K Number
K151698

Validate with FDA (Live)

Date Cleared
2015-07-27

(33 days)

Product Code
Regulation Number
880.5200
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

An intravascular catheter is a device that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure.

Device Description

Insyte™, Insyte-N™, and Insyte-W™ are available in 14-24 gauge and are made of BD Vialon™ Material. Angiocath™ is available in 14-24 gauge and is made of FFP Polymer material

AI/ML Overview

This is an IV catheter from Becton Dickinson, which is a class II device under 21 CFR 880.5200. This device, named BD Angiocath™ and Insyte™ IV Catheters, did not have any AI/ML components in 2015. So, the request for acceptance criteria for such components cannot be fulfilled.

However, I can provide the acceptance criteria and performance data for the mechanical and physical tests performed on the device.

1. Table of Acceptance Criteria and Reported Device Performance

Characteristic / Test PerformedTest MethodAcceptance CriteriaReported Device Performance
BiocompatibilityPer ISO 10993-1Meet biocompatibility standardsPass
Liquid leakagePer ISO 594-1, -2No liquid leakagePass
Air leakagePer ISO 594-1, -2No air leakagePass
Separation forcePer ISO 594-1, -2Withstand specified separation forcePass
Stress crackingPer ISO 594-1, -2No stress crackingPass
Unscrewing torquePer ISO 594-2Withstand specified unscrewing torquePass
Ease of assemblyPer ISO 594-2Easy assemblyPass
Resistance to overridingPer ISO 594-2Resistant to overridingPass
GeneralPer ISO 10555-1Meet general requirementsPass
Radio-detectabilityPer ISO 10555-1Radio-detectablePass
Biocompatibility (repeated)Per ISO 10555-1Meet biocompatibility standardsPass
SurfacePer ISO 10555-1Acceptable surface finishPass
Corrosion resistancePer ISO 10555-1Corrosion-resistantPass
Peak tensile forcePer ISO 10555-1Withstand specified peak tensile forcePass
Freedom from leakagePer ISO 10555-1Freedom from leakagePass
HubsPer ISO 10555-1Meet hub requirementsPass
FlowratePer ISO 10555-1Meet specified flowratePass
Power injectionPer ISO 10555-1Suitable for power injectionPass
Side holesPer ISO 10555-1N/A (Not applicable / no side holes)NA
Distal tipPer ISO 10555-1Meet distal tip requirementsPass
Color codePer ISO 10555-5Correct color codingPass
Catheter unitPer ISO 10555-5Meet catheter unit requirementsPass
Needle pointPer ISO 10555-5Meet needle point requirementsPass
Needle hubPer ISO 10555-5Meet needle hub requirementsPass
Strength of union between needle hub and needle tubePer ISO 10555-5Withstand specified union strengthPass
Vent fittingPer ISO 10555-5Meet vent fitting requirementsPass
Needle (Cannula) penetration forceInternal design input, identical to predicatePenetration force specified by internal design inputPass
Catheter penetration forceInternal design input, identical to predicatePenetration force specified by internal design inputPass
Catheter average drag forceInternal design input, identical to predicateAverage drag force specified by internal design inputPass
Minimal catheter and cannula tip adhesionInternal design input, identical to predicateMinimal adhesion specified by internal design inputPass

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the exact sample sizes used for each individual test. It states that "design verification tests were performed based on the risk analysis." The data provenance is internal, as these were tests conducted by the manufacturer, Becton Dickinson Infusion Therapy Systems Inc., as part of their design control process for regulatory submission. The document defines these as "design verification tests."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This device is a physical medical device, and its performance criteria are established through standardized international and internal engineering test methods, not through expert consensus on medical images or clinical outcomes.

