K Number
K160538

Validate with FDA (Live)

Date Cleared
2016-12-22

(300 days)

Product Code
Regulation Number
864.5220
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sysmex XN-L analyzer is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XN-L analyzer classifies and enumerates the following parameters in whole blood: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, NEUT%H, LYMPH%4, MONO%#, EO%#, BASO%#, IG%#, RDW-CV, RDW-SD, MPV, RET%#, IRF, RET-He and has a Body Fluid mode for body fluids. The Body Fluid mode enumerates the WBC-BF, RBC-BF, MN%#, PMN%#, and TC-BF# parameters in cerebrospinal, pleural, and synovial fluids. Whole blood should be collected in K2 or K3EDT A anticoagulant and peritoneal, pleural, and synovial fluids in K2EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants with CSF specimens is neither required nor recommended.

The performance of this device has not been established in pediatric patients under the age of 2 years.

Device Description

The Sysmex XN-L analyzer is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XN-L analyzer classifies and enumerates whole blood and bodyfluid parameters by means of electrical impedance, laser light scattering, and fluorescentlabeling. Tests are performed on venous and capillary whole blood samples collected in K2 or K3 EDTA anticoagulant and body fluids (peritoneal, pleural and synovial) collected in K2EDTA anticoagulant and cerebrospinal fluid that is not collected in anticoagulant. The instrument consists of two principal units: (1) Main Unit which will aspirate, dilute, mix, and analyze whole blood and body fluid samples and (2) Pneumatic Unit which supplies pressure and vacuum to the analyzer. The XN-L analyzer performs analysis using the following methods: DC Sheath Flow Detection Method, Flow Cytometry Methods using a Semiconductor Laser, and SLS- hemoglobin Method. Particle characterization and identification is based on detection of forward scatter, fluorescence, and adaptive cluster analysis. The XN-L analyzer automatically classifies cells from whole blood and body fluids and carries out all processes automatically from aspiration of the sample to outputting the results.

The XN-L has an external monitor with touch screen capability that is used to operate the instrument and process data from the Main Unit. The monitor also allows operator interfacing with the instrument by use of a panel keyboard.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the Sysmex® XN-L Automated Hematology Analyzer, asserting its substantial equivalence to a predicate device, the Sysmex XN-Series (XN-10, XN-20) Automated Hematology Analyzer (K112605).

The text does not provide a table of acceptance criteria with reported device performance values, nor does it detail specific sample sizes, expert qualifications, adjudication methods, or MRMC study results for the test sets. It broadly states that "predetermined acceptance criteria were met" and that "Performance, verification, and validation testing were conducted to characterize the performance of the XN-L analyzer."

Therefore, based only on the provided text, I cannot fulfill all parts of your request with specific details. However, I can extract the information that is present and highlight the missing components.


Acceptance Criteria and Study for Sysmex® XN-L Automated Hematology Analyzer

Based on the provided FDA 510(k) summary, the device's performance was evaluated for substantial equivalence to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

The document states that "predetermined acceptance criteria were met" and describes the types of performance evaluations conducted. However, it does not provide a table with specific quantitative acceptance criteria or the numerical results of these tests.

Performance Evaluations Conducted (Details not provided beyond type):

Performance MetricAcceptance Criteria (Not Detailed)Reported Device Performance (Not Detailed)
Whole Blood AnalysisSubstantial equivalence to predicate device (Sysmex XN-Series (XN-10, XN-20) K112605) based on performance characteristics. Acceptance criteria details not provided.Accuracy, Precision (Reproducibility & Repeatability), Linearity, Carryover, Stability, Reference Intervals (Verification), Limits of Blank, Detection, and Quantitation. The document states these evaluations were conducted and demonstrate substantial equivalence. Specific numerical results are not provided.
Body Fluid AnalysisSubstantial equivalence to predicate device (Sysmex XN-Series (XN-10, XN-20) K112605) based on performance characteristics. Acceptance criteria details not provided.Accuracy, Precision (Reproducibility & Repeatability), Linearity, Carryover, Stability, Limits of Blank, Detection, and Quantitation. The document states these evaluations were conducted and demonstrate substantial equivalence. Specific numerical results are not provided.

