AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

FastPack® High Sensitivity C-Reactive Protein Immunoassay is to be used for evaluation of conditions thought to be associated with inflammation, in otherwise healthy individuals. The FastPack® High Sensitivity C-Reactive Protein Immunoassay is intended for use with the FastPack® Analyzer. Not intended for Point-of-Care use.

FastPack® High Sensitivity C-Reactive Protein Calibrators are used for calibrating the quantitative FastPack® High Sensitivity C-Reactive Protein Immunoassay on the FastPack® Analyzer.

FastPack® High Sensitivity C-Reactive Protein Controls are used for quality control of the FastPack® High Sensitivity C-Reactive Protein Immunoassay on the FastPack® Analyzer.

FastPack® High Sensitivity C-Reactive Protein Verifiers are used in the quantitative verification of calibration and assay range of the quantitative FastPack® High Sensitivity C-Reactive Protein Immunoassay on the FastPack® Analyzer.

Device Description

The FastPack® High Sensitivity C-Reactive Protein Immunoassay employs a Sandwich immunoassay principle. Endogenous CRP in a patient sample, calibrator, control, or verifier is dispensed into a FastPack® reagent pack. In the reagent pack, the sample binds with a monoclonal anti-CRP antibody covalently linked to alkaline phosphatase (ALP) and a different monoclonal anti-CRP antibody linked to biotin. After incubation, immunoreacted complex (Monoclonal anti-CRP antibody-ALP conjugate and anti-CRP antibody linked to biotin reacted with CRP in the sample) is mixed with streptavidin coated paramagnetic particles. After washing steps (using a Tris buffer containing detergents) to separate bound from unbound anti-CRP monoclonal antibody-ALP, a chemiluminogenic substrate mixture is added to the system. This mixture contains indoxyl-3-phosphate, a substrate for ALP, and lucigenin (N,N'-dimethyl-9,9'-biacridinium dinitrate). ALP dephosphorylates indoxyl-3-phosphate to indol-3-ol, which subsequently undergoes oxidation. As a result, lucigenin is reduced to form a dioxetane structure that is cleaved to yield Nmethylacridone. This compound produces a sustained luminescent glow following excitation. The raw relative luminescence units (RLUs) generated are measured by a photomultiplier tube in the FastPack® Analyzer and are directly proportional to the concentration of CRP in the sample. The entire reaction sequence takes place at 37 ± 0.5 ℃ and is protected from external light.

AI/ML Overview

The provided document describes the FastPack® High Sensitivity C-Reactive Protein Immunoassay (device) and its performance to demonstrate substantial equivalence to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly list "acceptance criteria" as a separate section with pass/fail values. Instead, it presents performance characteristics and compares them to the predicate device or established guidelines (e.g., CLSI). Based on the provided sections, a table can be constructed as follows:

Acceptance Criteria (Implied)FastPack® High Sensitivity C-Reactive Protein Immunoassay PerformancePredicate Device Performance (Olympus CRP Latex reagent K051564)
Precision
Within-run %CV≤ 1.0% (across various CRP levels)≤ 3.2%
Between-run %CV≤ 5.2% (across various CRP levels)N/A (Total: ≤ 3.8%)
Total %CV≤ 9.0% (across various CRP levels)≤ 3.8%
Limits (LOB, LOD, LOQ)
LOB0.005 mg/LNot explicitly stated, implied to be comparable as predicates are established assays.
LOD0.032 mg/L
LOQ0.063 mg/L
Linearity Range0.063 mg/L to 15.0 mg/L0.2 - 160 mg/L (High Sensitivity Application: 0.2 - 15.0 mg/L is comparable to FastPack's application range)
InterferencesNo interference from specified endogenous and exogenous substances at noted concentrations.No interference from high levels of bilirubin, hemoglobin, and triglycerides.
Cross-reactivityNo cross-reactivity from Rheumatoid factor (up to 1000 U/mL), human anti-mouse IgG (up to 4 ug/mL), and heterophile samples (up to 3641 ng/mL).Not explicitly detailed for all substances, but implied to have established specificity.
Serum and Plasma Equivalence (vs. Serum)N/A (Device-specific validation)
EDTA Plasma Slope0.94
EDTA Plasma R0.984
Lithium-Heparin Plasma Slope1.00
Lithium-Heparin Plasma R0.993
Method Comparison (vs. Predicate)
Correlation Coefficient (R)0.990.999 (AU400 vs. AU640, a different comparison for predicate)
Slope0.98 (95% CI: 0.95-1.02)1.025 (Predicate's own method comparison)
y-intercept-0.12 (95% CI: -0.23 to 0.00)-0.022 (Predicate's own method comparison)

