K Number
K161928
Manufacturer
Date Cleared
2016-12-21

(160 days)

Product Code
Regulation Number
866.5510
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CAPI 3 IMMUNOTYPING kit is designed for the qualitative detection and the characterization of monoclonal proteins (immunotyping) in human serum with the CAPILLARYS 3 TERA instrument, SEBIA, for capillary electrophoresis. It is used in conjunction with the CAPI 3 PROTEIN(E) 6 kit, SEBIA, designed for proteins separation into 6 major fractions in alkaline buffer (pH 9.9).

The CAPILLARYS 3 TERA instrument performs all procedural sequences automatically to obtain a profile for qualitative analysis. Each serum sample is mixed with individual antisera that are specific against gamma (Ig G), alpha (Ig A) and mu (Ig M) heavy chains, and kappa (free and bound) light chains and lambda (free and bound) light chains, respectively.

The proteins, separated in silica capillaries, are directly detected by their absorbance at 200 mm.

The electrophoregrams are evaluated visually to detect the presence of specific reactions with the suspect monoclonal proteins.

For In Vitro Diagnostic Use.

The IT / IF Control is designed to qualitative detection and characterization of human monoclonal immunoglobulins (Ig G, Ig A, Ig M, Kappa and Lambda) with the electrophoresis methods :

  • Immunotyping performed using capillary electrophoresis on SEBIA CAPILLARYS 2, CAPILLARYS 2 FLEX-PIERCING and CAPILLARYS 3 TERA instruments and on SEBIA MINICAP instrument,

  • Immunofixation methods : SEBIA HYDRAGEL IF, HYDRAGEL IF Penta, HYDRAGEL BENCE JONES (Standard mask and Dynamic mask) performed using the HYDRASYS and HYDRASYS 2 instruments and the K20 electrophoresis chamber.

The IT / IF Control is designed for laboratory use. It should be used (with its barcode label for CAPILLARYS and MINICAP procedures) like a human serum sample.

The electrophoretic pattern obtained is specific for each batch of IT / IF control.

For In Vitro Diagnostic Use.

Device Description

The CAPILLARYS 3 TERA instrument uses the principle of capillary electrophoresis in liquid solution. With this technique, charged molecules are separated by their electrophoretic mobility in an alkaline buffer with a specific pH. Separation occurs according to the electrolyte pH and electroosmotic flow. The CAPILLARYS 3 TERA instrument has silica capillaries functioning in parallel allowing 12 simultaneous analyses.

In CAPI 3 MMUNUOTYPING kit contains the sample diluent and specific antisera against gamma (Ig G), alpha (Ig A), mu (Ig M) heavy chains, and free and bound Kappa and Lambda light chains. A sample dilution is prepared and injected at the anodic end of six capillaries. The reference pattern (ELP pattern), which is a complete electrophoretic pattern of the sample's proteins, is obtained by mixing the sample with the ELP solution and injection into the 1st capillary. The antisera patterns are obtained by sample aspiration into the 5 subsequent capillaries separation is performed in a high voltage electrical field and detected using absorbance at 200 nm.

The IT / IF control contains three monoclonal proteins which can be used as a qualitative control with the CAPI IMMUNOTYPING kit on the CAPILLARYS 3 TERA instrument.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state quantitative acceptance criteria in a dedicated table. However, the method comparison study provides the performance benchmark against the predicate device.

Acceptance Criteria CategoryAcceptance Criteria (Implicit)Reported Device Performance (CAPI 3 IMMUNOTYPING on CAPILLARYS 3 TERA)
Qualitative Detection & Characterization of Monoclonal Proteins100% agreement with predicate device (CAPILLARYS IMMUNOTYPING)100% complete agreement for both normal and pathological serum samples.
RepeatabilityConcordant results within run and between capillaries.All samples gave concordant results within run and between capillaries.
Reproducibility (between lots & instruments)Concordant results across different instruments and lots.All samples gave concordant results for all runs on 3 instruments and with 3 lots.
SensitivityAble to detect monoclonal components at clinically relevant concentrations.Detected Ig G, L at 30.9 mg/dL; Ig A, K at 13.3 mg/dL; Ig M, K at 25.0 mg/dL.
InterferenceNo interference from common interfering factors at specified levels.No interference detected from high concentrations of Triglycerides (3.59 g/dL), Bilirubin (20 mg/dL), Rheumatoid factor (981 IU/mL), and Hemoglobin (0.2 g/dL).

