K Number
K151325

Validate with FDA (Live)

Date Cleared
2015-08-04

(78 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Bovie J-Plasma Handpiece is used for the delivery of helium gas plasma for cutting, coagulation, and ablation of soft tissue during open and laparascopic surgical procedures.

The Bovie J-Plasma Handpiece is compatible only with Bovie J-Plasma generators.

Device Description

The Bovie J-Plasma Handpiece is a sterile, single use electrosurgical accessory intended to be used in conjunction with Bovie J-Plasma electrosurgical for the delivery of helium gas plasma for cutting, coagulating, and ablating soft tissue.

The Bovie J-Plasma Handpiece is available with a retractable cutting blade tip for excising tissue or a needle tip. Both tip configurations serve as electrodes to generate helium plasma. The device is available in 15cm, 33cm, and 45cm lengths. The device can be used in laparascopic or open surgical procedures.

AI/ML Overview

The provided text describes a 510(k) summary for the Bovie J-Plasma Handpiece (K151325). The document focuses on demonstrating substantial equivalence to a predicate device, rather than presenting a detailed study with specific acceptance criteria and detailed device performance metrics in a table format.

Therefore, much of the requested information regarding acceptance criteria, reported device performance, sample sizes for test and training sets, data provenance, expert ground truth establishment, MRMC studies, standalone performance, and ground truth types is not explicitly available in the provided document.

However, I can extract the information that is present:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not explicitly present a table of acceptance criteria with corresponding reported device performance metrics. Instead, it lists various performance tests and implies their successful completion to demonstrate equivalence.

Test TitleObjectiveResults/Performance (Implied)
Mechanical Verification and FunctionalityVerify the mechanical functionality of the handpiece.Met performance requirements. (Implied by conclusion of SE)
Electrical VerificationVerify the electrical functionality and safety of the handpiece.Met performance requirements. (Implied by conclusion of SE)
Plasma CharacteristicsMeasure effect of generator settings on the plasma stream, plasma stream characterization.Met performance requirements. (Implied by conclusion of SE)
Performance EvaluationConfirm device performance on various tissue types.Met performance requirements. (Implied by conclusion of SE)
Usability EvaluationVerify and validate the usability of the J-Plasma Pistol Grip Handpiece as it relates to safety.Met performance requirements. (Implied by conclusion of SE)
Shipping StudyVerify the device meets requirements after shipping and handling.Met performance requirements. (Implied by conclusion of SE)

2. Sample size used for the test set and the data provenance:
This information is not provided in the document. The document mentions "Performance testing" but does not detail the sample sizes for any of the tests. Data provenance (country of origin, retrospective/prospective) is also not stated.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not provided in the document. The tests described are primarily bench tests and engineering assessments, not clinical or diagnostic studies that would typically involve expert ground truth.

4. Adjudication method for the test set:
This information is not provided in the document. As there is no mention of expert ground truth, an adjudication method is not applicable in this context.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not provided in the document. The device in question is an electrosurgical handpiece, not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study with human readers and AI assistance would not be relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This information is not provided in the document. The device is a surgical instrument and does not involve an algorithm for standalone performance evaluation in the context of AI.

7. The type of ground truth used:
For the listed performance tests (mechanical, electrical, plasma characteristics, performance evaluation, usability, shipping), the "ground truth" would likely be based on engineering specifications, industry standards (e.g., IEC standards listed), and predefined functional requirements. For "Performance Evaluation on various tissue types," the ground truth would be the observed effect of the device (cutting, coagulation, ablation) on those tissues, assessed against desired outcomes or predicate device performance. However, the specific methodology for establishing this ground truth (e.g., histological analysis, expert observation) is not detailed.

8. The sample size for the training set:
This information is not provided in the document. The document describes a medical device, not an AI model that requires a training set.

9. How the ground truth for the training set was established:
This information is not provided in the document. Not applicable as this is not an AI model.

