K Number
K162287
Date Cleared
2017-04-06

(234 days)

Product Code
Regulation Number
892.2050
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Synapse 3D Perfusion Analysis is medical imaging software used with Synapse 3D Base Tools that is intended to provide trained medical professionals, with tools to aid them in reading, interpreting, and treatment planning. Synapse 3D Perfusion Analysis accepts DICOM compliant medical images acquired from CT and MR. This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.

Addition to Synapse 3D Base Tools, Synapse 3D Perfusion Analysis provides the parameter images by post-processing of dynamic scanned CT arteriography and magnetic resonance (MR) images acquired with contrast agents to aid the assessment of cerebral (CT and MR), myocardial (CT) and abdomen (CT) blood flows. The parameter images are Blood Volume (BV), Blood Flow (BF), Mean Transit Time (MTT), and Time To Peak (TTP),

Device Description

Synapse 3D Perfusion Analysis is medical device software for Perfusion Analysis that provides the parameter images by post-processing of dynamic scanned CT arteriography and MR images acquired with contrast agents to aid the assessment of cerebral (CT and MR), myocardial (CT) and abdomen (CT) blood flows. The parameter images are Blood Volume (BV), Blood Flow (BF), Mean Transit Time (MTT), and Time to Peak (TTP). The perfusion parameter images can be presented to the trained medical professionals as the time-density curves (TDC) and perfusion characteristics maps (in parametric and summary) to assist them in assessing the blood flows. Synapse 3D Perfusion Analysis includes the following perfusion analysis applications.

  • (1) Brain Perfusion (CT and MR) (unchanged from the FDA-cleared version K120637) is an application that analyzes the cerebral blood flow using the contrast-enhanced dynamic scanned CT and MR cerebral arteriography images. The Brain Perfusion post-processes the cerebral arteriography CT and MR images, and generates the parameter images of Cerebral Blood Volume (CBV), Cerebral Blood Flow (CBF), MTT, and TTP.
  • (2) 4D Perfusion (CT) is an application that analyzes the changes in the cerebral blood flow over time (in 4D) using the contrast-enhanced multi-phase 3D whole-brain images. The 4D Perfusion post-processes the cerebral arteriography CT and MR images, and generates the 4D (over time) parameter images of CBV, CBF, MTT, and TTP.
  • (3) Abdominal Perfusion (CT) is an application that analyzes the blood flow of abdominal organs over time (in 4D) using the contrast-enhanced multi-phase 3D abdomen images. The Abdominal Perfusion post-processes the abdominal CT images, and generates the parameter images of Tissue Blood Volume (TBV), Tissue Blood Flow (TBF), MTT, and TTP.
  • (4) Cardiac Perfusion (CT) is an application that analyzes the myocardial blood flow using the multi-phase 3D heart images. The Cardiac Perfusion post-processes the contrast-enhanced Myocardial CT images, and generates the parameter images of Myocardial Blood Volume (MBV), Myocardial Blood Flow (MBF), MTT, and TTP.

Synapse 3D Perfusion Analysis runs on Windows standalone and server/client configuration installed on a commercial general-purpose Windows-compatible computer. Synapse 3D Perfusion Analysis is an optional software module that works with Synapse 3D Base Tools (cleared by CDRH via K120361 on 04/06/2012) which is connected through DICOM standard to medical devices such as CT, MR, CR, US, NM, PT, XA, etc. and to a PACS system storing data generated by these medical devices, and it retrieves image data via network communication based on the DICOM standard. The retrieved image data are stored on the local disk managed by Synapse 3D Base Tools, and the associated image-related information of the image data is registered in the database and is used for display, image processing, analysis, etc. The software can display the images on a display monitor, or printed them on a hardcopy using a DICOM printer or a Windows printer.

AI/ML Overview

The provided text describes the Fujifilm Synapse 3D Perfusion Analysis device and its 510(k) submission. Here's a breakdown of the acceptance criteria and the study information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Error RateThe error is within ± 1%.

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size: Not explicitly stated as a number of cases or patients. The document refers to "all of the test cases" but does not quantify them.
  • Data Provenance: "actual clinical images" were used for benchmark performance testing. No specific country of origin or whether it was retrospective or prospective is mentioned.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not explicitly stated. The ground truth ("Reference data (Pt)") was derived from a perfusion calculation model using arbitrary CT signal values, not directly from expert consensus on clinical images in the performance testing section.

4. Adjudication method for the test set

Not applicable. The ground truth for performance testing was established computationally through a perfusion calculation model, not through expert adjudication of cases.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study is mentioned. The submission focuses on the performance of the algorithm itself.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, a standalone performance test was done. The "benchmark performance testing" calculated the "Result data (Pc)" by Synapse 3D Perfusion Analysis using input data (Pi) and compared it to "Reference data (Pt)" derived from a perfusion calculation model. This evaluates the algorithm's accuracy in isolation.

7. The type of ground truth used

For the benchmark performance testing, the ground truth (Reference data (Pt)) was derived from a perfusion calculation model using input data (Pi: an image created using an arbitrary CT signal value).

