(141 days)
No
The summary describes a device for non-invasive pressure measurements and waveform recordings using air plethysmography. There is no mention of AI or ML in the intended use, device description, or any other section. The technology described is based on standard physiological measurement principles.
No
The device is used for non-invasive Pulse Volume waveform Recordings (PVR) and to measure and control pressure in a blood pressure cuff, which are diagnostic and monitoring functions, not therapeutic.
Yes
The device performs "Pulse Volume waveform Recordings (PVR) on extremities of patients with vascular diseases," which involves collecting physiological data to assess a patient's condition.
No
The device description clearly outlines hardware components such as a modular unit, a display, a balloon pump, a pressure transducer, and the need for a standard blood pressure cuff. It is part of a larger hardware system (PeriFlux System 5000).
Based on the provided information, the PF 5050 Pressure Unit is not an In Vitro Diagnostic (IVD).
Here's why:
- IVD Definition: In Vitro Diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- PF 5050 Function: The PF 5050 Pressure Unit performs non-invasive measurements on the extremities of patients. It uses air plethysmography via a blood pressure cuff to record Pulse Volume waveforms (PVR) and measure/control cuff pressure. These are physical measurements taken directly from the patient's body, not tests performed on samples outside the body.
The device is used for in vivo measurements, meaning measurements taken within a living organism.
N/A
Intended Use / Indications for Use
The PF 5050 Pressure Unit is intended for Pulse Volume waveform Recordings (PVR) on limbs and digits.
It is also intended to measure the mean pressure in a blood pressure cuff to simplify simultaneous measurements of perfusion and pressure when used in conjunction with a laser Doppler perfusion monitor. This simplifies some typical laser Doppler measurements that include occlusions with blood pressure cuffs. such as: digital systolic blood pressure, Post Occlusive Reactive Hyperemia (PORH) and Skin Perfusion Pressure (SPP).
When installed in the PeriFlux System 5000, the PF 5050 Pressure Unit is intended to monitor Pulse Volume waveform Recordings (PVR).
When used together with PF 5010 LDPM Unit (laser Doppler unit) it is also possible to record Skin Perfusion Pressure.
Use of the PF 5050 Pressure Unit is indicated for non-invasive Pulse Volume waveform Recordings (PVR) on extremities of patients with vascular diseases.
It is also intended to measure and control the pressure in a blood pressure cuff to simplify simultaneous measurements of perfusion and pressure when used in conjunction with a laser Doppler perfusion monitor.
Product codes (comma separated list FDA assigned to the subject device)
DPW
Device Description
The PF 5050 Pressure Unit is a modular unit that can be installed in the PeriFlux System 5000. The device provides Pulse Volume waveform Recordings (PVR) on various limb and digit extremities made possible through air plethysmography. The PF 5050 Pressure Unit includes a small display where the mean cuff pressure is displayed.
The PVR tracing is accomplished by attaching a standard blood pressure cuff on various areas of the extremities. To provide skin contact, atmospheric air is inflated to approximately 50 - 65 mmHg using the balloon pump supplied with a standard blood pressure gauge. A pressure transducer records the pressure change in the cuff (Pulse Volume waveforms) secondary to change in v the extremity segment under the cuff during repeated cardiac cycles.
The PeriFlux System 5000 is a multi-channel, multi-functional system capable of hosting one, two, three or four function units including for example a PF 5010 LDPM Unit for blood perfusion measurements. The PeriFlux 5001 Main Unit accommodates up to four different function units of the same type or of different types enabling simultaneous measurements of several parameters. There are three different function unit types available at the moment.
PF 5010 LDPM Unit for blood perfusion measurements.
PF 5020 Temp Unit for temperature measurements.
PF 5040 tcpO2 / pCO2 for measurements of transcutaneous O2 and CO2.
The PeriFlux System 5000 with the Modular laser Doppler system for blood perfusion measurement was cleared for marketing under 510(k) application K974285 on May 28, 1998.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
limbs and digits, extremities
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Not Found
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.2100 Cardiovascular blood flowmeter.
