K Number
K143477
Date Cleared
2016-04-14

(496 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nitrile Patient Examination gloves. Powderfree. Blue color, is a disposable device intended for medical purposes that is worn on the examiners' hand or finger to prevent contamination between patient and examiner

Device Description

The subject device is a patient examination glove made of synthetic nitrile latex compound. It is non-sterile, powderfree and is Blue in color. The device is ambidextrous and can be worn on either the left or right hand. The device meets ASTM 06319-10: Standard specification for Nitrile Examination Gloves for Medical Application. The subject device is substantially equivalent to legally marketed Nitrile examination gloves identified as Product code LZA.

The device is for over-the counter single use.

AI/ML Overview

The provided document is a 510(k) premarket notification for Nitrile Patient Examination gloves. It does not describe an AI/ML medical device. Therefore, it is not possible to extract information related to AI/ML device acceptance criteria or studies proving device performance, as the requested fields (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, ground truth establishment, training set size) are not applicable to this type of medical device submission.

The document focuses on demonstrating substantial equivalence to a predicate device (K121947) through performance data based on established ASTM standards and biocompatibility tests for examination gloves.

Here's the relevant information that directly addresses the "acceptance criteria and the study that proves the device meets the acceptance criteria" for this non-AI/ML medical device:

1. A table of acceptance criteria and the reported device performance:

The document provides a detailed table comparing the subject device (Nitrile Patient Examination gloves, Powderfree, Blue color) to a legally marketed predicate device (ETS Blue Powderfree Nitrile patient exam glove, K121947) across various characteristics. The "Acceptance criteria/Standard" column outlines the benchmarks, typically ASTM standards or FDA guidance. The "SUBJECT DEVICE : K143477" column indicates the subject device's performance against these criteria, consistently stating "PASS" or meeting the specified physical properties.

CharacteristicsAcceptance Criteria/StandardReported Device Performance (Nitrile Patient Examination gloves, Powderfree, Blue color)
Freedom from HolesASTM D6319-10 / ASTM D5151-06 (FDA 1000 ml water leak test, ISO 2859-1 / G11 AQL 2.5)Meets / PASS
DimensionsASTM D6319-10 (ISO 2859-1 / S2 AQL 4.0) - Length: min 230 mm (> 240-400mm met); Width: 70±10 mm to 120±10 mm (sizes XS to XL met); Thickness: > 0.05 mm (palm & finger met)PASS
Physical PropertiesBefore aging: ASTM D6319-10 and ASTM D412-06 (ISO 2859-1 / S2 AQL 4.0) - Tensile strength: > 14 Mpa; Ultimate Elongation: > 500%After accelerated aging: ASTM D6319-10 and ASTM D412-06 (ISO 2859-1 / S2 AQL 4.0) - Tensile strength: > 14 Mpa; Ultimate Elongation: > 400%PASS (Tensile strength: > 14 Mpa, Ultimate Elongation: > 500% before aging; Tensile strength: > 14 Mpa, Ultimate Elongation: > 400% after accelerated aging)
Powder-free residueASTM D6319-10 / ASTM D6124-06 (N=5) - Less than 2 mg per gloveMeets / PASS (Less than 2 mg per glove)
BiocompatibilityISO 10993-10 - Primary skin irritation (non-irritant); Skin/Dermal Sensitization (non-sensitizer)Non-irritant; Non-sensitizer
Expiration dating/Shelf lifeASTM D7160-05Three years
Intended UseMedical Glove Guidance Manual (Intended for medical purpose to prevent contamination between patient and examiner)Meets
Indication for UseMedical Glove Guidance Manual (Disposable device for medical purposes, worn on hand/finger to prevent contamination)Meets
MaterialASTM D6319-10, LZA product code (Nitrile synthetic latex)Meets (Nitrile synthetic latex)
Non-sterile or sterileSterility (Non Sterile)Non Sterile
Single UseDisposable/Single useYes
AmbidextrousASTM D6319-0Yes

2. Sample sized used for the test set and the data provenance:

The document refers to sampling plans and inspection levels for certain tests:

  • Freedom from Pin holes: ISO 2859-1 / G11 AQL 2.5
  • Dimensions: ISO 2859-1 / S2 AQL 4.0
  • Physical Properties (Before and After aging): ISO 2859-1 / S2 AQL 4.0
  • Powder-free residue: N=5 gloves

