(177 days)
Not Found
No
The provided text describes a standard pulse oximeter and its performance characteristics. There is no mention of AI, ML, or any related technologies in the device description, intended use, or performance studies.
No.
The device is for monitoring physiological parameters (SpO2 and pulse rate) and does not provide therapy or treatment.
Yes
Explanation: The device is described as a "continuous non-invasive monitoring" system for SpO2 and pulse rate, and it provides "digital values of SpO2 and pulse rate." This falls under the definition of a diagnostic device as it measures and displays physiological parameters to assess a patient's health status.
No
The device description explicitly states it is a "pulse oximeter" and mentions being powered by an internal power supply (AC or battery), indicating it is a hardware device with integrated software, not a software-only medical device.
No, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are medical devices used to examine specimens taken from the human body, such as blood, urine, or tissue, to provide information for diagnosis, monitoring, or screening. This testing is performed outside of the living body (in vitro).
- Device Function: The Nellcor Bedside SpO2 Patient Monitoring System is a pulse oximeter. It works by shining light through tissue (typically a finger or toe) and measuring the absorption of different wavelengths of light to determine the oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. This is a non-invasive measurement performed on the living body.
- Intended Use: The intended use clearly states "continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate." This is a physiological measurement taken directly from the patient, not a test performed on a bodily specimen.
Therefore, based on its function and intended use, the Nellcor Bedside SpO2 Patient Monitoring System is a non-invasive physiological monitoring device, not an In Vitro Diagnostic.
N/A
Intended Use / Indications for Use
The Nellcor Bedside SpO2 Patient Monitoring System is indicated for prescription use only for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The Nellcor Bedside SpO2 Patient Monitoring System is intended for use with neonatal, pediatric, and adult patients during both no motion and motion conditions and for patients who are either well or poorly perfused, in hospitals, hospital-type facilities, intra hospital transport, or home environments.
Product codes
DQA
Device Description
The Nellcor Bedside SpO2 Patient Monitoring System is a pulse oximeter designed for continuous, non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The Nellcor Bedside SpO2 Patient Monitoring System is intended to be used with one of a range of Nellcor OxiMax oxygen transducers (sensors). The subject device displays digital values of SpO2 and pulse rate. Pulse amplitude is displayed by means of a 'blip bar' presentation or plethysmographic waveform. The Nellcor Bedside SpO2 Patient Monitoring System can be powered by an internal power supply operating on AC from a standard electrical utility receptacle (from 80V AC to 264V AC) or alternately by an integral sealed 7.2V, 83W/hr rechargeable lithium-ion battery.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
neonatal, pediatric, and adult patients
Intended User / Care Setting
hospitals, hospital-type facilities, intra hospital transport, or home environments.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Bench testing was conducted on the Nellcor Bedside SpO2 Patient Monitoring System to ensure the subject oximeter met design specifications. The device was tested to medical electrical equipment standards, environmental testing standards, medical device software standards, and usability engineering standards. These standards include: ISO 9919, ISO 80601-2-61, IEC 60601-1: 3rd Edition, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11, IEC 60068-2-1, IEC 60068-2-2, IEC 60068-2-6, IEC 60068-2-13, IEC 60068-2-27, IEC 60068-2-30, IEC 60068-2-64, IEC 62304, IEC 62366.
The results of this testing show that the subject device meets the product requirements and specifications including: SpO2 accuracy without motion of ±2% from 70-100%, SpO2 accuracy with motion of ±3 from 70-100%, and pulse rate of ±3% from 20 to 250 BPM, and is substantially equivalent to the identified predicate devices.
There was no new clinical testing conducted on the subject Nellcor Bedside SpO2 Patient Monitoring System. Clinical testing was conducted on a similar Nellcor pulse oximetry device utilizing the same PCBA module to demonstrate the accuracy of the oximetry performance during both motion and no motion conditions.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
SpO2 accuracy without motion of ±2% from 70-100%, SpO2 accuracy with motion of ±3 from 70-100%, and pulse rate of ±3% from 20 to 250 BPM.
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.2700 Oximeter.
(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 26, 2015
Covidien, LLC Megan Fessenden Senior Regulatory Affairs Specialist 6135 Gunbarrel Avenue Boulder, CO 80301
Re: K142865
Trade/Device Name: Nellcor Bedside SpO2 Patient Monitoring System Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: February 24, 2015 Received: February 27, 2015
Dear Ms. Fessenden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
1
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
K142865 510(k) Number, if known: _
De vice Name : Nellcor Bedside SpO2 Patient Monitoring System
Indications for Use:
The Nellcor Bedside SpO2 Patient Monitoring System is indicated for prescription use only for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The Nellcor Bedside SpO2 Patient Monitoring System is intended for use with neonatal, pediatric, and adult patients during both no motion and motion conditions and for patients who are either well or poorly perfused, in hospitals, hospital-type facilities, intra hospital transport, or home environments.
