K Number
K152748
Date Cleared
2016-03-01

(160 days)

Product Code
Regulation Number
880.2910
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Braun Thermoscan® PRO 6000 Ear Thermometer is indicated for the intermittent measurement and monitoring of human body temperature for patients of all ages in a professional use environment. The probe cover is used as a sanitary barrier between the infra red thermometer and the ear canal.

Device Description

The Braun Thermoscan® PRO 6000 Ear Thermometer is a hand held instrument that measures human body temperature through the opening of the auditory canal. It is a single mode ear thermometer that measures the natural thermal infrared radiation emitted from the tympanic membrane and adjacent surfaces with a built-in correction algorithm to compensate the influence of ambient temperature using a heated tip. The Braun Thermoscan® PRO 6000 Ear Thermometer also includes a technique compensation feature to detect depth of insertion to compensate for common methods of misuse, this feature uses capacitive rings mounted in the speculum of the thermometer to detect insertion and depth and measurements of the outer ear temperature during insertion to detect misuse and correct accordingly. The Braun Thermoscan® PRO 6000 Ear Thermometer is meant for professional use in hospitals and healthcare or professional office settings.

AI/ML Overview

The provided document describes the KAZ USA, Inc. Braun Thermoscan® PRO 6000 Ear Thermometer and its clinical study to demonstrate substantial equivalence to its predicate device, the Braun Thermoscan® PRO 4000 Ear Thermometer, in accordance with ASTM E1965-03.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Criterion (from ASTM E1965-03)Acceptance Criteria for Braun Thermoscan® PRO 6000Reported Device Performance (Reference Thermoscan® PRO 4000 and SureTemp Plus)
Bias (vs. Reference)"within clinical acceptability" (specifically, bias less than predicate device when compared to reference)Bias reported as "within clinical acceptability" and "less than predicate device when compared to reference" against the SureTemp Plus Oral/Rectal and Axillary Contact Thermometer

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of three human profiles facing right, arranged in a stacked formation. The profiles are simple and abstract, with flowing lines suggesting hair or head coverings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the profile graphic.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 1, 2016

KAZ USA, Inc (a Subsidiary of Helen of Troy, Inc) Mr. Raj Kasbekar Vice President, Regulatory Affairs 250 Turnpike Road Southborough, Massachusetts 01772

Re: K152748

Trade/Device Name: Braun Thermoscan® PRO 6000 Ear Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: December 4, 2016 Received: December 8, 2016

Dear Mr. Kasbekar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina Kiang -

for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use Statement Section 4:

Indications for Use

510(k) Number (if known): K152748

Braun Thermoscan® PRO 6000 Ear Thermometer Device Name:

Indications for Use:

The Braun Thermoscan® PRO 6000 Ear Thermometer is indicated for the intermittent measurement and monitoring of human body temperature for patients of all ages in a professional use environment. The probe cover is used as a sanitary barrier between the infra red thermometer and the ear canal.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use X (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Kaz USA, Incorporated. The logo is a stylized "kaz" in white letters against a blue background. Below the logo is the text "Kaz USA, Incorporated" followed by the address "250 Turnpike Road, Southborough, MA 01772".

K152748 510(k) Summary

Braun Thermoscan® PRO 6000 Ear Thermometer

1.0 Preparation Date:

February 19, 2016

2.0 Submitted By:

KAZ USA, Inc., a Helen of Troy Company 400 Donald Lynch Boulevard, Suite 300 Marlborough, MA 01752

Primary Contact Person/Prepared by:

Raj S Kasbekar, Global VP Regulatory & Clinical Affairs

KAZ USA, Inc., a Helen of Troy Company 400 Donald Lynch Boulevard, Suite 300 Marlborough, MA 01752

3.0 Device Identification:

3.1 Trade Name

Braun Thermoscan® PRO 6000 Ear Thermometer

Models:

US: PRO6000SCMN (Thermometer with Small Cradle), PRO6000LCMN (Thermometer with Large Cradle), PRO6000BSNA (Charge Dock) [Braun Thermoscan® PRO 6000 Ear Thermometer]

3.2 Common Name

Infrared Ear Thermometer

3.3 Classification Name

Thermometer, Clinical, Electronic

{4}------------------------------------------------

(21CFR 880.2910: Product code - FLL)

