K Number
K150386

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-11-05

(260 days)

Product Code
Regulation Number
882.5890
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Tyece OTC TENS Device, Model SEM44 is to be used for temporary relief of pain associated with sore and aching muscles in the lower back, arms, or legs due to strain from exercise or normal household activities.

Device Description

The Tyece OTC TENS Device, Model SEM44 is a dual channel TENS device operated by DC 4.5V (3 AAA batteries). It is made up of one main unit, two electrode cables, and two pairs of electrode pads. There are 12 selectable, pre-programmed output waveforms (modes) to choose from, blus 3 programmable modes. The two channels can be individually adjusted for intensity from 0 to 50. Running time can be selected from 5 to 100 minutes. All parameters (Menu, Program Number, Impulse intensity for channels, Jow battery, key-lock, frequency, pulse width and timer) are displayed on a full screen display. The only patient contacting components are the Electrode Pads. The patient contacting materials are FDA cleared under K132588.

AI/ML Overview

This is a premarket notification (510(k)) for a Transcutaneous Electrical Nerve Stimulator (TENS) device. The purpose of a 510(k) submission is to demonstrate that the new device is substantially equivalent to a legally marketed predicate device, meaning it is as safe and effective. In this context, the "acceptance criteria" and "device performance" usually refer to the equivalence demonstrated through engineering and non-clinical testing against recognized standards and comparison to the predicate device. Clinical studies for proving effectiveness in terms of pain relief are typically not required for 510(k) submissions of TENS devices if substantial equivalence can be shown through non-clinical means.

Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

For a 510(k) submission of a TENS device, the acceptance criteria are generally defined by compliance with recognized consensus standards for electrical safety, electromagnetic compatibility, and performance, as well as demonstrating comparable technical characteristics to the predicate device. The "device performance" reported is its compliance with these standards and its specified output characteristics.

Acceptance Criteria (Set by Recognized Standards & Predicate Comparison)Reported Device Performance (Tyece OTC TENS Device, Model SEM44)
Electrical Safety Standards:Complies with IEC 60601-1:2005+Corr.1(2006)+Corr.2(2007)+AM1(2012), EN 60601-1:2006/A1:2013
Basic Safety & Essential Performance of Nerve & Muscle Stimulators:Complies with IEC 60601-2-10 Edition 2.0 2012-08
Electromagnetic Disturbances Requirements & Tests (EMC):Complies with IEC 60601-1-2 Edition 2007-03
Home Healthcare Environment Requirements:Complies with IEC 60601-1-11 First Edition: 2010
Biocompatibility of Patient Contacting Parts:Electrode pads FDA cleared under K132588; comply with ISO 10993 (Class A ≤ 24 hrs contact)
Software Performance:Software performs within specifications, is safe for intended use, permanent hazard analysis implemented, clear software architecture. Level of concern: Moderate.
Cleaning Effectiveness:Cleaning instructions in Instruction Manual tested and found sufficient.
Maximum Output Voltage (Vp-p):70Vp-p@500 Ω, 110Vp-p@2K Ω, 190Vp-p@10K Ω (Predicate: 69.2Vp-p@500 Ω, 86.4Vp-p@2K Ω, 88.8Vp-p@10K Ω)
Maximum Output Current (mA):86mA@500 Ω, 23.3mA@2K Ω, 3.75mA@10K Ω (Predicate: 69.2mA@500 Ω, 21.6mA@2K Ω, 4.4mA@10K Ω)
Pulse Duration (μS):50-360μS (Predicate: 200μS)
Frequency (Hz):1-150Hz (Predicate: 1-136Hz)
Net Charge (micro coulombs μC) per pulse:0.001μC@500 Ω (Predicate: 0.0013μC@500 Ω)
Maximum Phase Charge (mC):0.0454mC@500 Ω (Predicate: 0.0135mC@500 Ω)
Maximum Current Density (mA/cm2, r.m.s):0.790mA/cm2@500 Ω (Predicate: 3.270mA/cm2@500 Ω)
Maximum Average Current (average absolute value), mA:16.0mA (Predicate: 2.218mA)
Maximum Average Power Density (W/cm2):0.00632W/cm2@500 Ω 1.0S (Predicate: 0.00116W/cm2@500 Ω 1.0S)
Number of Output Modes:15 Modes (01-15) (Predicate: 7 modes)
Intensity Levels:0 to 50 (Predicate: 0 to 16, but stated maximum intensity settings of both units produce similar outputs)
Timer Range:5-100min (Predicate: 20-40 min)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document states: "Savia did not conduct, nor rely upon, clinical tests to determine substantial equivalence of the SEM44 vs. the predicate." Therefore, there is no test set of clinical data from human subjects in the traditional sense. The testing performed was non-clinical bench testing to validate design and ensure conformance with voluntary design standards. This means there is no data provenance related to human subjects (country of origin, retrospective/prospective).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. As noted above, no clinical test set was used for determining substantial equivalence, hence no experts were involved in establishing ground truth for such a set. The "ground truth" was established by the requirements of the recognized consensus standards and direct comparison of technical specifications to the predicate device.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as no clinical test set involving human subjects requiring adjudication was used.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a TENS unit, not an AI-assisted diagnostic or therapeutic device that would typically involve human readers or MRMC studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. The device is a Transcutaneous Electrical Nerve Stimulator; it's a physical device that delivers electrical impulses, not an algorithm in the context of AI or imaging. Performance was evaluated as a standalone product against technical specifications and standards.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this 510(k) submission is primarily established by:

