AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Prevena Plus Incision Management System is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment, and removing exudate via the application of negative pressure wound therapy.

Device Description

The Prevena Plus Incision Management System provides surgical incision management via the application of negative pressure wound therapy over an incision site that has been surgically closed with sutures or staples. The system is applied to the incision site immediately after surgery for a minimum of 2 days up to a maximum of 7 days depending on the surgeon's preference. The Prevena Plus Incision Management System consists of: A Prevena Dressing (Prevena Plus Peel & Place Dressing or Prevena Plus Customizable Dressing) and One of the following KCI Negative Pressure Wound Therapy Units: Prevena Plus 125 Therapy Unit*, ActiV.A.C. Therapy Unit, V.A.C. Freedom Therapy Unit, InfoV.A.C. Therapy Unit, V.A.C. ATS Therapy Unit, V.A.C. Ulta Therapy Unit. * The Prevena Plus 125 Therapy Unit is a slightly modified V.A.C.Via Therapy Unit, previously cleared under K132741, that provides continuous negative pressure at -125mmHg only. The V.A.C.Via Therapy Unit serves as a reference device for 510(k) K153199.

AI/ML Overview

This document describes a 510(k) premarket notification for the Prevena Plus Incision Management System. It's a submission to the FDA to demonstrate substantial equivalence to a legally marketed predicate device. This type of submission relies on showing that the new device is as safe and effective as a previously approved device, often through a comparison of technological characteristics and performance data, rather than extensive clinical trials for new medical devices.

Therefore, the typical acceptance criteria and study design elements you'd find for an AI/ML device (e.g., sample sizes for test/training sets, expert ground truth, MRMC studies) are not directly applicable in the same way here. This is a conventional medical device submission focused on physical and functional equivalence to an existing device.

Here's an interpretation based on the provided text, addressing your points where possible within the context of a 510(k) submission for a non-AI device:

1. A table of acceptance criteria and the reported device performance

For this 510(k) submission, the "acceptance criteria" revolve around demonstrating the new device performs equivalently to the predicate device in key functional areas. The "reported device performance" is the result of non-clinical bench testing.

Acceptance Criteria (Equivalence to Predicate/Reference)Reported Device Performance
Negative Pressure Delivery: Maintain negative pressure within specification at -125 mmHg, continuous mode for 7 days, under wet conditions and maximum air leak, across the full length of the dressing.The average distribution of negative pressure across the full length of the Prevena Plus Customizable Dressing on a simulated wound bed under wet conditions and at maximum air leak demonstrated that the Prevena Plus 125 Therapy Unit and tubing set were able to provide negative pressure within specification at -125 mmHg, continuous mode for 7 days, equivalent to both the predicate and reference therapy units.
Alarm Functionality: Provide alarms for leak, blockage, and full canister within specification.The Prevena Plus 125 Therapy Unit was shown to provide alarms for leak, blockage and full canister within specification when connected to the Prevena Plus Customizable Dressing and tubing set.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: The document does not specify a "sample size" in the context of a patient-based test set, as this was a non-clinical bench study. The "test set" consisted of simulated wound beds and the components of the device (Prevena Plus Customizable Dressing, Prevena Plus 125 Therapy Unit, tubing set).
  • Data Provenance: The studies were non-clinical tests (bench testing/simulations) conducted to support the substantial equivalence determination. The country of origin is not explicitly stated, but the submission is from KCI USA, Inc. in San Antonio, TX, suggesting the testing was likely conducted in the US or by KCI-affiliated entities. It is prospective in the sense that the testing was performed to support this specific submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. As a non-clinical bench study, ground truth in terms of expert clinical assessment is not established in the same way as an AI/ML diagnostic or prognostic device. The "ground truth" for the bench tests would be the established engineering specifications and validation methods for negative pressure delivery and alarm functionality.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. This was a non-clinical bench study testing physical and functional specifications. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving interpretation of data (e.g., radiology images) by multiple experts.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is a negative pressure wound therapy system, not an AI/ML diagnostic or interpretive device that would involve human readers. Therefore, an MRMC study is not relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable directly in the AI/ML sense. However, the performance data presented is essentially "standalone" device performance, as it evaluates the device's ability to meet its technical specifications (-125 mmHg pressure, alarm function) without human patient interaction as part of the primary measurement.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For the non-clinical tests, the "ground truth" would be the engineering specifications and established validation protocols for negative pressure wound therapy systems. This includes:
    • Target negative pressure level (-125 mmHg).
    • Duration of therapy (7 days).
    • Conditions for testing (simulated wound bed, wet conditions, maximum air leak).
    • Functional requirements for alarms (leak, blockage, full canister).

