(262 days)
Not Found
No
The summary describes a standard I.V. catheter and does not mention any AI or ML capabilities. The "Mentions AI, DNN, or ML" field is explicitly marked as "Not Found".
Yes
The device is used to administer fluid intravenously, which is a therapeutic intervention.
No
The device is described as an I.V. Catheter intended to withdraw blood samples or administer fluid, which are procedures, not diagnostic functions.
No
The device description clearly states it is a physical I.V. Catheter, a sterile, single-use hardware device intended for insertion into the vascular system. There is no mention of software as the primary or sole component.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is to "be inserted in to a patient's vascular system for short term use (no more than 72 hours) to withdraw blood samples or administer fluid intravenously." This describes a device used on the patient for direct medical intervention (administering fluids) and sample collection from the patient.
- IVD Definition: In vitro diagnostics are devices intended for use in the examination of specimens derived from the human body to provide information for diagnostic, monitoring, or compatibility purposes. They are used outside the body (in vitro).
While the device is used to collect blood samples, the device itself is not used to examine those samples or provide diagnostic information. The examination of the blood samples would be performed using separate IVD devices.
N/A
Intended Use / Indications for Use
The proposed device, I.V. Catheter, is intended to be inserted in to a patient's vascular system for short term use (no more than 72 hours) to withdraw blood samples or administer fluid intravenously.
Product codes
FOZ
Device Description
The proposed devices, I.V. Catheters, are sterile, single use devices intended to be inserted in to a patient's vascular system for short term use to withdraw blood samples or administer fluid intravenously.
The I.V. Catheter is a closed system, and further available in two combinations of configurations. All of them are sterilized by Ethylene Oxide to achieve a SAL of 10^6, and packed in a sterility maintenance package. The shelf life of the product is three years.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
vascular system
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- ASTM F88-09, Standard Test Method for Seal Strength of Flexible Barrier Materials
- ASTM F 1140-07 Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Package for Medical Applications
- ASTM F1929-98(2004) Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration
- ISO 11135-1:2007 Sterilization of health care products- Ethylene oxide- Part 1: Requirements for development, validation and routing control of a sterilization process for medical devices;
- USP Bacterial Endotoxin Limit
- ISO 10993-3:2003, Biological evaluation of medical devices- Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity.
- ISO 10993-4:2002 A1:2006, Biological evaluation of medical devices- Part 4: Selection of tests for interactions with blood.
- ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro Cytotoxicity;
- ISO 10993-6:2007, Biological evaluation of medical devices- Part 6: Tests for local effects after implantation:
- ISO 10993-7:2008, Biological evaluation of medical devices- Part 7: Test of Ethylene Oxide Residues;
- ISO 10993-10:2010, Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization;
- ISO 10993-11:2006, Biological evaluation of medical devices Part 11: Tests for systemic toxicity;
- ASTM F756-08. Standard practice for assessment of hemolytic properties of material:
- ISO 10555-1:1995/AMD.1:1999/AMD.2:2004(E), Sterile, single-use intravascular catheters-Part 1: General requirements;
- ISO10555-5: 1996/AMD.1:1999(E), Sterile, single-use intravascular catheters- Part 5: Over-needle peripheral catheters;
- ISO594-2:1998. Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment -Part 2: Lock fittings
- ISO9626:1991/AMD-1:2001, Stainless steel needle tubing for the manufacture of medical devices
Key Metrics
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 880.5200 Intravascular catheter.
(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, stylized to resemble a bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 18, 2015
Jiangxi Sanxin Medtec Co., Ltd c/o Ms. Diana Hong Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai 200120 China
Re: K142797
Trade/Device Name: I.V. Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: May 13, 2015 Received: May 18, 2015
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
1
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K142797
Device Name I.V. Catheter
Indications for Use (Describe)
The proposed device, I.V. Catheter, is intended to be inserted in to a patient's vascular system for short term use (no more than 72 hours) to withdraw blood samples or administer fluid intravenously.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CER 801 Subpart D)
- Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BECOW THIS LINE - CONTINUE ON A SEPARATE PAGE IE NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM IO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspecti of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staffl PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Section 3 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K142797
-
- Date of prepare the summary: June 18, 2015
-
- Sponsor Identification
Jiangxi Sanxin Medtec Co., Ltd. No. 999, Fushan Road, Xiaolan Economic Development, Nanchang, Jiangxi, 330200, China
Establishment Registration Number: Not yet registered
Contact Person: Bingrong Liu Position: Technical Department Manager Tel: +86-791-85950238 Fax: +86-791-85988030 Email: tengyun2005@163.com
-
- Submission Correspondent
Ms. Diana Hong& Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net
- Submission Correspondent
4
- Proposed Device Identification 4.
Proposed Device Name: I.V. Catheter Proposed Device Common Name: Intravascular Catheter
Regulatory Information for I.V. Catheter Classification Name: Intravascular Catheter; Classification: II; Product Code: FOZ; Regulation Number: CFR 880.5200; Review Panel: General Hospital;
Intended Use Statement:
The proposed device, I.V. Catheter, is intended to be inserted in to a patient's vascular system for short term use (no more than 72 hours) to withdraw blood samples or administer fluid intravenously.
