K Number
K150393
Date Cleared
2015-06-09

(111 days)

Product Code
Regulation Number
872.3690
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

– As initial layer / first increment in Class I and II composite restorations in permanent teeth

  • Restorations in deciduous teeth
Device Description

Tetric EvoFlow® Bulk Fill is a flowable, light-curing radiopaque composite. As its opacity increases during polymerization, it is also suitable for discolored tooth structure. It is applied in increments of up to 4 mm as an initial layer in Class I and II restorations. Tetric EvoFlow Bulk Fill is intended for use as a dental filling material.

AI/ML Overview

The provided text is a 510(k) summary for a dental device, Tetric EvoFlow Bulk Fill. It details the device's characteristics and its comparison to a predicate device, Tetric EvoCeram Bulk Fill, to demonstrate substantial equivalence.

Here's an analysis of the provided information regarding acceptance criteria and supporting studies:

  1. A table of acceptance criteria and the reported device performance

The document does not specify general "acceptance criteria" for the device as a whole beyond demonstrating substantial equivalence to the predicate. Instead, it lists specific properties that were tested and compared to the predicate, often referencing ISO 4049.

PropertyAcceptance Criteria (Implicit from ISO 4049 and predicate comparison)Reported Device Performance (Tetric EvoFlow Bulk Fill)
BiocompatibilityIn accordance with ISO 10993-1:2009 and EN ISO 7405:2008Tested for cytotoxicity and Mutagenicity, photoinitiator tested in vitro; assessed via comparison to predicate and literature review. Performance not significantly different from predicate.
Flexural StrengthCompared to predicate (K111958)Tested and compared to predicate
Layer thickness/Curing DepthMax 4mm, compared to predicate (K111958)Max 4mm, tested and compared to predicate
Sensitivity to ambient lightRaised specification criteria (above ISO 4049)Achieves more than required by ISO 4049
Water SorptionCompared to predicate (K111958)Tested and compared to predicate
SolubilityCompared to predicate (K111958)Tested and compared to predicate
RadiopacityRaised specification criteria (above ISO 4049)Achieves more than required by ISO 4049
TransparencyCompared to predicate (K111958)Reduced (due to indication as dentin replacement)
Flexural ModulusCompared to predicate (K111958)Difference found (reflects consistency difference)
Vickers hardness profile (at 4mm layer)Equivalent to predicate (K111958)Results equivalent
  1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not provide details on the sample sizes used for the testing, nor does it specify the data provenance (e.g., country of origin, retrospective or prospective). It simply states that the device was "evaluated" and "tested."

  1. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This section is not applicable as the studies described are laboratory-based physical and chemical property tests, not clinical studies requiring expert ground truth for interpretation.

  1. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This section is not applicable for the same reason as point 3.

  1. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable. The device is a dental filling material, not an AI-assisted diagnostic or imaging device. Therefore, no MRMC study, human readers, or AI assistance is relevant.

  1. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This section is not applicable. The device is a physical dental material, not an algorithm.

  1. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the physical and chemical property tests, the "ground truth" or reference points are defined by established international standards (ISO 4049, ISO 10993-1, EN ISO 7405) and the characteristics of the legally marketed predicate device (Tetric EvoCeram Bulk Fill). For biocompatibility, it also includes literature review.

  1. The sample size for the training set

This section is not applicable as the device is a physical material and does not involve AI or machine learning that would require a training set.

  1. How the ground truth for the training set was established

This section is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a symbol. The symbol is a stylized representation of human profiles facing right, suggesting a sense of community and care. The text is arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 9, 2015

Ivoclar Vivadent AG c/o Ms. Donna Marie Hartnett Ivoclar Vivadent, INC 175 Pineview Drive Amherst. NY 14228

Re: K150393

Trade/Device Name: Tetric EvoFlow Bulk Fill Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: II Product Code: EBF Dated: March 4, 2015 Received: March 10, 2015

Dear Ms. Hartnett:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina
Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K150393

Device Name TETRIC EVOFLOW BULK FILL

Indications for Use (Describe)

  • As initial layer / first increment in Class I and II composite restorations in permanent teeth

  • Restorations in deciduous teeth
    Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

Tetric EvoFlow Bulk Fill

Rev. 3.6.15

Image /page/3/Picture/3 description: The image is a logo for Ivoclar Vivadent. The logo is blue and features the company name in a stylized font. Above the company name are green and blue dots and squares. Below the company name is the slogan "passion vision innovation" in a smaller font.

Contact:Donna Marie Hartnett, Ivoclar Vivadent, Inc.175 Pineview Drive, Amherst, NY 14228
Company:Ivoclar Vivadent, AGBendererstrasse 2, SchaanFL-9494, Liechtenstein+423-235-3535
Date Prepared:June 3, 2015
Proprietary Name:Tetric EvoFlow Bulk Fill
Classification Name:Material, Tooth Shade, Resin (872.3690)(Classification Code EBF)
Predicate Device:Tetric EvoCeram Bulk Fill (K111958)

Device Description: Tetric EvoFlow® Bulk Fill is a flowable, light-curing radiopaque composite. As its opacity increases during polymerization, it is also suitable for discolored tooth structure. It is applied in increments of up to 4 mm as an initial layer in Class I and II restorations. Tetric EvoFlow Bulk Fill is intended for use as a dental filling material.

