(117 days)
The SpotLight CT Computed Tomography X-ray is intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles. The system has the capability to image whole organs, including the heart, in a single rotation. The system may acquire data using Axial, Cine and Cardiac CT scan techniques from patients of all ages. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.
This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes.
The system is indicated for X-ray Computed Tomography imaging of organs that fit in a 25cm field of view, including cardiac and vascular CT imaging. The device output is useful for diagnosis of disease or abnormality and for planning of therapy procedures.
The SpotLight CT is a third generation rotate-rotate CT scanner, designed and built based on technologies and principles of operation of the predicate device and other legally marketed CT scanners. The SpotLight CT is a multi-slice (192 detector rows), dual tube CT scanner consisting of a gantry, patient table, operator console, power distribution unit (PDU) and interconnecting cables. The system includes image acquisition hardware, image acquisition and reconstruction software for operator interface and image handling.
The provided document describes the Arineta Ltd. SpotLight CT device and its substantial equivalence to a predicate device. Here's a breakdown of the acceptance criteria and the study details:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" in a tabular format with corresponding "reported device performance." Instead, it describes various performance specifications and how the device performed against them. I've re-framed the reported performance metrics as if they were acceptance criteria.
| Performance Metric (Implied Acceptance Criteria) | Reported Device Performance |
|---|---|
| Coverage (Z-direction) | Up to 140 mm in a single axial scan |
| Field of View (FOV) - Diagnostic | 250 mm (radiation outside 250mm or 160mm FOV is attenuated, providing diagnostic image quality up to 250mm FOV) |
| Gantry Rotation Speed | Up to 0.24 seconds per rotation |
| Temporal Resolution | 120 msec (at 0.24 second rotation speed) |
| Spatial Resolution | 0.31 mm |
| Detector Rows | 192 detector rows |
| Number of X-ray sources | Two ("Gemini" X-ray tubes) |
| Image Quality Evaluation | Evaluated for artifacts, spatial resolution, low contrast detectability, noise, and uniformity and CT number accuracy (details on specific pass/fail not provided, but generally stated as meeting specifications). |
| Dose Performance | Evaluated as meeting specifications |
| Ability to Image Whole Organs | Capable of imaging whole organs, including the heart, in a single rotation. |
| Diagnostic Quality (Animal Testing) | Images were evaluated for diagnostic quality with positive results. |
| Clinical Diagnostic Value & Image Quality (Human Testing) | Demonstrated diagnostic image quality performance. |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set (Clinical Testing): 38 subjects
- Data Provenance: Not explicitly stated, but the study was conducted at "one site," and the readers were "US certified." This suggests the data was collected in the US.
- Retrospective or Prospective: The clinical testing describes "data were collected," which could mean either. However, the phrase "The study protocol was designed to test the scanner across different patient populations, clinical scenarios and scan techniques" implies a prospective study design for collecting new data specifically for this evaluation.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Number of Experts: Four (4)
- Qualifications: "US certified readers who are qualified radiologists or cardiologists." Specific years of experience are not mentioned.
4. Adjudication Method for the Test Set
The document states, "The images were evaluated and rated by four US certified readers." It does not specify an adjudication method like 2+1 or 3+1 for resolving discrepancies. It implies independent evaluation by each reader.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was an MRMC study done? No, a multi-reader multi-case (MRMC) comparative effectiveness study (comparing human readers with and without AI assistance) was not done. The clinical testing focused on evaluating the device's diagnostic image quality for standalone performance.
- Effect size of human readers with AI vs without AI assistance: Not applicable, as no such study was conducted or reported.
6. Standalone (Algorithm Only) Performance
- Was a standalone study done? Yes, the described "Non-clinical Performance Testing" and "Clinical Testing" primarily focus on the standalone performance of the SpotLight CT system. The image quality, temporal resolution, dose performance, and diagnostic quality evaluations are all measures of the device's inherent capabilities without human intervention during the image generation or initial analysis phase. The "clinical diagnostic value and image quality" evaluated by the readers are also assessing the output of the device itself.
7. Type of Ground Truth Used
- Non-clinical/Phantom Testing: The ground truth for these tests would be objective physical measurements and known parameters of the phantoms used to evaluate image quality metrics (e.g., spatial resolution targets, known low contrast objects, CT number uniformity).
