K Number
K152715
Date Cleared
2016-02-25

(157 days)

Product Code
Regulation Number
880.6250
Panel
HO
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Nitrile Powder Free Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Nitrile films form a barrier to body fluids and bloodborne Pathogens. The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. Nitrile glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151 requirements.

AI/ML Overview

The provided document describes a 510(k) premarket notification for "Nitrile Powder Free Patient Examination Gloves, Blue Color," which is a Class I medical device. The document primarily focuses on demonstrating substantial equivalence to a predicate device, rather than describing a study to prove performance against acceptance criteria for an AI/ML powered device.

Therefore, many of the requested fields are not applicable to this document. I will extract the available information and indicate where information is not present or not relevant to an AI/ML context.


1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
DimensionASTM standard D 6319-10Meets
Physical PropertiesASTM standard D 6319-10Meets
Freedom from pinholes21 CFR 800.20 and ASTM D5151-06 (Reapproved 2011) (Holes at Inspection Level I AQL 2.5)Meets ASTM D5151-06 (Reapproved 2011), AQL 2.5
Powder ResidualASTM standard D 6319-10 and D6124-06 (Reapproved 2011)Meets (

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.