K Number
K151752
Date Cleared
2015-11-12

(136 days)

Product Code
Regulation Number
892.1750
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SOMATOM Emotion 6/16 CT systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles.

(*spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)

Device Description

The SOMATOM Emotion 6 and the SOMATOM Emotion 16 are whole body X-ray Computed Tomography Systems. The SOMATOM Emotion 6 and the SOMATOM Emotion 16 produce CT images in DICOM format, which can be used by post-processing applications commercially distributed by Siemens and other vendors.

The system software is a command-based program used for patient management, X-ray scan control, image reconstruction, and image archive/evaluation. The new version of system software, SOMARIS/5 VC30 supports the following modifications: 1) Localized language support of scan protocols, 2) Easy UI improvement, 3) Study Split Improvement, 4) FAST kV, 5) syngo. via client, 6) online help based on knowledge gateway, 7) new software field update concept 8) Temporal-MIP, 9) TrueD-4D viewer, 10) FAST 3D Align, 11) Tube Protection.

AI/ML Overview

Here's a breakdown of the requested information, based on the provided document. It's important to note that this document is a 510(k) summary for a CT scanner's software update, not an AI/CADe device. Therefore, many standard AI/CADe study components (like expert ground truth establishment, MRMC studies, and effect sizes) are not explicitly detailed in the way they would be for an AI device. The focus here is on the system's performance and safety.

Acceptance Criteria and Device Performance for SOMATOM Emotion 6/16 (syngo® CT VC30-easyIQ version)

The document primarily relies on non-clinical testing and conformance to international standards to demonstrate substantial equivalence to predicate devices, rather than a specific set of acceptance criteria tied to clinical endpoints with a test set of patient data. The "acceptance criteria" are implicitly the device's ability to meet the requirements of the listed standards and perform its intended functions without introducing new risks or compromising image quality/safety.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria (Derived from standards/claims)Reported Device Performance (Summary from doc)
Device Safety & Electrical CompatibilityConformance to IEC 60601 series (safety, radiation protection)Designed to fulfill requirements of IEC 60601-2-44, IEC 60601-1, IEC 60601-1-3, IEC 60601-1-6. Risk analysis completed, risk control implemented.
Image Quality & PerformanceConformance to IEC 61223 (evaluation & routine testing of imaging performance) and NEMA XR-25/29 (dose)Designed to fulfill requirements of IEC 61223-3-5, IEC 61223-2-6. Performance data demonstrates continued conformance. Non-clinical tests conducted during product development; modifications supported with verification/validation testing.
Software Functionality & ReliabilityConformance to IEC 62304 (medical device software life cycle processes), FDA Software Guidance (Moderate Level of Concern)Designed to fulfill requirements of IEC 62304. Software Documentation for a Moderate Level of Concern per FDA guidance included. Testing results support that all software specifications have met the acceptance criteria. Verification/validation testing found acceptable. The software supports listed modifications (e.g., Localized language support, Easy UI improvement, Study Split Improvement, FAST kV, syngo.via client, Online help, etc.) as intended.
Data Format & InteroperabilityConformance to NEMA PS 3.1 - 3.14 (DICOM)Designed to fulfill requirements of NEMA PS 3.1 - 3.14. Produces CT images in DICOM format.
Risk ManagementConformance to ISO 14971 (application of risk management)Risk analysis completed and risk control implemented to mitigate identified hazards.
Substantial Equivalence (Overall)Comparable intended use, indications for use, technological characteristics (image acquisition, operating platform, image manipulation) to predicate devices. No new safety/effectiveness concerns.Has the same intended use and comparable indication for use as predicate devices. Technological characteristics are similar to predicate devices. Non-clinical performance data and software validation demonstrate the subject device is as safe and effective. Testing results found acceptable to support the claim of substantial equivalence.