2. Sample Size Used for the Test Set and Data Provenance

  • The document does not specify the exact "sample size used for the test set." It generically states "Clinical testing was conducted on the XN-L analyzer."
  • The document does not specify the data provenance (e.g., country of origin, retrospective or prospective). It implies the data was collected as part of clinical testing for this submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

  • This information is not provided in the document. The device is an "Automated Differential Cell Counter," which typically relies on established methods and comparative performance to a predicate device or reference method, rather than subjective expert consensus for ground truth on individual cases.

4. Adjudication Method for the Test Set

  • This information is not provided in the document. Given the nature of an automated cell counter, adjudication methods as typically used in imaging or diagnostic interpretation studies are unlikely to be directly applicable in the same way. Performance is likely assessed by comparing automated counts to reference methods.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • An MRMC study is not applicable to this device. This device is an automated hematology analyzer that performs quantitative measurements of blood and body fluid parameters. It is not an AI-assisted diagnostic tool that helps human readers interpret medical images or perform similar tasks. Therefore, there is no "human-in-the-loop" performance improvement to measure.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Yes, implicitly. The entire study described is a "standalone" performance evaluation of the Sysmex XN-L analyzer. Its function is to autonomously classify and enumerate cells and parameters in blood and body fluids. The "Performance Testing" section details various evaluations (accuracy, precision, linearity, etc.) of the algorithm/device's direct output.

7. The Type of Ground Truth Used

  • The document does not explicitly state the "type of ground truth" (e.g., pathology, outcomes data). For an automated hematology analyzer, ground truth is typically established by:
    • Reference Methods: Such as manual microscopy (for cell counts and differentials), flow cytometry, or spectrophotometry (for hemoglobin), performed by highly trained medical technologists.
    • Reference Instruments: Comparison to an established, highly accurate reference instrument.
    • Predicated Device Comparison: The primary method of demonstrating substantial equivalence here is by showing comparable performance to the legally marketed predicate device (Sysmex XN-Series).

8. The Sample Size for the Training Set

  • This device is an automated analyzer (a hardware/software system) for clinical laboratory measurements, not a machine learning model that undergoes explicit "training" in the typical sense with a "training set." Therefore, the concept of a "training set" as understood in AI/ML is not applicable here. The device's algorithms are pre-programmed and validated through the described performance studies.

9. How the Ground Truth for the Training Set was Established

  • As explained in point 8, the concept of a "training set" for this type of medical device is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 22, 2016

Sysmex America, Inc. Ms. Sharita Brooks Sr. Manager, Regulatory Affairs 577 Aptakistic Road Lincolnshire, IL 60069

Re: K160538

Trade/Device Name: Sysmex® XN-L Automated Hematology Analyzer Regulation Number: 21 CFR 864.5220 Regulation Name: Automated Differential Cell Counter Regulatory Class: Class II Product Code: GKZ Dated: December 12, 2016 Received: December 13, 2016

Dear Ms. Brooks:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Leonthena R. Carrington, MS, MBA, MT(ASCP) Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K160538

Device Name

Sysmex ® XN-L Automated Hematology Analyzer

Indications for Use (Describe)

The Sysmex XN-L analyzer is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XN-L analyzer classifies and enumerates the following parameters in whole blood: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, NEUT%H, LYMPH%4, MONO%#, EO%#, BASO%#, IG%#, RDW-CV, RDW-SD, MPV, RET%#, IRF, RET-He and has a Body Fluid mode for body fluids. The Body Fluid mode enumerates the WBC-BF, RBC-BF, MN%#, PMN%#, and TC-BF# parameters in cerebrospinal, pleural, and synovial fluids. Whole blood should be collected in K2 or K3EDT A anticoagulant and peritoneal, pleural, and synovial fluids in K2EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants with CSF specimens is neither required nor recommended.

The performance of this device has not been established in pediatric patients under the age of 2 years.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Sysmex. The logo features a stylized graphic element in shades of blue and green, resembling a wave or abstract shape. To the right of the graphic is the word "sysmex" in a bold, sans-serif font, colored in a matching shade of blue.

