K Number
K141416

Validate with FDA (Live)

Date Cleared
2014-10-09

(133 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

It is intended for measuring adult blood pressure and pulse rate with circumference ranging from 135 mm to 215 mm.

Device Description

Blood pressure monitor is based on pressure vibration method. Blood pressure cuff use the air pump to inflate, then the arteries are extruded by the cuff with pressure. Pressure sensor collects the pressure in the cuff, and then converts it to digital signal to the CPU. Then the software calculates the systolic and diastolic blood pressure and pulse rate.

AI/ML Overview

The provided document describes a 510(k) premarket notification for a Blood Pressure Monitor (K141416). The acceptance criteria and the study proving the device meets these criteria are detailed, particularly in the section on "Testing data and clinical study" and the comparison table.

Here's a breakdown of the requested information:

1. Table of acceptance criteria and the reported device performance

Acceptance Criteria (Measurement Accuracy)Reported Device Performance (K141416)Predicate Device (K123498) Performance
Static pressure: ±3mmHg (±0.4kPa)±3mmHg (±0.4kPa)±3mmHg (±0.4kPa)
Pulse accuracy: ±5%±5%±5%

2. Sample size used for the test set and the data provenance

The document states that "Clinical study has been evaluated according to ANSI/AAMI SP10 and ANSI/AAMI/ISO 81060-2."

  • ANSI/AAMI SP10 (Manual, electronic, or automated sphygmomanometers) typically requires a minimum of 85 subjects for clinical validation.
  • ANSI/AAMI/ISO 81060-2 (Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type) specifies a clinical investigation with a minimum of 85 participants meeting certain age, gender, and blood pressure distribution requirements.

Based on the reference to these standards, it is highly probable that the study included at least 85 subjects. However, the exact sample size is not explicitly stated in the provided text.

The data provenance (country of origin, retrospective/prospective) is not explicitly stated in the provided text. It is a clinical study, implying it was prospective, but this is not confirmed.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not explicitly state the number of experts or their qualifications for establishing ground truth. Compliance with ANSI/AAMI SP10 and ANSI/AAMI/ISO 81060-2 for clinical validation typically involves reference measurements performed by trained observers (auscultatory method), but the specific details are not provided here.

4. Adjudication method for the test set

The document does not explicitly state any adjudication method for the test set. Compliance with ANSI/AAMI SP10 and ANSI/AAMI/ISO 81060-2 for clinical validation typically involves comparing device measurements against reference measurements taken by trained observers, where differences are analyzed. However, the specific method of handling discrepancies or consensus building is not detailed here.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The device is a "Blood Pressure Monitor," which is an automated measurement device, not an AI-assisted diagnostic tool requiring human reader interpretation. Therefore, an MRMC study or evaluation of human reader improvement with AI assistance would not be relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the device is standalone. It is a "Blood Pressure Monitor" that measures adult blood pressure and pulse rate automatically using an oscillometric method. Its performance is evaluated directly against reference measurements in clinical studies, without human intervention in the measurement process itself.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth for the clinical study would be established through reference blood pressure measurements obtained by trained observers using the auscultatory method, as per the requirements of ANSI/AAMI SP10 and ANSI/AAMI/ISO 81060-2. This is a form of expert reference measurement.

8. The sample size for the training set

The document does not provide any information regarding a specific "training set" or its size. For a medical device like a blood pressure monitor, manufacturers typically calibrate and develop their algorithms using internal data, but this is usually not referred to as a "training set" in the context of regulatory submissions the way it would be for an AI/ML device. The regulatory focus is on the clinical validation (test set) against established standards.

9. How the ground truth for the training set was established

Since there is no explicit mention of a "training set" in the context of this regulatory submission, there is no information on how its ground truth might have been established. Any internal development and calibration would have used methods consistent with blood pressure measurement best practices.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines extending from the faces.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 9, 2014

Shenzhen Pump Medical System Co., Ltd. c/o Ms. Xie Qiongyu Deputy General Manager 2/f West, M-7 Sinosteel Building, Maqueling Estate Hi-tech Industrial Park Nanshan District, Shenzhen, Guangdong, 518057 CN

Re: K141416

Trade/Device Name: Blood Pressure Monitor, Model No. BF2200, BF2210, BF2202, BF2212, BF2203, BF2213, BF2204, BF2214, BF2205, BF2215, BF2206, BF2216, BF2207, BF2217 Regulation Number: 21 CFR 870.1130 Regulation Name: Non-Invasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: Not Dated Received: August 11, 2014

Dear Ms. Xie Qiongyu,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Melissa A. Torres -S

For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known): _ K141416

Device Name: _ Blood pressure monitor

Indications for Use:

It is intended for measuring adult blood pressure and pulse rate with circumference ranging from 135 mm to 215 mm..

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use V (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

K141416

Page 1 Of 6

510(k) Summary

1. Applicant

Applicant Name: SHENZHEN PUMP MEDICAL SYSTEM CO., LTD.

2/F West, M-7 Sinosteel Building, Maqueling Estate, Address:

Hi-Tech Industrial Park, Nanshan District,

Shenzhen 518057, China

Contact person:

Xie Qiongyu Name:

Phone numbers: 86-0755-26710795

Fax numbers: 86-0755-26012025

E-mail: xieqy@bpump.com.cn

Date Prepared: 2014-3-30

2. Device information

  • Trade name: Blood pressure monitor
  • Model No.: BF2200, BF2210, BF2201, BF2211, BF2202, BF2212, BF2203, ● BF2213, BF2204, BF2214, BF2205, BF2215, BF2206, BF2216, BF2207 and BF2217
  • Regulation Description: Noninvasive blood pressure measurement system. ●
  • Regulation Number: 21CFR 870.1130 ●
  • Product Code: DXN
  • Class: Il
  • Review Panel: Cardiovascular
  • Indications for Use: It is intended for measuring adult blood pressure and pulse ● rate with circumference ranging from 135 mm to 215 mm.
    1. Predicate Devices
  • Noninvasive blood pressure measurement system ●

{4}------------------------------------------------

  • K-number: K123498
  • Product Code: DXN
  • Intended User: Home user

● Patient Population: This device is intended for use on adults.