4.0 Predicate Device:

PredicateManufacturerDocketNumber
Braun Thermoscan® PRO 4000 EarThermometerKaz USA, IncContract: KeytronicsK103800

5.0 Device Description:

The Braun Thermoscan® PRO 6000 Ear Thermometer is a hand held instrument that measures human body temperature through the opening of the auditory canal. It is a single mode ear thermometer that measures the natural thermal infrared radiation emitted from the tympanic membrane and adjacent surfaces with a built-in correction algorithm to compensate the influence of ambient temperature using a heated tip. The Braun Thermoscan® PRO 6000 Ear Thermometer also includes a technique compensation feature to detect depth of insertion to compensate for common methods of misuse, this feature uses capacitive rings mounted in the speculum of the thermometer to detect insertion and depth and measurements of the outer ear temperature during insertion to detect misuse and correct accordingly. The Braun Thermoscan® PRO 6000 Ear Thermometer is meant for professional use in hospitals and healthcare or professional office settings.

6.0 Indications for Use:

The Braun Thermoscan® PRO 6000 Ear Thermometer is indicated for the intermittent measurement and monitoring of human body temperature for patients of all ages in a professional use environment. The probe cover is used as a sanitary barrier between the infra red thermometer and the ear canal.

7.0 Validation Results:

Clinical Study to show substantial equivalence:

A comparison study and clinical repeatability testing was performed on the following four age groups: 0-12 months, 12 months- <5 years, <18 years, and 18 years and older in accordance with ASTM E1965-03 to compare the Braun Thermoscan® PRO 6000 Ear Thermometer (test thermometer) with the predicate Braun Thermoscan® PRO 4000 Ear Thermometer (K103800). The reference or the gold standard used was the SureTemp Plus Oral/Rectal and Axillary Contact Thermometer in the monitoring mode (K030580). This clinical comparison study demonstrated that the Braun Thermoscan® PRO 6000 Ear Thermometer is as good as (non-inferior or substantially equivalent to) the previously approved Braun Thermoscan® PRO 4000 Ear Thermometer in all age groups with respect to the bias and standard deviation in comparison to the Reference SureTemp Plus Oral/Rectal and Axillary Contact Thermometer in the monitoring mode (K030580). The temperatures obtained with the test Infrared Ear Thermometer were highly related when compared to the predicate device, where temperatures were

{5}------------------------------------------------

measured in the oral mode for children above 5 years) and axillary mode (for children under 5 years) . The clinical bias with stated uncertainty and clinical repeatability as defined in the ASTM E1965-03 standard were within clinical acceptability (bias less than predicate device when compared to reference). The clinical repeatability of the Infrared Ear Thermometer was statistically and clinically acceptable (less than 0.3 deg C or 0.58 deg F).

8.0 Device Comparison with Predicate

8.1 Intended Use

The predicate device, the Braun Thermoscan® PRO 4000 Ear Thermometer is intended for the intermittent determination of the human's body temperature for people of all ages. The intended use and indications for use of the predicate and the Braun Thermoscan® PRO 6000 Ear Thermometer are similar. The predicate device as well as the KAZ Braun Thermoscan® PRO 6000 Ear Thermometer use the measurement site.

8.2 Materials

Materials used in the manufacture of the Braun Thermoscan® PRO 6000 Ear Thermometer are similar (derived from similar resins) to the predicate device. All skin contacting materials used in the new thermometer have been tested for biocompatibility (cyto-toxicity in accordance with ISO 10993-5 as well as irritation and sensitization in accordance with ISO 10993-10) and FDA Blue book memo G 95-1 for both thermometers and the corresponding test reports are included in this submission.

8.3 Design

The industrial design of the KAZ Braun Thermoscan® PRO 6000 Ear Thermometer is identical to the predicate device except for a new outer shell.

8.4 Operational Principles

The Braun Thermoscan® PRO 6000 Ear Thermometer is a handheld device, containing an On/Off switch, sensor area, microcontroller and LCD screen to control the device and take measurements. The operating principle is based on detection of infrared energy and use of predictive algorithms to estimate the body temperature.