4. Adjudication Method for the Test Set

Not applicable. Performance testing is based on objective measurements against defined standards and specifications, not on subjective interpretations requiring an adjudication method.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI/ML diagnostic tools that humans interact with, which is not the case for this physical medical device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This device is a physical medical product, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the performance tests presented are established international standards (ISO standards like ISO 10993-1, ISO 594-1, ISO 594-2, ISO 10555-1, ISO 10555-5) and internal design inputs that set specific thresholds for physical and mechanical properties.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/ML device, so there is no training set in the context of machine learning.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 27, 2015

Becton Dickinson Infusion Therapy Systems, Incorporated c/o Mr. Mark Job Regulatory Technology Services, Inc. 1394 25th Street, NW Buffalo, MN 55313

Re: K151698

Trade/Device Name: BD Angiocath™ and Insyte™ Catheters Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: July 15, 2015 Received: July 16, 2015

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Mark Job

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Tina
Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICESFood and Drug AdministrationForm Approved: OMB No. 0910-0120Expiration Date: January 31, 2017See PRA Statement below.
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Indications for Use

510(k) Number ( if known )K151698
Device NameBD Angiocath IV Catheter and Insyte IV Catheter

Indications for Use (Describe)

An intravascular catheter is a device that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure.

Type of Use ( Select one or both, as applicable )
----------------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5

510(k) Summary

510(k) Summary

21 CFR §807.92

BD Angiocath™ and Insyte™ IV Catheters

SubmitterInformationSubmitter Name:Becton Dickinson Infusion Therapy Systems Inc.
Submitter Address:9450 South State StreetSandy, UT 84070
Contact Person:Henry BolandSenior Regulatory Affairs SpecialistPhone: 801.565.2550Fax: 801.304.3963Email: henry_boland@bd.com
Date of Preparation:14 May 2015
SubjectDevicesTrade Name:BD Angiocath™ IV Catheter
Common Name:Peripheral Intravascular or IV Catheter
Classification Name:FOZ - Intravascular Catheter
CFR Reference:21 CFR 880.5200 - Class II
Classification Panel:General Hospital
Trade Name:BD Insyte™ IV Catheter
Common Name:Peripheral Intravascular or IV Catheter
Classification Name:FOZ - Intravascular Catheter
CFR Reference:21 CFR 880.5200 - Class II
Classification Panel:General Hospital
PredicateDevicesTrade Name:BD Angiocath™ and Insyte™ IV Catheter
510(k) Reference:K013800
Common Name:Peripheral Intravascular or IV Catheter
Classification Name:FOZ - Intravascular Catheter
CFR Reference:21 CFR 880.5200 - Class II
Classification Panel:General Hospital
DeviceDescriptionInsyte™, Insyte-N™, and Insyte-W™ are available in 14-24 gauge and are madeof BD Vialon™ Material. Angiocath™ is available in 14-24 gauge and is made ofFFP Polymer material

{4}------------------------------------------------

Indications for An intravascular catheter is a device that is inserted into the patient's vascular Use system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure.

Technological characteristics of the subject devices are equivalent to the Technological predicates. The BD Angiocath™ and Insyte™ IV Catheters achieve their intended Characteristics use based on the same technology and principles of operation. The subject devices have been modified from the predicate devices as listed below. The results of design verification demonstrate that these changes are substantially equivalent to the predicate devices. All other aspects of the subject device are identical to those of the predicate device.

· Change to lubricants used in the assembly on the device

AttributeSUBJECT DEVICEPREDICATE DEVICE
Indications forUseAn intravascular catheter is a devicethat is inserted into the patient'svascular system for short term use(less than 30 days) to sample blood,monitor blood pressure, or administerfluids intravenously. These cathetersmay be used for any patient populationwith consideration given to adequacyof vascular anatomy andappropriateness of procedure.An intravascular catheter is a devicethat is inserted into the patient'svascular system for short term use(less than 30 days) to sample blood,monitor blood pressure, or administerfluids intravenously. These cathetersmay be used for any patient populationwith consideration given to adequacyof vascular anatomy andappropriateness of procedure.
DesignA fluorinated ethylene propylene (FEP)or polyurethane peripheral IV catheterA fluorinated ethylene propylene (FEP)or polyurethane peripheral IV catheter
ComponentSubject DeviceComponentPredicate Device
MaterialsCatheter TubingAngiocath™ IV CatheterFluorinated ethylenepropylene (FEP)Insyte™ IV CatheterBD Vialon™ polyurethaneCatheter TubingAngiocath™ IV CatheterFluorinated ethylenepropylene (FEP)Insyte™ IV CatheterBD Vialon™ polyurethane
Catheter HubPolypropyleneCatheter HubPolypropylene
Metal WedgeStainless SteelMetal WedgeStainless Steel
NeedleStainless SteelNeedleStainless Steel
Needle HubAngiocath™ IV CatheterPropionateInsyte™ IV CatheterPropionateNeedle HubAngiocath™ IV CatheterPropionateInsyte™ IV CatheterPropionate
Needle CoverPolypropyleneNeedle CoverPolypropylene