2. Sample Size Used for the Test Set and the Data Provenance:

  • Precision Study: 6 serum patient samples were used, each with 320 replicate determinations.
  • Limits of blank, detection, and quantitation: Information not provided about the specific "samples" used for these determinations.
  • Linearity Study: A high patient sample was intermixed with a low sample to generate 8 concentration levels, each tested in triplicate.
  • Interferences: Specific concentrations of various substances were tested. The number of samples/replicates isn't explicitly stated, but implies a controlled study.
  • Cross-reactivity: Specific concentrations of substances like Rheumatoid factor, human anti-mouse IgG, and heterophile samples were tested.
  • Serum and Plasma Equivalence: Blood collections from 41 volunteers.
  • Expected Values/Reference Intervals: Serum samples from 211 subjects.
  • Method Comparison: 131 samples.

Data Provenance:
The document states that the reference interval study employed serum samples from "211 subjects representing 4 different geographic regions of the United States." This indicates a prospective or retrospective collection of patient samples within the US for this specific study. For other studies (precision, linearity, interferences, serum/plasma equivalence, method comparison), specific provenance details like country of origin or retrospective/prospective nature are not explicitly stated, but are generally assumed to be carried out under controlled laboratory conditions often using commercially available or anonymized clinical samples.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

This information is not applicable to this type of device (in-vitro diagnostic immunoassay). The "ground truth" for a C-reactive protein immunoassay is typically the actual concentration of CRP in the sample, which is determined by established reference methods or by the device's own calibrated measurement. There are no human "experts" establishing ground truth in the sense of image interpretation or clinical diagnosis for this kind of analytical device. The "ground truth" for method comparison is the result obtained from the predicate device or a reference method.

4. Adjudication Method for the Test Set:

This information is not applicable for this type of device. Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies where multiple human readers interpret data (e.g., medical images) and their interpretations need to be resolved for ground truth. For an immunoassay, the results are quantitative measurements, not subjective interpretations requiring adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable to this device. An MRMC study is relevant for AI-assisted diagnostic tools where human readers are part of the diagnostic workflow. The FastPack® Immunoassay is an automated analytical device that directly measures CRP concentration, not an AI system assisting human interpretation.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:

The device described, the FastPack® High Sensitivity C-Reactive Protein Immunoassay, is an automated immunoassay system. Its performance is inherently "standalone" in the sense that the analyzer performs the measurement without human intervention influencing the quantitative result. The performance data presented (precision, linearity, limits, interference, method comparison) is the standalone performance of the algorithm/system. There is no "human-in-the-loop" component in terms of result generation for this type of device.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for the performance studies of this immunoassay primarily relies on:

  • Reference Methods/Predicate Device Results: For method comparison, the results obtained from a legally marketed predicate device (Olympus AU2700 for the method comparison) served as the reference for comparison.
  • Known Concentrations: For studies like linearity, limits of detection/quantitation, and precision, samples with known, spiked, or characterized CRP concentrations are used.
  • Clinical Samples: For serum and plasma equivalence and reference interval studies, collected human serum and plasma samples are used. The CRP concentrations in these samples, as measured by the device itself or a reference method, serve as the "true" or observed values for evaluating performance parameters.

8. The Sample Size for the Training Set:

This information is not applicable in the context of this document. This device is an immunoassay, not an AI/machine learning system that requires a "training set" in the computational sense. The device's operational parameters, calibration curves, and analytical procedures are established through manufacturing and analytical validation processes, not through machine learning training on a dataset.