2. Sample Sizes Used for the Test Set and Data Provenance

  • Sample Size for Method Comparison (Test Set): 115 serum samples (12 normal, 103 pathological).
  • Data Provenance: Not explicitly stated as country of origin. The manufacturing information (Sebia, FRANCE) implies the device developer is French, but the source of the samples is not provided. The study is retrospective as it compares results from existing clinical samples.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document does not specify the number of experts or their qualifications for establishing ground truth. The comparison is made against a "predicate device" (CAPILLARYS IMMUNOTYPING procedure), which implies the predicate device's results are considered the reference or "ground truth" for this study. The visual evaluation of electrophoregrams is mentioned, which typically involves trained laboratory personnel or pathologists, but specifics are missing.

4. Adjudication Method

Not explicitly stated. The study reports 100% agreement, suggesting no significant discrepancies arose that would require an adjudication process.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly done based on the provided text. The study compares the device's performance to a predicate device, not human readers with and without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the method comparison study appears to be a standalone performance evaluation of the CAPI 3 IMMUNOTYPING procedure (algorithm + instrument system) against the predicate device's procedure. While the "electrophoregrams are evaluated visually," the 100% agreement is attributed to the techniques rather than individual human interpretations.

7. The Type of Ground Truth Used

The ground truth for the method comparison study was established by the results obtained from the predicate device (CAPILLARYS IMMUNOTYPING procedure). This means the predicate device's ability to detect and characterize monoclonal proteins served as the reference standard.

8. The Sample Size for the Training Set

The document does not mention a training set nor does it explicitly describe an AI/machine learning component that would require a distinct training set. The device appears to be a diagnostic instrument based on capillary electrophoresis, not an AI-driven interpretation tool.

9. How the Ground Truth for the Training Set Was Established

Since no training set or AI component is described, this information is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with three abstract shapes above them that resemble wings or flowing ribbons.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 21, 2016

SEBIA, INC. Karen Anderson, MT (ASCP) Director of Technical Training and Regulatory Suite 400, 1705 Corporate Drive Norcross, GA 30093

Re: K161928

Trade/Device Name: CAPI 3 IMMUNOTYPING, CAPILLARYS 3 TERA, IT/IF Control Regulation Number: 21 CFR § 866.5510 Regulation Name: Immunoglobulins (A, G, M, D, E) Immunological Test Systems Regulatory Class: Class II Product Code: CFF - Immunoelectrophoretic, Immunoglobulins (G, A, M) Dated: December 15, 2016 Received: December 19, 2016

Dear Ms. Karen Anderson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Image /page/1/Picture/8 description: The image shows the name "Kelly Oliner -S" in black font. The text is centered on the page and is easy to read. In the background, there is a faded blue logo that appears to be the letters "FDA".

For

Leonthena Carrington, MBA, MS, MT(ASCP) Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161928

Device Name CAPI 3 IMMUNOTYPING

Indications for Use (Describe)

The CAPI 3 IMMUNOTYPING kit is designed for the qualitative detection and the characterization of monoclonal proteins (immunotyping) in human serum with the CAPILLARYS 3 TERA instrument, SEBIA, for capillary electrophoresis. It is used in conjunction with the CAPI 3 PROTEIN(E) 6 kit, SEBIA, designed for proteins separation into 6 major fractions in alkaline buffer (pH 9.9).

The CAPILLARYS 3 TERA instrument performs all procedural sequences automatically to obtain a profile for qualitative analysis. Each serum sample is mixed with individual antisera that are specific against gamma (Ig G), alpha (Ig A) and mu (Ig M) heavy chains, and kappa (free and bound) light chains and lambda (free and bound) light chains, respectively.

