(33 days)
The Delta TT Acetabular System is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- Non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis and hip dysplasia;
- Rheumatoid arthritis;
- Post-traumatic arthritis,
- Correction of functional deformity;
- Fractures, dislocation of the hip and unsuccessful cup arthroplasty;
- Revisions in cases of good remaining bone stock.
The Delta TT cup is intended for cementless use.
The Delta TT Acetabular System consists of a Delta TT cup, a liner and a modular metal femoral head. Bone screws can also be used as additional fixation for the cup.
The Delta TT cup is manufactured using an EBM (Electron Beam Melting) process with titanium alloy powder (Ti6Al4V, ASTM F1472 - ISO 5832-3). The Delta TT cup consists of a non porous bulk interior surface and a Trabecular Titanium structure on the external surface.
The Delta TT cup has an external hemi-spherical shape. A polar threaded hole is used for introduction of the cup and mates with the polar peg of the liner when the two components are assembled. Three (3) threaded holes in lateral positions allow additional fixation of the cup using bone screws: polyethylene plugs are inserted into these holes when bone screws are not required. The Delta TT cup is designed to be coupled with ultra high molecular weight polyethylene (UHMWPE) liners by means of a taper coupling.
Delta TT cups are available with external diameters of 44, 46 and 48 mm (for Small liners), 50 and 52 mm (for Medium liners) and 54, 56, 58, 60, 62 and 64 mm (for Large liners). The new sizes of the Delta TT Cups, the Delta TT Jumbo Cups, are Delta TT Cups with external diameters of 66, 68, 70, 72, 74 and 76mm and they are all intended to be used with Large Delta TT cup liners that were cleared in K112898. The liners are manufactured from standard UHMWPE (ASTM F648 - ISO 5834-2) or from cross-linked UHMWPE (X-Lima).
Liners are coupled with the Delta TT cup by means of a taper coupling. Two features are intended to give stability to the coupling, a peripheral ring and a polar peg. The peripheral ring is manufactured from Ti6A14V (ASTM F1472 - ISO 5832-3) and surrounds the taper of the liner circumferentially, enhancing the rotational stability of the coupling. The polar peg fits into the polar hole of the cup, increasing the lever-out stability of the top of the peg is resurfaced by a Ti6Al4V (ASTM F1472 - ISO 5832-3) plug to avoid direct contact between polyethylene and bone.
The polyethylene liners are available in two (2) versions: neutral and protruded. The protruded design provides greater coverage of the femoral head and is intended to reduce the risk of dislocation. For both the designs, the following sizes are available:
- Cross-linked UHMWPE liners: Small (for femoral heads Ø 28 mm), Medium (for femoral heads Ø 28 and 32 mm) and Large (for femoral heads Ø 28, 32 and 36 mm).
- Standard UHMWPE liners: Small, Medium and Large liners for femoral heads Ø 28 mm.
Bone screws are manufactured from Ti6Al4V (ASTM F1472 -- ISO 5832-3) and can be used to provide additional initial stability to the cup. Bone screws have a diameter of 6.5 mm and are available in lengths of 20, 25, 30, 35, 40, 45, 50, 55 and 60 mm.
This document describes the regulatory submission for design modifications to a hip replacement system, specifically the Delta TT Acetabular System. As such, it does not contain information about software algorithms, AI, or clinical studies that gather performance metrics like sensitivity/specificity or human reader improvement.
The information provided focuses on demonstrating substantial equivalence to previously cleared devices through non-clinical mechanical testing.
Here's a breakdown based on the categories requested, with explanations for why certain information is not present in the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Test | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Wear test | "Applicable" to previously cleared device (K112898) | Results of tests performed on Delta TT Cups cleared via K112898 are applicable. |
| Push-out, Lever-out, Axial torque | "Applicable" to previously cleared device (K112898) | Results of tests performed on Delta TT Cups cleared via K112898 are applicable. |
| Cup deformation | "Applicable" to previously cleared device (K112898) | Results of tests performed on Delta TT Cups cleared via K112898 are applicable. |
| Fatigue/Fretting test | "Applicable" to previously cleared device (K112898) | Results of tests performed on Delta TT Cups cleared via K112898 are applicable. |
Explanation: The acceptance criteria are not explicitly stated as numerical values. Instead, the document relies on the concept of "substantial equivalence" to predicate devices. The performance of the new components (Delta TT Jumbo Cups) is deemed acceptable because they do not represent a "worst case for mechanical testing" compared to the previously cleared Delta TT Cups (K112898). Therefore, the existing test results from K112898 are considered sufficient to demonstrate acceptable performance for the new sizes, implying they meet the same criteria as the original device.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not specified for the mechanical tests. For mechanical testing of medical devices, sample sizes are typically determined by engineering and statistical standards for the specific test (e.g., ISO or ASTM standards) and depend on the variability of the material and design. However, the document does not disclose the exact number of units tested.
- Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). The tests were conducted to support a regulatory submission for a device manufactured by Limacorporate S.p.A. in Italy, so it's likely the testing was conducted either internally by the manufacturer or by a contracted laboratory. The testing described is non-clinical (i.e., not involving human patients or clinical data).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Not applicable. This document describes mechanical testing of a physical medical device, not a diagnostic or prognostic algorithm that requires expert "ground truth" for interpretation of data.
4. Adjudication Method for the Test Set
- Not applicable. Adjudication methods like 2+1 or 3+1 are used in studies where human interpretation of data (e.g., medical images) needs consensus. This document pertains to mechanical engineering tests.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No. This is not a study involving human readers or AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- No. This is not a study involving an algorithm.
7. The Type of Ground Truth Used
- Not applicable. For mechanical testing, the "ground truth" is defined by the physical behavior of the device under specific load conditions, measured by instruments, and compared against engineering specifications or results from predicate devices. There is no subjective human interpretation involved in establishing ground truth for these types of tests.
8. The Sample Size for the Training Set
- Not applicable. This document describes a physical medical device, not a machine learning model that requires a training set.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. As above, no training set or ground truth in the context of machine learning.
Summary of the Study Proving Device Meets Acceptance Criteria:
The "study" in this context is a series of non-clinical mechanical tests detailed in the K112898 submission for the original Delta TT Acetabular System. The current submission (K141395) argues that because the new "Jumbo" cup sizes (66-76mm) do not represent a "worst case for mechanical testing" compared to the existing cleared sizes (44-64mm), the results from the K112898 tests are directly applicable and demonstrate substantial equivalence.
The tests conducted for K112898 (and deemed applicable for K141395) include:
- Wear test
- Push-out test
- Lever-out test
- Axial torque test
- Cup deformation test
- Fatigue/Fretting test
The underlying premise is that if the smaller, original cups performed acceptably under these mechanical tests, and the larger "Jumbo" cups are structurally similar and not more highly stressed, then they will also perform acceptably. The acceptance criteria for these tests would have been established based on recognized industry standards (e.g., ASTM, ISO) for hip prostheses at the time of the K112898 submission. The document relies on the previous clearance to establish that these criteria were met.
Clinical Testing: The document explicitly states, "Clinical testing was not necessary to demonstrate substantial equivalence of the Delta TT Jumbo Cups to the Delta TT Acetabular System." This further emphasizes that the basis for clearance was non-clinical mechanical performance.
{0}------------------------------------------------
JUN 3 0 2014
510(k) Summary
Date: June 25, 2014
Manufacturer: Limacorporate S.p.A. Via Nazionale, 52 33038 - Villanova di San Daniele Udine - Italy
U.S. Contact Person: Dr. Stephen J. Peoples Principal Consultant Phone: 260-645-0327 FAX: +39 0432945512
| Product | CommonName | ProductCode | Regulation and Classification Name |
|---|---|---|---|
| Delta TT AcetabularSystem | Total HipProsthesis | LPH, MBL,JDI, LZO | Hip joint metal/polymer/metal semi-constrained porous-coated uncementedprosthesis per 21 CFR 888.3358 |
Description:
The Delta TT Acetabular System consists of a Delta TT cup, a liner and a modular metal femoral head. Bone screws can also be used as additional fixation for the cup.
The Delta TT cup is manufactured using an EBM (Electron Beam Melting) process with titanium alloy powder (Ti6Al4V, ASTM F1472 - ISO 5832-3). The Delta TT cup consists of a non porous bulk interior surface and a Trabecular Titanium structure on the external surface.
The Delta TT cup has an external hemi-spherical shape. A polar threaded hole is used for introduction of the cup and mates with the polar peg of the liner when the two components are assembled. Three (3) threaded holes in lateral positions allow additional fixation of the cup using bone screws: polyethylene plugs are inserted into these holes when bone screws are not required. The Delta TT cup is designed to be coupled with ultra high molecular weight polyethylene (UHMWPE) liners by means of a taper coupling.
Delta TT cups are available with external diameters of 44, 46 and 48 mm (for Small liners), 50 and 52 mm (for Medium liners) and 54, 56, 58, 60, 62 and 64 mm (for Large liners). The new sizes of the Delta TT Cups, the Delta TT Jumbo Cups, are Delta TT Cups with external diameters of 66, 68, 70, 72, 74 and 76mm and they are all intended to be used with Large Delta TT cup liners that were cleared in K112898. The liners are manufactured from standard UHMWPE (ASTM F648 - ISO 5834-2) or from cross-linked UHMWPE (X-Lima).
