K Number
K152715

Validate with FDA (Live)

Date Cleared
2016-02-25

(157 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Nitrile Powder Free Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Nitrile films form a barrier to body fluids and bloodborne Pathogens. The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. Nitrile glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151 requirements.

AI/ML Overview

The provided document describes a 510(k) premarket notification for "Nitrile Powder Free Patient Examination Gloves, Blue Color," which is a Class I medical device. The document primarily focuses on demonstrating substantial equivalence to a predicate device, rather than describing a study to prove performance against acceptance criteria for an AI/ML powered device.

Therefore, many of the requested fields are not applicable to this document. I will extract the available information and indicate where information is not present or not relevant to an AI/ML context.


1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
DimensionASTM standard D 6319-10Meets
Physical PropertiesASTM standard D 6319-10Meets
Freedom from pinholes21 CFR 800.20 and ASTM D5151-06 (Reapproved 2011) (Holes at Inspection Level I AQL 2.5)Meets ASTM D5151-06 (Reapproved 2011), AQL 2.5
Powder ResidualASTM standard D 6319-10 and D6124-06 (Reapproved 2011)Meets (<2mg/glove)
Biocompatibility (Primary Skin Irritation)ISO 10993-10: Third Edition 2010-08-01 (rabbits)Passes (Not a Primary Skin Irritation)
Biocompatibility (Dermal Sensitization)ISO 10993-10: Third Edition 2010-08-01 (guinea pig)Passes (Not a Dermal sensitization)
Dimensions -- LengthMeets ASTM D6319-10 (≥230mm min)230mm min for all sizes
Dimensions -- WidthMeets ASTM D6319-10 Small 70-90 mm Medium 85-105mm Large 100-120mm Xlarge 110-130 mmSmall 84-86 mm Medium 94-96mm Large 104-106mm X large 114-118 mm
Dimensions -- ThicknessMeets ASTM D6319-10 Finger 0.05mm min. Palm 0.05mm min.Thickness (mm) min. Finger 0.10 mm min. Palm 0.06 mm min.
Physical Properties (Elongation)ASTM D6319-10 (Before aging/after aging: Elongation ≥500%)Elongation: 520-580%
Physical Properties (Tensile Strength)ASTM D6319-10 (Before aging/after aging: Tensile Strength ≥ 14MPa)Tensile Strength: 22-34 MPa

2. Sample size used for the test set and the data provenance

The document does not specify a distinct "test set" in the context of an AI/ML algorithm. The performance data presented (e.g., dimensions, physical properties, biocompatibility) are based on testing a sample of the manufactured gloves to ensure they meet the established standards. The size of these samples is not explicitly stated in the provided text. The data provenance is implied to be from the manufacturer's testing, Haining Medical Products Co., Ltd., which is located in China. The testing is assumed to be prospective relative to the 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable. The device is a patient examination glove, not an AI/ML diagnostic tool requiring expert ground truth for interpretation. Performance is assessed against international standards and regulations (e.g., ASTM, ISO, CFR).

4. Adjudication method for the test set

This information is not applicable. There is no expert review or adjudication process described for the performance testing of these gloves.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. The device is a physical medical glove and does not involve AI/ML or human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This information is not applicable. The device is a physical medical glove and does not involve an algorithm.

7. The type of ground truth used

The "ground truth" for the glove's performance is based on established international standards and regulatory requirements for medical gloves, such as ASTM D6319-10, ASTM D6124-06, ASTM D5151-06, 21 CFR 800.20, and ISO 10993-10. These standards define the acceptable physical, mechanical, and biological properties.

8. The sample size for the training set

This information is not applicable. The device is a physical medical glove and does not involve an AI/ML training set.

9. How the ground truth for the training set was established

This information is not applicable. The device is a physical medical glove and does not involve an AI/ML training set or its associated ground truth establishment.


Summary of the Study Proving Device Meets Acceptance Criteria:

The document describes non-clinical performance testing conducted by Haining Medical Products Co., Ltd. (the manufacturer) to demonstrate that the "Nitrile Powder Free Patient Examination Gloves, Blue Color" meet the requirements outlined in various ASTM standards (D6319-10, D6124-06, D5151-06), 21 CFR 800.20, and ISO 10993-10 (for biocompatibility). These standards set the acceptance criteria for properties like dimensions, physical strength (tensile strength, elongation), freedom from pinholes (waterleak test), powder residue, and biocompatibility (skin irritation and sensitization).