9. How the Ground Truth for the Training Set was Established:

This information is not applicable as there is no "training set" for this type of immunoassay device in the context of an AI/machine learning system. The "ground truth" for calibrating immunoassays is established using reference materials with assigned values traceable to international standards (e.g., ERM-DA474/IFCC for CRP, as mentioned in the device's traceability). These reference materials are meticulously characterized using highly accurate and precise analytical methods.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

QUALIGEN, INC. MICHAEL POIRIER SENIOR VICE PRESIDENT, CSO/CTO 2042 CORTE DEL NOGAL CARLSBAD CA 92011

January 7, 2015

Re: K141689

Trade/Device Name: Fastpack High Sensitivity C-reactive Protein Immunoassay Fastpack High Sensitivity C-reactive Protein Calibrator Kit Fastpack High Sensitivity C-reactive Protein Controls, Fastpack High Sensitivity C-reactive Protein Method Verification Kit Regulation Number: 21 CFR 866.5270 Regulation Name: C-reactive protein immunological test system Regulatory Class: II Product Code: DCK, JIT, JJX Dated: December 15, 2014 Received: December 18, 2014

Dear Mr. Poirier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -S

For : Courtney H. Lias, Ph.D.

Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K141689

Device Name

FastPack® High Sensitivity C-Reactive Protein Immunoassay, FastPack® High Sensitivity C-Reactive Protein Calibrator Kit, FastPack® High Sensitivity C-Reactive Protein Controls . High Sensitivity C-Reactive Protein Method Verification Kit

Indications for Use (Describe)

FastPack® High Sensitivity C-Reactive Protein Immunoassay is to be used for evaluation of conditions thought to be associated with inflammation, in otherwise healthy individuals. The FastPack® High Sensitivity C-Reactive Protein Immunoassay is intended for use with the FastPack® Analyzer. Not intended for Point-of-Care use.

FastPack® High Sensitivity C-Reactive Protein Calibrators are used for calibrating the quantitative FastPack® High Sensitivity C-Reactive Protein Immunoassay on the FastPack® Analyzer.

FastPack® High Sensitivity C-Reactive Protein Control of the FastPack® High Sensitivity C-Reactive Protein Immunoassay on the FastPack® Analyzer.

FastPack® High Sensitivity C-Reactive Protein Verifiers are used in the quantitative verification and assay range of the quantitative FastPack® High Sensitivity C-Reactive Protein Immunoassay on the FastPack® Analyzer.

Type of Use (Select one or both, as applicable)

[X] Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

510(k) SUMMARY

{4}------------------------------------------------

REVISED 510(k) SUMMARY

This 510(k) Summary information is submitted in accordance with the requirements of 21 CFR § 807.92.

510(k) Number: K141689

Submitter:

Qualigen, Inc. 2042 Corte Del Nogal, Suite B Carlsbad, CA 92009

Telephone:(760) 918-9165
Facsimile:(760) 918-9127

Contact Person:

Mr. Michael Poirier Senior Vice President, Chief Scientific & Technical Officer Telephone: (760) 918-9165 x227 Facsimile: (760) 918-9127 Email: mpoirier@qualigeninc.com

Device Identification

Trade Names:FastPack® High Sensitivity C-Reactive Protein Immunoassay
FastPack® High Sensitivity C-Reactive Protein Calibrator Kit
FastPack® High Sensitivity C-Reactive Protein Controls
FastPack® High Sensitivity C-Reactive Protein Method
Verification Kit

Common Names: C-reactive protein Assay C-reactive protein Calibrator C-reactive protein Controls C-reactive protein Verifiers

  • Classification names: Immunological Test Systems Calibrator Quality control material (assayed and unassayed)
    Classifications: Class II (assay)

{5}------------------------------------------------

Class II (calibrators)
Class I, reserved (controls)
Class I, reserved (verifiers)
Panel:Chemistry (75)
Product Codes:DCK - C-Reactive Protein, Antigen, Antiserum
JIT - Calibrator, Secondary
JJX - Quality control material (Assayed and Unassayed)
Regulation Numbers: 21 CFR § 866.5270 - C-reactive protein immunological test system
21 CFR § 862.1150 - Calibrator
21 CFR § 862.1660 - Quality control material (Assayed and
Unassayed)

Devices to Which Substantial Equivalence is Claimed

Olympus CRP Latex reagent Olympus America, Inc. 3500 Corporate Parkway PO Box 610 Center Valley, PA K051564