2. Sample Size Used for the Test Set and Data Provenance

The document does not describe a "test set" in the sense of a distinct dataset of patient scans used for evaluating clinical performance of an AI model. Instead, it refers to:

  • Non-clinical tests: These were conducted during product development to verify and validate the software modifications. The document does not specify the sample size of these tests (e.g., number of test cases, simulated data, phantom scans).
  • Data Provenance: Not applicable as there is no mention of a specific patient data test set. The document indicates that the predicate devices were cleared based on "non-clinical supportive information and clinical images," but these are not for the subject device's testing.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not provided in the document as it is not a study evaluating an AI/CADe system using expert-labeled ground truth patient data. The ground truth for this device's performance is established through adherence to engineering standards and functional testing, not clinical expert consensus on patient images.

4. Adjudication Method for the Test Set

This information is not provided as there is no described clinical test set with adjudicated ground truth.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and Effect Size

No, an MRMC comparative effectiveness study was not explicitly mentioned or described. This document is a 510(k) summary for a CT scanner's software update, not an AI-assisted reading device that would typically undergo such a study. The focus is on demonstrating that the updated software does not alter the fundamental safety or effectiveness of the CT system.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The device is a CT scanner, which inherently involves human operation. Its "performance" is evaluated against engineering standards and its functional specifications, not as a standalone diagnostic algorithm in the way an AI would be. So, while independent component testing would have occurred, it's not "standalone algorithm performance" in the typical AI/CADe sense.

7. The Type of Ground Truth Used

The "ground truth" for this submission is based on:

  • Engineering Standards and Specifications: Adherence to international standards (IEC, NEMA, ISO) for CT systems, software, and risk management.
  • Functional Requirements: The new software features (e.g., FAST kV, Temporal-MIP, Easy UI improvement) perform as designed and intended.
  • Predicate Device Equivalence: The ultimate "ground truth" for this 510(k) is that the device is substantially equivalent to legally marketed predicate devices, meaning it is as safe and effective.

8. The Sample Size for the Training Set

This information is not applicable and not provided. This document describes a software update for a CT scanner, not a machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable and not provided as there is no machine learning model or training set described.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other, with flowing lines connecting them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 12, 2015

Siemens Medical Solutions USA, Inc. % Mr. John Urtz Regulatory Affairs Specialist 40 Liberty Boulevard, Mail Code 65-1A MALVERN PA 19355

Re: K151752

Trade/Device Name: SOMATOM Emotion 6/16 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: October 23, 2015 Received: October 23, 2015

Dear Mr. Urtz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Robert Oakes

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The word is written in a light blue color. The letters are all capitalized and evenly spaced. The background is white.

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

Indications for Use

Image /page/2/Picture/5 description: The image shows the text "510(k) Number (if known)" followed by empty boxes. The text indicates that the image is part of a form or document where the 510(k) number, if known, should be entered. The empty boxes suggest that the user should write the number in the provided spaces.

Image /page/2/Picture/6 description: The image shows the text "K151752" above the words "Device Name". There is a line separating the text from the words below. The text is in a simple, sans-serif font and is easy to read. The image is likely a label or identifier for a device.

SOMATOM Emotion 6/16

Indications for Use (Describe)

SOMATOM Emotion 6/16 CT systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles.

(*spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)

Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1

PSC Publishing Services (301) 443-6740

.

{3}------------------------------------------------

510(K) SUMMARY FOR SOMATOM Emotion 6/16 Systems

Submitted by: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355

November 9, 2015

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

1. General Information:

Importer/Distributor Establishment:

Registration Number: 2240869 Siemens Medical Solutions, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355

Manufacturing Facility (1):

SIEMENS SHANGHAI, MEDICAL EQUIPMENT LTD 278 Zhou Zhu Rd Shanghai, CHINA, 201318 Establishment Registration Number: 3003202425

Manufacturing Facility (2):

Siemens AG; Medical Solutions Siemensstrasse 1 91301 Forchheim, GERMANY Establishment Registration Number: 3004977335

2. Contact Person:

John Urtz Regulatory Affairs Specialist Siemens Medical Solutions, Inc. USA 40 Liberty Boulevard Mail Code 65-1A Malvern, PA 19355-1406 Phone: (610) 448-6002 Fax: (610) 640-4481 Email: john.urtz@siemens.com

