K Number
K123498
Date Cleared
2013-03-21

(128 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5 1/4 inches to 8 1/2 inches (13.5 cm to 21.5 cm). The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.

Device Description

The device is an automatic non-invasive blood pressure system. The device is battery powered by 2 "AAA" batteries, there is no connection to external power. The device inflates a wrist cuff with an integral pump, then deflates the cuff via an electronically controllable valve. During inflation the cuff pressure is monitored and pulse waveform data is extracted. The extracted pulse waveform data is then analyzed by software which determines pulse rate, as well as systolic and diastolic pressure. The algorithm used to determine pulse rate, systolic and diastolic pressure is a minor modification of the predicate algorithm. The device has a memory function that automatically stores up to 100 of the latest measurements. It can also display an average of the last three values. The device also detects the appearance of irregular heartbeats during measurement.

AI/ML Overview

The Omron BP652N (HEM-6300-Z) blood pressure monitor underwent non-clinical testing to ensure its accuracy and compliance with relevant standards. Here's a breakdown of the acceptance criteria and the study details:

1. Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (from predicate K042505)Reported Device Performance (BP652N)
Accuracy of Pressure+/- 3 mmHg+/- 3 mmHg (or 2% of reading)
Accuracy of Pulse Rate+/- 5%+/- 5%

2. Sample Size and Data Provenance for Test Set

The provided document does not explicitly state the sample size used for the clinical accuracy test set, nor does it specify the country of origin or whether the data was retrospective or prospective. It only mentions that testing was done "to insure clinical accuracy of the device in accordance with ANSI/AAMI/ISO 81060-2." This standard typically outlines requirements for non-invasive sphygmomanometers, including patient selection criteria for clinical validation.

3. Number of Experts and Qualifications for Ground Truth Establishment (Test Set)

The document does not provide information on the number of experts used or their qualifications for establishing ground truth for the test set. For clinical accuracy studies of blood pressure monitors, ground truth is typically established by trained observers (often medical professionals) using a reference sphygmomanometer following a standardized protocol.

4. Adjudication Method for the Test Set

The document does not detail any adjudication method for the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done. The submission focuses on the standalone performance of the device and its substantial equivalence to a predicate device.

6. Standalone Performance Study (Algorithm Only)

Yes, a standalone performance study was done to assess the clinical accuracy of the device. This is implied by the statement "Testing to insure clinical accuracy of the device in accordance with ANSI/AAMI/ISO 81060-2." This standard specifies the clinical investigation of automated sphygmomanometers, which involves comparing the device's readings against a reference measurement. The results of this testing were that the device met the requirements for pressure accuracy (+/- 3 mmHg) and pulse rate accuracy (+/- 5%).

7. Type of Ground Truth Used

Based on the nature of blood pressure monitor accuracy testing and the reference to ANSI/AAMI/ISO 81060-2, the ground truth would have been established by expert reference measurements (e.g., using a mercury sphygmomanometer or validated electronic reference device) obtained by trained observers.

8. Sample Size for the Training Set

The document does not specify a separate training set. For blood pressure monitors, the "algorithm" is often based on established oscillometric principles, and its performance is validated rather than "trained" in the machine learning sense. The device's algorithm is described as a "minor modification of the predicate algorithm," suggesting an evolution rather than a de novo machine learning model requiring a distinct training set.

9. How Ground Truth for the Training Set was Established

As no explicit training set is mentioned in the context of machine learning, this information is not applicable. The device's algorithm would have been developed and refined based on engineering principles and data from prior validation efforts of similar oscillometric devices.

{0}------------------------------------------------

Premarket Notification 510(k) Section 5 - 510(k) Summary

Omron BP652N (HEM-6300-Z)

510(k) SummaryPage 1 of 511/12/2012
Omron Healthcare, Inc.
1925 West Field CourtTel - 847-247-5626
Lake Forest, IL 60045 USAFax- 847-680-5626
Official Contact:Renee Thornborough - Director QA/RA
Proprietary or Trade Name:Model BP652N(HEM-6300-Z)
Common/Usual Name:Noninvasive blood pressure measurement system.
Classification Name/Code:system.DXN - Noninvasive blood pressure measurement
21CFR 870.1130
Class II
Device:Model BP652N(HEM-6300-Z)

Predicate Device:

Omron – HEM-609N (HEM-6001-Z) - K042505

Device Description:

The device is an automatic non-invasive blood pressure system. The device is battery powered by 2 "AAA" batteries, there is no connection to external power. The device inflates a wrist cuff with an integral pump, then deflates the cuff via an electronically controllable valve. During inflation the cuff pressure is monitored and pulse waveform data is extracted. The extracted pulse waveform data is then analyzed by software which determines pulse rate, as well as systolic and diastolic pressure. The algorithm used to determine pulse rate, systolic and diastolic pressure is a minor modification of the predicate algorithm.

The device has a memory function that automatically stores up to 100 of the latest measurements. It can also display an average of the last three values

The device also detects the appearance of irregular heartbeats during measurement.

PDF Page 15 of 810

Intended User Home user

Patient Population

This device is intended for use on adults.

MAR 2 1 2013

{1}------------------------------------------------

Premarket Notification 510(k) Section 5 - 510(k) Summary

Omron BP652N (HEM-6300-Z)

510(k) Summary Page 2 of 5 11/12/2012

Indications for Use:

The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5 1/4 inches to 8 1/2 inches (13.5 cm to 21.5 cm).