Summary of what is available:

The document primarily serves as a 510(k) summary for a medical device and describes various performance tests (bench tests) conducted to ensure the device meets requirements and is substantially equivalent to a predicate. It does not provide detailed results or specific acceptance criteria in a quantitative, tabular format beyond the general statement that the device "meets performance requirements." The "study" here refers to the collection of these performance tests. There is no mention of clinical studies, human readers, AI, or large datasets for algorithm training or testing.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines above them. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Bovie Medical Corporation Mr. Brian Kunst Vice President, Quality Assurance/Regulatory Affairs 5115 Ulmerton Road Clearwater, Florida 33760

August 4, 2015

Re: K151325

Trade/Device Name: Bovie J-Plasma Handpiece Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: May 15, 2015 Received: May 18, 2015

Dear Mr. Kunst:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerelv vours.

Joshua C. Nipper -S

For

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151325

Device Name Bovie J-Plasma Handpiece

Indications for Use (Describe)

The Bovie J-Plasma Handpiece is used for the delivery of helium gas plasma for cutting, coagulation, and ablation of soft tissue during open and laparascopic surgical procedures.

The Bovie J-Plasma Handpiece is compatible only with Bovie J-Plasma generators.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "Bovie" in a stylized, cursive font. The word is written in a light blue color. There is a line that underlines the word. The registered trademark symbol is located to the right of the letter "e".

510(k) SUMMARY

K151325

(Submitted As Required per 21 CFR 807.92)

GENERAL INFORMATION:

Submitter Name:Bovie Medical Corporation
Establishment RegistrationNumber:3007593903
Submitter Address:5115 Ulmerton RoadClearwater, Florida 33760-4004United States of America
Submitter Telephone Number:(727) 803-8617
Submitter FAX Number:(727) 322-4465
Contact Person:Brian KunstVice President, Regulatory Affairs and QualityAssurance
Date Prepared:May 15, 2015
DEVICE IDENTIFICATION:
Proprietary Name:Bovie J-Plasma Handpiece
Common Name:Electrosurgical Generator Accessory
Classification Name:Electrosurgical Cutting and Coagulation Device andAccessories
Model Numbers:Multiple
Classification:21CFR 878.4400; Class II; Product Code GEI
Legally MarketedPredicate Device(s):510(k) Number: K112233
Predicate Device Name: Bovie J-Plasma Handpiece
Manufacturer: Bovie Medical Corporation

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Bovie" in a stylized, cursive font. The color of the text is a light blue. A registered trademark symbol is located to the right of the word.

510(k) SUMMARY

INTENDED USE/INDICATIONS

The Bovie J-Plasma Handpiece is used for the delivery of helium gas plasma for cutting, coagulation, and ablation of soft tissue during open and laparascopic surgical procedures.

The Bovie J-Plasma Handpiece is compatible only with Bovie J-Plasma generators.

DEVICE DESCRIPTION

The Bovie J-Plasma Handpiece is a sterile, single use electrosurgical accessory intended to be used in conjunction with Bovie J-Plasma electrosurgical for the delivery of helium gas plasma for cutting, coagulating, and ablating soft tissue.

The Bovie J-Plasma Handpiece is available with a retractable cutting blade tip for excising tissue or a needle tip. Both tip configurations serve as electrodes to generate helium plasma. The device is available in 15cm, 33cm, and 45cm lengths. The device can be used in laparascopic or open surgical procedures.

Image /page/4/Figure/8 description: The image shows a medical device that is white and blue. The device has a long, thin metal rod extending from the front. The handle of the device is shaped like a pistol grip, and there are several buttons and switches on the device.

The primary differences between the proposed J-Plasma handpieces and predicate J-Plasma handpieces are:

  • Pistol grip shape
  • Available in 15cm, 33cm, and 45cm shaft lengths for open and laparoscopic procedures, compared to the predicate 18cm and 27cm lengths
  • . Available with blade or needle electrode instead of only blade electrode

There are no technological differences or changes to principle of operation. There are no changes to intended use, design, material, chemical composition, or energy source. Both devices require a Bovie J-Plasma generator and channel helium gas flow over an energized inner electrode to create a plasma stream.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "Bovie" in a stylized, cursive font. The word is a light blue color, and there is a line extending from the bottom of the "B" that underlines the rest of the word. There is a registered trademark symbol to the right of the "e".