{4}------------------------------------------------

3. Device Name and Classification

Product Name:SOMATOM Emotion 6
Propriety Trade Name:SOMATOM Emotion 6 (with syngo® CT VC30-easyIQ version)
Classification Name:Computed Tomography X-ray System
Classification Panel:Radiology
CFR Section:21 CFR §892.1750
Device Class:Class II
Product Code:JAK

Device Name and Classification

Product Name:SOMATOM Emotion 16
Propriety Trade Name:SOMATOM Emotion 16 (with syngo® CT VC30-easyIQ version)
Classification Name:Computed Tomography X-ray System
Classification Panel:Radiology
CFR Section:21 CFR §892.1750
Device Class:Class II
Product Code:JAK

Legally Marketed Primary Predicate Device:

Product Name:SOMATOM Emotion 6/16
Propriety Trade Name:SOMATOM Emotion 6/16 (with syngo® CT 2013A ((SOMARIS/5 VC20))
Classification Name:Computed Tomography X-ray System
Classification Panel:Radiology
CFR Section:21 CFR §892.1750
Device Class:Class II
Product Code:JAK
Clearance Number:K133424

Legally Marketed Secondary Predicate Device:

Product Name:SOMATOM Definition Edge
Propriety Trade Name:SOMATOM Definition Edge
Classification Name:Computed Tomography X-ray System
Classification Panel:Radiology
CFR Section:21 CFR §892.1750
Device Class:Class II
Product Code:JAK
Clearance Number:K143401

4. Substantial Equivalence:

Siemens SOMATOM Emotion 6 and SOMATOM Emotion 16 configured with software version syngo® CT VC30-easyIQ version (SOMARIS/5 VC30) is substantially equivalent to the following medical devices in commercial distribution:

{5}------------------------------------------------

Table 1 Predicate Devices

ManufacturerPrimary Predicate Device510(k)NumberClearance Date
Siemenssyngo® CT 2013A (SOMARIS/5 VC20B) systemsoftware for SOMATOM Emotion 6 CT andSOMATOM Emotion 16 sysyemsK13342401/17/2014
ManufacturerSecondary Predicate Device510(k)NumberClearance Date
SiemensSOMATOM Definition Edge with SOMARIS/7 VA48K14340104/06/2015

ട. Device Description:

The SOMATOM Emotion 6 and the SOMATOM Emotion 16 are whole body X-ray Computed Tomography Systems. The SOMATOM Emotion 6 and the SOMATOM Emotion 16 produce CT images in DICOM format, which can be used by post-processing applications commercially distributed by Siemens and other vendors.

The system software is a command-based program used for patient management, X-ray scan control, image reconstruction, and image archive/evaluation. The new version of system software, SOMARIS/5 VC30 supports the following modifications: 1) Localized language support of scan protocols, 2) Easy UI improvement, 3) Study Split Improvement, 4) FAST kV, 5) syngo. via client, 6) online help based on knowledge gateway, 7) new software field update concept 8) Temporal-MIP, 9) TrueD-4D viewer, 10) FAST 3D Align, 11) Tube Protection.

6. Indications for Use

The SOMATOM Emotion 6/16 systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles.

(*spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)

7. Summary of Technical Characteristics of the Subject Device as Compared with the Predicate Device:

SOMATOM Emotion 6/16 configured with software version syngo® VC30-easyIQ version does not have significant changes in materials, energy source, or technological characteristics when compared to the predicate devices. Both the subject device and predicate devices are computed tomography scanners that support various visualization and evaluation tools. The intended use and fundamental scientific technology remain unchanged from the predicate devices; therefore Siemens believes that they are substantially equivalent to the predicate devices. Table 2 below provides a comparison of the primary features of the subject device in comparison to the predicate device.