8. The sample size for the training set

  • Not applicable. This device is a physical medical device, not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established

  • Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, composed of three stylized human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 29, 2016

Kinetic Concepts Incorporated USA Ms. Margaret Marsh Technical Director, Regulatory Affairs 6203 Farinon Drive San Antonio, Texas 78249

Re: K153199

Trade/Device Name: Prevena Plus Incision Management System Kit With Peel & Place Dressing, Prevena Plus Peel & Place Dressing 5-pack, Prevena Plus Incision Management System Kit With Customizable Dressing. Prevena Plus Customizable Dressing 5-pack

Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: February 1. 2016 Received: February 3, 2016

Dear Ms. Marsh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

David Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K153199

Device Name

Prevena Plus Incision Management System with Peel & Place and Customizable Dressings

Indications for Use (Describe)

The Prevena Plus Incision Management System is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment, and removing exudate via the application of negative pressure wound therapy.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY Prevena Plus Incision Management System

Submitter Information [21 CFR 807.92(a)(1)]
NameKCI USA, Inc. (Kinetic Concepts, Inc.)
Address6203 Farinon DriveSan Antonio, TX 78249
Phone number210-255-6481
Fax number210-255-6727
EstablishmentRegistration Number3005178245
Name of contact personMargaret Marsh, Technical Director, Regulatory Affairs
Date preparedFebruary 22, 2016
Name of the device [21 CFR 807.92(a)(2)]
Trade or proprietarynamePrevena Plus Incision Management System with Peel & Place andCustomizable Dressings
Common or usual nameNegative Pressure Wound Therapy System
Classification nameNegative Pressure Wound Therapy Powered Suction Pump (andcomponents)
Classification panelGeneral and Plastic Surgery
Regulation878.4780
Regulatory ClassII
Product Code(s)OMP
Legally marketeddevice(s) to whichequivalence is claimed[21 CFR 807.92(a)(3)]Prevena Incision Management System with Peel & Place and CustomizableDressings (as cleared under K141017)
Device description[21 CFR 807.92(a)(4)]The Prevena Plus Incision Management System provides surgical incisionmanagement via the application of negative pressure wound therapy over anincision site that has been surgically closed with sutures or staples. Thesystem is applied to the incision site immediately after surgery for a minimumof 2 days up to a maximum of 7 days depending on the surgeon's preference.The Prevena Plus Incision Management System consists of:A Prevena Dressing (Prevena Plus Peel & Place Dressing or Prevena Plus Customizable Dressing) and One of the following KCI Negative Pressure Wound Therapy Units: Prevena Plus 125 Therapy Unit* ActiV.A.C. Therapy Unit V.A.C. Freedom Therapy Unit InfoV.A.C. Therapy Unit V.A.C. ATS Therapy Unit V.A.C. Ulta Therapy Unit * The Prevena Plus 125 Therapy Unit is a slightly modified V.A.C.Via TherapyUnit, previously cleared under K132741, that provides continuous negativepressure at -125mmHg only. The V.A.C.Via Therapy Unit serves as areference device for 510(k) K153199.

{4}------------------------------------------------

:take on the street to MIN.:2008/02/2..**********:
11. PersonalComments of the consisted to the first to the first:Contraction of the ResearchThe submit the submit the subject to the..
.Control Concession ComparisSchool proportunity. .Coupon Children
1994/04. .BARRAND MARKETthe successful the production of the first of the country of the county of.Concession Comprehensive ChildrenComments of Children.Production.PARTAL PROPERTY.:..Actually as the first to the.
000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000.were and the state the state of the specifically and the first offor any the promise for the first from the first· September 1988 - 1988 - 1998.******************************************************************************************************************************************************************************:Company of Children Star:1933 - 1SchoolstrapperChildren Children Children ChildrenCallery College.

The Prevena Plus Incision Management System is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudate via the application of negative pressure wound therapy.

Summary of the technological characteristics of the device compared to the predicate device [21 CFR 807.92(a)(6)]

Negative Pressure Wound Therapy is the technological principal for both the subject and predicate Negative Produce Wound Though is the tosting on incision site that is closed via staples or sutures helps draw the incision edges together and removes fluid from the incision site. The occlusive drape of the dressing provides a negative pressure environment and protects the incision from external contamination.

