K Number
K160492
Device Name
Hospira Sapphire Sets
Manufacturer
Date Cleared
2016-12-06

(288 days)

Product Code
Regulation Number
880.5725
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Hospira Sapphire Sets are indicated for the delivery of fluids from a container to a patient's vascular system.
Device Description
Hospira Sapphire Sets are intended for use with the Sapphire Infusion System. Hospira Sapphire sets are comprised of various components including the following: male luer adapter with cap, female luer with cap, piercing pin connector, sapphire cassette, tubing, flow control device, in-line adapter, injection site assembly, check valve, and burette chamber. Hospira Sapphire sets are configured to ensure the intended use of the device is met. The sets are disposable devices for single patient use.
More Information

No
The summary describes a standard infusion set with mechanical components and does not mention any AI/ML capabilities or related performance metrics.

No
This device is for fluid delivery, not for treating a disease or condition.

No

The device is indicated for the delivery of fluids from a container to a patient's vascular system, which is a therapeutic function, not a diagnostic one.

No

The device description explicitly lists multiple hardware components (male luer adapter, female luer, piercing pin connector, sapphire cassette, tubing, flow control device, etc.) and describes it as a disposable device for single patient use, indicating it is a physical medical device, not software only.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is "for the delivery of fluids from a container to a patient's vascular system." This describes a device used for administering substances directly into the body, not for testing samples taken from the body.
  • Device Description: The components listed (tubing, luer adapters, cassette, etc.) are all consistent with an infusion set used for delivering fluids, not for performing diagnostic tests on biological samples.
  • No mention of testing biological samples: There is no indication that this device is used to analyze blood, urine, tissue, or any other biological sample for diagnostic purposes.

IVD devices are specifically designed to perform tests on samples taken from the human body to provide information for diagnosis, monitoring, or screening. This device's function is purely for the delivery of fluids.

N/A

Intended Use / Indications for Use

Hospira Sapphire Sets are indicated for the delivery of fluids from a container to a patient's vascular system.

Product codes (comma separated list FDA assigned to the subject device)

MRZ

Device Description

Hospira Sapphire Sets are intended for use with the Sapphire Infusion System. Hospira Sapphire sets are comprised of various components including the following: male luer adapter with cap, female luer with cap, piercing pin connector, sapphire cassette, tubing, flow control device, in-line adapter, injection site assembly, check valve, and burette chamber. Hospira Sapphire sets are configured to ensure the intended use of the device is met. The sets are disposable devices for single patient use.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

patient's vascular system

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

All materials of construction for Hospira Sapphire Sets remain the same as the predicate product. Additional data has been generated demonstrating that Hospira Sapphire sets, which are comprised of a combination of Hospira IV set components with the Sapphire Cassette, meet the applicable material test requirements for ISO 10993 as follows:

ISO Standard Biological Effect
10993-4 Hemocompatibility
10993-5 Cytotoxicity
10993-10 Sensitization
10993-11 Intracutaneous Irritation
Systemic Toxicity
Subacute Toxicity
Pyrogenicity

New performance data has been generated to ensure the Hospira Sapphire Sets, a combination of Hospira IV set components with the Sapphire Cassette, perform as intended in accordance with:
ISO 8536-4, 6.1 Particulate Contamination
ISO 8536-4, 6.2 Leakage
ISO 8536-4, 6.3 Tensile Strength
ISO 1135-4, 5.3 Tensile Strength

Additionally, flow rate accuracy testing using the proposed Hospira Sapphire Sets and the Sapphire Infusion System has been conducted. All testing is acceptable. The product Sterility Assurance Level is 10-6.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

Q Core Medical Sapphire Sets - K123049

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Hospira Plum Sets - K141789, Life Care PCA Infusion System - K043256

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 880.5725 Infusion pump.

