K Number
K160530

Validate with FDA (Live)

Date Cleared
2016-11-28

(277 days)

Product Code
Regulation Number
870.1120
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Med-link Disposable Blood Pressure Cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

Device Description

The proposed device, Med-link Disposable Blood Pressure Cuff is rectangle soft inelastic sleeve. There is a single-tube or twin-tube connected to a sealed gas chamber. It is wrapped around the patient's limb/thigh and secured by hook and loop closure. The device tubing is connected to a non-invasive blood pressure measurement system. It is available in various sizes for different limb range. It is provided non-sterile.

AI/ML Overview

This document is a 510(k) premarket notification for the Med-link Disposable Blood Pressure Cuff. It aims to demonstrate substantial equivalence to legally marketed predicate devices.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are primarily based on the FDA recognized consensus standard ISO 81060-1:2007/ AAMI/ANSI/ISO81060-1:2007/(R)2013 Non-Invasive Sphygmomanometers -Part 1: Requirements And Test Methods For Non-Automated Measurement Type, and ISO 10993 (Biological evaluation of medical devices) series for biocompatibility.

Acceptance Criteria CategorySpecific Acceptance Criteria (Standard Reference)Reported Device Performance (Summary)
PerformanceCompliance with ISO 81060-1:2007/ AAMI/ANSI/ISO81060-1:2007/(R)2013, Part 1: Requirements And Test Methods For Non-Automated Measurement Type"Performance testing was performed on the subject device and results were compared with predicate device. Tests were conducted following applicable procedures outlined in the FDA recognized consensus standard of AAMI/ANSI/ISO81060-1:2007/(R)2013, and results met all relevant requirements in the test standard."
BiocompatibilityISO 10993-1: Biological evaluation of medical devices-Part 1: Evaluation and testing"A series of safety, essential performance and in vitro biocompatibility tests were performed... The tests listed below were conducted in accordance with ISO 10993-1, ISO 10993-5, ISO 10993-10."
ISO 10993-5: Test for in vitro cytotoxicity"Cytotoxicity Test" performed. Data not explicitly detailed in summary but implied to meet criteria.
ISO 10993-10: Tests for Irritation And Skin Sensitization"Skin irritation Test" and "Skin Sensitization Test (the Guinea Pig maximization test)" performed. Data not explicitly detailed in summary but implied to meet criteria.
Pressure Limit0-300mmHg matching predicate devicesProposed device has a pressure limit of "0-300mmHg," which is identical to the primary and secondary predicate devices.
SterilityNon-sterile matching predicate devicesProposed device is "Non-sterile," which is identical to the primary and secondary predicate devices.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size used for the performance testing. It mentions "Performance testing was performed on the subject device," but does not specify the number of devices or measurements.

The data provenance is not specified regarding country of origin or specific study design (retrospective or prospective). The tests are described as "non-clinical testing," which suggests laboratory-based evaluations rather than human subject trials for the core performance and biocompatibility aspects.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This document describes the testing of a basic medical accessory (blood pressure cuff). The "ground truth" for its performance is established by objective measurements against a recognized standard (ISO 81060-1). There is no mention of experts establishing a "ground truth" in the way one might for diagnostic imaging or a subjective clinical assessment. The ground truth for biocompatibility is established through standardized laboratory tests.

4. Adjudication Method for the Test Set

Not applicable. The performance and biocompatibility tests are objective measurements against established standards, not subjective assessments requiring adjudication by multiple experts.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. This is a 510(k) submission for a blood pressure cuff, which is a physical accessory, not an AI or imaging-based diagnostic device. Therefore, an MRMC comparative effectiveness study, which typically evaluates human readers' performance with and without AI assistance, is not relevant or applicable here.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable. This device is a passive accessory and does not involve an algorithm. Its function is to interface with a blood pressure measurement system.

7. Type of Ground Truth Used

The ground truth used for performance testing (accuracy and reliability of pressure transmission, cuff inflation/deflation, etc.) is the adherence to the ISO 81060-1 standard's objective requirements and test methods.