The subject device system (Prevena Plus Incision Management System) is comprised of previously cleared components as shown below:

ComponentPredicate Device:Prevena IncisionManagement System(K141017)Reference Device:V.A.C. Via NegativePressure Wound TherapySystem(K132741)Reference Device:V.A.C. Negative PressureWound Therapy Systems(K120033)
Indicationsfor UseIdentical to PredicateNot applicableNot applicable
DressingsIdentical to the PredicatePrevena Peel & Place andCustomizable DressingsNot applicableNot applicable
TherapyUnits andCanistersIdentical to the PredicateActi V.A.C., Info V.A.C.,V.A.C.Ulta and V.A.C.Freedom Therapy UnitsThe Prevena Plus 125Therapy Unit is identical tothe reference V.A.C. ViaTherapy Unit clearedunder (K132741), except forremoval of options forintermittent and -75 mmHgtherapyNot applicable
Tubing SetNot applicableNot applicableSensaT.R.A.C. tubing set
LabelingIncludes PredicatePrevena IncisionManagement Systemsafety information anddressing applicationinstructionsIncludes ReferenceV.A.C. Via Therapy Unitoperating instructions andalarm resolutioninformation.Not applicable

{5}------------------------------------------------

Summary of the technological characteristics of the device compared to the predicate device, continued

At a high level, the subject device and predicate device are based on the following same technological elements:

  • The dressings that are applied over the incision site in the operating room are identical. One of the . following dressings may be selected by the surgeon, based on incision length and geometry:
    • The Prevena Plus Peel & Place Dressing which can be used for linear incisions up to 8 inches, or o
    • The Prevena Plus Customizable Dressing which can be configured for non-linear incisions or linear o incisions longer than 8 inches
  • A negative pressure pump (therapy unit) is required that can provide -125 mmHg of negative pressure . continuously to the dressing for a maximum of 7 days.
  • The dressings are connected to the therapy unit via a disposable canister and tubing set. .
  • Incision fluid is collected into the disposable canister .
  • The therapy unit provides alarms that indicate when negative pressure wound therapy may be . compromised (e.g., visual and audible alarms indicating an air leak in the system or when the canister is full).

The following technological differences exist between the subject and predicate devices:

  • Connection between the dressings and therapy unit canisters is via a SensaT.R.A.C. tubing set rather . than by a Prevena tubing set. This allows for pressure sensing at the incision site for all of the indicated therapy units.
  • Use of the Prevena Plus 125 Therapy Unit requires a connector (Prevena Plus Connector) to connect the . SensaT.R.A.C. tubing to the canister. No tubing connector is required for use of all of the other indicated therapy units.
  • Comparing the Prevena Plus 125 Therapy Unit to the Prevena 125 Therapy Unit, there are the following . technological differences:
    • The Prevena Plus 125 Therapy Unit is both AC and battery powered, whereas the Prevena 125 o Therapy Unit is powered by batteries only.
    • The Prevena Plus 125 Therapy Unit is provided with either a 150 or 250 cc canister, whereas the O Prevena 125 Therapy Unit has a 45 cc canister.
    • The Prevena Plus 125 Therapy Unit provides a shared blockage/canister full alert whereas the O Prevena 125 Therapy Unit has a canister full alert.

Performance Data [21 CFR 807.92(b)]

Summary of non-clinical tests conducted for determination of substantial equivalence [21 CFR 807.92(b)(1)]

The following performance data were provided in support of the substantial equivalence determination.

  • The average distribution of negative pressure across the full length of the Prevena Plus Customizable . Dressing on a simulated wound bed under wet conditions and at maximum air leak demonstrated that the Prevena Plus 125 Therapy Unit and tubing set were able to provide negative pressure within specification at -125 mmHg, continuous mode for 7 days, equivalent to both the predicate and reference therapy units.
  • The Prevena Plus 125 Therapy Unit was shown to provide alarms for leak, blockage and full canister . within specification when connected to the Prevena Plus Customizable Dressing and tubing set.

Summary of clinical tests conducted for determination of substantial equivalence or of clinical information [21 CFR 807.92(b)(2)]

No clinical tests were necessary

Conclusions drawn [21 CFR 807.92(b)(3)]

The Prevena Pius Incision Management System and its predicate, the Prevena Incision Management System (K141017) are substantially equivalent in terms of safety, function and indications for use.

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.