(213 days)
No reference devices were used in this submission.
No
The document describes a modular system for measuring micro-vascular perfusion and pressure using established technologies like Laser Doppler and pressure cuffs. There is no mention of AI, ML, or any related concepts in the intended use, device description, or performance studies.
No.
The device is intended for measuring, monitoring, and evaluating physiological parameters for diagnosis, and not for treating or providing therapy.
Yes
The device is explicitly stated to be "intended for Pulse Volume Recording (PVR) on human limbs and digits for diagnosis and evaluation of vascular disease in adults" and is described as offering "several tests for diagnosis and evaluation of vascular disease."
No
The device description clearly states it is a "modular multichannel system" consisting of a "Main Unit (PF 6001)" and various "function units" (PF 6010, PF 6050, PF 6040). It also describes the use of "optical fibre" and a "laser" for the Laser Doppler technology, indicating significant hardware components.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The PeriFlux 6000 system, with its various units, measures physiological parameters directly on or within the body (in vivo).
- PF 6010 measures micro-vascular perfusion in skin, muscle, and animal tissues using Laser Doppler technology applied to the tissue.
- PF 6050 measures pressure cuffs and performs Pulse Volume Recording (PVR) on limbs and digits.
- PF 6040 monitors transcutaneous partial pressures of oxygen and carbon dioxide through the skin.
- Lack of Sample Analysis: The device does not analyze samples taken from the body. It interacts directly with the body to obtain measurements.
Therefore, the PeriFlux 6000 system falls under the category of in vivo diagnostic devices rather than in vitro diagnostic devices.
N/A
Intended Use / Indications for Use
PeriFlux 6000 equipped with PF 6010 is intended for measuring micro-vascular perfusion in skin and muscle in humans. It is also intended for measuring micro-vascular perfusion in all tissues in animals for research purposes. The PF 6010 is also intended for evaluating tissue response in skin to local heating and providing temperature stabilization of skin at blood perfusion measurement.
PeriFlux 6000 equipped with PF 6050 is intended for measuring the pressure cuffs, to simplify simultaneous measurements of perfusion and pressure when used in conjunction with a laser Doppler perfusion monitor. PeriFlux 6000 equipped with PF 6050 is also intended for Pulse Volume Recording (PVR) on human limbs and digits for diagnosis and evaluation of vascular disease in adults. It is not intended for use on neonates and pregnant women.
PeriFlux 6000 equipped with PF 6040 is intended for continuous non-invasive transcutaneous monitoring of the partial pressures of oxygen and/or carbon dioxide. It is intended for use on neonates, pediatrics, and adults not under gas anesthesia.
Product codes
DPW
Device Description
The PeriFlux 6000 is a modular multichannel system that offers several tests for diagnosis and evaluation of vascular disease in one instrument. It consists of a Main Unit (PF 6001), which accommodates up to eight different function units of the same type or of different types enabling simultaneous measurements of several parameters. Currently, one type of unit –PF 6040 tcpO2/pCO2 Unit– is available for the US market. Two more units have been developed:
- PF 6010 LDPM/Temp Unit
- PF 6050 Pressure Unit
The PeriFlux 6000 equipped with PF 6010 function units measures microcirculatory blood flow using Laser Doppler technology. In combination with the PF 6050 function unit to inflate blood pressure cuffs and to register the cuff pressure, it enables a range of different tests, including ankle pressure, toe pressure, skin perfusion pressure (SPP), pulse volume recording (PVR), segmental pressures, heat provocations and post-occlusive reactive hyperemia (PORH).
In Laser Doppler Perfusion Monitoring, an optical fibre leads light generated by a laser to the LDPM probe tip, which rests against the tissue. The beam of light will enter the tissue and become scattered. Blood cells moving within the volume illuminated by the beam will cause the light to change frequency. This change in frequency is called a Doppler shift and it is used to calculate the blood flow in the illuminated tissue.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
skin, muscle, human limbs and digits
Indicated Patient Age Range
PeriFlux 6000 equipped with PF 6050: adults. Not intended for use on neonates and pregnant women.
PeriFlux 6000 equipped with PF 6040: neonates, pediatrics, and adults not under gas anesthesia.
Intended User / Care Setting
professional use in a hospital environment, laboratory environment and in a hospital intensive care unit environment.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Performance Testing - Bench: Laser Doppler Perfusion Monitoring is an intricate measurement, involving more electronics and calculations than measurements of the other parameters. Also, the PF 6010 measures over a higher frequency bandwidth than the PF 5010. We have therefore tested how the clinically relevant parameter, blood perfusion, correlates between these two systems. The tests show that the signals correlate very well between the two systems and that they can be regarded as equivalent.
Performance Testing - Animal: No performance tests on animals are submitted.
Performance Testing - Clinical: No clinical performance tests are submitted.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
K974285, K932068, K011899, K131253
Reference Device(s)
No reference devices were used in this submission
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.2100 Cardiovascular blood flowmeter.
