(12 days)
The Hospira LifeCare® PCA3 Infusion System is intended for accurate, volumetric, infusion of analgesic drugs by continuous or patient-demanded (PCA) intravenous administration. The LifeCare PCA® Infusion System is also indicated for short-term (less than 96 hours) continuous epidural administration of analgesic drugs.
The Hospira LifeCare® PCA3 Infusion System is a microprocessor controlled, pole mounted, standalone, electromechanical infusion pump that allows a patient to self administer, analgeric usines, a patient pendant, within physician prescribed, programmed parameters. A stepper motor exerts pressure on an inserted drug vial to control the infusion of analgesic into a patient. The infuser is powered from an AC power source and has an internal battery to maintain operation for short periods when an AC power source is not available. The infuser will accept a pre-filled drug vial manufactured by Haspire and includes one sterile empty vial that can be filled by a hospital's pharmacy.
The provided text is a 510(k) summary for the Hospira LifeCare® PCA3 Infusion System. It details the device, its intended use, and a comparison to predicate devices to establish substantial equivalence. However, this document does not contain information about specific acceptance criteria, a study proving the device meets those criteria, or details regarding sample sizes for testing or training sets, ground truth establishment, expert involvement, or MRMC studies.
The document states that the device "meets the functional claims and intended use as described in the product labeling" and that "The proposed modifications do not raise new issues of safety and/or effectiveness." It also makes claims of substantial equivalence based on:
- Same intended use
- Same indication for use
- Same fundamental technology and operating principle
- Same physical, operational, environmental and performance attributes
- Same materials of construction for infuser components and administration sets
It lists "Performance Features" such as "Delivery Rates, Dose Units, Delivery Accuracy, Delivery Modes, Occlusion Pressure Limits, Alarm Types and Conditions" as being "Same" as the predicate devices, implying these are the performance criteria that were considered. However, explicit numerical acceptance criteria and the results of a study demonstrating these are not present in this summary.
Therefore, I cannot provide the requested information for the following points as they are not present in the provided text:
- A table of acceptance criteria and the reported device performance
- Sample size used for the test set and the data provenance
- Number of experts used to establish the ground truth for the test set and their qualifications
- Adjudication method for the test set
- If a multi reader multi case (MRMC) comparative effectiveness study was done, and its effect size
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The sample size for the training set
- How the ground truth for the training set was established
{0}------------------------------------------------
DEC - 6 2004
1043256
Confidential
510(K) Summary
- Name of Submitter:
Hospira, Incorporated 275 North Field Drive Lake Forest, Illinois 60045
Owner/Operator # 9063339
2. Manufacturer and Establishment Registration Number:
Hospira, Inc. - Morgan Hill 755 Jarvis Drive Morgan Hill, CA 95037
Establishment Registration # 2921482
-
- Proprietary or Trade Name of Proposed Device: Hospira LifeCare® PCA3 Infusion System
-
- Common Name: Infusion Pump
- Device Classification, Pancode and ProCode: Class II, 80-FRN (Infusion Pump) Class II, 80-FPA (Administration Sets)
-
- Performance Standards: No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for intravenous infusion pumps. Infusion pumps are listed in 21 CFR 880.5725.
-
- Intended Use: The Hospira LifeCare® PCA3 Infusion System is intended for accurate, volumetric, infusion of analgesic drugs by continuous or patient-demanded (PCA) intravenous administration.
8. Indications for Use:
The LifeCare® PCA3 Infusion System is also indicated for short-term (less than 96 hours) continuous epidural administration of analgesic drugs. The epidural route can be used to provide anestbessa or analgesia. The pump must be used with sterile, dedicated, LifeCare PCA administrations sufer
9. Proposed Device Description:
The Hospira LifeCare® PCA3 Infusion System is a microprocessor controlled, pole mounted, standalone, electromechanical infusion pump that allows a patient to self administer, analgeric usines, a patient pendant, within physician prescribed, programmed parameters. A stepper motor exerts pressure on an inserted drug vial to control the infusion of analgesic into a patient. The infuser is powered from an AC power source and has an internal battery to maintain operation for short periods when an AC power source is not available. The infuser will accept a pre-filled drug vial manufactured by Haspire and includes one sterile empty vial that can be filled by a hospital's pharmacy.
As of May 03, 2004, both the infuser and the dedicated LifeCare® PCA administration sets are manufactured and distributed by Hospira Incorporated, formerly the Hospital Products Division of Abbott Laboratories.