- Animal Testing: The "diagnostic quality" evaluation in animal models likely used established veterinary diagnostic criteria and potentially post-mortem examination or other correlative imaging as ground truth.
- Clinical Testing: The "clinical diagnostic value and image quality" in human subjects were evaluated by "qualified radiologists or cardiologists" against established clinical diagnostic criteria. While not explicitly stated as "expert consensus ground truth," the assessment by multiple, qualified experts serves as the de-facto ground truth for evaluating diagnostic utility. It does not mention pathology or long-term outcomes data for ground truth.
8. Sample Size for the Training Set
The document does not provide any information about a training set since this is a hardware device (CT scanner) with associated imaging software, not typically an AI/ML algorithm that requires a distinct "training set" in the common sense. The image reconstruction algorithm is described as "Stereo CT reconstruction algorithm based on common algorithms used in single source scanners that are modified to combine the data acquired from the two sources." This implies an engineered algorithm, not one trained on a dataset.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no mention of an AI/ML algorithm requiring a training set in the document. The device uses an engineered image reconstruction algorithm.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the logo is a stylized image of three human profiles facing to the right, with flowing lines representing hair or movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Auqust 10, 2016
Arineta Ltd. % Mr. Ehud Dafni CEO 15 Halamish Street Caesarea, 3088900 ISRAEL
Re: K161066
Trade/Device Name: SpotLight CT Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: July 11, 2016 Received: July 14, 2016
Dear Mr. Dafni:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K161066
Device Name SpotLight CT
Indications for Use (Describe)
The SpotLight CT Computed Tomography X-ray is intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles. The system has the capability to image whole organs, including the heart, in a single rotation. The system may acquire data using Axial, Cine and Cardiac CT scan techniques from patients of all ages. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.
This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes.
The system is indicated for X-ray Computed Tomography imaging of organs that fit in a 25cm field of view, including cardiac and vascular CT imaging. The device output is useful for diagnosis of disease or abnormality and for planning of therapy procedures.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) SUMMARY
Arineta Ltd.'s SpotLight CT
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92
- Date prepared: April 12, 2016
- Submitted by: Arineta Ltd. 15 Halamish Street Caesarea 3088900 Israel Tel: +972-4-6374000 Fax: +972-4-6277006
PRODUCT IDENTIFICATION
| Device Trade Name: | SpotLight CT |
|---|---|
| Common/Usual Name: | Computed Tomography X-ray System |
| Classification Name: | Computed Tomography X-ray System(21CFR 892.1750) |
| Product Code: | 90-JAK |
| Predicate Device: | GE Healthcare Revolution CT K133705 |
| References Devices: | Siemens Somatom Definition Flash (K121072),Elscint Excel CT Twin (K915738), Philips Mx8000(K010817), Toshiba Equilion One Vision (K142465)Philips Brilliance iCT (K131773) |
INTENDED USE
The Spotlight CT Computed Tomography X-Ray system is intended for head, body, cardiac and vascular X-ray Computed Tomography applications.
INDICATIONS FOR USE
The SpotLight CT Computed Tomography X-ray is intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission projection data taken at different angles. The system has the capability to image whole organs,
{4}------------------------------------------------
including the heart, in a single rotation. The system may acquire data using Axial, Cine and Cardiac scan techniques from patients of all ages. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.
This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes.
The system is indicated for X-ray Computed Tomography imaging of organs that fit in a 25cm field of view, including cardiac and vascular CT imaging. The device output is useful for diagnosis of disease or abnormality and for planning of therapy procedures.
DEVICE DESCRIPTION
The SpotLight CT is a third generation rotate-rotate CT scanner, designed and built based on technologies and principles of operation of the predicate device and other legally marketed CT scanners. The SpotLight CT is a multi-slice (192 detector rows), dual tube CT scanner consisting of a gantry, patient table, operator console, power distribution unit (PDU) and interconnecting cables. The system includes image acquisition hardware, image acquisition and reconstruction software for operator interface and image handling.
Technological Characteristics / Principles of Operation
The system generates images through computed reconstruction of data acquired at different view angles of the rotating gantry, while irradiating the scanned subject by two alternating X-ray sources. The gantry rotates at up to 0.24 seconds per rotation and can acquire up to 240 slices of image data in a single axial scan with a maximum total coverage of 140 mm in the z direction. While the system covers field of view (FOV) of 450 mm. the radiation outside 250 mm (or 160 mm) FOV is attenuated, providing diagnostic image quality up to 250 mm FOV. The system can be operated in axial (partial or full scan), cine, cardiac and ECG gated modes.