The proteins, separated in silica capillaries, are directly detected by their absorbance at 200 mm.

The electrophoregrams are evaluated visually to detect the presence of specific reactions with the suspect monoclonal proteins.

For In Vitro Diagnostic Use.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K161928

Device Name

IT / IF Control

Indications for Use (Describe)

The IT / IF Control is designed to qualitative detection and characterization of human monoclonal immunoglobulins (Ig G, Ig A, Ig M, Kappa and Lambda) with the electrophoresis methods :

  • Immunotyping performed using capillary electrophoresis on SEBIA CAPILLARYS 2, CAPILLARYS 2 FLEX-PIERCING and CAPILLARYS 3 TERA instruments and on SEBIA MINICAP instrument,

  • Immunofixation methods : SEBIA HYDRAGEL IF, HYDRAGEL IF Penta, HYDRAGEL BENCE JONES (Standard mask and Dynamic mask) performed using the HYDRASYS and HYDRASYS 2 instruments and the K20 electrophoresis chamber.

The IT / IF Control is designed for laboratory use. It should be used (with its barcode label for CAPILLARYS and MINICAP procedures) like a human serum sample.

The electrophoretic pattern obtained is specific for each batch of IT / IF control.

For In Vitro Diagnostic Use.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510K SUMMARY (Summary of Safety and Effectiveness)

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

Submitter NameSebia, Inc.
Address1705 Corporate Drive Suite 400Norcross, Georgia 30093, USA
Karen Anderson, Dir of Technical and QAPhone: 1-800-835-6497Fax: 770-446-8511Email: karen.anderson@sebia-usa.com
ContactAigars Brants, Ph.D, Scientific Affairs OfficerPhone 1-800-835-6497Fax 770-446-8511Email: aigars.brants@sebia-usa.com
Date PreparedOctober 6, 2016 / Revision Dec 14, 2016
ManufacturingSebiaParc Technologique Léonard de VinciRue Léonard de Vinci,CP 8010 LISSES, 91008 EVRY CedexFRANCEPhone: (33) 1 69 89 80 80Fax: (33) 1 69 89 78 78
Product NameCAPI 3 IMMUNOTYPING (PN 2600)IT / IF Control (PN 4788)using CAPILLARYS 3 TERA instrument (PN 1246)
Common NameMonoclonal Immunoglobulins by CapillaryElectrophoresis
Product Regulation No.21 CFR Part 866.5510, 866.5550, 866.1630,862.1660
Product CodesCFF, DFH, DEH, CEF, JJY
Device classificationClass II (Test System), Class I (ControlsElectrophoretic Protein Fractionation)
Establishment Registration No.8023024

{5}------------------------------------------------

Predicate Device NamePredicate Device 510(k) number
CAPILLARYS IMMUNOTYPINGK130500
CAPILLARYS 2 / CAPILLARYS 2 FLEX-PIERCINGK130500
IT / IF CONTROLK130500

1. DEVICE DESCRIPTION

CAPILLARYS 3 TERA INSTRUMENT

The CAPILLARYS 3 TERA instrument uses the principle of capillary electrophoresis in liquid solution. With this technique, charged molecules are separated by their electrophoretic mobility in an alkaline buffer with a specific pH. Separation occurs according to the electrolyte pH and electroosmotic flow. The CAPILLARYS 3 TERA instrument has silica capillaries functioning in parallel allowing 12 simultaneous analyses.

CAPI 3 IMMUNOTYPING KIT

In CAPI 3 MMUNUOTYPING kit contains the sample diluent and specific antisera against gamma (Ig G), alpha (Ig A), mu (Ig M) heavy chains, and free and bound Kappa and Lambda light chains. A sample dilution is prepared and injected at the anodic end of six capillaries. The reference pattern (ELP pattern), which is a complete electrophoretic pattern of the sample's proteins, is obtained by mixing the sample with the ELP solution and injection into the 1st capillary. The antisera patterns are obtained by sample aspiration into the 5 subsequent capillaries separation is performed in a high voltage electrical field and detected using absorbance at 200 nm.