510(k) SUMMARY

PERFORMANCE TESTING

Performance testing to assure that the Bovie J-Plasma Handpiece meets performance requirements was performed and is summarized in the following table:

TestObjectiveProtocols
Mechanical Verification andFunctionalityVerify the mechanical functionality ofthe handpiece.VR-1191VR-1323
Electrical VerificationVerify the electrical functionality andsafety of the handpiece.VR-1191
Plasma CharacteristicsMeasure effect of generator settings onthe plasma stream, plasma streamcharacterizationVR-1323
Performance EvaluationConfirm device performance onvarious tissue types.VR-1323
Usability EvaluationThe purpose of this protocol is toverify and validate the usability of theJ-Plasma Pistol Grip Handpiece as itrelates to safety.VR-1216
Shipping StudyVerify the device meets requirementsafter shipping and handling.VR-1191

The handpiece, and the predicate handpiece, was designed in accordance with the following standards:

International StandardDescription
IEC-60601-1, Edition 3.1Medical Electrical Equipment - Part 1: General RequirementsFor Safety
IEC 60601-1-2:2007Medical electrical equipment - Part 1-2: General requirementsfor basic safety and essential performance - Collateral Standard:Electromagnetic disturbances - Requirements and tests
IEC-60601-2-2 : 2009Particular requirements for the safety of high frequency surgicalequipment
ANSI / AAMI / ISO 10993-1:2009Biological Evaluation of Medical Devices – Part 1: Evaluationand Testing
ANSI / AAMI / ISO 11135-1: 2007Sterilization of Healthcare Products – Ethylene Oxide - Part 1:Requirements for the development, validation, and routinecontrol of a sterilization process for medical devices.
ANSI / AAMI / ISO 10993-7:2008Biological Evaluation of Medical Devices - Part 7: EthyleneOxide Residuals

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "Bovie" in a stylized, cursive font, with a swooping line underlining the word. The text "510(k) SUMMARY" is located below the word "Bovie" and the swooping line. The text "510(k) SUMMARY" is underlined.

SUBSTANTIAL EQUIVALENCE

Feature/CharacteristicBovie J-Plasma HandpieceCurrent SubmissionBovie J-Plasma Handpiece(Predicate K112233)
Intended Use /Indications for UseThe Bovie J-Plasma Handpiece isused for the delivery of helium gasplasma for cutting, coagulation, andablation of soft tissue during open andlaparascopic surgical procedures.The Bovie J-Plasma Handpiece iscompatible only with Bovie J-PlasmageneratorsThe Bovie® J-Plasma Handpiece withRetractable Cutting Blade is anelectrosurgical device that utilizeshelium gas and a retractable blade forcutting and coagulation of soft tissue foruse during both open and laparoscopicsurgical procedures.
ProceduresOpen and LaparascopicOpen and Laparascopic
Energy TypeHelium gas plasmaHelium gas plasma
OutputMonopolarMonopolar
User InterfacePistol gripStraight
Shaft Working Lengths15cm, 33cm, 45cm18cm, 27cm
Shaft Outer Diameter5mm5mm
Tip ConfigurationBlade, NeedleBlade
Blade ExtensionMaximum 10mmMaximum 10mm
Blade Width x Thickness0.4mm x 0.08mm0.4mm x 0.08mm
CompatibilityOnly with Bovie J-Plasma GeneratorsOnly with Bovie J-Plasma Generators
ConnectorBovie ProprietaryBovie Proprietary

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image features the word "Bovie" in a stylized, cursive font. The color of the text is a bright, light blue. A curved line underlines the word, adding to the dynamic and flowing feel of the design. A registered trademark symbol is present to the right of the word.

CONCLUSION

The Bovie J-Plasma Handpiece has the same intended use, principle of operation, and utilizes the same technology as the predicate device. There is no new technology and no difference that would raise new or different questions of safety or efficacy. The differences between the subject device and the predicate device are mainly dimensional. Substantial equivalence was demonstrated through comparative bench testing.

Although there are slight difference in the wording of the indications statement between the subject device and the predicate device, the differences are not critical to the intended therapeutic, diagnostic, prosthetic, or surgical use of the device, and the differences do not affect the safety and effectiveness of the device when used as labeled. The devices are still used for the delivery of helium gas plasma for application on soft tissue during open and laparascopic procedures.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.