(a)
Identification. A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, with three overlapping profiles suggesting a sense of community and support. The profiles are rendered in a dark color, creating a strong contrast against the white background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 5, 2016
Perimed AB Jimmy Bakker Regulatory Affairs Officer Datavägen 9A SE-175 43 Järfälla. Sweden
Re: K152930
Trade/Device Name: PeriFlux 6000 Regulation Number: 21 CFR 870.2100 Regulation Name: Cardiovascular Blood Flowmeter Regulatory Class: Class II Product Code: DPW Dated: April 5, 2016 Received: April 13, 2016
Dear Jimmy Bakker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
1
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Shawn W. Forrest -S 2016.05.05 23:12:25 -04'00'
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K152930
Device Name PeriFlux 6000
Indications for Use (Describe)
PeriFlux 6000 equipped with PF 6010 is intended for measuring micro-vascular perfusion in skin and muscle in humans. It is also intended for measuring micro-vascular perfusion in all tissues in animals for research purposes. The PF 6010 is also intended for evaluating tissue response in skin to local heating and providing temperature stabilization of skin at blood perfusion measurement.
PeriFlux 6000 equipped with PF 6050 is intended for measuring the pressure cuffs, to simplify simultaneous measurements of perfusion and pressure when used in conjunction with a laser Doppler perfusion monitor. PeriFlux 6000 equipped with PF 6050 is also intended for Pulse Volume Recording (PVR) on human limbs and digits for diagnosis and evaluation of vascular disease in adults. It is not intended for use on neonates and pregnant women.
PeriFlux 6000 equipped with PF 6040 is intended for continuous non-invasive transcutaneous monitoring of the partial pressures of oxygen and/or carbon dioxide. It is intended for use on neonates, pediatrics, and adults not under gas anesthesia.
(Note: PeriFlux 6000 equipped with PF6040, with an intended use as in the last paragraph, was previously cleared in K131253 and is therefore not part of this submission)
Type of Use (Select one or both, as applicable) |
---|
------------------------------------------------- |
Prescription Use (Part 21 CFR 801 Subpart D) | |
---|---|
Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
PERIMED | |
---|---|
Document Type | Traditional 510(k) Notification |
Submission Date | 2015-10-02 |
Section-Page | 6-2/5 |
Object/Subject |
PeriFlux 6000 – 510(k) Summary
Submitter
Perimed AB Datavägen 9A SE-175 43 Järfälla, Sweden
Phone: (011) 46 8 580 119 90 Fax: (011) 46 8 580 100 28
Official Correspondent: Maria Prans Liljevret Contact Person for this submission: Jimmy Bakker
Date prepared: | 2 October 2015 | |
---|---|---|
Revised: | 5 May 2016 |
Device
Trade Name: | PeriFlux 6000 |
---|---|
Common Name: | Modular multichannel system for diagnosis and |
evaluation of vascular disease | |
Classification Name: | Cardiovascular blood flowmeter |
Classification Regulation: | 870.2100 |
Product Code: | DPW |
Panel: | Cardiovascular |
Regulatory Class: | II |
Type of product: | Finished product |
Predicate device
Predicate device No 1 (Primary)
| Trade name: | PeriFlux System 5000
PF 5010 LDPM Unit
PF 5020 Temp Unit
PF 5050 Pressure Unit |
|------------------|-----------------------------------------------------------------------------------------|
| 510(k) No: | K974285, K932068, K011899 |
| Type of product: | Finished product |
| Panel: | Cardiovascular |
Predicate device No 2
Trade name: | PeriFlux 6000 |
---|---|
PF 6040 LDPM Unit | |
510(k) No: | K131253 |
Type of product: | Finished product |
Panel: | Anesthesiology |
4
Image /page/4/Figure/0 description: This image is a document header. It includes the logo for "PERIMED" on the left. The document type is listed as "Traditional 510(k) Notification", the submission date is "2015-10-02", and the section-page is "6-3/5". The object/subject is "PeriFlux 6000 - 510(k) Summary".
Reference device
No reference devices were used in this submission
Device Description
The PeriFlux 6000 is a modular multichannel system that offers several tests for diagnosis and evaluation of vascular disease in one instrument. It consists of a Main Unit (PF 6001), which accommodates up to eight different function units of the same type or of different types enabling simultaneous measurements of several parameters. Currently, one type of unit –PF 6040 tcpO2/pCO2 Unit– is available for the US market. Two more units have been developed:
- PF 6010 LDPM/Temp Unit ●
- PF 6050 Pressure Unit
The PeriFlux 6000 equipped with PF 6010 function units measures microcirculatory blood flow using Laser Doppler technology. In combination with the PF 6050 function unit to inflate blood pressure cuffs and to register the cuff pressure, it enables a range of different tests, including ankle pressure, toe pressure, skin perfusion pressure (SPP), pulse volume recording (PVR), segmental pressures, heat provocations and post-occlusive reactive hyperemia (PORH).