IT/IF CONTROL

The IT / IF control contains three monoclonal proteins which can be used as a qualitative control with the CAPI IMMUNOTYPING kit on the CAPILLARYS 3 TERA instrument.

{6}------------------------------------------------

Reagents:

CAPI 3 IMMUNOTYPING KIT

ITEMSPN 2600
Sample diluent (ready to use)1 vial, 60 mL
Pierceable cap for the Sample diluent vial1 cap
Rack with ELP solution and antiserum tubes
ELP solution (ready to use)1 vial, 1.2 mL
Mammalian immunoglobulins antihumangamma heavy chains (ready to use)1 vial, 1.2 mL
Mammalian immunoglobulins antihumanalpha heavy chains (ready to use)1 vial, 1.2 mL
Mammalian immunoglobulins antihumanmu heavy chains (ready to use)1 vial, 1.2 mL
Mammalian immunoglobulins antihumankappa (free and bound) light chains(ready to use)1 vial, 1.2 mL
Mammalian immunoglobulins antihumanlambda (free and bound) light chains(ready to use)1 vial, 1.2 mL

Additional reagents not included in the CAPI 3 IMMUNOTYPING KIT

ITEMSPNCOMPONENTS
CAPI 3 PROTEIN(E) 625033 vials buffer, 700 mL each4 filters
CAPICLEAN CAPILLARYS 320601 vial, 25 mL
CAPILLARYS 3 WASH SOLUTION20621 vial, 75mL
CAPI 3 DISPOSABLES KIT258010 packs of 14 reagent cups5 bins for used reagent cups
TEST TUBES9214200 of 100mm-tubes
CAPI 3 BINS FOR USED REAGENTCUPS25815 units

{7}------------------------------------------------

2. INDICATIONS FOR USE

The CAPI 3 IMMUNOTYPING kit is designed for the qualitative detection and the characterization of monoclonal proteins (immunotyping) in human serum with the CAPILLARYS 3 TERA instrument, SEBIA, for capillary electrophoresis. It is used in conjunction with the CAPI 3 PROTEIN(E) 6 kit, SEBIA, designed for proteins separation into 6 maior fractions in alkaline buffer (pH 9.9).

The CAPILLARYS 3 TERA instrument performs all procedural sequences automatically to obtain a protein profile for qualitative analysis. Each serum sample is mixed with individual antisera that are specific against gamma (lg G), alpha (lg A) and mu (lg M) heavy chains, and kappa (free and bound) light chains and lambda (free and bound) light chains, respectively. The proteins, separated in silica capillaries, are directly detected by their absorbance at 200 nm.

The electrophoregrams are evaluated visually to detect the presence of specific reactions with the suspect monoclonal proteins.

For In Vitro Diagnostic Use.

The IT / IF Control is designed to qualitative detection and characterization of human monoclonal immunoglobulins (lg G, lg A, lg M, Kappa and Lambda) with the electrophoresis methods:

Immunotyping performed using capillary electrophoresis on SEBIA CAPILLARYS 2, -CAPILLARYS 2 FLEX-PIERCING and CAPILLARYS 3 TERA instruments and on SEBIA MINICAP instrument,

Immunofixation methods: SEBIA HYDRAGEL IF, HYDRAGEL IF Penta, HYDRAGEL . BENCE JONES (Standard mask and Dynamic mask) performed using the HYDRASYS and HYDRASYS 2 instruments and the K20 electrophoresis chamber.

The IT / IF Control is designed for laboratory use. It should be used (with its barcode label for CAPILLARYS and MINICAP procedures) like a human serum sample.

The electrophoretic pattern obtained is specific for each batch of IT / IF control.

For In Vitro Diagnostic Use.