The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.

Environment of Use: Home

Contraindications

There are no known contraindications.

Predicate Device Comparison

The BP652N(HEM-6300-Z) was compared to the predicate HEM-609N (K042505) as in the device comparison table below.

PDF Page 16 of 810

{2}------------------------------------------------

remarket Notification 510(

·

mron BP652N (HEM-6300-2

·

.

: :

Device ComparisonOmron BP652N (HEM-6300-Z)Omron HEM-609N 510(k) K042505Comment
Indications for UseThe device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5 1/4 inches to 8 1/2 inches (13.5 cm to 21.5 cm).The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5 1/4 inches to 8 1/2 inches (13.5 cm to 21.5 cm).The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.Identical
Patient PopulationAdultAdultIdentical
Environment of UseHomeHomeIdentical
PrescriptiveOTCNoIdentical
Patient ConnectionYes via cuffYes via cuffIdentical
TechnologyOscillometricOscillometricIdentical
Measurement rangePressure: 0-299 mmHgPulse rate: 40 to 180 bpmPressure: 0-299 mmHgPulse rate: 40 to 180 bpmIdentical
Pressure sensorPiezo resistance sensorSilicone capacitive sensor rSimilar
Accuracy or pressure indicator$+/-3 mmHg$ or 2% of reading$+/-3 mmHg$Similar
Accuracy Pulse Rate$+/-5%$$+/-5%$Identical
Inflation MethodPiezo-electric pumpDC pumpSimilar
Deflation MethodInternal valveInternal valveIdentical
Display TypeLCDLCDIdentical
Irregular pulse detectionYesYesIdentical
Power SourceAAA batteriesAAA batteriesIdentical
Operating ConditionsTemperature: 10° to 40° CHumidity: 15 to 85% RHTemperature: 10° to 40° CHumidity: 30 to 85% RHSimilar
Storage ConditionsTemperature: -20° to 60° CHumidity: 10 to 95% RHTemperature: -20° to +60°CHumidity: 10 to 95% RHIdentical
Dimensions89(W) x 61(D) x 13(H) mm70(W) x 54(D) x 37(H) mmSimilar
Weight80g110gSimilar

:

·

.

·

.

.

Page 5.3

PDF Page 17 of 81

K123498

. .

{3}------------------------------------------------

Omron BP652N (HEM-6300-Z)

510(k) Summary Page 4 of 5 11/12/2012

Differences Between Other Legally Marketed Predicate Devices

The Omron BP652N is viewed as substantially equivalent to the predicate device because: The BP652N uses the exact same technology and has identical indications for use. The differences that exist between the devices are insignificant in the terms of safety or effectiveness.

Indications -

The indications for use are identical.

Prescriptive - The BP652N and predicate are both OTC.

Design and Technology - The BP652N has equivalent design and features as the predicate and has the identical technology to the predicate.

Performance and Specifications - The BP652N has equivalent specifications of performance as the predicate.

Compliance with standards - The BP652N and predicate device declare compliance with SP10, IEC 60601-1 and IEC 60601-1-2. 100 - 100 -

Materials -

The patient contacting materials of the cuffs has been tested in accordance with ISO 10993-1 and FDA Guidance. The tests included Cytotoxicity, Sensitization, and Intracutaneous Reactivity.

Patient Population -

The BP652N and predicate are indicated for adults

Non-Clinical Testing Summary:

We have performed bench tests and found that the BP652N met all requirements specifications and standards requirements and were found to be equivalent in comparison to the predicate. Testing includes the following:

  • . Verification Testing
  • Testing for compliance to IEC 60601-1
  • Testing for compliance to IEC 60601-1-2
  • Testing for compliance to AAMI SP10
  • Comparative Testing to the Predicate

PDF Page 18 of 810

{4}------------------------------------------------

Premarket Notification 510(k) Section 5 - 510(k) Summary

Omron BP652N (HEM-6300-Z)

510(k) Summary Page 5 of 5 11/12/2012

Testing to insure clinical accuracy of the device in accordance with ANSI/AAMI/ISO 81060-2.

Substantial Equivalence Conclusion

Omrop maintains that the RD652N is substantially equivalent to the predicate

Omron maintains that the BP652N is substantially equivalent to the predicate HEM-609N (K042505) in indications for use, patient population, and environment for use, technology characteristics, specifications / performance and compliance with international standards

PDF Page 19 of 810

.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white and appears to be a scanned image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

March 21, 2013

Omron Healthcare, Inc. c/o Mr. Paul Dryden, President ProMedic, Inc. 24301 Woodsage Drive Bonita Springs, FL 34134

Re: K123498

Trade/Device Names: BP652N (HEM-6300-Z) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: February 14, 2013 Received: February 15, 2013

Dear Mr. Dryden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2 - Mr. Paul Dryden, President

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Owen Pyraris -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

K123498

Indications for Use Statement

Page 1 of 1

510(k) Number:

Device Name:

Omron BP652N (HEM-6300-Z)

(To be assigned)

Indications for Use:

The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5 1/4 inches to 8 1/2 inches (13.5 cm to 21.5 cm).

The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.

Prescription Use (Part 21 CFR 801 Subpart D)

or

Over-the-counter use _XX (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Owen P. Faris -
2013.03.21
08:54:03 -04'00

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).