(a)
Identification. A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.(b)
Classification. Class II (performance standards).
0
|--|--|
PERIMED
Document Type | Traditional 510(k) |
---|---|
Section-Page | 6-2 |
Object/Subject | PF 5050 Pressure Unit - 510(k) Summary |
Date | NOV 06 2001 |
Koll894 p.1/3
510(k) Summary as required by section 807.92(c)
6.1 Submitter's Name & Address:
Perimed AB Datavägen 9A SE-175 26 Järfälla. Sweden . Tel: (011) 46 8 580 119 90 Fax: (011) 46 8 580 100 28 Official Correspondent: Biörn Bakken Contact Person for this submission: Biörn Bakken
6.2 Trade Name:
Device name: PF 5050 Pressure Unit
Common name, Classification number, Class & Regulation Number:
Common Name | Classification Number | Class | Requlation Number |
---|---|---|---|
Air Plethvsmograph/Pulse Volume | Unknown, | 870.2100 | |
Recorder | IN FINAL PRODUCTION COLLECTION COLLECTION LECTURES LECTURAL LECTURAL LECTURAL CONTROLLERS COURTES COLLECTION CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR |
6.3 Predicate Device Identification:
Legally marketed devices to which equivalence is being claimed | 510(k) |
---|---|
MODEL PV2000 VASCULAR MICROLABORATORY. VASAMEDICS | K951486 |
CHECK CONTRACT CONTROLLERS CONSULTION CONSULTION CONSULTERS CONSULTERS CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONS | |
A month annual consumer considere considere consideration and consideration of the consideration of the consistences of |
6.4 Device Description (for detailed description see Section 12.4):
The PF 5050 Pressure Unit is a modular unit that can be installed in the PeriFlux System 5000. The device provides Pulse Volume waveform Recordings (PVR) on various limb and digit extremities made possible through air plethysmography. The PF 5050 Pressure Unit includes a small display where the mean cuff pressure is displayed.
The PVR tracing is accomplished by attaching a standard blood pressure cuff on various areas of the extremities. To provide skin contact, atmospheric air is inflated to approximately 50 - 65 mmHg using the balloon pump supplied with a standard blood pressure gauge. A pressure transducer records the pressure
COMPANY CONFIDENTIAL
1
Document Type | Traditional 510(k) |
---|---|
Section-Page | 6-3 |
Object/Subject | |
Date |
PF 5050 Pressure Unit - 510(k) Summary
change in the cuff (Pulse Volume waveforms) secondary to change in v the extremity segment under the cuff during repeated cardiac cycles.
The PeriFlux System 5000 is a multi-channel, multi-functional system capable of hosting one, two, three or four function units including for example a PF 5010 LDPM Unit for blood perfusion measurements. The PeriFlux 5001 Main Unit accommodates up to four different function units of the same type or of different types enabling simultaneous measurements of several parameters. There are three different function unit types available at the moment.
PF 5010 LDPM Unit for blood perfusion measurements.
PF 5020 Temp Unit for temperature measurements.
PF 5040 tcpO2 / pCO2 for measurements of transcutaneous O2 and CO2.
The PeriFlux System 5000 with the Modular laser Doppler system for blood perfusion measurement was cleared for marketing under 510(k) application K974285 on May 28, 1998.
6.5 Intended Use of the Device:
The PF 5050 Pressure Unit is intended for Pulse Volume waveform Recordings (PVR) on limbs and digits.
It is also intended to measure the mean pressure in a blood pressure cuff to simplify simultaneous measurements of perfusion and pressure when used in conjunction with a laser Doppler perfusion monitor. This simplifies some typical laser Doppler measurements that include occlusions with blood pressure cuffs. such as: digital systolic blood pressure, Post Occlusive Reactive Hyperemia (PORH) and Skin Perfusion Pressure (SPP).