3. TECHNOLOGICAL CHARACTERISTICS

The CAPILLARYS 3 TERA instrument uses the principle of capillary electrophoresis in liquid solution. With this technique, charged molecules are separated by their electrophoretic mobility in an alkaline buffer with a specific pH. Protein separation is performed in a high voltage electrical field. The separated proteins are directly detected using absorbance at 200 nm at the cathodic end of the capillary. Separation occurs according to the electrolyte pH and is driven by electroosmotic flow. The CAPILLARYS 3 TERA instrument has silica capillaries functioning in parallel allowing 12 simultaneous analyses.

With the CAPI 3 IMMUNOTYPING procedure, the immunotyping using specific antibodies is performed to identify abnormal fractions in serum profiles. The immunotyping is performed in four automated steps:

{8}------------------------------------------------

Dilution of serum samples with a specific diluent in the pre-dilution well of the 1. reagent cup. This dilution is made according to the sample's immunoglobulin concentration.

  1. Mixing diluted serum sample with specific antisera. The antigen - antibody complex is formed rapidly in liquid medium.

  2. Injection of prepared samples with simultaneous aspiration into 6 capillaries at their anodic end. Protein separation occurs when a high voltage field is applied to the alkaline buffer. The separated proteins are detected using absorbance at 200 nm at the cathodic end of the capillary.

Overlay of the ELP pattern on the antisera patterns (Ig G, Ig A, Ig M, Kappa 4. and Lambda) allows characterization of suspected monoclonal component.

2. SUBSTANTIAL EQUIVALENCE INFORMATION:

Predicate Device NamePredicate Device 510(k) number
CAPILLARYS IMMUNOTYPINGK130500
CAPILLARYS 2 / CAPILLARYS 2 FLEX-PIERCINGK130500
IT / IF CONTROLK130500

The technological characteristics of the CAPI 3 IMMUNOTYPING procedure using the CAPILLARYS 3 TERA instrument (candidate device) utilizes the same principles of capillary electrophoresis in an alkaline buffer reading at a wavelength of 200nm as the CAPILLARYS IMMUNOTYPING procedure (predicate device). The CAPILLARYS 3 TERA using the CAPI 3 IMMUNOTYPING procedure follow the FDA guidance document, Replacement Reagent and Instrument Family Policy.

{9}------------------------------------------------

Table A.

Similarities and differences between the candidate device (CAPI 3 IMMUNOTYPING and the predicate device (CAPILLARYS IMMUNOTYPING. Both are members of the Sebia CAPILLARYS family of instruments.

