AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BioScanner Plus Glucose Test Systems are intended for the quantitative determination of glucose in human whole blood for use by healthcare professionals in both physicians' office and convalescent care facility bedside testing. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia and idiopathic hypoglycemia and of pancreatic islet cell carcinoma.

The Chol+Glu Test Panel system is intended to measure cholesterol and glucose in whole blood on a BioScamer Plus (CardioChek brand) analyzer. The test strips are intended to be used by healthcare professionals to measure blood cholesterol and glucose. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipoprotein metabolism disorders. Glucose measurements are used in the management of carbohydrate metabolism disorders.

The Lipid Panel Test system is intended to measure cholesterol and triglycerides in whole blood on a BioScanner Plus (CardioChek brand) analyzer. The test strips are intended to be used by healthcare professionals to measure three blood analytes: cholesterol and triglycerides. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipoprotein metabolism disorders. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.

The PTS PANELS HDL Cholesterol Test system is intended to measure high density lipoprotein cholesterol in whole blood. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders mellitus), atherosclerosis, and various liver and renal diseases.

The PTS PANELS CHOL+HDL Panel Test system is intended to be used by medical professionals to measure cholesterol and high density lipoprotein cholesterol in whole blood. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipoprotein metabolism disorders. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis and various liver and renal diseases.

The PTS PANELS CHOL+HDL+GLU Panel Test system is intended to be used by medical professionals to measure cholesterol, high density lipoprotein cholesterol and glucose in whole blood. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipoprotein metabolism disorders. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. Glucose measurements are used by individuals with diabetes to measure glucose in fingerstick whole blood at home for the management of carbohydrate metabolism disorders.

The PTS PANELS Metabolic Chemistry Panel Test system is intended to be used by medical professionals to measure glucose, high density lipoprotein cholesterol and triglycerides in fingerstick whole blood. Glucose measurements are used in the management of carbohydrate metabolism disorders. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipoprotein metabolism disorders. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver diseases involving lipid metabolism or various endocrine disorders.

Device Description

The above named test systems are for in vitro diagnostic use with the prescription use BioScanner Plus (CardioChek PA) reflectance photometer.

The above named test systems are modified in this submission by the addition of a CardioChek ChekMate strip for use only with the prescription use analyzer. CardioChek ChekMate strips are dry strips that mimic the use of the PTS PANELS test strips to check the CardioChek analyzer system optics, calibration and result handling algorithms. ChekMate strips should not be used as a substitute for liquid quality control materials. There is no change to any of the test strips in any of the above named systems.

AI/ML Overview

The provided document describes the BioScanner Plus Glucose Test System and related PTS PANELS test systems. The primary focus of the performance characteristics section is on the CardioChek ChekMate strips, which are a modification to the system. While the document mentions several test systems and their intended uses (glucose, cholesterol, triglycerides, HDL cholesterol), the performance data specifically addresses the ChekMate strips, which are used to check the analyzer's optics, calibration, and result handling. The information regarding the various analyte test systems largely defers to previously cleared strips and an unmodified analyzer, indicating their performance was established in prior submissions.

Here's a breakdown of the acceptance criteria and study information for the CardioChek ChekMate Strips:

Table of Acceptance Criteria and Reported Device Performance (for CardioChek ChekMate Strips)

Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
Value Assignment: Range AppropriatenessRange (MAX – MIN) must be at least 5% R but not more than 10% R.Not explicitly stated what the calculated ranges were, but the document states "The range is acceptable if it is at least 5% R, but not more than 10% R."
Value Assignment: Individual Strip VerificationEvery ChekMate strip, when measured on a different CardioChek PA meter, must pass the established acceptable ranges.The document states, "Every ChekMate strip is then measured...to verify that every strip will pass these ranges." Implies compliance.
Real-time Stability: Percent RecoveryPercent recovery ((Result - Result baseline) * 100) must be 100±20%.All four lots at each of the two levels passed the criterion.
Re-use Stability: Failure Rate on First TestLess than 2% failures on the first test.The acceptance criteria was met for 1,253 uses. (Implying <2% failures).
Re-use Stability: Failure Rate on Re-testNo failures upon re-test, if a test result did not meet the range initially.The acceptance criteria was met for 1,253 uses. (Implying no failures on re-test).

