(54 days)
The Bovie J-Plasma Precise 360 Handpiece is used for the delivery of helium gas plasma for cutting, coagulation, and ablation of soft tissue during open and laparascopic surgical procedures.
The Bovie J-Plasma Precise 360 Handpiece is compatible only with Bovie J-Plasma generators.
The Bovie J-Plasma Precise 360 Handpiece is a sterile, single-use electrosurgical accessory intended to be used in conjunction with Bovie J-Plasma electrosurgical generators for the delivery of helium gas plasma for cutting, coagulating and ablating of soft tissue.
The Bovie J-Plasma Precise 360 Handpiece is available with a retractable needle tip, or with a retractable cutting blade tip for excising tissue. Both tip configurations serve as electrodes to generate helium plasma. A 30 degree angled shaft tip provides improved visibility during use, and a rotating knob is used to orient the angled shaft tip to the optimal position. The device is available with 33cm and 45cm shaft lengths. The device can be used in laparoscopic procedures using a standard 5mm trocar.
The provided text describes performance testing for the Bovie J-Plasma Precise 360 Handpiece. Based on the information given, here's a breakdown of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" for each test with specific numerical targets. Instead, it lists objectives for performance tests, implying that meeting these objectives constitutes acceptance. The reported device performance is described as the device fulfilling these objectives, as evidenced by the corresponding test reports.
| Test | Objective (Implied Acceptance Criteria) | Reported Device Performance (as per document) |
|---|---|---|
| Mechanical Verification & Functionality | Verify the mechanical functionality of the device. | Met (Reports VR-1318, VR-1324) |
| Electrical Verification | Verify the electrical functionality and safety of the device. | Met (Report VR-1318) |
| Plasma Characteristics | Measure effect of generator settings on the plasma stream, plasma stream characterization. | Met (Report VR-1324) |
| Performance Evaluation | Confirm device performance on various tissue types. | Met (Report VR-1324) |
| Usability Evaluation | To verify and validate the usability of the J-Plasma Precise 360 as it relates to safety. | Met (Report VP-1387) |
| Shipping Study | Verify the device meets requirements after shipping and handling. | Met (Report VP-1384) |
| Evaluation of Rotation Effects | Evaluation of Shaft Rotation Effects on Plasma stream for Bovie J-Plasma Precise 360 Handpieces. | Met (Report VR-1383) |
| Biocompatibility Testing | Verify material safety. | Met (Report VR-1335) |
| Shelf Life Testing | Confirm product and package performance at labeled shelf life. | Met (Report VP-1326) |
Additionally, the device was designed in accordance with several international standards (IEC-60601-1, IEC 60601-1-2, IEC-60601-2-2, ANSI/AAMI/ISO 10993-1, ANSI/AAMI/ISO 11135-1, ANSI/AAMI/ISO 10993-7), implying that compliance with these standards (including their safety and performance requirements) served as acceptance criteria.
2. Sample size used for the test set and the data provenance
The document does not specify the sample sizes used for each test set (e.g., number of devices tested, number of tissue samples, number of users for usability). It also does not mention the provenance of the data (e.g., country of origin, retrospective or prospective). The testing appears to be primarily bench and laboratory testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The document does not provide information regarding the number of experts used or their qualifications for establishing ground truth, as the testing described appears to be primarily technical and engineering-focused rather than involving clinical judgment for ground truth.
4. Adjudication method for the test set
The document does not specify any adjudication method. This is consistent with the type of technical performance and safety testing described, which typically relies on objective measurements and established protocols rather than expert adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
The document does not describe an MRMC comparative effectiveness study, nor does it mention any AI assistance. The device is an electrosurgical handpiece, not an AI-powered diagnostic or assistive tool for human readers.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
The device is a medical instrument (electrosurgical handpiece) that requires human operation. Therefore, the concept of "standalone (algorithm only)" performance without human-in-the-loop is not applicable to this device.
7. The type of ground truth used
The ground truth for the performance tests was established through objective measurements and adherence to engineering and safety standards. For example:
- Mechanical and Electrical Verification: Ground truth would be the device meeting specified mechanical tolerances, electrical safety parameters, and functional requirements.
- Plasma Characteristics: Ground truth would involve quantitative measurements of plasma stream properties under various generator settings.
- Performance Evaluation (on tissue): Ground truth would relate to the observed effects on tissue (cutting, coagulation, ablation depth, etc.) as per predefined expectations for safety and efficacy.
- Usability Evaluation: Ground truth would be based on user feedback and demonstrated safe and effective operation by representative users.
- Biocompatibility: Ground truth from standard tests to ensure material safety.
- Shelf Life: Ground truth from measurements demonstrating maintained performance and integrity over time.
8. The sample size for the training set
The document does not mention a training set. This is because the Bovie J-Plasma Precise 360 Handpiece is a physical medical device (electrosurgical instrument) and not an algorithm or AI system that requires a training set.