8. The sample size for the training set

Not mentioned in the provided text.

9. How the ground truth for the training set was established

Not mentioned in the provided text.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The seal is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

FUJIFILM Medical Systems U.S.A., Inc. % Jyh-Shyan Lin, Ph.D. Senior Manager, Regulatory, Quality and Clinical Affairs 419 West Avenue STAMFORD CT 06902

Re: K162287

Trade/Device Name: Synapse 3D Perfusion Analysis Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ, JAK Dated: March 13, 2017 Received: March 15, 2017

Dear Dr. Lin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

April 6, 2017

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162287

Device Name Synapse 3D Perfusion Analysis

Indications for Use (Describe)

Synapse 3D Perfusion Analysis is medical imaging software used with Synapse 3D Base Tools that is intended to provide trained medical professionals, with tools to aid them in reading, interpreting, and treatment planning. Synapse 3D Perfusion Analysis accepts DICOM compliant medical images acquired from CT and MR. This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.

Addition to Synapse 3D Base Tools, Synapse 3D Perfusion Analysis provides the parameter images by post-processing of dynamic scanned CT arteriography and magnetic resonance (MR) images acquired with contrast agents to aid the assessment of cerebral (CT and MR), myocardial (CT) and abdomen (CT) blood flows. The parameter images are Blood Volume (BV), Blood Flow (BF), Mean Transit Time (MTT), and Time To Peak (TTP),

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

FUJIFILM MEDICAL SYSTEMS U.S.A., Inc.

Synapse 3D Perfusion Analysis

510(k) Summary

Date Prepared:

March 11, 2017

Submitter's Information:

FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, Connecticut 06902 Telephone: (301) 251-1092 Fax: (203) 602-3785 Contact: Jyh-Shyan Lin

Device Trade Name:

Synapse 3D Perfusion Analysis

Device Common Names:

Picture Archiving and Communications System (PACS)

Device Classification Name:

System, Image Processing, Radiological

Product Code:

LLZ

Regulation Number:

21 CFR 892.2050

Device Class:

Class II

Panel:

Radiology

Predicate Devices:

  • Synapse 3D Brain Perfusion (K120637), FUJIFILM Medical Systems U.S.A., Inc. o
  • syngo CT Myocardial Perfusion (K150713), Siemens Medical Solutions USA, Inc. ●
  • syngo Volume Perfusion-CT Body (K092013), Siemens Medical Solutions USA, Inc. o

{4}------------------------------------------------

FUJIFILLM MEDICAL SYSTEMS U.S.A., Inc.

Synapse 3D Perfusion Analysis

1. Description of the Device

Synapse 3D Perfusion Analysis is medical device software for Perfusion Analysis that provides the parameter images by post-processing of dynamic scanned CT arteriography and MR images acquired with contrast agents to aid the assessment of cerebral (CT and MR), myocardial (CT) and abdomen (CT) blood flows. The parameter images are Blood Volume (BV), Blood Flow (BF), Mean Transit Time (MTT), and Time to Peak (TTP). The perfusion parameter images can be presented to the trained medical professionals as the time-density curves (TDC) and perfusion characteristics maps (in parametric and summary) to assist them in assessing the blood flows. Synapse 3D Perfusion Analysis includes the following perfusion analysis applications.

  • (1) Brain Perfusion (CT and MR) (unchanged from the FDA-cleared version K120637) is an application that analyzes the cerebral blood flow using the contrast-enhanced dynamic scanned CT and MR cerebral arteriography images. The Brain Perfusion post-processes the cerebral arteriography CT and MR images, and generates the parameter images of Cerebral Blood Volume (CBV), Cerebral Blood Flow (CBF), MTT, and TTP.
  • (2) 4D Perfusion (CT) is an application that analyzes the changes in the cerebral blood flow over time (in 4D) using the contrast-enhanced multi-phase 3D whole-brain images. The 4D Perfusion post-processes the cerebral arteriography CT and MR images, and generates the 4D (over time) parameter images of CBV, CBF, MTT, and TTP.
  • (3) Abdominal Perfusion (CT) is an application that analyzes the blood flow of abdominal organs over time (in 4D) using the contrast-enhanced multi-phase 3D abdomen images. The Abdominal Perfusion post-processes the abdominal CT images, and generates the parameter images of Tissue Blood Volume (TBV), Tissue Blood Flow (TBF), MTT, and TTP.
  • (4) Cardiac Perfusion (CT) is an application that analyzes the myocardial blood flow using the multi-phase 3D heart images. The Cardiac Perfusion post-processes the contrast-enhanced Myocardial CT images, and generates the parameter images of Myocardial Blood Volume (MBV), Myocardial Blood Flow (MBF), MTT, and TTP.