The SpotLight CT is a powerful, volumetric, high resolution CT that is designed to provide whole organ coverage, high image quality and uncompromised dose performance with the following features:
- 140 mm detector coverage;
- 120 msec temporal resolution (at 0.24 second rotation speed); ●
- 0.31 mm spatial resolution: .
- 250 mm diagnostic FOV: ●
- Reduced cone beam artifacts. 0
The SpotLight CT features two "Gemini" X-ray tubes to provide two X-ray sources displaced in the z axis (parallel to rotation axis). The sources are alternating using electrostatic grid control, such that the scanned subject is irradiated alternatively by two
{5}------------------------------------------------
overlapping cone X-ray beams while the gantry rotates and attenuation data is acquired by a single array detector. Image reconstruction is accomplished by a Stereo CT reconstruction algorithm based on common algorithms used in single source scanners that are modified to combine the data acquired from the two sources.
The detector array comprises of three regions, center part covering FOV 250 mm with high pitch detector array and two peripheral parts with detector arrays. The data acquired in the peripheral parts is useful for patient positioning (scout scan) and for elimination of truncation artifacts from the images. The detector array features antiscatter collimator (post patient) to reduce scatter artifacts, ensure CT number uniformity and reduce beam hardening artifacts.
The gantry features a short source to axis distance (SAD) of 450 mm, resulting in higher usable X-ray flux at a given source voltage and current as compared to scanners with longer SAD.
DEVICE TESTING
The SpotLight CT has completed testing and is in compliance with AAMI/ES 60601-1. IEC 60601-1 Ed. 3 and its associated collateral standards and particular standards, 21CFR Subchapter J and NEMA XR-25, XR-28 and XR-29.
Testing included:
- Testing on unit level; ●
- Integration testing (system verification); 0
- Performance testing (verification): o
- Safety testing (verification); ●
- o Simulated use testing (validation);
- . Animal testing (validation);
- o Clinical testing (validation).
The device was developed under a rigorous quality system and has successfully completed design control activities, including risk management, verification and validation.
Non-clinical Performance Testing
The performance evaluation used a variety of test methods, phantoms and scan conditions. Various mathematical, physics and statistical analyses were performed to demonstrate that performance specifications are met. The tests evaluated image quality, temporal resolution and dose performance. Image quality evaluation included evaluation of artifacts, spatial resolution, low contrast detectability, noise, and uniformity and CT number accuracy.
{6}------------------------------------------------
Animal Testing
The scanner was tested on a porcine model with and without contrast injection, with and without ECG gating at different heart rates. The trial included three cardiac CTA studies under different conditions, abdomen (renal) scan and neck (carotids) CTA. The images were evaluated for diagnostic quality with positive results.
Clinical Testing
Clinical data were collected on 38 subjects at one site with the approval of appropriate ethics committee and in accordance with applicable regulatory requirements as well as Arineta quality system procedures. The study protocol was designed to test the scanner across different patient populations, clinical scenarios and scan techniques. Image data sets included:
- . Cardiac - Coronary CTA, cardiac function, calcium burden, preparation for TAVI
- Body Abdomen (at a limited field of view), spine ●
- Head brain, inner ear 0
The images were evaluated and rated by four US certified readers who are qualified radiologists or cardiologists for clinical diagnostic value and image quality.
The results of this clinical testing demonstrate the diagnostic image quality performance of the SpotLight CT, as well as the safety and efficacy of the device.
SUBSTANTIAL EQUIVALENCE
Based on the extensive testing as described above, the SpotLight CT is as safe and effective as the Revolution CT. The SpotLight CT has the same intended uses and similar indications, technological characteristics, energy type and principles of operation as its predicate device. The technological differences between the SpotLight CT and its predicate devices raise no new issues of safety or effectiveness. Thus, the SpotLight CT is substantially equivalent.
CONCLUSION
Based on the conformance to standards, as well as the bench and clinical testing provided, Arineta Ltd. believes that the SpotLight CT is substantially equivalent to the predicate device, GEHC Revolution CT (K133705).
§ 892.1750 Computed tomography x-ray system.
(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.