability is marked by a trendline with a slope better than 0.90, preferably 0.95. Re values of 0.85 or better confirm consistency of results.

In both HIV positive and healthy donors, markers CD3, CD4 and CD8 are recovered well within the acceptance criteria for absolute cell count. CD19 shows some deterioration at 14 days but additional patient sampling would probably improve the results.

HIV positive blood specimens collected in Cyto-Chex BCT can be stored/transported for analysis of HIV markers CD3, CD4 and CD8 as long as 14 days after collection. This will provide a benefit where specimens need to be transported over long distances for analysis such as in resource-poor countries.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three horizontal lines above it, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Streck c/o Kerrie Oetter Quality Assurance Coordinator 7002 South 109th Street Omaha, NE 68128

Re: K080552

Trade/Device Name: Cyto-Check BCT Regulation Number: 21 CFR 862.1675 Regulation Name: Blood specimen collection device Regulatory Class: Class II Product Code: GIM Dated: July 14, 2008 Received: July 16, 2008

Dear Ms. Oetter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in intestate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, youg, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, listo recei devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

JUL 3 1 2008

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements aft the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing pratis f requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket

{2}------------------------------------------------

Page 2 - Streck

notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-041), preso, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRice of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may blain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 633-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html .

Sincerely yours.

Mama In Chan

Maria M. Chan, Ph.D. Acting Division Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indication for Use

510(k) Number (if known): 长050552

Ko80552

Device Name: Cyto-Chex® BCTTM

Indication For Use:

Cyto-Chex® BCT™ is intended for collection and storage of blood specimens for immunophenotyping of WBC by flowcytometry. Recovery of lymphocyte subset cell markers of the HIV panel can be accomplished over a 14-day period following collection.

Prescription Use X (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use __ (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)

Josephine Bautista
Division Sign Off

Division Sig Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K080552

§ 862.1675 Blood specimen collection device.

(a)
Identification. A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.(b)
Classification. Class II.