Study Details for CardioChek ChekMate Strips:

  1. Sample sizes used for the test set and the data provenance:

    • Value Assignment Test Set: 45 strips evaluated on a single CardioChek PA meter, and then "Every ChekMate strip" (up to 400 pieces per lot) was measured on a different CardioChek PA meter.
    • Real-time Stability Test Set: Four (4) lots of ChekMate strips.
    • Re-use Stability Test Set: Individual ChekMate strips were tested for 1,253 uses across 13 CardioChek PA analyzers.
    • Data Provenance: Not explicitly stated, but typically such studies are conducted by the manufacturer (Polymer Technology Systems, Inc.) in the USA. The document does not specify if the data is retrospective or prospective, but stability studies are inherently prospective.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This device (CardioChek ChekMate strips) is a quality control/verification strip, not a diagnostic device that requires expert interpretation for ground truth. The "ground truth" for the ChekMate strips is based on instrument readings (% R values) against predefined acceptable ranges and stability curves. Therefore, no human experts were involved in establishing the ground truth for these strips.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. The evaluation criteria for the ChekMate strips are objective, numerical measurements of % R values and their adherence to predefined ranges and stability criteria. There is no need for expert adjudication.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is an in vitro diagnostic device for quantitative measurements of analytes or, in the case of ChekMate strips, for verification of an instrument. It is not an AI-assisted diagnostic tool that would involve human readers or an MRMC study.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • The ChekMate strips themselves do not operate as an algorithm in isolation. They are physical strips used with an analyzer. The document describes the performance of the strips in verifying the analyzer's functions (optics, calibration, result handling algorithms). The analyzer itself (CardioChek PA) operates in a "standalone" fashion to generate a result from the strip, but the performance evaluation here is focused on the strips and their ability to consistently produce expected values within the analyzer. The algorithms referred to are integral to the analyzer's function, not a separate AI system.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the ChekMate strips, the ground truth is established through internal scientific and engineering processes, likely involving highly calibrated reference instruments and defined chemical/optical properties. The "acceptable ranges" and "baseline results" are the ground truth for testing these verification strips.
  7. The sample size for the training set:

    • Not applicable. This document describes the performance of an in vitro diagnostic device (and its quality control strips). It does not involve machine learning or AI models that require a separate "training set."
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no "training set" in the context of this device's performance evaluation.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a stylized image of an eagle with three human profiles incorporated into its design. The eagle's wings are represented by curved lines that also form the silhouettes of human faces.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 2, 2014

POLYMER TECHNOLOGY SYSTEMS, INC. JACK ROGERS DIRECTOR, REGULATORY AFFAIRS 7736 ZIONSVILLE ROAD INDIANAPOLIS IN 46268

Re: K142302

Trade/Device Name: BioScanner Plus Glucose Test System PTS PANELS Chol+Glu Test Panel System PTS PANELS Lipid Panel Test System PTS PANELS HDL Cholesterol Test System PTS PANELS CHOL+HDL Panel Test System PTS PANELS CHOL+HDL+GLU Panel Test System PTS PANELS Metabolic Chemistry Panel Test System

Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: CGA, CHH, JGY, LBR, MRR Dated: August 15, 2014 Received: August 18, 2014

Dear Mr. Jack Rogers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Courtney

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K142302

Device Name

BioScanner Plus Glucose Test System, PTS PANELS Chol+Glu Test Panel System, PTS PANELS Lipid Panel Test System, PTS PANELS HDL Cholesterol Test System, PTS PANELS CHOL+HDL Panel Test System, PTS PANELS CHOL+HDL+GLU Panel Test System, PTS PANELS Metabolic Chemistry Panel Test System

Indications for Use (Describe)

The BioScanner Plus Glucose Test Systems are intended for the quantitative determination of glucose in human whole blood for use by healthcare professionals in both physicians' office and convalescent care facility bedside testing. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia and idiopathic hypoglycemia and of pancreatic islet cell carcinoma.