The study involved testing samples of the manufactured gloves against these specified standards. For instance, dimensions (length, width, thickness), physical properties (elongation, tensile strength), and powder residue were measured and found to "Meet" or exceed the requirements of the ASTM standards. Freedom from pinholes was assessed against 21 CFR 800.20 and ASTM D5151-06 with an Acceptable Quality Level (AQL) of 2.5% for holes. Biocompatibility testing (Primary Skin Irritation in rabbits and Dermal Sensitization in guinea pigs) was performed according to ISO 10993-10, and the gloves "Passed," demonstrating they were not irritants or sensitizers.

The results of these tests, as summarized in the table and the "6.0 Summary of the Technological Characteristics of the Device" and "7.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data" sections, directly demonstrate that the device performs according to the established acceptance criteria. No clinical data was deemed necessary for this type of device according to the document. The overall conclusion is that the device is "as safe, as effective, and performs as well as the predicate device," thereby establishing substantial equivalence.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 25, 2016

Haining Medical Products Co., LTD Mr. Chu Xiaoan Room 1606 Bldg, 1. Jianxiang Yuan No. 209 Bei Si Huan Zhong Road Haidian District, Beijing, 100083 China

Re: K152715

Trade/Device Name: Nitrile Powder Free Patient Examination Gloves, Blue Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA Dated: January 29, 2016 Received: February 1, 2016

Dear Mr. Chu Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K152715

Device Name

Nitrile Powder Free Patient Examination Gloves, Blue Color

Indications for Use (Describe)

Nitrile Powder Free Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Section 6 510(k) Summary

510(K) Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is: =

Premarket Notification [510(k)] Summary

1.0 Submitter:

Submitter's name :Haining Medical Products Co.,Ltd.
Submitter's address :West of Dianchang Road, LuannanCounty, Tangshan City,063500 HebeiProvince,China
Phone number :(86)315-4169201
Fax number :(86)315-4167535
Name of contact person:Zhang Liang
Date of preparation :2016-01-29

2.0 Name of the Device

Device Name:Nitrile Powder Free Patient ExaminationGloves, Blue Color
Proprietary/Trade name:Haining Nitrile Powder Free PatientExamination Gloves, Blue Color
Common Name:Exam gloves
Classification Name:Patient examination glove
Device Classification:I
Regulation Number:21 CFR 880.6250
Panel:General Hospital (80)
Product Code:LZA

3.0 Predicate device

Device Name:Nitrile Powder Free Patient Examination Gloves,Blue Color
Company name:Tangshan Zhonghong Pulin Plastic Co.,Ltd.
510(K) Number:K120970

4.0 Device Description:

  • 4.1 How the device functions: Nitrile films form a barrier to body fluids and bloodborne Pathogens
  • 4.2 Scientific concepts that form the basis for the device The nitrile rubber is water tight under normal conditions of use. It's tensile

{4}------------------------------------------------

properties cause it to conform to the hand, allowing movements necessary for a medical procedure.

4.3 Physical and performance characteristics such as design, materials and physical properties:

Nitrile glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151 requirements.

5.0 Device Intended Use (Indication for use):

Nitrile Powder Free Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

6.0 Summary of the Technological Characteristics of the Device:

The Nitrile Powder Free Patient Examination Gloves, Blue Color, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

CharacteristicsStandardDevice performance
DimensionASTM standard D 6319-10.Meets
Physical PropertiesASTM standard D 6319-10.Meets
Freedom from pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 6319-10 and D6124-06(Reapproved 2011).Meets<2mg/glove
BiocompatibilityPrimary Skin Irritation in rabbitsISO 10993-10: Third Edition 2010-08-01.Dermal sensitization in the guinea pig ISO 10993-10: Third Edition 2010-08-01.PassesNot a Primary Skin IrritationPassesNot a Dermal sensitization

7.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data:

Nitrile Powder Free Patient Examination Gloves, Blue Color, meet requirements per ASTM D6319-10.per ASTM D6124-06(Reapproved 2011), per 21 CFR 800.20 and ISO 10993-10: Third Edition 2010-08-01.

The performance test data of the non clinical tests that support a determination of substantial equivalent is the same as mentioned immediately above.

