Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K180105
    Manufacturer
    Date Cleared
    2018-03-08

    (51 days)

    Product Code
    Regulation Number
    870.5150
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INDIGO Aspiration Catheters and Separators

    As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral and venous systems.

    INDIGO Aspiration Tubing

    As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.

    Penumbra Aspiration Pump

    The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

    Device Description

    The Indigo Aspiration System ("Indigo System") is intended for the removal of fresh, soft emboli and thrombi from vessels of the peripheral and venous systems using the Indigo Aspiration Catheter, Indigo Separator, Indigo Aspiration Tubing, and Penumbra Aspiration Pump. The Indigo System was most recently cleared under K161523.

    The Indigo System is designed to remove thrombus from the vasculature using continuous aspiration. The Aspiration Catheter targets aspiration from the Aspiration Pump directly to the thrombus. The Separator may be used to clear the lumen of the Aspiration Catheter should it become blocked with thrombus. The use of the Separator may not be necessary when using an Aspiration Catheter with an I.D. of 0.054in or larger. The Aspiration Catheter is introduced through a guide catheter or long femoral sheath and into the site of the primary occlusion. The Aspiration Catheter is used with the Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, a Separator may be deployed from the Aspiration Catheter to assist with thrombus removal. The Separator is advanced and retracted through the Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the Aspiration Catheter tip. For the aspiration source, the Aspiration Catheter is used in conjunction with the Aspiration Pump, which is connected using the Aspiration Tubing and Canister. The Aspiration Catheter is provided with a steam shaping mandrel and rotating hemostasis valve, and a peelable sheath. The Separator is provided with an introducer and torque device. The Aspiration Catheters and Separators are visible under fluoroscopy.

    The Penumbra Engine Pump and Canister

    The Penumbra Engine Pump ("Engine Pump") is an electromechanical device designed to create vacuum pressure for use with the Indigo System. The Engine Pump maintains vacuum pressure for 3 hours of continuous use and has an operating life of ≥ 500 hours. The pump can be used in environments with 100 - 240 Vac and 50/60 Hz. The Engine Pump is intended for use in operating rooms or interventional catheterizations laboratories. The Engine Pump and Canister do not come into contact with the patient. The Engine Pump Canister is a 1000 mL minimum volume reservoir which is an accessory to the Engine Pump. The Canister has a lid with a stop-flow filter to prevent excess fluid from entering the pump. The patient port on the lid is sized to accept the Suction Connector on the Indigo Aspiration Tubing.

    AI/ML Overview

    The provided document describes the Indigo Aspiration System - Penumbra Engine Pump and Canister. It is a 510(k) premarket notification for a medical device. The document primarily focuses on demonstrating substantial equivalence to a legally marketed predicate device (K161523 cleared on July 1, 2016, also called Indigo Aspiration System, and a reference device K122756 Penumbra Pump MAX).

    Here's an analysis of the provided information regarding acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document provides a table for the "Design Verification (Bench-top Testing and Electrical Safety/EMC Testing)" of the Penumbra Engine Pump and Canister.

    AttributeSpecificationAcceptance CriteriaReported Device Performance
    Pump - Electrical SafetyThe Pump conforms to IEC 60601-1 and IEC 60601-1-2 requirements including international worldwide variants (CB Scheme). The Pump is compliant with EN ISO 10079-1.100% Pass100% Pass
    Pump - Environmental TestingPump performance specifications under environmental conditions100% Pass100% Pass
    Pump - Dimensional InspectionDimensional specifications per Product Specification100% Pass100% Pass
    Pump - Inspection of Design FeaturesDesign specifications per Product Specification100% Pass100% Pass
    Pump - PerformancePerformance specifications per Product Specification100% Pass100% Pass
    Pump - Performance at variable voltage and frequencyPerformance specifications per Product Specification100% Pass100% Pass
    Pump and Canister - 500 hour Use TestingPerformance specifications after 500 hour use100% Pass100% Pass
    Pump Canister - Inspection of Design FeaturesDesign specifications per Product Specification100% Pass100% Pass
    Pump Canister - PerformancePerformance specifications per Product Specification100% Pass100% Pass
    Pump and Canister - Simulated UsePump and Canister use specifications100% Pass100% Pass

    2. Sample Size Used for the Test Set and Data Provenance:

    The sample size for the test set is indicated in the "Sample Size" column of the table above:

    • For "Pump - Electrical Safety": N=1
    • For all other tests related to the Pump and Canister: N=5

    The data provenance is not explicitly stated in terms of country of origin or retrospective/prospective nature. However, the tests are described as "Bench-top testing" and "Electrical safety and EMC testing," performed by "Intertek, a nationally recognized test laboratory." This implies laboratory-based testing rather than clinical data from human subjects.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. The tests described are engineering and performance verification tests for a mechanical/electrical device, not diagnostic or interpretive tasks where human experts would establish ground truth.

    4. Adjudication Method for the Test Set:

    This information is not applicable as the tests are objective engineering performance measures, not subjective interpretations requiring adjudication. The "100% Pass" acceptance criteria suggest direct measurement against specifications.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    An MRMC comparative effectiveness study was not done. This device is a pump and canister, which are components of an aspiration system, not a diagnostic imaging device or an AI algorithm intended to assist human readers. The clinical performance of the entire Indigo Aspiration System (catheters and separators) for removing emboli and thrombi is separate from the performance of the vacuum pump itself.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    A standalone study was done, but it refers to the standalone performance of the device (pump and canister) itself, not an algorithm. The tests listed in the table demonstrate the performance of the Penumbra Engine Pump and Canister without human intervention for the specific parameters being tested (e.g., electrical safety, dimensional inspection, performance specs, 500-hour use). This is a technical performance assessment, not an AI algorithm assessment.

    7. Type of Ground Truth Used:

    For the tests described, the "ground truth" is based on engineering specifications, performance standards (e.g., IEC 60601-1, EN ISO 10079-1), and product design specifications. These are objective, measurable criteria, not expert consensus, pathology, or outcomes data typically associated with diagnostic or therapeutic device efficacy.

    8. Sample Size for the Training Set:

    This information is not applicable. The device is an electromechanical pump and canister, not an artificial intelligence or machine learning device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1