For biocompatibility testing, the ground truth is established by the results of standardized laboratory tests (Cytotoxicity, Skin Irritation, Skin Sensitization) as defined by the ISO 10993 series.

8. Sample Size for the Training Set

Not applicable. This device is not an AI/machine learning algorithm and therefore does not have a "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is a symbol of medicine and healing. The logo also includes the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement around the caduceus symbol.

Public Health Service

November 28, 2016

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Shenzhen Med-link Electronics Tech Co., Ltd. Mina Cheng Regulatory Affairs Specialist 4th Fl, Bldg A, Yingtailong Industrial Prk, Dalang South Road, Longhua District Shenzhen, 518109 CN

Re: K160530

Trade/Device Name: Med-link Disposable Blood Pressure Cuff Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXQ Dated: December 22, 2015 Received: February 25, 2016

Dear Mina Cheng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Mude Jellm
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160530

Device Name Med-link Disposable Blood Pressure Cuff

Indications for Use (Describe)

Med-link Disposable Blood Pressure Cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for MedLinket. The logo has the word "MED" in red, with a blue swoosh underneath. To the right of "MED" is a red circle, followed by the word "LINKET" in blue, with a "TM" symbol in the upper right corner.

Shenzhen Med-link Electronics Tech Co., Ltd.

510(k) Summary

This 510(k) summary is being submitted in accordance with the requirements of 21 CFR§807.92. Type of submission: Traditional The assigned 510(k) number is:K160530

1. Submitter information

Manufacturer Name: Shenzhen Med-link Electronics Tech Co., Ltd. Address: 4th Fl, Bldg A, Yingtailong Industrial Park, Dalang South Road, Longhua District, Shenzhen, Guangdong, CHINA, 518109 Tel: 0086-755-61120299 Fax: 0086-755-61120055 Establishment Registration Number: 3006636961

2. Contact Person

Mina Cheng (Regulatory Affairs Specialist) E-mail: user22@med-linket.com

3. Date of preparation

Nov. 19, 2016

4. I Identification of the Device

Trade Name: Med-link Disposable Blood Pressure Cuff

Models as listed below:

Device Name/TypeDevice Model
Disposable BloodPressure CuffY000DSN1-1#, Y000DSN2-1#, Y000DSN1-2#, Y000DSN2-2#,Y000DSN1-3#, Y000DSN2-3#, Y000DSN1-4#, Y000DSN2-4#,Y000DSN1-5#, Y000DSN2-5#, Y000DC1, Y000DC2, Y000DSA1,Y000DSA2, Y000DA1, Y000DA2, Y000DAL1, Y000DAL2,Y000DLA1, Y000DLA2, Y000DT1, Y000DT2

Common Name: Non-invasive Blood Pressure Cuff

II

Classification Name: Blood Pressure Cuff

Device Classification:

Classification Regulation: 21 CFR 870.1120

K160530

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for MedLinket. The logo has the word "MED" in red, with the letters stacked on top of each other. There is a blue swoosh underneath the word "MED", and a red circle to the right of the swoosh. The word "LINKET" is in blue, and is to the right of the red circle, with the letters in a bold font. There is a trademark symbol to the right of the word "LINKET".

Product Code: DXQ

5. I Identification of the Predicate Device

No.Device NameCommonNameManufactureClassificationand CodeClassificiationregulation510(k)number
PrimaryPredicateDisposable/ReusableBlood Pressure CuffNon-invasiveBloodpressure cuffXUZHOUMAICUFFTECHNOLOGYCO.,LTD.Class II,DXQ21 CRF870.1120K151290
SecondaryPredicateUnimed DisposableBlood Pressure CuffNon-invasiveBloodpressure cuffUnimed Medicalsupplies Inc.Class II,DXQ21 CRF870.1120K120364

Table 2 Predicate Devices Information

Intended Use and Indications for Use of the Subject device 6.