Table ASEBIA CAPILLARYS IMMUNOTYPING(K) 130500SEBIA CAPI 3 IMMUNOTYPING
Intended UseThe CAPILLARYS IMMUNOTYPING kit isdesigned for the detection and thecharacterization of monoclonal proteins(immunotyping) in human urine and serum withthe CAPILLARYS, the CAPILLARYS 2 and theCAPILLARYS 2 FLEX-PIERCING, SEBIA, forcapillary electrophoresis. It is used in conjunctionwith the SEBIA CAPILLARYS PROTEIN(E) 6 kitdesigned for proteins separation into 6 majorfractions in alkaline buffer (pH 9.9).The CAPILLARYS, CAPILLARYS 2 and theCAPILLARYS 2 FLEX-PIERCING perform allprocedural sequences automatically to obtain aprotein profile for qualitative analysis. Each urineor serum sample is mixed with individual antiserathat are specific against gamma (Ig G), alpha (IgA) and mu (Ig M) heavy chains, and kappa (freeand bound) light chains and lambda (free andbound) light chains, respectively.The proteins, separated in silica capillaries, aredirectly detected by their absorbance at 200 nm.The electrophoregrams are evaluated visually todetect the presence of specific reactions with thesuspect monoclonal proteins.The CAPI 3 IMMUNOTYPING kit is designed forthe qualitative detection and the characterizationof monoclonal proteins (immunotyping) in humanserum with the CAPILLARYS 3 TERA instrument,SEBIA, for capillary electrophoresis. It is used inconjunction with the CAPI 3 PROTEIN(E) 6 kit.SEBIA, designed for proteins separation into 6major fractions in alkaline buffer (pH 9.9).The CAPILLARYS 3 TERA instrument performsall procedural sequences automatically to obtaina protein profile for qualitative analysis. Eachserum sample is mixed with individual antiserathat are specific against gamma (Ig G), alpha (IgA) and mu (Ig M) heavy chains, and kappa (freeand bound) light chains and lambda (free andbound) light chains, respectively.The proteins, separated in silica capillaries, aredirectly detected by their absorbance at 200 nm.The electrophoregrams are evaluated visually todetect the presence of specific reactions with thesuspect monoclonal proteins.
For In Vitro Diagnostic Use.For In Vitro Diagnostic Use.
Separation systemFree solution capillary electrophoresis (FSCE);protein separation in an alkaline buffer (pH 9.9)according to their charge, to the electrolyte pHand electroosmotic flow.Fast separation and good resolution.Electrophoregrams show separated fractionsaccording to their charge.Same
CAPILLARYS IMMUNOTYPING Kit (PN 2100) :CAPI 3 IMMUNOTYPING Kit (PN 2600) :
ReagentAntisera segment containing :ELP solutionmammalian immunoglobulins anti-human gamma heavy chainsmammalian immunoglobulins anti-human alpha heavy chainsmammalian immunoglobulins anti-human mu heavy chainsmammalian immunoglobulins anti-human kappa (free and bound) lightchainsmammalian immunoglobulins anti-human lambda (free and bound) lightchainsSample diluentRack with ELP solution tube and antiserumtubes :mammalian immunoglobulins anti-human gamma heavy chainsmammalian immunoglobulins anti-human alpha heavy chainsmammalian immunoglobulins anti-human mu heavy chainsmammalian immunoglobulins anti-human kappa (free and bound) lightchainsmammalian immunoglobulins anti-human lambda (free and bound) lightchains
Table ASEBIA CAPILLARYS IMMUNOTYPING (K) 130500SEBIA CAPI 3 IMMUNOTYPING
InstrumentSEBIA CAPILLARYS 2 FLEXPIERCING instrument, PN 1227SEBIA CAPILLARYS 2 instrument, PN 1222SEBIA CAPILLARYS 3 TERA instrument, PN 1246
Sample throughput8 analyses / hour21 (samples) analysis/ 2 hours
InterfacePC interfacePC interface + touch screen
Temperature ControlBy Peltier deviceSame
Detection systemDeuterium lampDeuterium lamp and LED
Software for data processingSEBIA PHORESIS™ softwareSame
FirmwareIncluded into the PHORESIS softwareIncluded into the instrument
Number of separation units8 parallel capillaries12 parallel capillaries
Samples tubesuncapped tubes or capped tubes depending on the procedureSame
Samples identificationYes (Bar code reading on both sample racks and tubes)Yes (Bar code reading on sample tubes and RFID labels on sample racks)
Reagent identificationNoYes (RFID labels on reagent vials)
Introduction of the samples into the automatic systemPrimary capacity of 13 tubes for IT technique (i.e. 13 sample racks), uninterrupted throughput on sample racks. Only position 1 on the sample rack contains sample tube.Primary maximal capacity of 120 tubes (i.e. 15 sample racks), uninterrupted throughput on sample racks (8 positions available).
Reagent bay : main compartementCAPILLARYS 2 : Contains one vial of water, wash solution and buffer container.CAPILLARYS 2 FLEX-PIERCING : Contains one vial of water, wash solution, hemolyzing solution (for Hb and Hb A1c techniques) and buffer container.Up to 4 analysis buffers or hemolysing solutions (identified by RFID labels); 1 waste container, 1 container for, 1 container for the wash solution
Reagent bay : secondary compartmentNAUp to 3 vials and 1 rack with immunotyping reagents (all RFID tagged) in temperature controlled environment (< 15 °C); 1 RFID labeled vial and three tubes (for maintenance solutions) at room temperature
DimensionsL. 95 cm x H. 39 cm x D. 63 cmL. 90 cm x H. 54 cm x D. 67 cm
Weight50 kg75 kg

510(k) Summary 6

{10}------------------------------------------------

{11}------------------------------------------------

Table B.