5. 510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K160538

Submitter's name, address, telephone number, a contact person, and date the summary was prepared:

Submitter's Name:Sysmex America, Inc.
Submitter's Address:577 Aptakistic RoadLincolnshire, IL 60069
Submitter's Telephone:(224) 543-9618
Submitter's FAX:(224) 543-9699
Submitter's Contact:Sharita Brooks
Date 510(k) Summary Prepared:February 25, 2016

Name of the device, including the trade or proprietary name, the common or usual name, and the classification name:

Proprietary Name:Sysmex® XN-L Automated Hematology Analyzer
Common Name:Automated Hematology Analyzer
Regulation Description:Automated Differential Cell Counter
Regulation Section:21 CFR 864.5220
Device Class:2
Product Code:GKZ

Related Items:

Product Code: 81GIF

CELLPACK®DCL (Diluent) CELLPACK® DST (Diluent) CELLPACK®DFL (Diluent)

Product Code: 81KJK

Fluorocell" WDF (Dye) Fluorocell™ RET (Dye)

Product Code: 81JPK

XN CHECK™ (Control) XN CHECK™ BF (Control) XN-L CHECK™ (Control)

Product Code: 81GGK

SULFOLYSER® (Lyse) Lysercell™ WDF (Lyse)

Product Code: 81KSA

XN CAL™ (Calibrator)

Product Code: 81JCB

CELLCLEAN™ AUTO

Predicate Device and 510(k) number: Sysmex XN-Series (XN-10, XN-20) Automated Hematology Analyzer, K112605

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the Sysmex logo. The logo features an abstract graphic in shades of blue and green, resembling a stylized wave or ribbon. To the right of the graphic, the word "sysmex" is written in lowercase, using a bold, sans-serif font and a blue color that matches the graphic.

Description of the Device:

The Sysmex XN-L analyzer is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XN-L analyzer classifies and enumerates whole blood and bodyfluid parameters by means of electrical impedance, laser light scattering, and fluorescentlabeling. Tests are performed on venous and capillary whole blood samples collected in K2 or K3 EDTA anticoagulant and body fluids (peritoneal, pleural and synovial) collected in K2EDTA anticoagulant and cerebrospinal fluid that is not collected in anticoagulant. The instrument consists of two principal units: (1) Main Unit which will aspirate, dilute, mix, and analyze whole blood and body fluid samples and (2) Pneumatic Unit which supplies pressure and vacuum to the analyzer. The XN-L analyzer performs analysis using the following methods: DC Sheath Flow Detection Method, Flow Cytometry Methods using a Semiconductor Laser, and SLS- hemoglobin Method. Particle characterization and identification is based on detection of forward scatter, fluorescence, and adaptive cluster analysis. The XN-L analyzer automatically classifies cells from whole blood and body fluids and carries out all processes automatically from aspiration of the sample to outputting the results.

The XN-L has an external monitor with touch screen capability that is used to operate the instrument and process data from the Main Unit. The monitor also allows operator interfacing with the instrument by use of a panel keyboard.

Statement of Intended Use:

The Sysmex XN-L analyzer is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XN-L analyzer classifies and enumerates the following parameters in venous and capillary whole blood: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, NEUT%/#, LYMPH%/#. MONO%/#, EO%/#, BASO%/#, IG%/#, RDW-CV, RDW-SD, MPV, RET%/#, IRF, RET-He and has a Body Fluid mode for body fluids. The Body Fluid mode enumerates the WBC-BF, RBC-BF, MN%/#, PMN%/#, and TC-BF# parameters in cerebrospinal, peritoneal, pleural, and synovial fluids. Whole blood should be collected in K2 or K3EDTA anticoagulant and peritoneal, pleural, and synovial fluids in K2EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants with CSF specimens is neither required nor recommended.

The performance of this device has not been established in pediatric patients under the age of 2 years.

Summary of Substantial Equivalence:

Table 5-1 compares the Sysmex XN-L Automated Hematology analyzer with the XN-Series (XN-10, XN-20) Automated Hematologyanalyzer.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Sysmex logo. The logo features a stylized graphic element in shades of blue and green, resembling a wave or abstract shape. To the right of the graphic is the word "sysmex" in lowercase, using a bold, sans-serif font and a blue color that matches the graphic element.