8.5 Technology

The technology of the Braun Thermoscan® PRO 6000 Ear Thermometer is identical to the predicate device. The technique compensation feature is additional technology that is integrated into the Braun Thermoscan® PRO 6000 Ear Thermometer using additional hardware (three capacitive rings) and software.

{6}------------------------------------------------

8.6Comparison table between testand Predicate Devices
SEComparisonsSubject (Braun Thermoscan® PRO6000 Ear Thermometer)Predicate (BraunThermoscan® PRO 4000)K103800Comment
Classification21CFR 880.291021CFR 880.2910Same
Product CodeFLLFLLSame
FDA ClassIIIISame
Intended UseA non-sterile, re-useable clinicalthermometer intended for the intermittentdetermination of the human's bodytemperature for people of all agesA non-sterile, re-useableclinical thermometerintended for thedetermination of thehuman's body temperaturefor people of all ages.Same
OperationHand held-Manually operatedHand held-ManuallyoperatedSame
SensorInfraredInfraredSame
MaterialsCommon Materials- including an impactresistant casing. Biocompatible metalsand resins.Common Materials-including an impactresistant casing.Biocompatible metals andresins.Similar - validatedfor cytotoxicity perISO10993-5 andirritation as well assensitization per ISO10993-10
Dimensions150mm x 60 mm x 35mm150mm x 45 mm x 30 mmSimilar
Accuracy0.3 deg C for <35 deg C0.2 deg C for 35 deg C to 42 deg C0.3 deg C for >42deg C0.3 deg C for <35.5 degC0.2 deg C for 35.5 deg Cto 42 deg C0.3 deg C for >42 deg CMeets ASTM E1965,EN12470-5 andISO 80601-2-56
Precision (SD)<0.6 deg C<0.6 deg CSame
Repeatability<0.3 deg C<0.3 deg CMeets EN 12470-5
Operatingtemperature10 to 40 deg C ambient temperature(amb) and up to 95% RH10 to 40 deg C ambientand up to 95% RHSame
MeasurementRange20 deg C to 42.2 deg C20 deg C to 42.2 deg CSame
DisplayLCDLCDSame
ResponseTime2-3 sec2-3 secSame
MeasurementSiteEarEarSame
ScaleDeg F/Deg CDeg F/Deg CSame
Power SupplyTwo AA Alkaline Batteries or CustomNickel Metal Hydride Battery PackTwo AA Alkaline Batteriesor custom rechargeableNickel Metal HydrideBattery PackSame
StandardsMet for Benchand ClinicalPerformance1) ASTM E1965-1998: 2009 InfraredThermometers for IntermittentDetermination of Patient temperature2) Clinical accuracy test requirementsestablished in the standard ASTME1965-09 (Clinical part only)- StandardSpecification for Infrared ThermometerSameAll standards verifiedand validated for testdevice and metacceptance criteria.

{7}------------------------------------------------

For Intermittent Determination of PatientTemperature;
3) IEC 60601-1 3rd edition:2005 MedicalElectrical Equipment: Generalrequirements for Safety, Requirementsand Tests.
4) IEC 60601-1-2: 2007 Medical ElectricalEquipment- Part 1: GeneralRequirements for Safety, ElectromagneticCompatibility- Requirements and Tests.
5) AAMI / ANSI / ISO 10993-1:2003(E),Biological evaluation of medical devices -- Part 1: Evaluation and testing- andappropriate Parts
6) ISO 10993-5: Biological Evaluation ofMedical Devices: Part 5: Tests for In-vitrocytotoxicity.
7) ISO 10993-10: Biological Evaluation ofMedical Devices: Part 10: Tests forIrritation and Sensitization.
8) ISO 14971:2012: Application of RiskManagement to Medical Devices
9) EN IEC 62304:2006 Medical DeviceSoftware – Software life-cycle processes
10) IEC EN62366:2007 Applicability ofUsability Engineering to Medical Devices– Collateral Standard: Usability
11) ISO 14155:2011 Clinical investigationof medical devices for human subjects –Good Clinical Practice (also includes21CFR812, parts 50 and 56)

9. Conclusion:

Based on the performance testing and compliance with acceptable voluntary standards, we believe that the Braun Thermoscan® PRO 6000 Ear Thermometer is substantially equivalent to its predicate device cited above.

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.