The Chol+Glu Test Panel system is intended to measure cholesterol and glucose in whole blood on a BioScamer Plus (CardioChek brand) analyzer. The test strips are intended to be used by healthcare professionals to measure blood cholesterol and glucose. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipoprotein metabolism disorders. Glucose measurements are used in the management of carbohydrate metabolism disorders.

The Lipid Panel Test system is intended to measure cholesterol and triglycerides in whole blood on a BioScanner Plus (CardioChek brand) analyzer. The test strips are intended to be used by healthcare professionals to measure three blood analytes: cholesterol and triglycerides. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipoprotein metabolism disorders. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.

The PTS PANELS HDL Cholesterol Test system is intended to measure high density lipoprotein cholesterol in whole blood. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders mellitus), atherosclerosis, and various liver and renal diseases.

The PTS PANELS CHOL+HDL Panel Test system is intended to be used by medical professionals to measure cholesterol and high density lipoprotein cholesterol in whole blood. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipoprotein metabolism disorders. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis and various liver and renal diseases.

The PTS PANELS CHOL+HDL+GLU Panel Test system is intended to be used by medical professionals to measure cholesterol, high density lipoprotein cholesterol and glucose in whole blood. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipoprotein metabolism disorders. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. Glucose measurements are used by individuals with diabetes to measure glucose in fingerstick whole blood at home for the management of carbohydrate metabolism disorders.

{3}------------------------------------------------

The PTS PANELS Metabolic Chemistry Panel Test system is intended to be used by medical professionals to measure glucose, high density lipoprotein cholesterol and triglycerides in fingerstick whole blood. Glucose measurements are used in the management of carbohydrate metabolism disorders. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipoprotein metabolism disorders. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver diseases involving lipid metabolism or various endocrine disorders.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

1. Applicant Information:

Polymer Technology Systems, Inc. 7736 Zionsville Road Indianapolis, IN 46268

Contact Person:Jack Rogers
Phone:317-870-5610
Fax:317-870-5608

2. Date of Preparation:

September 26, 2014

3. Trade Name:

BioScanner Plus Glucose Test System PTS PANELS Chol+Glu Test Panel System PTS PANELS Lipid Panel Test System PTS PANELS HDL Cholesterol Test System PTS PANELS CHOL+HDL Panel Test System PTS PANELS CHOL+HDL+GLU Panel Test System PTS PANELS Metabolic Chemistry Panel Test System

4. Device Description:

The above named test systems are for in vitro diagnostic use with the prescription use BioScanner Plus (CardioChek PA) reflectance photometer.

The above named test systems are modified in this submission by the addition of a CardioChek ChekMate strip for use only with the prescription use analyzer. CardioChek ChekMate strips are dry strips that mimic the use of the PTS PANELS test strips to check the CardioChek analyzer system optics, calibration and result handling algorithms. ChekMate strips should not be used as a substitute for liquid quality control materials. There is no change to any of the test strips in any of the above named systems.

5. Classification Names:

Glucose Test System Panel: Clinical Chemistry 75 Product Codes: CGA Device Classification: Class 2 (Regulation 21CFR 862.1345)

Cholesterol Test System Panel: Clinical Chemistry 75 Product Codes: СНН Device Classification: Class I (Regulation: 21 CFR 862.1175*) * Meets limitations of exemption as per 21 CFR 862.9 (c)(4)

Triglyceride Test System Panel: Clinical Chemistry 75 Product Codes: JGY Device Classification: Class I (Regulation: 21 CFR 862.1705*) * Meets limitations of exemption as per 21 CFR 862.9 (c)(4)

{5}------------------------------------------------

Lipoprotein Test System Panel: Clinical Chemistry 75 Product Codes: LBR, MRR Device Classification: Class I (Regulation: 21 CFR 862.1475*) * Meets limitations of exemption as per 21 CFR 862.9 (c)(4)

6. Intended Use

The BioScanner Plus Glucose Test Systems are intended for the quantitative determination of glucose in human whole blood for use by healthcare professionals in both physicians' office and in acute and convalescent care facility bedside testing. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia and idiopathic hypoglycemia and of pancreatic islet cell carcinoma.