  • Recognized consensus standards: Electrical safety, EMC, performance standards (e.g., IEC 60601 series).
  • Technical specifications of the predicate device: The Tyece device's technical characteristics are compared directly to the cleared predicate device to ensure substantial equivalence.
  • Design requirements: Internal company design requirements for the device.

There is no "expert consensus," "pathology," or "outcomes data" ground truth in the clinical sense, as clinical studies were not performed for this submission.

8. The sample size for the training set

Not applicable. Since no clinical studies were performed and the device is not an AI/machine learning model, there is no concept of a "training set" of data.

9. How the ground truth for the training set was established

Not applicable, as no training set was used.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, with three curved lines above them, possibly representing the flow of information or services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 5, 2015

Tyece Ltd % Guenter Ginsberg Consultant Media Trade Corporation 11820 Red Hibiscus Drive Bonita Springs, Florida 34135

Re: K150386

Trade/Device Name: Tyece OTC TENS Device, Model SEM44 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: NUH Dated: April 4, 2015 Received: September 30, 2015

Dear Mr. Ginsberg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael J.Hoffmann -A

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K150386

Device Name Tyece OTC TENS Device, Model SEM44

Indications for Use (Describe)

The Tyece OTC TENS Device, Model SEM44 is to be used for temporary relief of pain associated with sore and aching muscles in the lower back, arms, or legs due to strain from exercise or normal household activities.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY (revised 6.30.2015)

Date of Summary Prepared: 02/05/2015

1. Submitter's Name: Tyece Ltd.

Unit 801, Block A, Po Lung Centre, 11 Wang Chiu Rd., Address: Kowloon Bay, Kowloon, Hong Kong, SAR

Contact person: Parshid Falahati

Tel: 852- 2 - 3497456

Fax: 852 -2 - 3499166

E-mail sales@tyece.com

2. Proposed New Device:

Trade Name: Tyece OTC TENS Device, Model SEM44 Classification Name: TENS Regulation Number: 882.5890 Product Code: NUH Device Class: II

3. Predicate (cleared) Device:

Counter TENS Device, Model PTS-IV

510(k) Number: K132993

Manufacturer: Counter Scientific Development Ltd.

{4}------------------------------------------------

4. Description of Proposed Device:

The Tyece OTC TENS Device, Model SEM44 is a dual channel TENS device operated by DC 4.5V (3 AAA batteries). It is made up of one main unit, two electrode cables, and two pairs of electrode pads. There are 12 selectable, pre-programmed output waveforms (modes) to choose from, blus 3 programmable modes. The two channels can be individually adjusted for intensity from 0 to 50. Running time can be selected from 5 to 100 minutes.

All parameters (Menu, Program Number, Impulse intensity for channels, Jow battery, key-lock, frequency, pulse width and timer) are displayed on a full screen display.

The only patient contacting components are the Electrode Pads. The patient contacting materials are FDA cleared under K132588.

The device has been tested to and meets the requirements of the following recognized consensus standards:

IEC 60601-1:2005+Corr.1(2006)+Corr.2(2007) + AM1(2012). EN 60601-1:2006/A1:2013 general requirements for basic safety and essential performance

IEC 60601-2-10 Edition 2.0 2012-08 basic safety and essential performance of nerve and muscle stimulators. (Neurology)

IEC 60601-1-2 Edition 2007-03 medical electrical equipment - part 1-2: general requirements for basic safety and essential performance - collateral standard: electromagnetic disturbances requirements and tests. (General II (ES/EMC)

IEC 60601-1-11 First Edition: 2010 medical electrical equipment - part 1-11: general requirement for basic safety and essential performance -- collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

5. Indication for Use:

This Tyece OTC TENS Device. Model SEM44 is to be used for temporary relief of pain associated with sore and aching muscles in the lower back, arms, or legs due to strain from exercise or normal household work activities.