Liquichek™ Cardiac Markers Plus Control Bio-Rad Laboratories 4000 Alfred Nobel Drive Hercules, CA 94547 K050537

VITROS Chemistry Products hsCRP Performance Verifier I, II, and III Ortho-Clinical Diagnostics, Inc. 100 Indigo Creek Drive Rochester, NY 14626-5101 K041799

Device Description

The FastPack® High Sensitivity C-Reactive Protein Immunoassay employs a Sandwich immunoassay principle. Endogenous CRP in a patient sample, calibrator, control, or verifier is dispensed into a FastPack® reagent pack. In the reagent pack, the sample binds with a monoclonal anti-CRP antibody covalently linked to alkaline

{6}------------------------------------------------

phosphatase (ALP) and a different monoclonal anti-CRP antibody linked to biotin. After incubation, immunoreacted complex (Monoclonal anti-CRP antibody-ALP conjugate and anti-CRP antibody linked to biotin reacted with CRP in the sample) is mixed with streptavidin coated paramagnetic particles. After washing steps (using a Tris buffer containing detergents) to separate bound from unbound anti-CRP monoclonal antibody-ALP, a chemiluminogenic substrate mixture is added to the system. This mixture contains indoxyl-3-phosphate, a substrate for ALP, and lucigenin (N,N'-dimethyl-9,9'-biacridinium dinitrate). ALP dephosphorylates indoxyl-3-phosphate to indol-3-ol, which subsequently undergoes oxidation. As a result, lucigenin is reduced to form a dioxetane structure that is cleaved to yield Nmethylacridone. This compound produces a sustained luminescent glow following excitation. The raw relative luminescence units (RLUs) generated are measured by a photomultiplier tube in the FastPack® Analyzer and are directly proportional to the concentration of CRP in the sample. The entire reaction sequence takes place at 37 ± 0.5 ℃ and is protected from external light.

Intended Use

FastPack® High Sensitivity C-Reactive Protein Immunoassay is to be used for evaluation of conditions thought to be associated with inflammation, in otherwise healthy The FastPack® High Sensitivity C-Reactive Protein Immunoassay is individuals. intended for use with the FastPack® Analyzer. Not intended for Point-of-Care use.

FastPack® High Sensitivity C-Reactive Protein Calibrators are used for calibrating the quantitative FastPack® High Sensitivity C-Reactive Protein Immunoassay on the FastPack® Analyzer.

FastPack® High Sensitivity C-Reactive Protein Controls are used for quality control of the FastPack® High Sensitivity C-Reactive Protein Immunoassay on the FastPack® Analyzer.

FastPack® High Sensitivity C-Reactive Protein Verifiers are used in the quantitative verification of calibration and assay range of the quantitative FastPack® High Sensitivity C-Reactive Protein Immunoassay on the FastPack® Analyzer.