The exact numerical sample sizes for each test are not explicitly stated as absolute numbers but rather by the sampling plans (e.g., ISO 2859-1). The data provenance is not explicitly mentioned as "country of origin" for the test data, nor is it categorized as "retrospective or prospective." However, given the nature of the device (examination gloves) and the standards used, it's implied that these are laboratory tests performed on samples of the manufactured gloves.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

Not applicable. This is a non-AI/ML medical device submitting for 510(k) clearance. The "ground truth" for the physical and chemical properties of the gloves is established through adherence to defined ASTM and ISO standards and laboratory testing protocols, not by expert consensus or interpretations in a diagnostic sense.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Not applicable, as this is not an AI/ML device involving human interpretation that would require an adjudication method.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This is a non-AI/ML medical device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This is a non-AI/ML medical device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

The "ground truth" for this device's performance is based on established industry standards and validated laboratory test methods. This includes:

  • ASTM International Standards: Such as ASTM D6319-10 (Standard Specification for Nitrile Examination Gloves for Medical Application), ASTM D5151-06 (Standard Test Method for Detection of Holes in Medical Gloves), ASTM D412-06 (Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers—Tension), ASTM D6124-06 (Standard Test Method for Residual Powder on Medical Gloves), and ASTM D7160-05 (Standard Practice for Determination of Expiration Dating for Medical Gloves).
  • ISO Standards: Specifically ISO 10993-10 (Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization) for biocompatibility.
  • FDA water leak test (1000ml).

8. The sample size for the training set:

Not applicable. This is a non-AI/ML medical device; therefore, there is no "training set" in the context of machine learning.

9. How the ground truth for the training set was established:
Not applicable. There is no training set for this non-AI/ML medical device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing right, with a wing-like shape above them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 14, 2016

Primus Gloves PVT Limited Jose Paul M Manager-QA & RA Plot No: 14-A, Cochin Special Economic Zone Kakkanad, Cochin, Kerala 682037 INDIA

Re: K143477

Trade/Device Name: Nitrile Patient Examination gloves, Powderfree, Blue Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: undated Received: April 12, 2016

Dear Mr. Jose Paul M:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K143477

Device Name

Nitrile Patient Examination gloves, Powderfree, Blue color

Indications for Use (Describe)

The Nitrile Patient Examination gloves. Powderfree. Blue color, is a disposable device intended for medical purposes that is worn on the examiners' hand or finger to prevent contamination between patient and examiner

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1

PSC Publishing Services (301) 443-6740 RE

{3}------------------------------------------------

ATTACHMENT V

K143477

1.0SUBMITTER
1.1Company NamePRIMUS GLOVES PRIVATE LIMITED
1.2AddressPlot No: 14-A, Cochin Special Economic Zone,
Kakkanad, Cochin, Kerala, India - 682037
1.3Telephone+ 91 484 2413063
1.4Fax+91 484 2413089
1.5Emailjosegaul@grimusgloves.com
1.6Contact PersonMr. JOSE PAUL M
MANAGER-QA & RA
2.0OFFICIAL CORRESPONDENT
2.1Company NamePRIMUS GLOVES PRIVATE LIMITED
2.2AddressPlot No: 14-A, Cochin Special Economic Zone,
Kakkanad, Cochin, Kerala, India - 682037
2.3Telephone+ 91 484 2413063
2.4Fax+91 484 2413089
2.5Emailjosegaul@grimusgloves.com
2.6Contact PersonMr. JOSE PAUL M
MANAGER-QA & RA
3.0Preparation date14th April, 2016
4.0Name of the device
4.1Device NameNITRILE PATIENT EXAMINATION
GLOVES, POWDERFREE, BLUE COLOR
4.2Trade NamePRIMUS NITRILE EXAMINATION GLOVES
4.3Common NamePATIENT EXAMINATION GLOVES
4.4Classification21 CFR 880.6250 PATIENT EXAMINATION GLOVES
4.5ClassCLASS I