9. How the ground truth for the training set was established
Since there is no training set mentioned or implied for this physical device, this question is not applicable.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is the logo for the U.S. Department of Health and Human Services. The logo features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around a central emblem. The emblem consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare, with three human profiles incorporated into the design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 2, 2015
Bovie Medical Corporation Mr. Brian Kunst Vice President, Regulatory Affairs and Quality Assurance 5115 Ulmerton Road Clearwater, Florida 33760
Re: K152570
Trade/Device Name: Bovie J-Plasma Precise 360 Handpiece Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: September 4, 2015 Received: September 10, 2015
Dear Brian Kunst:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the
{1}------------------------------------------------
electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 . and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Joshua C. Nipper -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K152570
Device Name Bovie J-Plasma Precise 360 Handpiece
Indications for Use (Describe)
The Bovie J-Plasma Precise 360 Handpiece is used for the delivery of helium gas plasma for cutting, coagulation, and ablation of soft tissue during open and laparascopic surgical procedures.
The Bovie J-Plasma Precise 360 Handpiece is compatible only with Bovie J-Plasma generators.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for the brand "Bovie". The logo is written in a cursive font and is a light blue color. The logo is slanted upwards from left to right. There is a registered trademark symbol to the right of the word.
510(k) SUMMARY – K152570
(Submitted As Required per 21 CFR 807.92)
GENERAL INFORMATION:
| Submitter Name: | Bovie Medical Corporation |
|---|---|
| Establishment RegistrationNumber: | 3007593903 |
| Submitter Address: | 5115 Ulmerton RoadClearwater, Florida 33760-4004United States of America |
| Submitter Telephone Number: | (727) 803-8617 |
| Submitter FAX Number: | (727) 322-4465 |
| Contact Person: | Brian KunstVice President, Regulatory Affairs and QualityAssurance5115 Ulmerton RoadClearwater, Florida 33760-4004United States of America |
| Contact email: | brian.kunst@boviemed.com |
| Date Prepared: | October 30, 2015 |
| DEVICE IDENTIFICATION: | |
| Proprietary Name: | Bovie J-Plasma Precise 360 Handpiece |
| Common Name: | Electrosurgical Generator Accessory |
| Classification Name: | Electrosurgical Cutting and Coagulation Device andAccessories |
| Model Numbers: | Multiple |
| Classification: | 21CFR 878.4400; Class II; Product Code GEI |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Bovie" in a stylized, cursive font. The word is rendered in a bright blue color. A horizontal line underlines the word, adding to the cursive effect. A registered trademark symbol is present to the right of the word.
Legally Marketed Predicate Device(s):
510(k) Number: K151325
Predicate Device Name: Bovie J-Plasma Handpiece Manufacturer: Bovie Medical Corporation
INTENDED USE/INDICATIONS
The Bovie J-Plasma Precise 360 Handpiece is used for the delivery of helium gas plasma for cutting, coagulation, and ablation of soft tissue during open and laparascopic surgical procedures.
The Bovie J-Plasma Precise 360 Handpiece is compatible only with Bovie J-Plasma generators.
DEVICE DESCRIPTION
The Bovie J-Plasma Precise 360 Handpiece is a sterile, single-use electrosurgical accessory intended to be used in conjunction with Bovie J-Plasma electrosurgical generators for the delivery of helium gas plasma for cutting, coagulating and ablating of soft tissue.
The Bovie J-Plasma Precise 360 Handpiece is available with a retractable needle tip, or with a retractable cutting blade tip for excising tissue. Both tip configurations serve as electrodes to generate helium plasma. A 30 degree angled shaft tip provides improved visibility during use, and a rotating knob is used to orient the angled shaft tip to the optimal position. The device is available with 33cm and 45cm shaft lengths. The device can be used in laparoscopic procedures using a standard 5mm trocar.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a medical device with a handle and a long, thin probe extending from it. The handle is white with blue accents, and the probe is black with a slight bend at the end. The device is likely used for minimally invasive surgical procedures, allowing doctors to access and treat areas inside the body without making large incisions. The brand name "Bovie" is visible in the upper left corner of the image.
The primary differences between the proposed J-Plasma Precise 360 Handpiece and the predicate J-Plasma Handpiece are:
- The J-Plasma Precise 360 Handpiece is available in 33cm and 45cm shaft lengths and the predicate J-Plasma Handpiece is available in 15cm, 33cm, and 45cm shaft lengths
- The J-Plasma Precise 360 Handpiece has a rotating knob that controls an angled shaft tip . which provides improved visibility and optimal positioning at the surgical site. The angled shaft tip uses a medical-grade which is not present in the predicate device. The rotating knob utilizes material which is present on the predicate device. The predicate J-Plasma Handpiece has a straight shaft tip.
The J-Plasma Precise 360 has the same intended use and energy source as the predicate device. There are no technological differences, no changes to the principle of operation or the method of application, and no changes to sterilization methods. Both devices require a Bovie J-Plasma electrosurgical generator and channel helium gas flow over an energized inner electrode to create a plasma stream.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Bovie" in a stylized, cursive font. The word is a bright, light blue color. The font is slanted slightly upwards from left to right. There is a registered trademark symbol to the right of the word.