{5}------------------------------------------------

Materials(cont.)Vent PlugAngiocath™ IV CatheterPolyethyleneAngiocath™ IV CatheterPolyethylene
Insyte™ IV CatheterPolypropylene with filterInsyte™ IV CatheterPolypropylene with filter
Catheter TippingLubricantPolydimethylsiloxane-basedLubricantPolydimethylsiloxane-basedLubricant
Catheter LubricantPolydimethylsiloxane-basedLubricantPolydimethylsiloxane-basedLubricant
Wedge LubricantPolydimethylsiloxane-basedLubricantPolydimethylsiloxane-basedLubricant
Needle (Cannula)LubricantTwo-PartPolydimethylsiloxane-basedLubricantSingle-PartPolydimethylsiloxane-basedLubricant
Physical /MechanicalSpecificationsSUBJECT DEVICESPREDICATE DEVICES
Angiocath™ IV CatheterAngiocath™ IV Catheter
Catheter Diameters:14G, 16G, 18G, 20G, 22G, 24GCatheter Diameters:14G, 16G, 18G, 20G, 22G, 24G
Catheter Lengths:0.75", 1.00", 1.16", 1.88"Non-winged onlyCatheter Lengths:0.75", 1.00", 1.16", 1.88"Non-winged only
Insyte™ IV CatheterInsyte™ IV Catheter
Catheter Diameters:14G, 16G, 18G, 20G, 22G, 24GCatheter Diameters:14G, 16 G, 18G, 20G, 22G, 24G
Catheter Lengths:0.56", 0.75", 1.00", 1.16", 1.77", 1.88"Winged or Non-wingedCatheter Lengths:0.56", 0.75", 1.00", 1.16", 1.77", 1.88"Winged or Non-winged

Summary of Performance Tests

Pursuant to 21 CFR 820.30, Design Controls, design verification tests were performed based on the risk analysis, and the results of these tests demonstrate that the BD Angiocath™ and Insyte™IV Catheters are substantially equivalent to the predicate devices. Design verification testing included the following:

Characteristic / Test PerformedTest MethodResults
BiocompatibilityPer ISO 10993-1Pass
Liquid leakagePer ISO 594-1, -2Pass
Air leakagePer ISO 594-1, -2Pass
Separation forcePer ISO 594-1, -2Pass
Stress crackingPer ISO 594-1, -2Pass
Unscrewing torquePer ISO 594-2Pass
Ease of assemblyPer ISO 594-2Pass
Resistance to overridingPer ISO 594-2Pass
GeneralPer ISO 10555-1Pass
Radio-detectabilityPer ISO 10555-1Pass
BiocompatiblityPer ISO 10555-1Pass
SurfacePer ISO 10555-1Pass
Corrosion resistancePer ISO 10555-1Pass
Peak tensile forcePer ISO 10555-1Pass
Freedom from leakagePer ISO 10555-1Pass
HubsPer ISO 10555-1Pass
FlowratePer ISO 10555-1Pass
Power injectionPer ISO 10555-1Pass
Side holesPer ISO 10555-1NA
Distal tipPer ISO 10555-1Pass
Color codePer ISO 10555-5Pass
Catheter unitPer ISO 10555-5Pass
Needle pointPer ISO 10555-5Pass
Needle hubPer ISO 10555-5Pass
Strength of union between needle hub and needle tubePer ISO 10555-5Pass
Vent fittingPer ISO 10555-5Pass
Needle (Cannula) penetration forceInternal design input, identical to predicatePass
Catheter penetration forceInternal design input, identical to predicatePass
Catheter average drag forceInternal design input, identical to predicatePass
Minimal catheter and cannula tip adhesionInternal design input, identical to predicatePass
Summary of Substantial EquivalenceBased on the indications for use, technological characteristics, and performance testing, the subject BD AngiocathTM and InsyteTM IV Catheters meet all predetermined requirements in accordance with 21 CFR 820.30, Design Controls, and demonstrates that the subject devices are substantially equivalent to the predicate devices.

{6}------------------------------------------------

{7}------------------------------------------------

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).