{4}------------------------------------------------

  • 4.6 Product code LZA

5.0 ldentification of the leqally marketed predicate device

5.1Device NameETS Blue Powderfree Nitrile patient examglove
5.2510(k) NumberK121947
5.3CompanyNorthstar Healthcare Holdings, 70 Sir JohnRogerson's Quay, Dublin 2, Ireland.
5.4Device DescriptionNon sterile Nitrile powderfree exam gloves
5.5ClassificationPATIENT EXAMINATION GLOVES
5.6ClassCLASS I
5.7Product codeLZA
5.8Classification Panel :General Hospital

Description of the Device 6.0

The subject device is a patient examination glove made of synthetic nitrile latex compound. It is non-sterile, powderfree and is Blue in color. The device is ambidextrous and can be worn on either the left or right hand. The device meets ASTM 06319-10: Standard specification for Nitrile Examination Gloves for Medical Application. The subject device is substantially equivalent to legally marketed Nitrile examination gloves identified as Product code LZA.

The device is for over-the counter single use.

7.0 Indications for use

The Nitrile Patient Examination gloves, Powderfree, Blue color, is a disposable device intended for medical purposes that is worn on the examiners' hand or finger to prevent contamination between patient and examiner.

{5}------------------------------------------------

8.0 Summary of performance data

There is no difference in technological characteristics compared to the predicate device. Gloves are made from Nitrile latex compound, Non sterile, Powderfree and Blue in color. The gloves have the same technological characteristics compared to ASTM or equivalent standards as given below,

CharacteristicsStandardsPerformance of Nitrilepatient examinationgloves, Powderfree, Bluecolor
Freedom from HolesASTM 06319-10 / ASTM05151-06Meets
DimensionsASTM D6319-10Meets
Physical PropertiesASTM D6319-10 / ASTM0412-06Meets
Powder-free residueASTM 06319-10Meets
Bio-compatibilityPrimary skin irritationISO 10993-10Non-irritant
Skin/Dermal SensitizationISO 10993-10Non-sensitizer
Expiration dating/Shelf lifeASTM D7160-05Three years

{6}------------------------------------------------

Performance data of gloves based on ASTM 06319-10 and FDA 1000ml water leak test

ASTM 06319-10 and FDA 1000 ml water leak test
CharacteristicsTestTest standardSampling plan/Inspection level/AQLNon sterile, powderfree, Non sterile Nitrile examination gloves-PRIMUSRESULT
Freedom from Pin holesFDA 1000 ml water leak testASTM D5151-06 (reap 2011)ISO 2859-1 / G11AQL 2.5PASSPASS
DimensionsLengthASTM D6319-10ISO 2859-1 / S21AQL 4.0> 230 mm ( 240-400mm)PASS
WidthASTM D6319-10ISO 2859-1 / S21AQL 4.070±10 mm to 120±10 mm (sizes XS to XL)PASS
ThicknessASTM D6319-10ISO 2859-1 / S21AQL 4.0> 0.05 mm (palm & finger)PASS
Physical propertiesBefore agingASTM D6319-10 and ASTM D412-06ISO 2859-1 / S21AQL 4.0Tensile strength : > 14 MpaUltimate Elongation : > 500%PASS

{7}------------------------------------------------

strength :> 14 Mpa
AfterAcceleratedagingASTM06319-10and ASTM0412-06ISO 2859-1 /S2/AQL 4.0UltimateElongation :> 400 %PASS
Powder-freeresiduePowder-freeresidueASTMD6319-10and ASTM06124-06N=5Less than 2mg per glovePASS
BiocompatibilityPrimary skinirritationISO10993-10Under the conditions of thestudy the device is not anirritantPASS
Skin/DermalSensitizationISO10993-10Under the conditions of thestudy the device is not asensitizerPASS

9.0 Summary of the technological characteristics of device compared to the legally marketed predicate device

CharacteristicsPREDICATE-510(K) :K121947SUBJECT DEVICE :K143477Acceptancecriteria/Standard
ManufacturerNorthstarHealthcareHoldings, 70 SirJohnRogerson'sQuay, Dublin 2,Ireland.PRIMUS GLOVESPRIVATE LIMITED, PlotNo: 14-A, CSEZ,Kakkanad, Cochin,Kerala, India -682037
Product NameETS BluePowder FreeNitrile PatientExamination Gloves,Patientexamination

