K Number
K151698
Date Cleared
2015-07-27

(33 days)

Product Code
Regulation Number
880.5200
Panel
HO
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

An intravascular catheter is a device that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure.

Device Description

Insyte™, Insyte-N™, and Insyte-W™ are available in 14-24 gauge and are made of BD Vialon™ Material. Angiocath™ is available in 14-24 gauge and is made of FFP Polymer material

AI/ML Overview

This is an IV catheter from Becton Dickinson, which is a class II device under 21 CFR 880.5200. This device, named BD Angiocath™ and Insyte™ IV Catheters, did not have any AI/ML components in 2015. So, the request for acceptance criteria for such components cannot be fulfilled.

However, I can provide the acceptance criteria and performance data for the mechanical and physical tests performed on the device.

1. Table of Acceptance Criteria and Reported Device Performance

Characteristic / Test PerformedTest MethodAcceptance CriteriaReported Device Performance
BiocompatibilityPer ISO 10993-1Meet biocompatibility standardsPass
Liquid leakagePer ISO 594-1, -2No liquid leakagePass
Air leakagePer ISO 594-1, -2No air leakagePass
Separation forcePer ISO 594-1, -2Withstand specified separation forcePass
Stress crackingPer ISO 594-1, -2No stress crackingPass
Unscrewing torquePer ISO 594-2Withstand specified unscrewing torquePass
Ease of assemblyPer ISO 594-2Easy assemblyPass
Resistance to overridingPer ISO 594-2Resistant to overridingPass
GeneralPer ISO 10555-1Meet general requirementsPass
Radio-detectabilityPer ISO 10555-1Radio-detectablePass
Biocompatibility (repeated)Per ISO 10555-1Meet biocompatibility standardsPass
SurfacePer ISO 10555-1Acceptable surface finishPass
Corrosion resistancePer ISO 10555-1Corrosion-resistantPass
Peak tensile forcePer ISO 10555-1Withstand specified peak tensile forcePass
Freedom from leakagePer ISO 10555-1Freedom from leakagePass
HubsPer ISO 10555-1Meet hub requirementsPass
FlowratePer ISO 10555-1Meet specified flowratePass
Power injectionPer ISO 10555-1Suitable for power injectionPass
Side holesPer ISO 10555-1N/A (Not applicable / no side holes)NA
Distal tipPer ISO 10555-1Meet distal tip requirementsPass
Color codePer ISO 10555-5Correct color codingPass
Catheter unitPer ISO 10555-5Meet catheter unit requirementsPass
Needle pointPer ISO 10555-5Meet needle point requirementsPass
Needle hubPer ISO 10555-5Meet needle hub requirementsPass
Strength of union between needle hub and needle tubePer ISO 10555-5Withstand specified union strengthPass
Vent fittingPer ISO 10555-5Meet vent fitting requirementsPass
Needle (Cannula) penetration forceInternal design input, identical to predicatePenetration force specified by internal design inputPass
Catheter penetration forceInternal design input, identical to predicatePenetration force specified by internal design inputPass
Catheter average drag forceInternal design input, identical to predicateAverage drag force specified by internal design inputPass
Minimal catheter and cannula tip adhesionInternal design input, identical to predicateMinimal adhesion specified by internal design inputPass

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the exact sample sizes used for each individual test. It states that "design verification tests were performed based on the risk analysis." The data provenance is internal, as these were tests conducted by the manufacturer, Becton Dickinson Infusion Therapy Systems Inc., as part of their design control process for regulatory submission. The document defines these as "design verification tests."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This device is a physical medical device, and its performance criteria are established through standardized international and internal engineering test methods, not through expert consensus on medical images or clinical outcomes.

4. Adjudication Method for the Test Set

Not applicable. Performance testing is based on objective measurements against defined standards and specifications, not on subjective interpretations requiring an adjudication method.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI/ML diagnostic tools that humans interact with, which is not the case for this physical medical device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This device is a physical medical product, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the performance tests presented are established international standards (ISO standards like ISO 10993-1, ISO 594-1, ISO 594-2, ISO 10555-1, ISO 10555-5) and internal design inputs that set specific thresholds for physical and mechanical properties.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/ML device, so there is no training set in the context of machine learning.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set.

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).