PERFORMANCE TESTING
Performance testing to assure that the Bovie J-Plasma Precise 360 meets performance requirements was performed and is summarized in the following table:
| Test | Objective | Reports |
|---|---|---|
| Mechanical Verificationand Functionality | Verify the mechanical functionality of thedevice. | VR-1318VR-1324 |
| Electrical Verification | Verify the electrical functionality and safety ofthe device. | VR-1318 |
| Plasma Characteristics | Measure effect of generator settings on theplasma stream, plasma stream characterization | VR-1324 |
| Performance Evaluation | Confirm device performance onvarious tissue types. | VR-1324 |
| Usability Evaluation | To verify and validate the usability of the J-Plasma Precise 360 as it relates to safety. | VP-1387 |
| Shipping Study | Verify the device meets requirements aftershipping and handling. | VP-1384 |
| Evaluation of RotationEffects | Evaluation of Shaft Rotation Effects on Plasmastream for Bovie J-Plasma Precise 360Handpieces | VR-1383 |
| Biocompatibility Testing | Verify material safety | VR-1335 |
| Shelf Life Testing | Confirm product and package performance atlabeled shelf life | VP-1326 |
The Bovie J-Plasma Precise 360 Handpiece, and the predicate Bovie J-Plasma Handpiece, were designed in accordance with the following standards:
| International Standard | Description |
|---|---|
| IEC-60601-1, Edition 3.1 | Medical Electrical Equipment - Part 1: General RequirementsFor Safety |
| IEC 60601-1-2:2007 | Medical electrical equipment - Part 1-2: General requirementsfor basic safety and essential performance - Collateral Standard:Electromagnetic disturbances - Requirements and tests |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Bovie" in a stylized, cursive font. The word is rendered in a light blue color. The "B" is capitalized and slightly larger than the other letters, with a curved flourish extending from the bottom of the "B" and underlining the rest of the word. A registered trademark symbol is located to the right of the "e".
| IEC-60601-2-2 : 2009 | Particular requirements for the safety of high frequency surgicalequipment |
|---|---|
| ANSI / AAMI / ISO 10993-1:2009 | Biological Evaluation of Medical Devices - Part 1: Evaluationand Testing |
| ANSI/AAMI/ISO 11135-1: 2007 | Sterilization of Healthcare Products - Ethylene Oxide - Part 1:Requirements for the development, validation, and routinecontrol of a sterilization process for medical devices. |
| ANSI / AAMI / ISO 10993-7:2008 | Biological Evaluation of Medical Devices – Part 7: EthyleneOxide Residuals |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "Bovie" in a stylized, cursive font. The word is in a bright blue color, and there is a registered trademark symbol to the right of the word. The font is thick and has a flowing, handwritten appearance.
510(k) SUMMARY
SUBSTANTIAL EQUIVALENCE
| Feature/Characteristic | Bovie J-Plasma Precise 360Current Submission | Bovie J-Plasma Handpiece(Predicate K151325) |
|---|---|---|
| Intended Use /Indications for Use | The Bovie J-Plasma Precise 360 isused for the delivery of helium gasplasma for cutting, coagulation, andablation of soft tissue during open andlaparascopic surgical procedures.The Bovie J-Plasma Precise 360 iscompatible only with Bovie J-Plasmagenerators. | The Bovie J-Plasma Handpiece is usedfor the delivery of helium gas plasma forcutting, coagulation, and ablation of softtissue during open and laparascopicsurgical procedures.The Bovie J-Plasma Handpiece iscompatible only with Bovie J-Plasmagenerators. |
| Energy Type | Helium gas plasma | Helium gas plasma |
| Output | Monopolar | Monopolar |
| User Interface | Pistol grip | Pistol grip |
| Shaft Design | Angled | Straight |
| Shaft Rotation | Yes | No |
| Shaft Working Lengths | 33cm and 45cm | 15cm, 33cm, 45cm |
| Shaft Configuration | Angled | Straight |
| Shaft Outer Diameter | 5mm | 5mm |
| Tip Configuration | Blade, Needle | Blade, Needle |
| Blade Extension | Maximum 10mm | Maximum 10mm |
| Blade Width x Thickness | 0.4mm x 0.08mm | 0.4mm x 0.08mm |
| Compatibility | Only with Bovie J-Plasma Generators | Only with Bovie J-Plasma Generators |
| Connector | Bovie Proprietary | Bovie Proprietary |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "Bovie" in a stylized, cursive font. The word is rendered in a bright, light blue color. A registered trademark symbol is located to the right of the word.
CONCLUSION
There is no difference between the J-Plasma Precise 360 Handpiece and the legally marketed predicate device Bovie J-Plasma handpiece described in K151325 in terms of intended use, principle of operation, and the technology used for device performance. All features of the Bovie J-Plasma Precise 360 were subjected to verification testing. There is no new technology and no difference that would raise new or different questions of safety or efficacy. The angled shaft tip and rotating knob components are mechanical in nature and do not affect the technology of the device. The purpose of the angled tip and rotating shaft is to allow better positioning of the tip and better visualization of the target tissue. The devices exhibit substantially equivalent performance during bench testing, including equivalent depth of tissue effect.
§ 878.4400 Electrosurgical cutting and coagulation device and accessories.
(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.