{8}------------------------------------------------

Nitrile PatientExam GlovesPowderfree Blue colorglovesgloves
Intended UseIntended formedical purposethat is worn onthe Examinershand to preventcontaminationbetween patientand examinerIntended for medicalpurpose that is worn onthe Examiners hand toprevent contaminationbetween patient andexaminerMedical GloveGuidance Manual
Indication for useThe examinationgloves is adisposabledevice intendedfor medicalpurposes that isworn on theexaminers handor finger topreventcontaminationbetween patientand examinerThe examination glovesis a disposable deviceintended for medicalpurposes that is worn onthe examiners hand orfinger to preventcontamination betweenpatient and examinerMedical GloveGuidance Manual
DescriptionNon sterilepowderfree ,examinationgloves made ofnitrile andcolored blue Thetextured glovesNon sterile powderfree ,examination glovesmade of nitrile andcolored blue Thetextured gloves areprovided in sizes ExtraSmall, Small, Medium,Medical GloveGuidance Manual
are provided insizes ExtraSmall, Small,Medium, ExtraLarge. Thesmooth glovesare provided inSizes Small,Medium, Largeand Extra LargeExtra Large. Thesmooth gloves areprovided in Sizes Small,Medium, Large andExtra Large
PresentationNon sterilegloves areprovided indispenser boxesNon sterile gloves areprovided in dispenserboxesMedical GloveGuidance Manual
MaterialNitrile syntheticlatexNitrile synthetic latexASTM 06319-10LZA product code
Non-sterile orsterileNon SterileNon SterileSterility
Single UseYesYesDisposable/Singleuse
AmbidextrousYesYesASTM D 6319-0
DimensionsMeets ASTM D6319-10Overall length min 240mm ,width varies from70 mm for XS size to120 mm for XL size,thickness in finger andpalm has a minimum0.05 mmMeets ASTM D6319-10-Overall lengthmin 240 mm,width varies from70 mm for XSsize to 120 mmfor XL size,
and palm has a
minimum 0.05
mm
Tensile strength 14 MpaMeets ASTM D
min for before aging and6319-10- Tensile
14 Mpa min for afterstrength 14 Mpa
agingmin for before
Aging done at 70 ±2 degaging and 14 Mpa
C for 166±2 hrs ormin for after
Tensile StrengthMeets ASTM D100±2deg C for 22±0.3aging
6319-10hrsUltimate
elongation 500 %
min for before
aging and 400 %
min for after
aging. Aging
done at 70 ±2
deg C for 166±2
hrs or 100±2deg
C for 22±0.3 hrs
Ultimate elongation 500Meets ASTM D
% min for before aging6319-10-
UltimateMeets ASTM Dand 400 % min for afterUltimate
Elongation6319-10aging. Aging done at 70elongation 500 %
±2 deg C for 166±2 hrsmin for before
or 100±2deg C foraging and 400%
22±0.3 hrsmin for after
aging. Aging
done at 70 ±2
deg C for 166±2hrs or 100±2degC for 22±0.3 hrs
Freedom frompinholesMeets ASTM D5151 -06 andASTM 06319-10Meets ASTM D 5151 -06 (2011) and ASTM06319-10ASTM D 5151 -06(2011) and ASTM06319-10
Residual PowderMeets ASTM D6124-06Less than 2 mg pergloveASTM O 6124-06(2011) : Lessthan 2 mg perglove
BiocompatibilityTestsISO 10993-10Non-irritant -Primary SkinIrritation InRabbitsNon-sensitizer -skin Sensitizationin Guinea pigsUnder the conditions ofthe study the device isnot an irritantUnder the conditions ofthe study the device isnot a sensitizerUnder theconditions of thestudy the deviceis not an irritantUnder theconditions of thestudy the deviceis not a sensitizer
LabelingPowderfree,Nitrile patientexam gloveNon sterileSingle use onlyAmbidextrousBlue colorManufactured forLot NoIntended useQuantityCountry of origin*Powderfree,*Nitrile Patient examglove*Non sterile*Single use only*Blue color*Ambidextrous*Manufactured for*Lot No*Intended use*Quantity*Country of originChapter 4-Labeling-Medical GloveGuidance Manual

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

10.0 CONCLUSION

The physical performance of the subject device is substantially equivalent to predicate K121947 and will perform according to the glove performance and biocompatibility standards referenced. Based on the intended use, physical properties and technological characteristics, the subject device is as safe, effective and performs as well as the legally marketed predicate device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.