Intended Use:

– As initial layer / first increment in Class I and II composite restorations in permanent teeth

  • Restorations in deciduous teeth

Comparison to Predicate: The predicate device to which Tetic EvoFlow Bulk Fill has been compared is Tetric EvoCeram Bulk Fill (K111958). For this application, Tetric EvoFlow Bulk Fill has been compared to its predicate with regard to chemical composition, physical properties, operating principles and indications for use. The comparison shows that Tetric EvoFlow Bulk Fill is substantially equivalent to the predicate device.

{4}------------------------------------------------

510(K) SUMMARY Tetric EvoFlow Bulk Fill

Rev. 3.6.15

Image /page/4/Picture/2 description: The image shows the logo for Ivoclar Vivadent. The logo is composed of the company name in blue, with the first word, "Ivoclar," stacked on top of the second word, "Vivadent." Above the company name is a semi-circle of green and blue dots and squares. The dots are on the left side of the semi-circle and the squares are on the right side.

passion vision innovation

Predicate - Tetric EvoCeram Bulk Fill(K111958)Subject Device
Tetric EvoCeram Bulk Fill is a light-curing, radiopaque dental filling composite.The special features are:Tetric EvoFlow Bulk Fill is a flowable, light-curing radiopaque dental filling composite.The special features are:
- It is applied in increments of up to 4 mm, saving the dentist time- it is applied in increments of up to 4 mm, saving the dentist time- that its opacity increases during polymerization. It is intended to be a dentin replacement and is also suitable for discolored tooth structure
Indications for UseIndications for Use
- Restorations of deciduous teeth- Restorations in the posterior region (Classes I and II)- Class V restorations (cervical caries, root erosion, wedge-shaped defects)- Extended fissure sealing in molars and premolars- As initial layer / first increment in Class I and II composite restorations in permanent teeth- Restorations in deciduous teethThe indications are reduced because the material has been designed with an opacity similar to dentin. Flowables are often used by dentists as luting composites and this is excluded by the subject device.
Working principle- Shade selection- Isolation- Cavity preparation- Pulp protection / base- Application of the matrix / interdental wedge- Conditioning / application of the bonding agent- Application of Tetric EvoCeram Bulk Fill in layers of maximum 4mm polymerizing each layer.- Finishing / checking the occlusion / polishingWorking principle remains unchanged from predicate. The consistency has been changed to flowable for easy application as the first layer of a restoration. An optimization of the device makes it possible to have a materials which is transparent enough to be polymerized in 4mm layer but them sufficiently opaque to achieve the esthetics of dentin.

Technological Characteristics and Testing Summary: FDA Guidance Document for Dental Composite Resin Devices dated October 26, 2005 was followed in connection with this application.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Ivoclar Vivadent, a dental company. The logo features the company name in two lines, with "Ivoclar" in a larger, bold, teal font above "Vivadent" in a similar font. Above the company name is a semi-circular arrangement of green and teal dots and squares. Below the company name is the company's slogan, "passion vision innovation" in a smaller, sans-serif font.

The device was evaluated for biocompatibility in accordance with ISO 10993-1:2009 and EN ISO 7405:2008. The device was tested for cytotoxicity and Mutagenicity. In addition, the photoinitiator was tested in a cytotoxicity assay in vitro. Biocompatibility was further assessed based on a comparison to the predicate device and review of the literature.

The device was tested in accordance with ISO 4049 Dentistry - Polymer-based restorative materials (ISO 4049:2009) as a Type 1, Class 2 restorative. The following properties were tested and compared to the predicate:

Flexural Strength Laver thickness/Curing Depth Sensitivity to ambient light Water Sorption Solubility Radiopacity Transparency Flexural Modulus

The specification criteria for sensitivity to light and radiopacity have been raised so that the product achieves more than required by ISO 4049. Transparency of the subject device is reduced because of its indication as a dentin replacement. A difference in flexural modulus was found and reflects the difference in consistency of the subject device from the predicate.

Both products are applied in a maximum layer thickness of 4 mm. The performance of the subject device compared to the predicate was also evaluated using the Vickers hardness profile and the results were equivalent.

Tetric EvoFlow Bulk Fill is a flowable version of the predicate device, Tetric EvoCeram Bulk Fill. The composition has been optimized to achieve the desired consistency and translucency. Overall the performance and biocompatibility are not significantly different.

CONCLUSION: In conclusion, Tetric EvoFlow Bulk Fill is substantially equivalent to the predicate device, Tetric EvoCeram Bulk Fill.

§ 872.3690 Tooth shade resin material.

(a)
Identification. Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.(b)
Classification. Class II.