(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three heads, representing the department's commitment to health, human services, and the future. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 6, 2016

Hospira, Inc. David Blonski Director Regulatory Affairs 375 North Field Drive Lake Forest, Illinois 60045

Re: K160492

Trade/Device Name: Hospira Sapphire Sets Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: MRZ Dated: December 1, 2016 Received: December 2, 2016

Dear David Blonski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

1

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/1/Picture/8 description: The image contains a signature and the name "Tina Kiang". There is also a "-s" below the name. The signature is a stylized loop with a dot above it.

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K160492

Device Name Hospira Sapphire Set

Indications for Use (Describe)

Hospira Sapphire sets are indicated for the delivery of fluids from a container to a patient's vascular system.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

Image /page/3/Picture/1 description: The image shows the Hospira logo. The logo features the company name "Hospira" in a sans-serif font, with the "H" slightly larger than the other letters. Above the name is an abstract graphic consisting of three curved lines that converge at the top, resembling a stylized representation of a plant or a medical symbol. The logo is presented in grayscale.

Section 5 510(k) Summary: K160492

A summary of 510(k) substantial equivalence information in accordance with the requirements of 21 CFR 807.92 for Hospira Sapphire Sets.

Submitter Information
NameHospira, Incorporated
AddressD-393, Bldg. H3
375 North Field Drive
Lake Forest, IL. 60046
Phone number24-212-5010
Mobile number224-515-6807
Fax number(224) 212-5401
Establishment Registration
Number3005579246 (Owner/Operator #9063339)
Name of contact personDavid Blonski, Director Regulatory Affairs
Date prepared2/19/2016
Name of device
Trade or proprietary nameHospira Sapphire Sets
Common or usual nameInfusion pump accessory
Classification nameInfusion Pump, 21 CFR 880.5725, Class II
Product Code(s)MRZ
Legally marketed device(s) to which
equivalence is claimedQ Core Medical Sapphire Sets - K123049 (Predicate 510(k))
Hospira Plum Sets - K141789 (reference 510(k) for technological characteristics)
Life Care PCA Infusion System - K043256 (reference 510(k) for technological characteristics)
Reason for 510(k) submissionNew IV administration sets to be used with Q Core Medical Ltd.'s Sapphire Infusion System.
Device descriptionHospira Sapphire Sets are intended for use with the Sapphire Infusion System. Hospira
Sapphire sets are comprised of various components including the following: male luer adapter
with cap, female luer with cap, piercing pin connector, sapphire cassette, tubing, flow control
device, in-line adapter, injection site assembly, check valve, and burette chamber. Hospira
Sapphire sets are configured to ensure the intended use of the device is met. The sets are
disposable devices for single patient use.
Intended Use of DeviceHospira Sapphire sets are intended for use with the Sapphire Infusion System to deliver fluids
from a container to a patient's vascular system. Hospira Sapphire Sets are indicated for the
delivery of fluids from a container to a patient's vascular system.
Indication for useHospira Sapphire Sets are indicated for the delivery of fluids from a container to a patient's
vascular system.

4

Image /page/4/Picture/1 description: The image shows the Hospira logo. The logo features the company name "Hospira" in a sans-serif font, with the "H" slightly larger than the other letters. Above the name is an abstract graphic consisting of three curved lines converging at the top, resembling a stylized plant or a modern architectural element. The logo is presented in grayscale.