{7}------------------------------------------------

Comparison of new nevice to preated devices
CHARACTERISTICQualigen FastPack® HighOlympus America, Inc. CRP
Sensitivity C-Reactive ProteinLatex reagent
ImmunoassayK051564
Intended Use/FastPack® High Sensitivity C-Olympus System Reagent and
Indications for UseReactive Protein Immunoassay iscalibrators for the quantitative
to be used for evaluation ofdetermination of C-Reactive
conditions thought to be associatedProtein on OLYMPUS
with inflammation, in otherwiseAnalyzers. Measurement of
healthy individuals. TheCRP is useful for the detection
FastPack® High Sensitivity C-and evaluation of infection,
Reactive Protein Immunoassay istissue injury, inflammatory
intended for use with thedisorders, and associated
FastPack® Analyzer. Not intendedfor Point-of-Care use.diseases. Measurements mayalso be used as an aid in the
identification of individuals at
risk of future cardiovascular
disease. High sensitivity CRP
(hsCRP) measurements, when
used in conjunction with
traditional clinical laboratory
evaluation of acute coronary
syndromes, may be useful as an
independent marker of prognosis
for recurrent events, in patients
with stable coronary disease or
acute coronary syndromes.
Sample TypeSerum or plasma(EDTAorSame
lithium heparin)
Sample PreparationStandard processing for serum orSame
plasma
Interpretation ofStandard CurveSame
Results
Reagent Storage2-8 °CSame
Temperature
Testing EnvironmentProfessional useSame
Precision (% CV)Within-run: ≤ 1.0%Within-run: ≤ 3.2%
Between-run: ≤ 5.2%Total: ≤ 3.8%
Total: ≤ 9.0%
LinearityAssay linear from 0.2 to 15.0 mg/LAssay linear from 0.2 - 160 mg/L
in High Sensitivity Application
InterferingNo interference from high levels ofNo interference from high levels
Substances/Specificityconjugated bilirubin, unconjugatedof bilirubin, hemoglobin, and
hemoglobin,bilirubin,triglycerides
triglycerides.humanserum
L-ascorbicalbumin,acid.
oxaloaceticacid,glutathione,
L-DOPA; noisoniazid,and -fromrheumatoidinterference
factor, human anti-mouse IgG, and
Comparative Testing vsEstablished MethodstransferrinFastPack® vs. Olympus AU2700N = 131Range of observations:0.21 to 15.0 mg/LDeming regression:Slope (95% CI): 0.98 (0.95-1.02)y (95% CI): -0.12 (-0.23 to 0.00)R (95% CI) = 0.99 (0.99-0.99)Sy\x = 0.629AU400 vs. AU640N = 109Range of observations:0.28-147.2 mg/LDeming regression:Slope = 1.025y = -0.022R = 0.999
ExpectedValues/ReferenceIntervals0.2 - 11.4 mg/LCardiac risk assessmentcategories:Low < 1 mg/LAverage 1.0 to 3.0 mg/LHigh > 3.0 mg/L

Comparison of new device to predicate devices

{8}------------------------------------------------

Differences between FastPack® and Olympus High-Sensitivity C-Reactive Protein

Assays
CHARACTERISTICQualigen FastPack®High-Sensitivity C-ReactiveProtein ImmunoassayOlympus America, Inc. CRPLatex reagentK051564
MethodologyThe FastPack® High Sensitivity C-Reactive Protein Immunoassay is a paramagnetic particle, chemiluminescent immunoassay employing specific monoclonal antibodies.The Olympus CRP Latex reagent is a turbidimetric assay employing rabbit antibodies coated on latex particles.
Assay principleChemiluminescenceTurbidimetry
Assay procedureAutomatedAutomated
Approximate assay time8 minutes3.5 minutes
Assay range0.2 - 15.0 mg/L in High Sensitivity Application0.2 - 160 mg/L (provides measurements both for “Normal Application” and “Highly Sensitive Application”)
TraceabilityTraceable to the ERM-DA474/IFCC reference which serves as the Primary Reference Material“…traceable to an external standard.” (K051564 510(k) Decision Summary)

{9}------------------------------------------------

CHARACTERISTICQualigen FastPack®High-Sensitivity C-Reactive ProteinImmunoassayOlympus America, Inc. CRPLatex reagentK051564
Intended Use/Indication for UseFor in-vitro diagnosticuse in calibratingFastPack® High-Sensitivity C-ReactiveProtein ImmunoassayOlympus System Reagent andcalibrators for the quantitativedetermination of C-ReactiveProtein on Olympus Analyzers.
Antigen used in calibratorsHuman CRPSame
MatrixLiquid human serummatrix containing apredetermined level ofhuman CRPLiquid human serum matrixcontaining predetermined levels ofhuman CRP
Storage temperature2-8 °CSame

Similarities between FastPack® and Olympus CRP Calibrators

Differences between FastPack® and Olympus CRP Calibrators

CHARACTERISTICQualigen FastPack®High-Sensitivity C-Reactive ProteinImmunoassayOlympus America, Inc. CRPLatex reagentK051564
Number of calibrators15 (additional calibrators providedat higher concentrations to enableCRP measurements in the “NormalApplication”)