6.6 Summary of technological characteristics of Device and Predicate Device:
The PF 5050 Pressure Unit is substantially equivalent to the following predicate device:
| Manufacture: | ставенные с правника страниции приводительные в Австранского в Автора с верника в с п. к. п. к. п. н. е. п. н. н. н. н. н. н. н. н. н. н. н. н. н. н. н. н. н. н. н. н. н. н.
All A. B man Religio Collection of children of charges and can be a markets manufacture a market a manufacture a comments of the many and
11 -- 1 -- 1 -- 1 -- 1 -- 1 -- 1 -- 1 -- 1 -- 1 -- 1 -- 1 -- 1 -- 1 1 -- 1 1 -- 1 1 -- 1 1 -- 1 1 -- 1 1 -- 1 1 -- 1 1 -- 1 1 -- 1 1 -- 1 1 1 -- 1 1 1 -- 1 1 1 -- 1 1 1 -- 1
marketed devices to which equivalence is being claimed
Children and addition and the minutes and results and results and religion in the first and the first and other for the different and other for the first in the first in the | |
|--------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| IACARIAN | and and the manufacturer of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of th
ODEL PV2000 VASCULAR MICROLAB'
MODE!
ACCCASS .
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 1051400 |
This conclusion is based on the following:
Indications for use:
2
| | Document Type
Traditional 510(k) | Section-Page
6-4 |
|-------------------------------------------------------------|-------------------------------------|---------------------|
| Object/Subject
PF 5050 Pressure Unit - 510(k) Summary | | Date |
When installed in the PeriFlux System 5000, the PF 5050 Pressure Unit is intended to monitor Pulse Volume waveform Recordings (PVR).
K011899 p.3/3
olume wavelorm Recordings (P When used together with PF 5010 LDPM Unit (laser Doppler unit) it is also possible to record Skin Perfusion Pressure.
Technological Comparison:
The Vasamedics MODEL PV2000 VASCULAR MICROLABORATORY has the . same indications for use as the PF 5050 Pressure Unit in that they both are measuring Pulse Volume Waveform Recordings (PVR). When the PeriFlux System 5000 also includes a PF5010 laser Doppler perfusion monitor it is possible in both instruments to measure Skin Perfusion Pressure.
The instruments provide similar types of tracings and have comparable output ranges. Both instruments can be connected to external devices for data analysis
A more detailed comparison of the design and performance features of these devices is provided in appendix B.
COMPANY CONFIDENTIAL
3
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract human profiles facing to the right, with flowing lines suggesting movement or connection.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 0 6 2001
Mr. Björn Bakken Vice President Quality Assurance & Regulatory Affairs Perimed AB Datavägen 9A SE-175 26 Järfälla, Sweden
Re: K011899
Trade Name: Perimed PF 5050 Pressure Unit Regulation Number: 21 CFR 870.2100 Regulation Name: Cardiovascular Blood Flowmeter Regulatory Class: Class II (two) Product Code: DPW Dated: October 8, 2001 Received: October 12, 2001
Dear Mr. Bakken:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
4
Page 2 - Mr. Björn Bakken
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4645. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
N. Dala Tella
James E. Dillard III Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
5
| PERIMED | Document Type
Traditional 510(k) | Section-Page
9-1 |
|--------------------------------------------------------------|-------------------------------------|---------------------|
| Object/Subject
PF 5050 Pressure Unit - Indication for Use | | Date |
NOV 0 6 200510(k) Number (if known): KO11899
- PF 5050 Pressure Unit Device Name:
- Use of the PF 5050 Pressure Unit is indicated for non-invasive Indications For Use: Pulse Volume waveform Recordings (PVR) on extremities of patients with vascular diseases.
It is also intended to measure and control the pressure in a blood pressure cuff to simplify simultaneous measurements of perfusion and pressure when used in conjunction with a laser Doppler perfusion monitor.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
✓
OR
Division of Cardiovascular & Respiratory Devices
510(k) Number K011899
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________