Med-link Disposable Blood Pressure Cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonatal, pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

7. Device Description

The proposed device, Med-link Disposable Blood Pressure Cuff is rectangle soft inelastic sleeve. There is a single-tube or twin-tube connected to a sealed gas chamber. It is wrapped around the patient's limb/thigh and secured by hook and loop closure. The device tubing is connected to a non-invasive blood pressure measurement system. It is available in various sizes for different limb range. It is provided non-sterile.

No.ModelIntended Limb Range(Unit: cm)
1Y000DSN1-1#3 - 6
2Y000DSN2-1#3 - 6
3Y000DSN1-2#4 - 8
4Y000DSN2-2#4 - 8
5Y000DSN1-3#6 - 11

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Medlinket. The logo is red and blue, with the word "MED" in red and the word "LINKET" in blue. There is a red circle in the middle of the logo. The logo also has the trademark symbol.

Shenzhen Med-link Electronics Tech Co., Ltd.
No.ModelIntended Limb Range(Unit: cm)
6Y000DSN2-3#6 - 11
7Y000DSN1-4#7 - 14
8Y000DSN2-4#7 - 14
9Y000DSN1-5#8 - 15
10Y000DSN2-5#8 - 15
11Y000DC115 - 22
12Y000DC215 - 22
13Y000DSA117 - 25
14Y000DSA217 - 25
15Y000DA124 - 32
16Y000DA224 - 32
17Y000DAL128 - 37
18Y000DAL228 - 37
19Y000DLA132 - 42
20Y000DLA232 - 42
21Y000DT142 - 50
22Y000DT242 - 50

Table 4 Contact Components and Relevant Materials

Device TypeContact ComponentsContact Materials
Med-link Disposable BloodPressure CuffInner face of hoseNonwoven Polyester with aPVC laminate Backing

8. Non-clinical Testing

A series of safety, essential performance and in vitro biocompatibility tests were performed to assess the safety and effectiveness of Med-link Disposable Blood Pressure Cuff.

The tests listed below were conducted in accordance with

ISO 10993-1 Biological evaluation of medical devices-Part 1: Evaluation and testing

Biological Evaluation of medical devices-Part 5: Test for in vitro cytotoxicity ISO 10993-5

ISO 10993-10 Biological evaluation of medical devices- Part 10: Tests for Irritation And Skin Sensitization

ISO 81060-1:2007/ AAMI/ANSI/ISO81060-1:2007/(R)2013 Non-Invasive Sphygmomanometers -Part 1: Requirements And Test Methods For Non-Automated Measurement Type

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Medlinket. The logo has the word "MED" in red, with a blue swoosh underneath. To the right of the word "MED" is a red circle, followed by the word "LINKET" in blue. The letters "TM" are in the upper right corner of the word "LINKET".

  • Cytotoxicity Test .
  • Skin irritation Test
  • . Skin Sensitization Test (the Guinea Pig maximization test )