Similarities and differences between the candidate device (IT / IF Control) and the predicate device (IT / IF Control).

TABLE BSEBIAIT / IF CONTROLK130500SEBIAIT / IF CONTROL
Intended UseThe IT / IF Control is designed toquality control the qualitative detectionand characterization of humanmonoclonal immunoglobulins (Ig G, IgA, Ig M, Kappa and Lambda) with theelectrophoresis methods:- Immunotyping performedusing capillary electrophoresis onSEBIA CAPILLARYS 2 andCAPILLARYS 2 FLEXPIERCINGinstruments and on SEBIA MINICAPinstrument,- Immunofixation methods :SEBIAHYDRAGEL IF, HYDRAGEL IF Penta,HYDRAGEL BENCE JONES(Standard mask and Dynamic mask)performed using the HYDRASYS andHYDRASYS 2 instruments and theK20 electrophoresis chamber. The IT /IF Control is designed for laboratoryuse. It should be used (with its barcodelabel for CAPILLARYS and MINICAPprocedures) like a human serumsample.The electrophoretic pattern obtained isspecific for each batch of IT / IFcontrol.For In Vitro Diagnostic Use.The IT / IF Control is designed toquality control the qualitative detectionand characterization of humanmonoclonal immunoglobulins (Ig G, IgA, Ig M, Kappa and Lambda) with theelectrophoresis methods:- Immunotyping performedusing capillary electrophoresis onSEBIA CAPILLARYS 2,CAPILLARYS 2 FLEXPIERCING,CAPILLARYS 3 TERA instrumentsand on SEBIA MINICAP instrument,- Immunofixation methods :SEBIAHYDRAGEL IF, HYDRAGEL IF Penta,HYDRAGEL BENCE JONES(Standard mask and Dynamic mask)performed using the HYDRASYS andHYDRASYS 2 instruments and theK20 electrophoresis chamber. The IT/ IF Control is designed for laboratoryuse. It should be used (with its barcodelabel for CAPILLARYS and MINICAPprocedures) like a human serumsample.The electrophoretic pattern obtained isspecific for each batch of IT / IFcontrol.For In Vitro Diagnostic Use
ProductNumber47884788
Format1 vial (1.0 mL)Same
PreparationReconstitute the lyophilized control vialwith 1.0 mL of distilled or deionizedwater.Same
StoragetemperatureBefore reconstitution, the lyophilizedcontrol must be stored between 2 °Cand 8 °C. They are stable until theexpiration date indicated on the viallabels.Same
Shelf life3 years at 2 - 8°C4 years at 2 - 8°C
TABLE BSEBIAIT / IF CONTROLK130500SEBIAIT / IF CONTROL
In use storageAfter reconstitution, the control isstable for 2 monthsmaximum between - 18 °C and- 30 °C.Do not freeze and thaw thereconstituted control more than 20times.After reconstitution, the control isstable for 6 monthsmaximum between - 18 °C and- 22 °C.Do not freeze and thaw thereconstituted control more than 20times.

510(k) Summary 8

{12}------------------------------------------------

3. Performance Data:

  • a. Repeatability
    Six (6) different serum samples, including 1 normal serum sample and 5 pathological serum samples with monoclonal components were run using the CAPI 3 IMMUNOTYPING procedure on all capillaries of the same CAPILLARYS 3 TERA instrument and with one lot of CAPI 3 IMMUNOTYPING kit.

Each sample was analyzed with each reagent (ELP solution, anti-Ig G, anti-Ig A, antilg M, anti-Kappa and anti-Lambda) on all capillaries, including 2 runs per reagent.

In this study, all dilution programs were tested.

For each tested reagent, all samples gave concordant results within run and between capillaries.