Table 5-1: Comparison of the Predicate XN-10 and the Proposed XN-L Automated Hematology Analyzers

ItemPredicate AnalyzerXN-Series (XN-10)aK112605Proposed AnalyzerXN-L
Similarities
Intended UseThe XN-Series modules (XN-10, XN-20)are quantitative multi-parameterautomated hematology analyzers intendedfor in vitro diagnostic use in screeningpatient populations found in clinicallaboratories. The XN-Series modulesclassify and enumerate the followingparameters in whole blood: WBC, RBC,HGB, HCT, MCV, MCH, MCHC, PLT,NEUT%/#, LYMPH%/#, MONO%/#,EO%/#, BASO%/#, IG%/#, RDW-CV,RDW-SD, MPV, NRBC%/#, RET%/#,IPF, IRF, RET-He and has a Body Fluidmode for body fluids. The Body Fluidmode enumerates the WBC-BF,RBC-BF, MN%/#, PMN%/# and TC-BFparameters in cerebrospinal fluid (CSF),serous fluids (peritoneal, pleural) andsynovial fluids. Whole blood should becollected in K2 or K3EDTA anticoagulantand, Serous and Synovial fluids inK2EDTA anticoagulant to preventclotting of fluid. The use ofanticoagulants with CSF specimens isneither required nor recommended.The Sysmex® XN-L analyzer is aquantitative multi-parameter automatedhematology analyzer intended for in vitrodiagnostic use in screening patientpopulations found in clinical laboratories.The XN-L analyzer classifies andenumerates the following parameters invenous and capillary whole blood: WBC,RBC, HGB, HCT, MCV, MCH, MCHC,PLT, NEUT%/#, LYMPH%/#, MONO%/#, EO%/#, BASO%/#, IG%/#, RDW-CV,RDW-SD, MPV, RET%/#, IRF, RET-Heand has a Body Fluid mode for bodyfluids. The Body Fluid mode enumeratesthe WBC-BF, RBC-BF, MN%/#, PMN%/#, and TC-BF# parameters in CSF,peritoneal, pleural and synovial fluids.Whole blood should be collected in K2 orK3 EDTA anticoagulant and peritoneal,pleural and synovial fluids in K2EDTAanticoagulant to prevent clotting of fluid.The use of anticoagulants with CSFspecimens is neither required norrecommended.The performance of this device has notbeen established in pediatric patientsunder the age of 2 years.
SpecimenTypeWhole Blood and Body Fluids (CSF andPeritoneal, Pleural, Synovial Fluids)SAME
TestPrinciplePerforms hematology analyses accordingto the Hydro Dynamic Focusing (DCDetection), flow cytometry method (usinga semiconductor laser), and SLS-hemoglobin method.SAME
ParametersWhole Blood Mode:WBC, RBC, HGB, HCT, MCV, MCH,MCHC, PLT, NEUT%/#, LYMPH%/#,MONO%/#, EO%/#, BASO%/#, RDW-CV, RDW-SD, MPV, RET%/#, IRF,IG%/#, RET-He#Body Fluid Mode:WBC-BF, RBC-BF, MN%/#, PMN%/#,TC-BF#SAME
Predicate AnalyzerXN-Series (XN-10)aK112605Proposed AnalyzerXN-L
Item
ReagentsK112605CELLPACK® DCL (Diluent)CELLPACK® DFL (Diluent)Lysercell™ WDF (Lyse)Fluorocell™ WDF (Stain)Fluorocell™ RET (Stain)SULFOLYSER® (Lyse)SAME
AnalysisModesSampler Analysis Mode(rack autoloader)Whole Blood ModeManual Analysis ModeWhole Blood ModeLWBC Analysis ModePre-Dilute Analysis ModeBody Fluid ModeSAME
SampleAspiration/FluidicPathwaySingle PathwaySAME
MeasuringChannelsRBC/PLT, HGB, RET, WDFSAME
Controls/Calibrators/LinearityMaterialWhole BloodXN CHECK™ 3 Levels (K120742)XN CAL™ (K120745)Body FluidXN CHECK™ BF 2 Levels (K120744)Whole Blood LinearityRange Check X III (K960557)Retic Chex (K000115)SAME
CleaningDetergentCELLCLEAN AUTO™SAME
Software/HardwareRules based rerun/reflexSAME
Differences
ParametersPLT (PLT-F), NRBC%/#, IPFNot Available
ReagentsK112605Lysercell™ WNR (Lyse)Fluorocell™ WNR (Stain)Fluorocell™ PLT (Stain)Not AvailableNot AvailableNot Available
MeasuringChannelsWNR, PLT-FNot Available
ItemPredicate AnalyzerXN-Series (XN-10)aK112605Proposed AnalyzerXN-L
Controls/CalibratorsXN CAL™ PF – (K120747)Not Available
Not AvailableXN-L CHECK™ b
ThroughputWhole Blood Mode100 samples/hour maximum dependingon mode used.Whole Blood Mode60 samples/hour maximum depending onmode used.
Body Fluid ModeBody Fluid Mode
40 samples/hour maximum30 samples/hour maximum
SampleAspirationVolumesSampler Mode - 88 μLSampler Mode - 25 μL
Manual (Closed Cap) Mode - 88 μLManual (Closed Cap) Mode - 25 μL
Manual (Open Cap) Mode - 88 μLManual (Open Cap) Mode - 25 μL
Dilution Mode - 70 μLDilution Mode - 70 μL
Body Fluid Mode - 88 μLBody Fluid Mode - 70 μL