Summary of the technological characteristics of the device compared to the predicate device
CharacteristicPredicateProposed Device
Indications for UseThe Q Core Sapphire infusion pump is
intended for the controlled delivery through
intravascular,
subcutaneous, intra-arterial and epidural
routes. The pump is designed to deliver
saline, Total Parenteral
Nutrition, lipids, IV medication, epidural
medication, blood and blood products.
The Sapphire pump includes the following
infusion modes for all intended uses:
Continuous, Intermittent,
TPN, PCA, Multi-step, and Epidural.
The pump is intended to be used by both
licensed health care professionals in a
clinical environment, and
home users in an ambulatory environment.
The Sapphire pump is designed to follow the
patient through
the various care areas, and in pre-hospital
medical ground transportation.
The dedicated Q Core administration sets for
the Sapphire pump are intended for single-
patient use and
single-use only.Hospira Sapphire Sets are indicated for the
delivery of fluids from a container to a
patient's vascular system.
Discussion of change in
indication for useThe differences between the predicate and proposed indications for use stated above are not
critical to the intended therapeutic, diagnostic, prosthetic, or surgical use of the proposed
device. The predicate indication for use stated above is primarily related to use of the
Sapphire Infusion pump with an inclusionary statement that the Q Core administration sets
for the Sapphire pump are intended for single-patient use and single-use only. The Hospira
Sapphire sets are intended for use with the Sapphire pump and as such are a component of
the predicate device system. The proposed sets are also intended for single patient use and
single-use only. The proposed Hospira Sapphire sets have the same intended use, and as
indicated below are tested to the same performance standards as the predicate sets and
therefore do not affect the safety and effectiveness predicate device system.
Design and Materials of
ConstructionThe design and materials of construction of
all components for the set, with the
exception of the Sapphire Cassette, are as
cleared under the 510(k)s K141789 and
K043256 which are reference 510(k)s for
technological characteristics.
The design and materials of construction of
the Sapphire Cassette are as cleared under
the predicate Q Core Medical Ltd. 510(k)
K123049.The design and materials of construction for
all components of the sets remain the same
as the predicate product.
Summary of non-clinical tests
for determination of substantial
equivalenceAll materials of construction for Hospira
Sapphire Sets meet the applicable material
test requirements for ISO 10993 as
demonstrated in the predicate 510(k)s.All materials of construction for Hospira
Sapphire Sets remain the same as the
predicate product. Additional data has been
generated demonstrating that Hospira
Sapphire sets, which are comprised of a
combination of Hospira IV set components
with the Sapphire Cassette, meet the
applicable material test requirements for
ISO 10993 as follows:
ISO Standard Biological Effect
10993-4                Hemocompatibility
10993-5                Cytotoxicity
10993-10               Sensitization
10993-11               Intracutaneous Irritation
Systemic Toxicity
Subacute Toxicity
Pyrogenicity
Performance testing for all Hospira Sapphire
set components, with the exception of the
Sapphire Cassette, was conducted as
indicated in the predicate 510(k)s to ensure
the device performs as intended in
accordance with applicable standards. All
testing is acceptable.

Performance testing for the Q Core Medical
Ltd. Sapphire Cassette , was conducted as
indicated in the predicate 510(k) to ensure
the cassette performs with the Sapphire
Pump as intended in accordance with
applicable standards.

All testing is acceptable. | New performance data has been generated
to ensure the Hospira Sapphire Sets, a
combination of Hospira IV set components
with the Sapphire Cassette, perform as
intended in accordance with:
ISO 8536-4, 6.1 Particulate Contamination
ISO 8536-4, 6.2 Leakage
ISO 8536-4, 6.3 Tensile Strength
ISO 1135-4, 5.3 Tensile Strength

Additionally, flow rate accuracy testing
using the proposed Hospira Sapphire Sets
and the Sapphire Infusion System has been
conducted.

All testing is acceptable.

The product Sterility Assurance Level |
| | The product Sterility Assurance Level is 10-6 | is 10-6 |

5

Image /page/5/Picture/1 description: The image shows the Hospira logo. The logo features the company name "Hospira" in a sans-serif font, with the "H" slightly larger than the other letters. Above the name is an abstract graphic consisting of three curved lines that converge at the bottom and spread out at the top, resembling stylized blades of grass or a stylized flower. The logo is presented in grayscale.

Conclusion

Hospira Sapphire Sets meet the functional claims and intended use as described in the product labeling. Hospira Sapphire Sets are substantially equivalent to the predicate device.