9. Technological Comparison with Predicate Device

ItemProposed DevicePrimary PredicateSecondary Predicate
Trade nameMed-link Disposable BloodPressure CuffDisposable Blood PressureCuffUnimed Disposable BloodPressure Cuff
510(K)SubmitterShenzhen Med-linkElectronics Tech Co., Ltd.XUZHOU MAICUFFTECHNOLOGY CO.,LTD.Unimed Medical supplies Inc.
510(K) NumberK151290K120364
ClassificationRegulation21CRF 870.112021CRF 870.112021CRF 870.1120
Classificationand CodeClass II,DXQClass II,DXQClass II,DXQ
Common nameNon-invasive Blood PressureCuffNon-invasive Blood PressureCuffNon-invasive Blood PressureCuff
Intended useMed-link Disposable BloodPressure Cuff is an accessoryused in conjunction withnoninvasive blood pressuremeasurement systems. Thecuff is non-sterile and forsingle-patient use. It isavailable in neonatal,pediatric and adult sizes. Thecuff is not designed, sold, orintended for use except asindicated.The disposable bloodpressure cuff is an accessoryused in conjunction withnoninvasive blood pressuremeasurement systems. Thecuff is non-sterile and forsingle-patient use. It isavailable in neonatal,pediatric and adult sizes.The Unimed disposable bloodpressure cuff is an accessoryused in conjunction withnoninvasive blood pressuremeasurement systems. Thecuff is non-sterile and forsingle-patient use. It isavailable in neonatal,pediatric and adult sizes. Thecuff is not designed, sold, orintended for use except asindicated.
PatientpopulationsAdults/PediatricsAdults/PediatricsAdults/Pediatrics
TubeconfigurationOne or two tubeOne or two tubeOne or two tube
SizeConform to AHA bladderConform to AHA bladderConform to AHAbladder
ItemProposed DevicePrimary PredicateSecondary Predicate
sizes recommendationssizes recommendationssizes recommendations
Neonatal 1# (3-6cm)Neonatal 1 (3-5.5 cm)Neonate 1 (3-6 cm)
Neonatal 2# (4-8cm)Neonatal 2 (4-8 cm)Neonate 2 (4-8 cm)
Neonatal 3# (6-11cm)Neonatal 3 (6-11 cm)Neonate 3 (6-11 cm)
Neonatal 4# (7-14cm)Neonatal 4 (7-13 cm)Neonate 4 (7-13 cm)
Infant (8-15cm)Infant (9-14.5 cm)Neonate 5 (8-15 cm)
Pediatric (15-22 cm)Pediatric (13-21.5 cm)Infant (9-14.8 cm)
Small Adult (17-25 cm)Small Adult (21-27 cm)Child (13.8-21 cm)
Adult (24-32cm)Adult (26-35.5 cm)Small Adult (20.5-28.5 cm)
Adult long(28-37cm)Adult (27.5-36.5 cm)
Large Adult (32-42cm)Large Adult(35-45 cm)Large Adult (35.5-46 cm)
Thigh (42-50cm)Large Adult Long(35.5-46cm)
Thigh (45-56cm)
Pressure limit0-300mmHg0-300mmHg0-300mmHg
SterilityNon-sterileNon-sterileNon-sterile
AAMI/ANSI/ISO81060-1:20
Non-clinicalRequirement07/(R)2013Comply with ISO 10993biocompatibility evaluationComply with ISO 10993biocompatibility evaluationComply with ISO10993biocompatibility evaluation

Table 4 Comparison to Primary Predicate Device

K160530

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Medlinket. The logo has the word "MED" in red, with a blue swoosh underneath it. To the right of the swoosh is a red circle, followed by the word "LINKET" in blue. There is a small "TM" symbol in the upper right corner of the word "LINKET".

Shenzhen Med-link Electronics Tech Co., Ltd.

Med-link Disposable Blood Pressure Cuff has the same intended use, basic construction and technology specification as the predicate device. They are wrapped around the patient's arm or thigh and secured by a hook and loop fastener commonly called Velcro.

We extend the size to satisfy some special patients such as different neonatal size. Based on the performance testing in this submission, the slight difference on the range of these blood pressure cuffs does not raise any safety or effectiveness issue.

10. Comparison of Performance with Predicate Device

Performance testing was performed on the subject device and results were compared with predicate device. Tests were conducted following applicable procedures outlined in the FDA recognized consensus standard of AAMI/ANSI/ISO81060-1:2007/(R)2013, and results met all relevant requirements in the test standard.

11. Conclusion

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Medlinket. The logo has the word "MED" in red, with a blue swoosh underneath it. To the right of the swoosh is a red circle, and then the word "LINKET" in blue. The letters "TM" are in the upper right corner of the word LINKET.

Shenzhen Med-link Electronics Tech Co., Ltd.

After analyzing safety, essential performance and biocompatibility testing data, it can be concluded that: Med-link Disposable Blood Pressure Cuff are safe and effective as the predicate device.

§ 870.1120 Blood pressure cuff.

(a)
Identification. A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.(b)
Classification. Class II (performance standards).