Liners are coupled with the Delta TT cup by means of a taper coupling. Two features are intended to give stability to the coupling, a peripheral ring and a polar peg. The peripheral ring is manufactured from Ti6A14V (ASTM F1472 - ISO 5832-3) and surrounds the taper of the liner circumferentially, enhancing the rotational stability of the coupling. The polar peg fits into the polar hole of the cup, increasing the lever-out stability of the top of the peg is resurfaced by a Ti6Al4V (ASTM F1472 - ISO 5832-3) plug to avoid direct contact between polyethylene and bone.
{1}------------------------------------------------
The polyethylene liners are available in two (2) versions: neutral and protruded. The protruded design provides greater coverage of the femoral head and is intended to reduce the risk of dislocation. For both the designs, the following sizes are available:
- . Cross-linked UHMWPE liners: Small (for femoral heads Ø 28 mm), Medium (for femoral heads Ø 28 and 32 mm) and Large (for femoral heads Ø 28, 32 and 36 mm).
- . Standard UHMWPE liners: Small, Medium and Large liners for femoral heads Ø 28 mm.
Bone screws are manufactured from Ti6Al4V (ASTM F1472 -- ISO 5832-3) and can be used to provide additional initial stability to the cup. Bone screws have a diameter of 6.5 mm and are available in lengths of 20, 25, 30, 35, 40, 45, 50, 55 and 60 mm.
Intended Use/Indications:
The Delta TT Acetabular System is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
-
. Non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis and hip dysplasia;
く -
. Rheumatoid arthritis;
-
. Post-traumatic arthritis,
-
. Correction of functional deformity;
-
Fractures, dislocation of the hip and unsuccessful cup arthroplasty;
-
. Revisions in cases of good remaining bone stock.
The Delta TT cup is intended for cementless use.
Predicate Devices:
- . Delta TT Acetabular System (Limacorporate, K112898)
- . Continuum Acetabular System (Zimmer, K091508)
Comparable Features to Predicate Device(s):
The new sizes of the Delta TT Acetabular Cups are identical to the intended use and indications for use, the design and the materials of the Delta TT Acetabular Cups cleared via K112898. The subject devices are also of similar design to the Zimmer Continuum Acetabular shells and are available in outer diameter sizes that are identical to the outer diameter sizes of the largest Zimmer Continuum Acetabular System shells.
Non-Clinical Testing:
The new sizes of the Delta TT Cups do not represent worst case for mechanical testing. Therefore, the results of the tests performed on the Delta TT Cups cleared via K112898 are applicable also to Delta TT Jumbo Cups object of this submission.
With the Delta TT Acetabular System submission the following tests were provided:
- . Wear test
- . Push-out, Lever-out, Axial torque
{2}------------------------------------------------
:
:
- Cup deformation .
- . Fatigue/Fretting test
:
. . . . . . . . .
..
:
.
Clinical Testing: Clinical testing was not necessary to demonstrate substantial equivalence of the Delta TT Jumbo Cups to the Delta TT Acetabular System.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is an abstract symbol that resembles an eagle or bird.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 30, 2014
Limacorporate S.p.A. % Stephen Peoples, VMD President Peoples and Associates Consulting LLC 5010 Lodge Pole Lane Fort Wayne, Indiana 46814
Re: K141395
Trade/Device Name: Delta TT Acetabular System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, MBL, JDI, LZO Dated: June 4, 2014 Received: June 5, 2014
Dear Dr. Peoples:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Dr. Stephen Peoples
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Lori A. Wiggins
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K141395 Pg.1/1
Indications for Use Statement
K141395 510(k) Number (if known):
Delta TT Acetabular System Indications for Use
The Delta TT Acetabular System is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- Non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis and � hip dysplasia;
- . Rheumatoid arthritis;
- Post-traumatic arthritis, .
- Correction of functional deformity; .
- Fractures, dislocation of the hip and unsuccessful cup arthroplasty; .
- Revisions in cases of good remaining bone stock. .
The Delta TT cup is intended for cementless use.
Prescription Use X (Part 21 CFR 801 Subpart D) ·
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth | Frank -S
Division of Orthopedic Devices
Page 1 of 1
Special 510(k) - Device Modification: Delta TT Acctabular System May 19, 2014
§ 888.3358 Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis.
(a)
Identification. A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.(b)
Classification. Class II.