Indications for Use: ●

The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5 1/4 inches to 8 1/2 inches (13.5 cm to 21.5 cm).

The device detects the appearance of irreqular heartbeats during measurement and gives a warning signal with readings.

● Manufactured by

Omron Healthcare, Inc.

4. Description of the device

Blood pressure monitor is based on pressure vibration method. Blood pressure cuff use the air pump to inflate, then the arteries are extruded by the cuff with pressure. Pressure sensor collects the pressure in the cuff, and then converts it to digital signal to the CPU. Then the software calculates the systolic and diastolic blood pressure and pulse rate.

The blood pressure monitor BF2200, BF2210, BF2202, BF2212, BF2203, BF2213, BF2204, BF2214, BF2205, BF2215, BF2206, BF2216, BF2207 and BF2217 have the same basic principles, main function, performance and intended use, and they are consistent in product structure and material.

5. Testing data and clinical study

Laboratory testing was conducted to validate and verify that Blood pressure monitor met all requirements of related international standards, including electrical safety, EMC, biocompatibility, specification. Results of these tests demonstrate compliance to the requirements of the bellow consensus standards. Clinical study has been evaluated according to ANSI/AAMI SP10 and ANSI/AAMI/ISO 81060-2.

{5}------------------------------------------------

K141416 Page 3 of 6

Applied Standard:
Electrical Safety and performance requirements:IEC 60601-1AAMI performance standard ANSI/AAMI SP10EN 81060-2-30
Home-used medical equipment requirements and Environmental test:IEC 60601-1-11
Electromagnetic Compatibility Requirements:EN 60601-1-2
Biocompatibility Evaluation for NIBP CuffISO 10993-1, ISO 10993-5, ISO 10993-10
Clinical Evaluation:ANSI/AAMI SP10ANSI/AAMI/ISO 81060-2

Conclusion:

According to above information, comparing to the predicate, the subject devices do not constitute a new intended use, and used same the Measuring Principle, same electrical noninvasive blood pressure measurement technologic and same measurement site (measuring on wrist).

The subject devices have been evaluated to meet IEC 60601-1, IEC 60601-12, AAMI/ANSI SP10, EN 80601-2-30, ISO 10993-5, and ISO 10993-10. The related test reports and clinical study reports demonstrate and do not raise different questions of safety and effectiveness than the predicate.

As a result, the subject device is Substantially Equivalent (SE) to the predicate device which is US legally market device.

{6}------------------------------------------------

K141416

Page 4 of 6

ParameterName: Blood Pressure MonitorModel No.:BF2200, BF2210, BF2201, BF2211, BF2202, BF2212, BF2203,BF2213, BF2204, BF2214, BF2205, BF2215, BF2206, BF2216,BF2207 and BF2217Name: Noninvasive Blood Pressure Measurement System (K123498)Model No.: BP652N (HEM-6300-Z)
Intended useMeasurement of human blood pressure and pulse rate for adults.The device is a digital monitor intended for use in measuring bloodpressure and pulse rate in adult patient population with wristcircumference ranging from 5 1/4 inches to 8 1/2 inches (13.5 cm to 21.5cm).The device detects the appearance of irregular heartbeats duringmeasurement and gives a warning signal with readings.
PatientPopulationAdultAdult
Environment ofUseHomeHome
TechnologyOscillometric methodOscillometric method
Anatomical sitesWristWrist
MeasurementrangeBlood pressure: 0mmHg280mmHg(0kPa37.3kPa)Pulse: 40 pulse/min ~180 pulse/minBlood pressure: 0mmHg299mmHg(0kPa39.9kPa)Pulse: 40 pulse/min ~180 pulse/min
MeasurementaccuracyStatic pressure: ±3mmHg(±0.4kPa)Pulse: ±5%Static pressure: ±3mmHg(±0.4kPa)Pulse: ±5%

6. Comparison to Predicate Devices

{7}------------------------------------------------

ParameterName: Blood Pressure MonitorModel No.:BF2200, BF2210, BF2201, BF2211, BF2202, BF2212, BF2203,BF2213, BF2204, BF2214, BF2205, BF2215, BF2206, BF2216,BF2207 and BF2217Name: Noninvasive Blood Pressure Measurement System (K123498)Model No.: BP652N (HEM-6300-Z)
DisplayLCDLCD
Power source2AAA Alkaline batteries2AAA Alkaline batteries
OperatingenvironmentTemperature: +5 °C~+40 °CHumidity: ≤93%Temperature: +10°C~+40°CHumidity: 15%~85%
StorageenvironmentTemperature: -25 °C~+70 °CHumidity: 10%~95%Temperature: -20 °C~+60 °CHumidity: 10%~95%

The subject device is Substantially Equivalent (SE) to the predicate device which is US legally market device.

{8}------------------------------------------------

K141416 Page 6 of 6

7. Conclusion

As stated above, the Blood pressure monitors (Models: BF2200, BF2210, BF2201, BF2211, BF2202, BF2212, BF2203, BF2213, BF2204, BF2214, BF2205, BF2215, BF2206, BF2216, BF2207 and BF2217) are safe and effective, and comply with the appropriate medical device standards. And they are substantially equivalent to the earlier identified predicate devices.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).