In Laser Doppler Perfusion Monitoring, an optical fibre leads light generated by a laser to the LDPM probe tip, which rests against the tissue. The beam of light will enter the tissue and become scattered. Blood cells moving within the volume illuminated by the beam will cause the light to change frequency. This change in frequency is called a Doppler shift and it is used to calculate the blood flow in the illuminated tissue.
Indications for Use
PF 6010 LDPM/Temp Unit
PeriFlux 6000 equipped with PF 6010 is intended for measuring micro-vascular perfusion in skin and muscle in humans. It is also intended for measuring micro-vascular perfusion in all tissues in animals for research purposes.
The PF 6010 is also intended for evaluating tissue response in skin to local heating and providing temperature stabilization of skin at blood perfusion measurement.
PF 6040 tcpO2/CO2 Unit
PeriFlux 6000 equipped with PF 6040 is intended for continuous non-invasive transcutaneous monitoring of the partial pressures of oxygen and/or carbon dioxide. It is intended for use on neonates, pediatrics, and adults not under gas anesthesia.
PF 6050 Pressure Unit
PeriFlux 6000 equipped with PF 6050 is intended for measuring the pressure in blood pressure cuffs, to simplify simultaneous measurements of perfusion and pressure when used in conjunction with a laser Doppler perfusion monitor. PeriFlux 6000 equipped with PF 6050 is also intended for Pulse Volume Recording
5
Image /page/5/Figure/0 description: The image shows a document with the title "Traditional 510(k) Notification". The document type is "Traditional 510(k) Notification", the submission date is "2015-10-02", and the section-page is "6-4/5". The document also includes the logo "PERIMED".
PeriFlux 6000 - 510(k) Summary
(PVR) on human limbs and digits for diagnosis and evaluation of vascular disease in adults.
It is not intended for use on neonates and pregnant women.
(Note: PeriFlux 6000 equipped with PF 6040, with an intended use as described under "PF 6040 tcpO2/CO2 Unit", was previously cleared in K131253 and is therefore not part of this submission)
Environment of use
The PeriFlux 6000 instrument is intended for professional use in a hospital environment, laboratory environment and in a hospital intensive care unit environment.
Comparison of technological characteristics with the predicate devices
There are no technological differences between the instrument approved in K131253 and the current Periflux 6000 when equipped with PF 6040 besides the ability to host the other function units. Predicate device No 2 is listed for completeness and will not be discussed further. The Periflux 6000 and the primary predicate device have Laser Doppler Perfusion Monitoring as a technological principle. It is based on measuring the Doppler shift of laser light scattered in the tissue as a measure of micro-vascular blood flow.
At high level, the Periflux 6000 and the primary predicate device are based on the following same technological elements:
- An infrared laser with laser class 1 (IEC 60825-1:2007, 21 CFR 1040.10 and ● 1040.11)
- Measurement through optical fibre probes. ●
- Temperature measurement and control in the probe possible ●
- Measurement and control of pressure in blood pressure cuffs ●
- Possibility to combine different function units in one device o
The following technological differences exist between the PeriFlux 6000 and the primary predicate device:
- Use of an SVGA display with touch screen operation ●
- Up to 8 function units in one device (compared to 4 in primary predicate) ●
- Laser Doppler Perfusion Monitoring and probe heating combined in one function ● unit
- o Perfusion calculated from a larger frequency bandwidth.
- o 6 individually controlled cuff connectors and a built-in pump.
Summary of Performance Tests
Performance Testing - Bench
Laser Doppler Perfusion Monitoring is an intricate measurement, involving more electronics and calculations than measurements of the other parameters. Also, the PF
6
Image /page/6/Figure/0 description: The image is a document header that includes the company logo, document type, submission date, and section-page information. The company logo is "PERIMED" with a red graphic to the left. The document type is "Traditional 510(k) Notification", the submission date is "2015-10-02", and the section-page is "6-5/5".
PeriFlux 6000 - 510(k) Summary
6010 measures over a higher frequency bandwidth than the PF 5010. We have therefore tested how the clinically relevant parameter, blood perfusion, correlates between these two systems. The tests show that the signals correlate very well between the two systems and that they can be regarded as equivalent.
Performance Testing - Animal
No performance tests on animals are submitted.
Performance Testing - Clinical
No clinical performance tests are submitted.
Conclusion
The discussion of similarities and differences between the new device and the predicate device shows that the new device is similar to the predicate device in most of the relevant properties. The intended use of the devices is the same. A few differences in technological characteristics have been found and the discussion concludes that these differences do not raise different questions of safety and effectiveness, nor do they adversely affect safety and effectiveness. The PeriFlux 6000 simply follows the state of the art in standards and available technology to create a safe, effective and convenient device. Temperature and pressure are measured using the same technology and use the same type of sensors in PeriFlux 6000 as in PeriFlux System 5000 and comparative testing shows that Laser Doppler Perfusion measurements on these two systems can also be regarded as equivalent.
The conclusion is that the new device PeriFlux 6000 is substantially equivalent to the predicate device PeriFlux System 5000.