The Chol+Glu Test Panel system is intended to measure cholesterol and glucose in whole blood on a BioScanner Plus (CardioChek brand) analyzer. The test strips are intended to be used by healthcare professionals to measure blood cholesterol and dlucose. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. Glucose measurements are used in the management of carbohydrate metabolism disorders.

The Lipid Panel Test system is intended to measure cholesterol and triglycerides in whole blood on a BioScanner Plus (CardioChek brand) analyzer. The test strips are intended to be used by healthcare professionals to measure three blood analytes: cholesterol and triglycerides. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipoprotein metabolism disorders. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.

The PTS PANELS HDL Cholesterol Test system is intended to measure high density lipoprotein cholesterol in whole blood. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.

The PTS PANELS CHOL+HDL Panel Test system is intended to be used by medical professionals to measure cholesterol and high density lipoprotein cholesterol in whole blood. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis and various liver and renal diseases.

The PTS PANELS CHOL+HDL+GLU Panel Test system is intended to be used by medical professionals to measure cholesterol, high density lipoprotein cholesterol and glucose in whole blood. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipoprotein metabolism disorders. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. Glucose measurements are used by individuals with diabetes to measure glucose in fingerstick whole blood at home for the management of carbohydrate metabolism disorders.

{6}------------------------------------------------

The PTS PANELS Metabolic Chemistry Panel Test system is intended to be used by medical professionals to measure glucose, high density lipoprotein cholesterol and triglycerides in fingerstick whole blood. Glucose measurements are used in the management of carbohydrate metabolism disorders. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.

7. Reason for 510(k):

Device Modification

8. Predicate Device Information

Device Trade Name:PTS PANELS CHOL+HDL+GLU Panel Test StripsPTS PANELS Metabolic Chemistry Panel Test Strips
Device Company:Polymer Technology Systems, Inc.
510(k) number:K071507 and K070017

Comparison to Predicate:

Each system is identical to the unmodified system with the exception of the ChekMate strip.

9. Performance Characteristics

The strips for each system are identical to the previously cleared strips. The prescription use BioScanner Plus (CardioChek PA) analyzer is unmodified compared to the predicate device with the exception of the addition of the CardioChek ChekMate strips.

CardioChek ChekMate Strips Value Assignment:

Following assembly of a lot of ChekMate strips (up to 400 pieces per lot), 45 strips are evaluated on a single CardioChek PA meter. The four results (% R) for each analyzer (C1 RED, C2 RED, C1 GREEN) are recorded and the average, minimum and maximum results are determined and evaluated for meeting acceptable ranges.

The acceptable value ranges to be printed on the ChekMate range card are calculated as the minimum result minus 1% R (MIN) to the maximum result plus 1% R (MAX). This range (MAX – MIN) is then evaluated for appropriateness and acceptability. The range is acceptable if it is at least 5% R, but not more than 10% R. Every ChekMate strip is then measured on a different CardioChek PA meter to verify that every strip will pass these ranges.

Each package of ChekMate strips contains a range card which lists the lot specific expected ranges for each of the four measurements for Level 1 and Level 2.

{7}------------------------------------------------

CardioChek ChekMate Strips Stability:

Real-time Stability:

A real-time stability study was performed on four (4) lots of ChekMate Strips. Prior to study initiation, each lot was tested 30 times to establish the baseline result (Result paseline). The strips were tested at predetermined intervals throughout the one hundred and four week study. The percent recovery ((Result = Result paseline) * 100) was calculated. Percent recovery was required to be 100±20% to consider the test as passed. All four lots at each of the two levels passed the criterion.

The study supports an 18 month stability claim.

Re-use Stability:

The stability of ChekMate for repeated re-use was verified over a 9 month study using 13 CardioChek PA analyzers. Individual ChekMate strips were tested for 1,253 uses.

Each result was evaluated against the acceptable performance range published for the lot. If a test result did not meet this range, the test was to be repeated. The acceptance criteria for the re-use stability study was lower than 2% failures on the first test and no failures upon re-test. The acceptance criteria was met for 1,253 uses and the claimed re-use stability is 500 uses for the ChekMate strips.

10. Conclusion:

The performance characteristics above demonstrate that the modification of the above named test systems by the addition of the CardioChek ChekMate strips are substantially equivalent to the unmodified systems.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.