{10}------------------------------------------------

CHARACTERISTICQualigen FastPack®High-Sensitivity C-ReactiveProtein ImmunoassayBio-Rad LaboratoriesLiquichek™ CardiacMarkers Plus ControlK050537
Intended Use/Indication for UseFor in-vitro diagnostic use tomonitor the precision andaccuracy of the FastPack®High-Sensitivity C-ReactiveProtein Immunoassay on theFastPack® Analyzer.Liquichek™ CardiacMarkers Plus Control areintended for use as qualitycontrol serum to monitorthe precision of laboratorytesting procedures listing inthe package insert.
Antigen used in controlsHuman CRPB-type Natriuretic Peptide,Creatine Kinase (Total), C-Reactive Protein,Homocysteine, Digitoxin,N-terminal pro-B-typeNatriuretic Peptide, CK-MB, Myoglobin, TroponinI, Troponin T
MatrixLiquid human serum matrixcontaining a predeterminedlevel of human CRPPrepared from humanserum with addedconstituents of human andanimal origin,preservatives, andstabilizers. The controls arein liquid form.
Storage temperature2-8 °C2-8 °C (Opened), or -20 to -70 °C (Unopened)

Similarities between FastPack® and Predicate Controls

Differences between FastPack® and Predicate Controls

CHARACTERISTICQualigen FastPack®High-Sensitivity C-ReactiveProtein ImmunoassayBio-Rad LaboratoriesLiquichek™ CardiacMarkers Plus ControlK050537
Number of levels23

{11}------------------------------------------------

CHARACTERISTICQualigen FastPack®High-Sensitivity C-ReactiveProtein ImmunoassayOrtho-ClinicalDiagnostics, Inc. VITROChemistry ProductshsCRP PerformanceVerifier, I, II, and IIIK041799
Intended Use/Indication for UseFor use in the quantitativeverification of calibration andassay range of the quantitativeFastPack® High-Sensitivity C-Reactive Protein Immunoassayon the FastPack® Analyzer.VITROS hsCRPPerformance Verifiers areassayed controls to monitorperformance of hsCRPReagent on VITROS 5,1 FSChemistry Systems
Antigen usedHuman CRPSame
Storage temperature2-8 °CSame
MatrixLow Verifier: HEPES bufferwith Bovine Serum Albumin(BSA) and Detergent(Cremophor A25)Mid and High Verifiers: Liquidhuman serum matrix containinga predetermined level of humanCRPA base matrix of humanplasma proteins to whichstabilizers and preservativehave been added.
Number of levels3Same

Similarities between FastPack® CRP Verifiers and Predicate Verifiers

Performance Summary

Precision

Precision was evaluated following the CLSI EP5-A2 guidance. Six serum patient samples with concentrations of ~0.5, 1.0, 2.0, 5.0, 7.5, and 12.5 mg/L CRP were tested in duplicate determinations in each of two runs per day on each of two FastPack® reagent lots, two FastPack® Analyzers per reagent lot, over a period of 20 days to yield 320 replicate determinations of each sample. A single FastPack® calibrator lot was utilized for all runs. Within-run, between-run, between-day, and residual components of variation were calculated as well as total imprecision using a general linear model. The table below presents the results:

{12}------------------------------------------------

Within-RunBetween-RunBetween-DayResidualTotal
SampleOverall MeanSD% CVSD% CVSD% CVSD% CVSD% CV
Serum Sample 10.380.0020.530.0143.680.0051.320.0297.740.0338.68
Serum Sample 21.000.0040.430.0383.800.0131.300.0777.710.0878.70
Serum Sample 32.060.0080.390.0713.450.0301.460.1627.890.1808.74
Serum Sample 45.010.0230.460.2064.110.0731.460.2885.740.3627.23
Serum Sample 57.670.0350.460.3144.090.1111.450.5967.780.6848.92
Serum Sample 612.540.0730.580.6495.180.2301.830.8436.721.0918.70

Limits of blank, detection, and quantitation

The Limit of Blank (LOB), the Limit of Detection (LOD), and the Limit of Quantitation (LOQ) of the FastPack® C-Reactive Protein Immunoassay were determined according to CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation. The following are the limits determined:

LOB = 0.005 mg/L LOD = 0.032 mg/L LOQ=0.063 mg/L

Linearity

Linearity was determined following CLSI EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: a Statistical Approach: Approved Guideline. A high patient sample was intermixed with a low sample to generate 8 concentration levels each tested in triplicate determinations. Linear results were compared to 2nd and 3th order polynomial fits against a pre-specified allowable error. The linearity range was found to extend from the LOQ (0.063 mg/L) to 15.0 mg/L. Samples recovering above the range may be diluted using Sample Diluent A.