Volume 1 of 1
{1}------------------------------------------------
10. Predicate Device Information:
Infusion pumps cleared for commercial distribution and determined to be appropriate for use as predicates are summarized in the following table.
| 510(k) # | Product Name | Clearance Date |
|---|---|---|
| K042800 | LifeCare PCA® Infusion System with Hospira MedNet™Software | 10/28/2004 |
| K022203 | Abbott LifeCare® PCA3 infusion System | 08/01/2002 |
11. Comparison to Legally Marketed Device(s)
| Factors | Subject Device(s)Hospira LifeCare® PCA3 Infusion System | Predicate Device(s)Abbott LifeCare® PCA3Infusion SystemandLifeCare® PCA InfusionSystem with Hospira™MedNet Software |
|---|---|---|
| Intended Use | Intended for volumetric, infusion of analgesic drugs bycontinuous or patient-demanded (PCA) intravenousadministration. It is indicated for short-term (less than96 hours) continuous epidural administration ofanalgesic drugs. | Same |
| Indications for Use | Hospital clinical settings using sterile, dedicated,LifeCare® PCA administration sets. | Same |
| Operating Principle | Advancement of a syringe-like drug vial by a steppermotor driven plunger to expel drug from the vial into anattached sterile intravenous administration set designed tobe used exclusively with LifeCare® PCA infusers. | Same |
| Administration Sets andFluid Contact Materials | Sterile, dedicated, non-pyrogenic, latex-free LifeCare®PCA administration sets. | Same |
| Physical Features | Materials, Size, Weight, Power Sources, Battery Type,Power Cord | Same |
| Environmental Features | Operating Temperature, Storage Temperature, RelativeHumidity, Pressure | Same |
| Performance Features | Delivery Rates, Dose Units, Delivery Accuracy,Delivery Modes, Occlusion Pressure Limits, AlarmTypes and Conditions | Same |
| Wireless MedNet™Software Compatibility | No wireless/network capability.Not compatible with MedNet™ Software. | Same as AbbottLifeCare® PCA3Infusion System |
{2}------------------------------------------------
12. Statement of Safety and Effectiveness
The Hospira LifeCare® PCA3 Infusion System meets the functional claims and intended use as described in the product labeling. The proposed modifications do not raise new issues of safety and/or effectiveness.
13. Statement of Substantial Equivalence:
The Hospira LifeCare® PCA3 infusion System is substantially equivalent in terms of safety and effectiveness to the predicate Abbott LifeCare® PCA3 Infusion System and the LifeCare PCA® Infusion System with Hospira MedNet™ Software based on the following characteristics.
-
- Same intended use,
-
- Same indication for use,
-
- Same fundamental technology and operating principle.
-
- Same physical, operational, environmental and performance attributes, and
-
- Same materials of construction for infuser components and administration sets.
The claim for substantial equivalence is supported by the information provided in this 510(k) submission.
Prepared and submitted by:
Patricia Meleski 11/23/04
Patricia Melerski Manager, Global Device Regulatory Affairs Hospira, Inc. 275 North Field Drive Lake Forest, IL 60045 Phone: 224/212-4880 Fax: 224/212-5401
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC - 6 2004
Ms. Patricia Melerski Manager, Global Device Regulatory Affairs Hospira, Incorporated 275 North Field Drive, Dept. 389, Bldg. H2 Lake Forest, Illinois 60045
Re: K043256
Trade/Device Name: Hospira LifeCare® PCA3 Infusion System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: MEA Dated: November 23, 2004 Received: November 24, 2004
Dear Ms. Melerski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Melerski
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Suxen Runner
(Chin Lin, Ph.D.
(Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
:
:
:
| 510(k) Number (if known) | K043256 |
|---|---|
| Device Name: | Hospira LifeCare® PCA3 Infusion System |
Indications for Use
The Hospira LifeCare® PCA3 Infusion System is indicated for accurate, volumetric, infusion of analgesic drugs by continuous or patient-demanded (PCA) intravenous administration. The LifeCare PCA® Infusion System is also indicated for short-term (less than 96 hours) continuous epidural administration of analgesic drugs.
| Prescription Use X(Part 21 801 Subpart D) | AND/OR | Over-The_Counter Use(Part 21 CFR 807 Subpart C) |
|---|---|---|
| ------------------------------------------------- | -------- | ----------------------------------------------------- |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
| 510(k) Number | K043256 |
|---|---|
| --------------- | --------- |
Volume 1 of 1
·
§ 880.5725 Infusion pump.
(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).