6. Environment of Use:

Clinics. Hospitals and home environments

7. Contraindications:

Do not use this device if you have a cardiac pacemaker, implanted defibrillator, or other implanted metallic or electronic device. Such use could cause electrical shock, burns, electrical interference, or death.

{5}------------------------------------------------

Do not use this device together with a life-supporting medical electronic device such as an artificial heart or lung.

Do not use this device together with a body-worn medical electronic device such as an ECG.

8. Technological Characteristics Compared to the Predicate Device

The Tyece OTC TENS Device, Model SEM44 is substantially equivalent to the Counter healthcare product OTC TENS for muscle pain relief, Model PTS-IV, The function, performance, and intended use of the two devices are very similar. Both systems are designed to relief pain in the muscle areas. Outputs are almost the identical, each can produce square electronic pulses and positive-going, reverse and bidirectional waveform output. Each can produce an asymmetrical biphasic pulse with peak voltages up to 60/70 volts on 500 Ohms Joad. Both devices have selection modes to allow users to select programs and adjust intensity level by pressing the keypad on the main unit. Both devices are transiting pulses by the use of Electrode Pads. Both are portable devices. Technological characteristics, features, specifications, materials and intended uses of the new device are substantially equivalent to the quoted predicate device. The differences that exist between the devices are insignificant in terms of safety or effectiveness.

Differences:

  1. Device SEM44 has 15 preset programs, while the predicate device has 7.

  2. Intensity levels of SEM44 are selectable from 0 to 50, while the predicate device is from 0 to 16, however, the maximum intensity settings of both units produce very similar outputs to the electrode pads. The SEM44 model can be adjusted in finer increments.

  3. All other parameters such as waveform, frequency, pulse width, output current and voltage, although not exactly similar to this particular predicate, they do not raise any new questions of safety or effectiveness.

Note: It is very difficult, if not impossible, to find predicate devices that completely match the specifications of the subject device.

9. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence is as follows:

Savia did not conduct, nor rely upon, clinical tests to determine substantial equivalence of the SEM44 vs. the predicate. Non-clinical testing was performed in order to validate the design according to the company's specified design requirements, and to assure conformance with the following voluntary design standards:

IEC 60601-1:2005+Corr.1(2006)+Corr.2(2007) + AM1 (2012), EN 60601-1:2006/A1:2013 general requirements for basic safety and essential performance

IEC 60601-2-10 Edition 2.0 2012-08 basic safety and essential performance of nerve and muscle stimulators. (Neurology) .. 1991 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 199

{6}------------------------------------------------

IEC 60601-1-2 Edition 2007-03 medical electrical equipment - part 1-2: general requirements for basic safety and essential performance - collateral standard: electromagnetic disturbances requirements and tests. (General II (ES/EMC)

IEC 60601-1-11 First Edition: 2010 medical electrical equipment - part 1-11: general requirement for basic safety and essential performance - collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

a. EMC and electrical safety

IEC 60601-1:2005+Corr.1(2006)+Corr.2(2007) + AM1 (2012), EN 60601-1:2006/A1:2013 general requirements for basic safety and essential performance

IEC 60601-2-10 Edition 2.0 2012-08 basic safety and essential performance of nerve and muscle stimulators. (Neurology)

IEC 60601-1-2 Edition 2007-03 medical electrical equipment - part 1-2: general requirements for basic safety and essential performance - collateral standard: electromagnetic disturbances requirements and tests. (General II (ES/EMC)

IEC 60601-1-11 First Edition: 2010 medical electrical equipment -- part 1-11: general requirement for basic safety and essential performance - collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

b. Biocompatibility

The electrode pads are the only body contacting parts. They have been FDA cleared for OTC under K132588. The patient contacting electrodes comply with ISO 10993 and are considered Class A ≤ 24 hrs contact.

c. Software

Based upon the test results it was concluded that the software performs within specifications and is safe to the stated intended use. The software of this device carries a MODERATE level of concern. Since a permanent hazard analysis is implemented in the software development process, and due to the clear software architecture, it is believed that the test protocol sufficiently verifies the software's main functional operation.

{7}------------------------------------------------

d. Cleaning

The cleaning instructions as described the Instruction Manual have been tested to be sufficient. Testing involved validation of the manual cleaning method as per the instructions. All testing concluded that that the Tyece OTC TENS Device, Model SEM44 can be cleaned by the use of the methods described in the Instruction Manual.