Interferences

The following substances normally present in blood were tested and found not to interfere in the FastPack® High Sensitivity C-Reactive Protein Immunoassay at the noted concentrations:

Bilirubin (conjugated)Tested to 40 mg/dL
Bilirubin (unconjugated)Tested to 40 mg/dL
HemoglobinTested to 750 mg/dL
LipidsTested to 1000 mg/dL

{13}------------------------------------------------

Human serum albumin Transferrin Human IgG

Tested to 7.7 g/dL Tested to 567 mg/dL Tested to 2961 ug/mL

The following exogenous substances potentially present in blood were tested and found not to interfere in the FastPack® High Sensitivity C-Reactive Protein Immunoassay at the noted concentrations:

L-ascorbic acidTested to 200 mg/dL
Oxaloacetic acidTested to 300 μΜ
GlutathioneTested to 300 μΜ
IsoniazidTested to 300 μΜ
L-DOPATested to 300 μΜ

Cross-reactivity

Rheumatoid factor at up to 1000 U/mL and human anti-mouse IgG at up to 4 ug/mL do not cross-react in the FastPack® High Sensitivity C-Reactive Protein Immunoassay. Additionally, heterophile samples with activity up to 3641 ng/mL do not interfere in the assay.

Serum and plasma equivalence

Blood collections were obtained from 41 volunteers and processed in parallel to serum, EDTA plasma, and lithium-heparin plasma. Measurements in FastPack® High Sensitivity C-Reactive Protein Immunoassay were compared via Deming regression and indicated equivalence between the three matrices.

ParameterResult
N compared41
Range of observations, mg/LSerum: 0.33 – 14.72EDTA Plasma: 0.29 – 14.76
Absolute bias, mg/L-0.225
% Bias-6.1
Deming regression results
Slope0.94
y-intercept0.0
R0.984
R20.967

Serum versus EDTA plasma

{14}------------------------------------------------

ParameterResult
N compared41
Range of observations, mg/LSerum: 0.33 - 14.72Lithium-Heparin Plasma: 0.31 – 14.86
Absolute bias, mg/L0.002
% Bias0.5
Deming regression results
Slope1.00
y-intercept0.00
R0.993
R20.986

Serum versus lithium-heparin plasma

Expected Values/Reference Intervals

A reference interval study employing serum samples from 211 subjects representing 4 different geographic regions of the United States yielded the results in the table below. The non-parametric 2.5th - 97.5th percentile of 0.2 - 11.4 mg/L provides the reference interval determined from this study.

Observed values
Mean (SD)3.2 (3.1) mg/L
Median (Min - Max)1.9 (0.2 - 13.1) mg/L
2.5th - 97.5th percentile0.2 - 11.4 mg/L

Newborns with no evidence of infection have CRP concentrations of < 2 mg/L (Soldin OP, Bierbower LH, Choi JJ, et al. Serum iron, ferritin, transferrin, total iron binding capacity, hs-CRP, LDL cholesterol and magnesium in children; new reference intervals using the Dade Dimension Clinical Chemistry System; Clin Chim Acta 2004;342:211-7.).

Method Comparison

Human serum samples were tested with the FastPack® High Sensitivity C-Reactive Protein Immunoassay and the obtained results were compared to the predicate method. A total of 131 samples ranging from 0.21 to 15.0 mg/L were tested in both assays. The FastPack® High Sensitivity C-Reactive Protein Immunoassay correlated well with the predicate method with correlation coefficient (R) of 0.99, slope = 0.98, and y-intercept = -0.12 mg/L.

{15}------------------------------------------------

ParameterResult
Slope (95% CI)0.98 (0.95-1.02)
y-intercept (95% CI)-0.12 (-0.23 to 0.00)
R (95% CI)0.99 (0.99 - 0.99)

SUMMARY

The information provided in this pre-market notification indicates that the FastPack® High Sensitivity C-Reactive Protein Immunoassay is substantially equivalent to the stated predicate device. The information further indicates that the FastPack® High Sensitivity C-Reactive Protein Immunoassay is safe and effective for its stated intended use.

§ 866.5270 C-reactive protein immunological test system.

(a)
Identification. A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.(b)
Classification. Class II (performance standards).