K Number
K141288

Validate with FDA (Live)

Date Cleared
2014-08-22

(98 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Voyant electrosurgical generator is indicated for use with Voyant devices in open and laparoscopic surgical procedures where the electrosurgical ligation of vessels or tissue bundles is desired.

The Voyant 5mm Sealer-Divider device is a bipolar, electrosurgical device indicated for use with the Voyant electrosurgical generator in laparoscopic procedures where the ligation and division of vessels and tissue bundles is desired.

The device can seal and divide vessels up to and including 7mm in diameter and tissue bundles that can be captured in the jaws of the device.

The device has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.

Device Description

The Applied Medical Voyant Electrosurgical Generator (ESG) is a tabletop radiofrequency (RF) power supply designed for use in electrosurgery. It operates outside the sterile field and is equipped with receptacles for Applied Medical bipolar devices. The ESG enclosure is constructed of various metals and polymers and houses all electrical hardware and software components. The front panel features an LCD and backlit buttons for the navigation, adjustment, and selection of ESG and device settings.

The Applied Medical Voyant 5mm Laparoscopic Tissue Sealer-Divider instrument is designed for use with the Voyant ESG. This device is an advanced bipolar instrument that uses RF energy, provided by the generator, to seal vessels up to and including 7mm in diameter. The device may also be used to seal tissue bundles that can be captured in the device jaws. The device features a mechanical, useractuated blade for the division of sealed tissue.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the Voyant® Electrosurgical Generator and Voyant® 5mm Sealer-Divider System. It outlines the device's indications for use, technological characteristics, and performance testing conducted to demonstrate substantial equivalence to predicate devices. However, the document does not contain a table of acceptance criteria and reported device performance in a way that directly answers the request. It states that testing was performed but does not list specific numerical criteria or outcomes in a clear tabular format for direct comparison.

Therefore, for the information requested in your prompt, I can only extract what is available and highlight what is missing.

Here's an attempt to answer based on the provided text, noting the limitations:

1. A table of acceptance criteria and the reported device performance

The document states its Indications for Use which implicitly sets a performance expectation: "The device can seal and divide vessels up to and including 7mm in diameter and tissue bundles that can be captured in the jaws of the device."

However, specific numerical acceptance criteria (e.g., minimum burst pressure, maximum thermal spread) and the precise reported performance values against these criteria are not explicitly presented in a table. The document generally concludes that the system is "substantially equivalent in performance to the predicate system with respect to vessel sealing performance and local tissue effects."

Acceptance Criteria (Implied from Indications for Use and Testing Goals)Reported Device Performance (General Statement in Document)
Ability to seal and divide vessels up to 7mm in diameter."substantially equivalent in performance to the predicate system with respect to vessel sealing performance"
Ability to seal and divide tissue bundles that can be captured in jaws."substantially equivalent in performance to the predicate system with respect to vessel sealing performance"
Acceptable local tissue effects (thermal damage)."substantially equivalent in performance to the predicate system with respect to... local tissue effects."
Electrical, mechanical, and functional capabilities.Tested in accordance with IEC 60601-1-2 and IEC 60601-2-2. Basic mechanical and functional capabilities were tested.
System safety and efficacy.Evaluated in preclinical bench and animal studies; chronic hemostasis evaluated.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Test Set Sample Size: The document mentions "preclinical bench (ex vivo porcine vessels/tissue)" and "acute animal study (in vivo canine model)," and "a chronic study (in vivo canine model)." No specific sample sizes (e.g., number of porcine vessels, number of canine subjects) are provided.
  • Data Provenance:
    • Country of Origin: Not specified.
    • Retrospective or Prospective: The preclinical and animal studies are prospective in nature, as they involve testing the device under controlled conditions.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This section is not applicable as the studies are device performance studies on animal/ex vivo models, not studies requiring expert interpretation of diagnostic images or clinical outcomes in humans to establish ground truth. The "ground truth" for these types of studies would be objective measurements (e.g., burst pressure measurements, histological assessment of thermal damage) performed by trained technicians or researchers. The document does not specify the number or qualifications of individuals performing these measurements.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This section is not applicable for the type of device performance studies described. Adjudication methods are typically used in clinical or diagnostic studies to resolve discrepancies in expert interpretation of data (e.g., radiology reads).

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The document describes the performance of an electrosurgical device, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was conducted or mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. The device is an electrosurgical instrument used by a surgeon, not a standalone algorithm. Its performance inherently involves human-in-the-loop operation.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the performance studies appears to be based on:

  • Objective physical measurements: Energy output, mechanical strength, durability, electrical safety, burst pressures of sealed vessels.
  • Histopathological assessment: Evaluation of thermal damage in tissue, and chronic hemostasis in the animal studies.
  • Comparative performance: Against a legally marketed predicate device to establish "substantial equivalence."

