AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Verrata® PLUS pressure guide wire is indicated for use to measure pressure in blood vessels, including both coronary and peripheral vessels, during diagnostic angiography and/or any interventional procedures. Blood pressure measurements provide hemodynamic information for the diagnosis and treatment of blood vessel disease.

Device Description

The Verrata® PLUS Pressure Guide Wire (hereafter referred to as the “Verrata PLUS”) is a steerable guide wire with a pressure transducer mounted 3 cm proximal to the tip. The Verrata® PLUS measures pressure when used with the SmartMap®, ComboMap®, s5 Series™, and CORE™ Series of systems. The Verrata PLUS has a diameter of 0.014" (0.36 mm) and is available in lengths of 185 cm or 300 cm and also in straight or pre-shaped tips. The Verrata PLUS is packaged attached to the connector with a torque device to facilitate navigation through the vasculature.

AI/ML Overview

The provided text describes the Verrata PLUS Pressure Guide Wire and its performance testing to demonstrate substantial equivalence to a predicate device. Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Visual Inspection (Cable Routing)Cable must be routed out the right side of the connector.Sample wires and connectors were visually inspected to confirm proper routing of the cable out of the connector. Implied: Complied with acceptance criteria.
Visual Inspection (Pressure Sensor)The sensor adhesive is not protruding out of the housing and the sensor is not damaged or misaligned.Sample wires were visually inspected to confirm that the pressure sensor within the pressure housing is properly aligned, positioned and potted in place. Implied: Complied with acceptance criteria.
Wire Outer DiameterThe sensor housing outer diameter must meet product specifications.The maximum diameter of the sensor housing is measured with a micrometer. Implied: Complied with acceptance criteria.
Zero Test and Connection DurabilityThe wire must successfully zero the first time and after 20 disconnect and re-connect cycles.Wires with connectors were connected to a Volcano Pressure System and "zeroed". The pressure value on the screen of the System should read "0". Disconnect the wire from the connector and re-connect up to 20 times. After 20 disconnect and re-connects, verified that the wire still "zeros". Implied: Complied with acceptance criteria.
Accuracy TestThe pressure measured by the wire must meet product specifications.The wire is connected to a test fixture and is placed in an air pressure chamber at 37℃. The pressure inside the chamber is varied and the pressure reading from the wire is compared to the pressure to which the chamber is set. Implied: Complied with acceptance criteria.
Drift TestThe drift over 10 minutes must meet product specifications.The wire is connected to a test fixture and placed in saline at 37℃. The pressure reading is recorded for 10 minutes. Implied: Complied with acceptance criteria.
Connector to Cable Tensile TestThe tensile strength must meet product specifications.The connector to cable connection was pull tested to assess its tensile strength. Implied: Complied with acceptance criteria.
Biocompatibility (ISO 10993)Per ISO 10993 and test protocols.Biocompatibility of the raw materials in the Verrata PLUS wire were assessed per ISO 10993. Implied: Complied with acceptance criteria.
Biocompatibility (Direct Hemolysis)The difference between the hemolytic indices of the test article and negative control must be

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.