Note:
- Hospital use typically covers such areas as general care floors (GCFs), operating rooms, ● special procedure areas, intensive and critical care areas within the hospital, and in hospital-type facilities.
- . Hospital-type facilities include physician office-based facilities, sleep labs, skilled nursing facilities, surgicenters, and sub-acute centers.
- . Intra-hospital transport includes transport of a patient within the hospital or hospital-type facility.
- . Home environment includes any environment other than a professional healthcare facility or clinical laboratory where a device may be used.
Prescription Use 区 (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use □ (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evalutaion (ODE)
3
Submitter | |
---|---|
Address | Covidien LLC |
6135 Gunbarrel Ave | |
Boulder, CO 80301 | |
Phone | 303-305-2291 |
Fax | 303-305-2212 |
Company Contact | Megan Fessenden, Sr. Regulatory Affairs Specialist |
Date Prepared | February 23, 2015 |
Device | |
Trade Name | Nellcor Bedside SpO2 Patient Monitoring System |
Common Name | Pulse Oximeter |
Classification Name | Oximeter (21 CFR 870.2700) |
Classification | Class II |
Product Code | DQA |
Branch | Anesthesiology |
Device Description
The Nellcor Bedside SpO2 Patient Monitoring System is a pulse oximeter designed for continuous, non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The Nellcor Bedside SpO2 Patient Monitoring System is intended to be used with one of a range of Nellcor OxiMax oxygen transducers (sensors). The subject device displays digital values of SpO2 and pulse rate. Pulse amplitude is displayed by means of a 'blip bar' presentation or plethysmographic waveform. The Nellcor Bedside SpO2 Patient Monitoring System can be powered by an internal power supply operating on AC from a standard electrical utility receptacle (from 80V AC to 264V AC) or alternately by an integral sealed 7.2V, 83W/hr rechargeable lithium-ion battery.
The intended patient population includes neonatal, pediatric, and adult patients during both motion and no motion conditions. The patient may be well or poorly perfused and undergoing monitoring in hospitals, hospital-type facilities, intra-hospital transport, or home environments.
Intended Use
The Nellcor Bedside SpO2 Patient Monitoring System is indicated for prescription use only for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The Nellcor Bedside SpO2 Patient Monitoring System is intended for use with neonatal, pediatric, and adult patients during both no motion conditions and for patients who are either well or poorly perfused and undergoing monitoring in hospitals, hospitaltype facilities, intra hospital transport, or home environments.
4
Note:
- Hospital use typically covers such areas as general care floors (GCFs), operating rooms, special . procedure areas, intensive and critical care areas within the hospital, and in hospital-type facilities.
- . Hospital-type facilities include physician office-based facilities, sleep labs, skilled nursing facilities, surgicenters, and sub-acute centers.
- . Intra-hospital transport includes transport of a patient within the hospital or hospital-type facility.
- Home environment includes any environment other than a professional healthcare facility or clinical laboratory where a device may be used.
Technological Characteristics
The subject and predicate devices measure functional oxygen saturation non-invasively via a light signal interacting with tissue, by utilizing the time-varying changes in tissue optical properties that occur with pulsatile blood flow.
At a high level, the subject and predicate devices are based on the following same technological characteristics:
- . Pulse Oximetry - Used to non-invasively measure functional oxygen saturation
- Regulation Number 870.2700 Oximeter ●
- . Product Code - DQA
- . Collect signal via light signal interacting with tissue - red & infrared light emitting diodes (LEDs) are used as light sources and a photodectector senses the signal strengths of the two wavelengths of light.
- . Use of software including the same oximetry algorithm to processes the information providing real-time values of SpO2, pulse rate, and pulse amplitude.
- . Use of an alarm management system – SatSeconds allows the caregiver to set the limits on desaturation and pulse rate including specifying acceptable duration and magnitude.
- Use of commonly used materials
- . Limited contact with skin, no contact with mucous membranes or non-intact skin designed cleaning and low-level disinfecting for multiple uses
- . Are electrically operated – come equipped with an internal battery supply and an external power cord
- . Operating and Storage Temperature and Humidity
- . Availability of Nurse Call connectivity
- . Compliance to Electrical Safety and EMC standards
- Intended patient population and sensor compatibility
- SpO2 and Pulse rate accuracy specifications
The following technological differences exist between the subject and predicate devices:
NOTE: When the difference does not apply to all predicates, the applicable predicate(s) are noted in parentheses.