10. Conclusions:

The Tyece OTC TENS Device, Model SEM44 has the same intended use and similar technological characteristics as the predicate device. Moreover, bench testing and safety report documentation demonstrate that the submitted device could maintain the same safety and effectiveness as that of predicate device. In the other words, the differences do not affect the intended use and do not raise any new questions of safety or effectiveness or alter the fundamental scientific technology of the device. Thus, the Tyece OTC TENS Device, Model SEM44 is substantially equivalent to the predicate device.

a managar and consistence and consistence and consistence and consistences and consistences and consistences and consistences and consistences and consistences and consistenc

... .. . . . . . . . . . . . . . . . . . . . . . . .

{8}------------------------------------------------

Output Specifications:

ParameterYour devicePredicate Device
Mode or Program NamePAIN THERAPY DEVICE,SEM44PAIN THERAPY DEVICE,PTS-IV
Waveform (e.g. positive-going,reverse, biphasic)BiphasicPositive-going, Reverse andBiphasic
Shape(e.g., spike, rectangular,square ware)Square wareSquare ware
Maximum Output Voltage (Volts)(+/-15%)70vp p@500 Ω110vp p@2K Ω190vp p@10K Ω69.2vp p@500 Ω86.4vp p@2K Ω88.8vp p@10K Ω
Maximum Output Current (+/-15%)86mA@500 Ω23.3mA@2K Ω3.75mA@10K Ω69.2mA@500 Ω21.6mA@2K Ω4.4mA@10K Ω
Duration of primary Phase00
Pulse Duration50-360μS200μS
Frequency1-150Hz1-136Hz
For multi SymmetricalProgram Programswaveforms Pulse durationonly:N/AN/A
Net Charge (micro coulombs μC)(per pulse) (if zero, State methodof achieving zero net charge)0.001μC@500 Ω0.0013μC@500 Ω
Maximum Phase Charge (mC)0.0454mC@500 Ω0.0135mC@500 Ω
Maximum Current Density(mA/cm2, r.m.s)0.790mA/cm2@500 Ω3.270mA/cm2@500 Ω
Maximum Average Current(average absolute value), mA16.0mA2.218mA
Maximum Average Power Density,(W/cm2)(using smallest electrodeconductive surface area)0.00632W/cm2@500 Ω 1.0S0.00116W/cm2@500 Ω 1.0S
Burst Mode(i.e.pulse trains)(a) Pulses per burst 4(b) Bursts per second 4/83Hz(c) Burst duration(seconds) 180μS(d) Duty Cycle 35ms/60ms4/230/136Hz800μS/400μS24ms/54.4ms
ON Time (seconds)2s0.5s
OFF Time (seconds)2s0.5s
Additional Features (specify, ifapplicable)N/AN/A
Electrode area:20.25sqcm x 4 (81sqcm)21.16sqcm x 4(84.64sqcm)
Basic Unit Characteristics:
ParameterYour DevicePredicate Device
510(k) NumberK150386K132993
Device Name and See attached photo ProgramPAIN THERAPY DEVICESEM44PAIN THERAPY DEVICEPTS-IV
ManufacturerSAVIA LTD.COUNTER Scientific Development (GZ) Limited
Power Source4.5V(batteries, 3x1.5V AAA)3.7v Lithium ( Power supply)
Method of Line Current IsolationN/AN/A
Patient Leakage CurrentN/AN/A
Normal Condition (µA)N/AN/A
Single Fault Condition (µA)N/AN/A
Average DC current through electrodes when device is on but no pulse are being applied (µΑ)0µA0µA
Numbers of Output Modes15 Modes(01-15)7 modes(A,B,C,D,E,F,ALL)
Number Synchronous or Alternating?2 Synchronous2 Synchronous
Channels Method of Channel IsolationPCB InsulationBoost IsolationBoost Isolation
Regulated current or regulated voltage?Regulated voltageRegulated voltage
Software/firmware/microprocessor control?YesYes
Automatic overload trip?NoNo
Automatic no-load trip?YesYes
Automatic shut off?YesYes
Users override control?YesYes
Indicator On/off status?YesYes
display Low battery?YesYes
Voltage/current level? Only voltage level displayYes for voltageYes for voltage
Timer range (minutes)5-100min20-40 min
Compliance with voluntary standards?YesYes
Compliance with 21CFR898-7?YesYes
DimensionsL135xW65xT20mmL120xW70xT18mm
Housing materials and constructionABSABS

:

6

: 多

.

:

{9}------------------------------------------------

Basic Unit Characteristics:

.

.

7

1 ---------------

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).