  • b. Reproducibility between lots and instruments
    Six (6) different serum samples, including 1 normal serum sample and 5 pathological serum samples with monoclonal components were run using the CAPI 3 IMMUNOTYPING procedure on 3 CAPILLARYS 3 TERA instruments and with 3 lots of CAPI 3 IMMUNOTYPING kit.

Each sample was analyzed on 18 runs over 5 working days (at 2 different times of the day).

In this study, all dilution programs were tested.

All samples gave concordant results for all runs on the 3 CAPILLARYS 3 TERA instruments and with the 3 lots of CAPI 3 IMMUNOTYPING kit.

{13}------------------------------------------------

  • c. Sensitivity
    Serial dilutions were prepared in normal serum with three pathological serum samples all exhibiting monoclonal components and analyzed using the CAPI 3 IMMUNOTYPING procedure on CAPILLARYS 3 TERA instrument.

The results are summarized below:

SampleNo.Monoclonal componentDetection limit (mg/dL)
TypeConcentration (g/dL)
1Ig G, L3.9630.9
GammaLambda30.9
2Ig A, K3.4013.3
AlphaKappa13.3
3Ig M, K1.6025.0
MuKappa25.0
  • d. Interferences
    Studies were performed to assess common or known substances that could interfere with the CAPI 3 IMMUNOTYPING procedure performed with the CAPILLARYS 3 TERA instrument. The interfering substances were evaluated in 4 different serum samples (1 normal serum sample and 3 pathological serum samples with monoclonal component).

Samples containing potential interferents were tested and the results were compared to those obtained from control samples containing no potential interfering substances.

No interference with the CAPI 3 IMMUNOTYPING procedure was detected due to the serum sample's high concentration of the following interfering factors tested at levels equal to the concentrations listed below:

Interfering factorConcentration
Triglycerides3.59 g/dL (41 mM)
Bilirubin20 mg/dL (342 µM)
Rheumatoid factor981 IU/mL
Hemoglobin0.2 g/dL

510(k) Summary 10

{14}------------------------------------------------

Inference studies were performed according to the Clinical Laboratory Standards Institute (CLSI - USA) EP7-A2 guideline "Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition".

  • e. Method Comparison
    A method comparison studies were conducted using a total of 115 serum samples (12 normal serum samples and 103 pathological serum samples).

Samples were analyzed in singlicate using the CAPI 3 IMMUNOTYPING procedure on the CAPILLARYS 3 TERA instrument (candidate device) and the CAPILLARYS IMMUNOTYPING procedure (predicate device).

510(k) Summary 11

{15}------------------------------------------------

  • i. Study
    A total of 115 serum samples (12 normal serum samples and 103 pathological serum samples).

This study demonstrated a 100 % agreement between the two techniques:

• For the 12 normal serum samples: complete agreement (concordance).

• For the 103 pathological serum samples: complete agreement (concordance).

In all cases, both techniques detected and characterized the monoclonal proteins (immunotyping) in human serum with complete agreement (see results below).

Qualitative ResultsNumber of serum samplesComplete Agreement
Normal12Complete Agreement
Ig G, Kappa37Complete Agreement
Ig G, Lambda27Complete Agreement
Ig A, Kappa6Complete Agreement
Ig A, Lambda4Complete Agreement
Ig M, Kappa14Complete Agreement
Ig M, Lambda4Complete Agreement
2 Ig A, Kappa3Complete Agreement
2 Ig M, Kappa2Complete Agreement
Ig G, Kappa + Ig G, Lambda2Complete Agreement
2 Ig A, Kappa + Ig G, Kappa1Complete Agreement
Ig G, Kappa + Ig A, Kappa + Ig M,Lambda1Complete Agreement
2 Lambda free1Complete Agreement
1 Kappa free1Complete Agreement
Grand Total115Complete Agreement

4. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 866.5510 Immunoglobulins A, G, M, D, and E immunological test system.

(a)
Identification. An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.(b)
Classification. Class II (performance standards).