Synapse 3D Perfusion Analysis runs on Windows standalone and server/client configuration installed on a commercial general-purpose Windows-compatible computer. Synapse 3D Perfusion Analysis is an optional software module that works with Synapse 3D Base Tools (cleared by CDRH via K120361 on 04/06/2012) which is connected through DICOM standard to medical devices such as CT, MR, CR, US, NM, PT, XA, etc. and to a PACS system storing data generated by these medical devices, and it retrieves image data via network communication based on the DICOM standard. The retrieved image data are stored on the local disk managed by Synapse 3D Base Tools, and the associated image-related information of the image data is registered in the database and is used for display, image processing, analysis, etc. Synapse 3D Perfusion Analysis can handle images of CT and MR. The software can display the images on a display monitor, or printed them on a hardcopy using a DICOM printer or a Windows printer.

{5}------------------------------------------------

FUJIFILLM MEDICAL SYSTEMS U.S.A., Inc.

Synapse 3D Perfusion Analysis

2. Indications for Use

Synapse 3D Perfusion Analysis is medical imaging software used with Synapse 3D Base Tools that is intended to provide trained medical professionals, with tools to aid them in reading, interpreting, reporting, and treatment planning. Synapse 3D Perfusion Analysis accepts DICOM compliant medical images acquired from CT and MR. This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.

Addition to Synapse 3D Base Tools, Synapse 3D Perfusion Analysis provides the parameter images by post-processing of dynamic scanned CT arteriography and magnetic resonance (MR) images acquired with contrast agents to aid the assessment of cerebral (CT and MR), myocardial (CT) and abdomen (CT) blood flows. The parameter images are Blood Volume (BV), Blood Flow (BF), Mean Transit Time (MTT), and Time To Peak (TTP).

3. Substantial Equivalence Comparison

Synapse 3D Perfusion Analysis has the same intended use, similar labeling, and clinical application tools as those of the cleared predicate devices. It assesses the blood flows in the same intended body parts of the predicate devices (Table 1).

Subject body partsPredicate devices510(k) #
Cerebral regionSynapse 3D Brain PerfusionK120637
Cardiac regionsyngo.CT Myocardial PerfusionK150713
Abdomen regionsyngo Volume Perfusion-CT BodyK092013

Table 1. Perfusion analysis in different body parts

Synapse 3D Perfusion Analysis provides the parameter images by post-processing of dynamic scanned CT arteriography and magnetic resonance (MR) images acquired with contrast agents to aid the assessment of cerebral (CT and MR), myocardial (CT) and abdomen (CT) blood flows. The parameter images are Blood Volume (BV), Blood Flow (BF), Mean Transit Time (MTT), and Time To Peak (TTP) (Table 2).

Table 2. Device features and technological characteristics comparison

Device ParametersSynapse 3DPerfusion Analysis(This submission)Synapse 3DBrain Perfusion(K120637)Syngo.CTMyocardial Perfusion(K150713)syngoVolume Perfusion-CT Body(K092013)
DICOM compliantCT images withcontrast imagehandlingYes*YesYesYes
DICOM compliantMR images withcontrast agenthandlingYesYesNoNo

{6}------------------------------------------------

J | FILM MEDICAL SYSTEMS U.S.A., Inc.

Synapse 3D Perfusion Analysis

Time-Signal CurveYesYesYesYes
Time to Peak imageYesYesYesUnknown
Parameter images:• Blood Volume• Blood Flow• Mean Transit TimeYesYesYesYes

*Yes: Equivalent features and technological characteristics

4. Safety Information

Synapse 3D Perfusion Analysis introduces no new safety or efficacy issues other than those already indentified with the predicate device. The Risk Management and the results of the Hazard Analysis combined with the appropriate preventive measures taken indicate that the device is of moderate concern as per the May 11, 2005 issue of the "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices." The Synapse 3D Perfusion Analysis labeling contains instructions for use and necessary cautions, warnings and notes to provide the safe and effective use of the device.

5. Testing and Performance Information

Synapse 3D Perfusion Analysis is tested successfully with reference to its product requirements, as well as design verification and validation documents and traceability matrix document. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the Synapse 3D Perfusion Analysis software, which is found to be safe and effective and substantially equivalent to the predicate devices.

Testing involved system level functionality test, component testing, verification testing, integration testing, usability testing, installation/upgrade testing, labeling testing, as well as the testing for risk mitigations associated with the risk management process. In addition, we conducted benchmark performance testing using actual clinical images to help demonstrate that the proposed device achieved the expected accuracy performance.

Pass/Fail criteria were based on the requirements and intended use of the product. Test results showed that all tests successfully passed.

The data used for performance testing are defined as follows.

  • . Input data (Pi): an image which is created using an arbitrary CT signal value.
  • Reference data (Pt): a value derived from perfusion calculation model using the input data (Pi). .
  • . Result data (Pc): a value calculated by Synapse 3D Perfusion Analysis using the input data (Pi).
  • Error: ((Pt Pc) / Pt)*(100%). o
  • Pass criteria: the error is within ± 1%. ●

Based on the above definitions for the performance testing data, all of the test cases were tested and the performance testing results (accuracy test results including uncertainties) were passed.

{7}------------------------------------------------

FUJIFILLM MEDICAL SYSTEMS U.S.A., Inc.

Synapse 3D Perfusion Analysis

6. Conclusion

This 510(k) premarket notification submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. We conclude the subject device to be as safe and effective as the predicate device.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).