{6}------------------------------------------------

Table 2 Predicate Device Comparison

Subject Device FeaturePredicate Device Feature
New Software version syngo® CTVC30-easyIQ versionSystem scanner softwaresyngo® CT 2013AK133424System scanner softwaresyngo® CT VA48K143401
Temporally MIP (t-MIP)N/ATemporally MIP (t-MIP)
FAST 3D AlignN/AFAST 3D Align
TrueD-4D ViewerN/ATrueD-4D Viewer
Fast kVN/AN/A
Localized Language Support ofScan ProtocolsN/AN/A
Easy UI ImprovementEasy UIEasy UI
Study Split ImprovementStudy SplitStudy Split
syngo.via clientN/Asyngo.via client
Online Help Based on KnowledgeGatewayOnline help availableOnline help available
New Software Field UpdateConceptN/AN/A
Tube ProtectionN/AN/A

8. Nonclinical Testing:

SOMATOM Emotion 6/16 configured with software version syngo® VC30-easyIQ version is designed to fulfill the requirements of the following standards:

  • IEC 60601-2-44: Medical electrical equipment – Part 2-44: Particular requirements for the safety of X-ray equipment for computed tomography - Ed. 3.1
  • IEC 61223-3-5: Evaluation and routine testing Evaluation and routine testing in medical imaging departments - Part 3-5: Acceptance tests - Imaging performance of computed tomography X-ray equipment CORRIGENDUM 1
  • . NEMA XR-25: Computed Tomography Dose Check
  • IEC 61223-2-6: Evaluation and routine testing in medical imaging departments Part 2-6: Constancy . tests - Imaging performance of computed tomography X-ray equipment
  • NEMA PS 3.1 - 3.14: Digital Imaging and Communications in Medicine DICOM) Set
  • IEC 62304 Ed. 1.0: Medical device software software life cycle processes .
  • IEC 60601-1: Medical electrical equipment Part 1: General requirements for Safety, 2005 .
  • IEC 60601-1-3: 2008, Medical electrical equipment Part 11-3: General requirements for radiation . protection in diagnostic X-ray equipment
  • IEC 60601-1-6: 2010; Medical electrical equipment Part 1-6: General requirements for basic safety . and essential performance -- Collateral Standard: Usability

{7}------------------------------------------------

  • ISO 14971: Medical devices – Application of risk management to medical devices
  • NEMA XR-29: Standard Attributes on CT Equipment Related to Dose Optimization and Management .
  • NEMA XR-26: Access Controls for Computed Tomography—Identification, Interlocks, and Logs ●
  • IEC/ISO 10918: Information Technology Digital Compression and Coding of Continuous-Tone Still . Images: Requirements and Guidelines [Including: Technical Corrigendum (2005)]

This submission contains performance data to demonstrate continued conformance with special controls for medical devices containing software. Non clinical tests were conducted for the SOMATOM Emotion 6/16 configured with software version syngo® VC30-easyIQ version during product development. The modifications described in this Premarket Notification were supported with verification/validation testing.

The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results supports that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.

Software Verification and Validation

Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission.

Summary

Performance tests were conducted to test the functionality of the SOMATOM Emotion 6/16 configured with software version syngo® CT VC30-easyIQ version. These tests have been performed to test the ability of the included features of the subject device. The results of these tests demonstrate that the subject device performs as intended. The result of all conducted testing was found acceptable to support the claim of substantial equivalence.

9. General Safety and Effectiveness Concerns:

The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device. Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.

10. Conclusion as to Substantial Equivalence:

The SOMATOM Emotion 6/16 configured with software version syngo® CT VC30-easyIQ version has the same intended use and comparable indication for use as the predicate devices. The technological characteristics such as image acquisition, operating platform, and image manipulation are similar to the predicate devices.

The predicate devices were cleared based on non-clinical supportive information and clinical images. The results of these tests demonstrate that syngo® CT 2013A (SOMARIS/5 VC20B) system software for SOMATOM Emotion 6 CT systems and SOMATOM Emotion 16 (K133424) are adequate for the intended use. The comparison of technological characteristics, non-clinical performance data, and software validation demonstrates that the subject device is as safe and effective when compared to the predicate devices that are currently marketed for the same intended use.

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.