- Updated compliance to current standards
- . Homecare use environment (Nellcor Bedside SpO2 Patient Monitoring System (Libra) and Nellcor Bedside Respiratory Patient Monitoring System only)
5
- Physical Dimensions and/or weight (All) ●
- . Operating Altitude (N-600x and Nellcor Bedside Respiratory Patient Monitoring System only)
- Storage Altitude (Nellcor Bedside Respiratory Patient Monitoring System only)
- Display size (N-600x and Nellcor Bedside Respiratory Patient Monitoring System only)
- . Printed Circuit Board Assembly (PCBA) (N-600x and Nellcor Bedside SpO> Patient Monitoring System only)
- . Ability to measure respiratory rate is only available on Nellcor Bedside Respiratory Patient Monitoring System
Substantial Equivalence
The Nellcor Bedside SpO2 Patient Monitoring System is considered to be substantially equivalent to the devices presented in Table 6-2.
Device Manufacturer | Device Name | 510(k) Number | Date of Clearance |
---|---|---|---|
Covidien LLC | Nellcor OxiMAX N-600x | ||
Pulse Oximetry System | K123581 | May 9, 2013 | |
Covidien LLC | Nellcor Bedside SpO2 | ||
Patient Monitoring | |||
System | K120773 | July 12, 2012 | |
Covidien LLC | Nellcor Bedside | ||
Respiratory Patient | |||
Monitoring System | K130320 | February 4, 2014 |
Table 6-2 – Substantial Equivalence
Performance Testing - Bench
Bench testing was conducted on the Nellcor Bedside SpO2 Patient Monitoring System to ensure the subject oximeter met design specifications. The device was tested to medical electrical equipment standards, environmental testing standards, medical device software standards, and usability engineering standards. These standards include:
Standard | Title |
---|---|
ISO 9919 | Medical Electrical Equipment - Particular Requirements for the Basic Safety and Essential Performance of Pulse Oximeter Equipment for Medical Use |
ISO 80601-2-61 | Medical Electrical Equipment - Part 2-61: Particular Requirements for the Basic Safety and Performance of Pulse Oximeter Equipment |
IEC 60601-1: 3rd Edition | Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance |
IEC 60601-1-2 | Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests |
IEC 60601-1-8 | Medical Electrical Equipment - Part 1-8: General Requirements for Basic Safety and Essential Performance - Collateral Standard: General Requirements, Tests and Guidance for Alarm Systems in medical Electrical Equipment and Medical Electrical Systems |
IEC 60601-1-11 | Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems used in the Home Healthcare Environment |
IEC 60068-2-1 | Environmental Testing - Part 2-1: Tests - Test A: Cold |
IEC 60068-2-2 | Environmental Testing - Part 2-2: Tests - Test B: Dry Heat |
6
Nellcor Bedside SpO2 Patient Monitoring System 510(k) Summary
Standard | Title |
---|---|
IEC 60068-2-6 | Environmental Testing - Part 2-6: Tests - Fc: Vibration (sinusoidal) |
IEC 60068-2-13 | Basic Environmental Testing Procedures - Part 2: Tests – Test M: Low Air Pressure |
IEC 60068-2-27 | Environmental Testing - Part 2-27: Tests Test Ea and Guidance: Shock |
IEC 60068-2-30 | Environmental Testing - Part 2-30: Tests – Test Db: Damp Heat, Cyclic |
IEC 60068-2-64 | Environmental Testing - Part 2-64: Tests - Test Fh: Vibration, Broadband Random |
and Guidance | |
IEC 62304 | Medical Device Software - Software Lifecycle Processes |
IEC 62366 | Medical Devices - Application of Usability Engineering to Medical Devices |
The results of this testing show that the subject device meets the product requirements and specifications including: SpO2 accuracy without motion of ±2% from 70-100%, SpO2 accuracy with motion of ±3 from 70-100%, and pulse rate of ±3% from 20 to 250 BPM, and is substantially equivalent to the identified predicate devices.
Performance Testing - Clinical
There was no new clinical testing conducted on the subject Nellcor Bedside SpO2 Patient Monitoring System. Clinical testing was conducted on a similar Nellcor pulse oximetry device utilizing the same PCBA module to demonstrate the accuracy of the oximetry performance during both motion and no motion conditions.
Summary
The Nellcor Bedside SpO2 Patient Monitoring System is a pulse oximeter that is substantially equivalent to the identified predicate pulse oximeters.