- રું. Predicate Device Identification
Predicate Device 510(k) Number: K083429 Product Name: I.V. Catheter for Single Use Manufacturer: Weihai Jierui Medical Products Co., Ltd.
6. Device Description
The proposed devices, I.V. Catheters, are sterile, single use devices intended to be inserted in to a patient's vascular system for short term use to withdraw blood samples or administer fluid intravenously.
The I.V. Catheter is a closed system, and further available in two combinations of configurations. All of them are sterilized by Ethylene Oxide to achieve a SAL of 106, and packed in a sterility maintenance package. The shelf life of the product is three years.
5
-
- Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- Non-Clinical Test Conclusion
-
ASTM F88-09, Standard Test Method for Seal Strength of Flexible Barrier Materials
-
A ASTM F 1140-07 Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Package for Medical Applications
-
ASTM F1929-98(2004) Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration
-
A ISO 11135-1:2007 Sterilization of health care products- Ethylene oxide- Part 1: Requirements for development, validation and routing control of a sterilization process for medical devices;
-
USP Bacterial Endotoxin Limit >
-
A ISO 10993-3:2003, Biological evaluation of medical devices- Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity.
-
A ISO 10993-4:2002 A1:2006, Biological evaluation of medical devices- Part 4: Selection of tests for interactions with blood.
-
A ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro Cytotoxicity;
-
A ISO 10993-6:2007, Biological evaluation of medical devices- Part 6: Tests for local effects after implantation:
-
ISO 10993-7:2008, Biological evaluation of medical devices- Part 7: Test of Ethylene Oxide Residues;
-
A ISO 10993-10:2010, Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization;
-
ISO 10993-11:2006, Biological evaluation of medical devices Part 11: Tests for systemic toxicity;
-
ASTM F756-08. Standard practice for assessment of hemolytic properties of material: ▲
-
ISO 10555-1:1995/AMD.1:1999/AMD.2:2004(E), Sterile, single-use intravascular catheters-Part 1: General requirements;
-
A ISO10555-5: 1996/AMD.1:1999(E), Sterile, single-use intravascular catheters- Part 5: Over-needle peripheral catheters;
-
A ISO594-2:1998. Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment -Part 2: Lock fittings
-
ISO9626:1991/AMD-1:2001, Stainless steel needle tubing for the manufacture of medical devices
6
8. Substantially Equivalent (SE) Comparison
Item | Proposed device | Predicate Device K083429 |
---|---|---|
Product Code | FOZ | FOZ |
Regulation Number | CFR 880.5200 | CFR 880.5200 |
Intended Use | The proposed device, I.V. Catheter, is | |
intended to be inserted in to a patient's | ||
vascular system for short term use (no more | ||
than 72 hours) to withdraw blood samples or | ||
administer fluid intravenously. | I.V. Catheter for Single Use, including Type | |
I, Type Y and Type Straight, is intended to be | ||
inserted in to a patient's vascular system for | ||
short term use to withdraw blood samples, | ||
administer fluid intravenously or through | ||
which to place monitoring equipment such as | ||
blood pressure monitors. | ||
Features | Closed System, Radio Detectable, Color | |
Coding, Single Use | Closed System, Radio Detectable, Color | |
Coding, Single Use | ||
Configuration | Catheter tube, Catheter hub, Needle, Needle | |
handle, Flexible tube and Infusion joint | Catheter tube, Catheter hub, Needle, Needle | |
handle, Flexible tube and Infusion joint | ||
Principle of Operation | Manual | Manual |
Sterility | EO Sterilized | EO Sterilized |
Performance | Comply with: | |
ISO 10555-1:1995/AMD.1:1999/ AMD. | ||
2:2004 | ||
ISO10555-5: 1996/AMD.1:1999(E) | ||
ISO 9626: 1991/AMD-1:2006 | ||
ISO 594-2:1998 | Comply with: | |
ISO 10555-1:1995/AMD.1:1999/ AMD. | ||
2:2004 | ||
ISO10555-5: 1996/AMD.1:1999(E) | ||
ISO 9626: 1991/AMD-1:2006 | ||
ISO 594-2:1998 | ||
Biocompatibility | Conforms to the requirement of ISO 10993 | |
series Standards | Conforms to the requirement of ISO 10993 | |
series | ||
Standards |
Table 1 Product Comparison
Substantially Equivalent (SE) Conclusion 9.
The propose device has the same intended use with the predicate device, except that the proposed device is not intended for placing monitoring equipment. The proposed device has less indication compared with the propose device, and the indications are provided on the label and user manual, therefore, this difference is not considered to affect the substantially equivalent between the proposed and predicate device.
The features, configurations, principle of operation, sterility of the proposed are the same as those of the predicate device.
7
In addition, the results of performance tests performed on the proposed device can also demonstrate the proposed device is complied with FDA recognized standards, which the proposed device was also complied with. The results of biocompatibility studies performed on the proposed device demonstrate that the patient materials used in proposed device are biocompatibile.
Based on the comparison above, the proposed device, I.V. Catheter, is determined to be Substantially Equivalent (SE) to the predicate devices, I.V. Catheter for Single Use (K083429).