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Sysmex logo. The logo features a stylized graphic element in shades of blue and green, resembling a wave or a flowing ribbon. To the right of the graphic is the word "sysmex" in a sans-serif font, with the letters in a dark blue color.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the Sysmex logo. The logo consists of a stylized graphic element above the word "sysmex". The graphic element is a combination of blue and green shapes, with the blue shapes resembling waves and the green shape resembling a mountain. The word "sysmex" is written in blue, sans-serif font.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the Sysmex logo. The logo consists of a stylized graphic element in shades of blue and green, resembling a wave or a stylized leaf. To the right of the graphic is the word "sysmex" in lowercase, using a sans-serif font and a blue color that matches the graphic element.

8 Intended use for the predicate analyzer was cleared in submission K112605. All information listed for the predicate analyzer refers to the XN-10 module.

b Control material specific for the XN-L analyzer.

The XN-L analyzer's Indications for Use statement is similar to the predicate device with minor variation. The XN-L analyzer also has similar technological characteristics as the predicate device with minor variation. Both devices measure similar parameters and utilize most of the same reagents, controls, calibrators, and cleaning detergent. The datacollection software functionality, communication method with data management software functionality, monitor software, connectivity, and communication are similar to the predicate device with minor variation.

The XN-L analyzer differs from the predicate with a slower throughput and smaller sample aspiration volumes, fewer measuring channels, and fewer parameters measured, it is very similar in all other electronic and technological characteristics related to automated hematology measurements of the predicate device to assure equivalence. In addition. XN-L CHECK™is a control material which is specific for the XN-L analyzer. The XN-LCHECK™ or the XN CHECK™ may be used with the XN-L analyzer. Performance data are provided to support this. The results of all performance testing demonstrate substantial equivalence.

Summary of Performance Testing:

Clinical testing was conducted on the XN-L analyzer to show equivalent performance to the XN-Series analyzers. Testing included:

  • . Whole Blood Analysis
    • o Accuracy Evaluation
    • Precision Evaluation: Reproducibility o
    • o Precision Evaluation: Repeatability
    • o Linearity Evaluation

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains the Sysmex logo. The logo features a stylized graphic element in shades of blue and green, resembling a wave or a leaf. To the right of the graphic is the word "sysmex" in a sans-serif font, with the letters in a blue color that matches the graphic element.

  • Carryover Evaluation O
  • Stability Evaluation O
  • Verification of Reference Intervals O
  • o Limits of Blank, Detection, and Quantitation
  • Body Fluid Analysis
    • o Accuracy Evaluation
    • Precision Evaluation: Reproducibility O
    • o Precision Evaluation: Repeatability
    • o Linearity Evaluation
    • o Carryover Evaluation
    • o Stability Evaluation
    • o Limits of Blank, Detection, and Quantitation

Evaluation of the performance characteristics establishes that the performance, functionality, and reliability of the XN-L analyzer are substantially equivalent to the predicate device. The evaluation included accuracy, precision, linearity, carryover, stability, and Limits of Blank, Detection and Quantitation on whole blood and body fluid and verification of reference intervals on whole blood.

Conclusions:

The XN-L Automated Hematology analyzer and its predicate device, XN-Series modules (XN-10, XN-20) Automated Hematology analyzers (K112605), have similar Indications for Use, fundamental technology, principles of operation, and comparable performance characteristics. The modifications consist of a smaller analyzer with a slower throughput and smaller sample aspiration volumes.

Performance, verification, and validation testing were conducted to characterize the performance of the XN-L analyzer and the predetermined acceptance criteria were met. Results of this testing have documented that the XN-L analyzer is substantially equivalent to the XN-Series analyzers and is suitable for the labeled indication for use. The XN-L analyzer and the predicate device do not raise any questions regarding safety and effectiveness.

§ 864.5220 Automated differential cell counter.

(a)
Identification. An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.(b)
Classification. Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”