{5}------------------------------------------------

8.0 Substantial Equivalent Based on Assessment of Clinical Performance Data:

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

9.0Features &DescriptionPredicate DeviceSubject DeviceResultofComparison
CompanyTangshan Zhonghong PulinPlastic Co.,Ltd.Haining Medical ProductsCo.,Ltd.--
510(K) NumberK120970K152715
Product nameNitrile Powder Free PatientExamination Gloves, BlueColorNitrile Powder Free PatientExamination Gloves, Blue Colorsame
Product CodeLZALZAsame
SizeSmall/ Medium/Large/X largeSmall/ Medium/Large/X largesame
Intend for useNitrile Powder Free PatientExamination Gloves, BlueColor is a non-steriledisposable device intendedfor medical purposes that isworn on the examiner's handor finger to preventcontamination betweenpatient and examiner.Nitrile Powder Free PatientExamination Gloves, Blue Color isa non-sterile disposable deviceintended for medical purposes thatis worn on the examiner's hand orfinger to prevent contaminationbetween patient and examiner.Substantiallyequivalent
DeviceDescription andSpecificationsMeets ASTM D6319-10Meets ASTM D6319-10Substantiallyequivalent
Dimensions--LengthMeets ASTMD6319-10≥230mm min230mm min for all sizesSubstantiallyequivalent
Dimensions-- WidthMeets ASTM D6319-10Small 70-90 mmMedium 85-105mmLarge 100-120mmXlarge 110-130 mmSmall 84-86 mmMedium 94-96mmLarge 104-106mmX large 114-118 mmSubstantiallyequivalent
Dimensions--ThicknessMeets ASTM D6319-10Finger 0.05mm min.Palm 0.05mm min.Thickness (mm) min.Finger 0.10 mm min.Palm 0.06 mm min.Substantiallyequivalent
PhysicalPropertiesMeets ASTM D D6319-10Before aging/after agingBefore aging/after agingSubstantiallyequivalent
Elongation ≥500%Tensile Strength≥ 14MPaElongation :520-580%Tensile Strength:22-34 MPa
FreedomfromPinholesMeets• 21 CFR 800.20• ASTM D6319-10• ASTM D 5151-06(Reapproved 2011)Meets ASTMD5151-06(Reapproved 2011)Holes atInspection Level IAQL2.5Substantiallyequivalent
ResidualMeets ASTMMeets ASTMSubstantially
PowderD 6124-06D 6124-06equivalent
(Reapproved 2011)(Reapproved 2011)
below2mgofresidualpowderResults generated values below2mg of residual powder
Materials usedNitrileNitrileSubstantially
to fabricate theequivalent
devices
Dusting orPUSurface Coating AgentSubstantially
Donningequivalent
Powder: name
CompareMeetsMeetsSubstantially
performanceASTM D5151-06ASTM D5151-06equivalent
data supporting(Reapproved 2011)(Reapproved 2011)
substantialASTM D6319-10ASTM D6319-10
equivalenceASTM D6124-06ASTM D6124-06
(Reapproved 2011)(Reapproved 2011)
SinglePatientSingle Patient UseSingle Patient UseSubstantially
Useequivalent
BiocompatibilitySKIN IRRITATIONIRRITATIONSKINDERMALSubstantially
DERMAL andand SENSITIZATION STUDIESequivalent
SENSITIZATIONMeetsISO 10993-10:Third
STUDIES Meets ISOEdition 2010-08-01.
10993-10:2002/Amd.1:2006
Under the conditions of the study,
not an irritant and under
conditions of the study, not a
sensitizer.
Labeling for the-Powder Free-Powder FreeSubstantially
legally marketed-Patient Examination Glove-Patient Examination Gloveequivalent
device to which-Single Use Only-Single Use Only
substantial- Manufactured For:- Manufactured For:
equivalence is- Lot- Lot
claimed.-Blue color-Blue color
- Non sterile- Non sterile

Substantial Equivalence Comparison: 9.0

{6}------------------------------------------------

10.0 Substantial Equivalence Comparison:

It can be concluded that the Nitrile Powder Free Patient Examination Gloves, Blue Color meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims.

It can be concluded that the Nitrile Powder Free Patient Examination Gloves, Blue Color is as safe, as effective, and performs as well as the predicate device, Nitrile Powder Free Patient Examination Gloves, Blue Color, Tangshan Zhonghong Pulin Plastic Co.,Ltd.. K120970.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.