8. The sample size for the training set

This is not applicable. The document describes a physical medical device (electrosurgical generator and sealer-divider), not a machine learning or AI model that requires a "training set."

9. How the ground truth for the training set was established

This is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 22, 2014

Applied Medical Resources Ms. Jessica Cho Regulatory Affairs Specialist 22872 Avenida Empresa Rancho Santa Margarita, California 92688

Re: K141288 Trade/Device Name: Voyant® Electrosurgical Generator and Voyant® 5mm Sealer-Divider System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: July 22, 2014 Received: July 23, 2014

Dear Ms. Cho:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

David Krause -S

  • for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{2}------------------------------------------------

INDICATIONS FOR USE

Device Name: Voyant® Electrosurgical Generator and Voyant® 5mm Sealer-Divider System

Indications for Use:

The Voyant electrosurgical generator is indicated for use with Voyant devices in open and laparoscopic surgical procedures where the electrosurgical ligation of vessels or tissue bundles is desired.

The Voyant 5mm Sealer-Divider device is a bipolar, electrosurgical device indicated for use with the Voyant electrosurgical generator in laparoscopic procedures where the ligation and division of vessels and tissue bundles is desired.

The device can seal and divide vessels up to and including 7mm in diameter and tissue bundles that can be captured in the jaws of the device.

The device has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

Concurrence of CDRH. Office of Device Evaluation (ODE)

{3}------------------------------------------------

510(k) Summary

510(k) Submitter:Applied Medical Resources Corp.22872 Avenida EmpresaRancho Santa Margarita, CA 92688
(949) 713 – 8000
Contact Person:Jessica ChoRegulatory Affairs SpecialistApplied Medical ResourcesTel: (949) 713 – 7958Fax: (949) 713 – 8205Email: jcho@appliedmedical.com
Preparation Date:May 14, 2014
Trade Name:Voyant® Electrosurgical Generator and Voyant® 5mm Laparoscopic Tissue Sealer-Divider Device
Common Name:Electrosurgical GeneratorBipolar Electrosurgical Sealer-Divider
Classification:Electrosurgical Cutting and Coagulation Device and Accessories, General and PlasticSurgery Devices, 21 CFR 878.4400, Product Code GEI
Predicate Devices:Trade Name: LigaSure™ 5mm Blunt Tip Laparoscopic Sealer-Divider510(k) #: K092879Product Code: GEI
Trade Name: ForceTriad™ Electrosurgical Generator510(k) #: K070162Product Code: GEI
Device Description:The Applied Medical Voyant Electrosurgical Generator (ESG) is a tabletopradiofrequency (RF) power supply designed for use in electrosurgery. It operatesoutside the sterile field and is equipped with receptacles for Applied Medicalbipolar devices. The ESG enclosure is constructed of various metals and polymersand houses all electrical hardware and software components. The front panelfeatures an LCD and backlit buttons for the navigation, adjustment, and selection

The Applied Medical Voyant 5mm Laparoscopic Tissue Sealer-Divider instrument is designed for use with the Voyant ESG. This device is an advanced bipolar instrument that uses RF energy, provided by the generator, to seal vessels up to and including 7mm in diameter. The device may also be used to seal tissue bundles that can be captured in the device jaws. The device features a mechanical, useractuated blade for the division of sealed tissue.

of ESG and device settings.

{4}------------------------------------------------

Intended Use:

The Voyant electrosurgical generator is indicated for use with Voyant devices in open and laparoscopic surgical procedures where the electrosurgical ligation of vessels or tissue bundles is desired.

The Voyant 5mm Sealer-Divider device is a bipolar, electrosurgical device indicated for use with the Voyant electrosurgical generator in laparoscopic procedures where the ligation and division of vessels and tissue bundles is desired.

The device can seal and divide vessels up to and including 7mm in diameter and tissue bundles that can be captured in the jaws of the device.

The device has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.

Summary of Technological Characteristics between Subject and Predicate:

The subject and predicate generators are technologically similar in that both are electrically isolated, microprocessor controlled, tabletop RF power supplies designed for use in electrosurgery. The generators are designed to provide RF energy to connected bipolar instruments with fuse energy modes. Both generators are equipped with a graphical user interface that allows the user to adjust the energy output settings. The generators are dissimilar in that the predicate generator supports electrosurgical cutting and coagulation as well as monopolar and standard bipolar instruments, where the subject device does not.

The subject and predicate instruments are technologically similar in size, configuration, and operation. They are designed to deliver RF energy to vessels up to 7mm and tissue captured between the jaws of the device for tissue fusion. Both instruments feature pistol-grip style handles with a trigger for jaw closure and button on the back of the handle for energy activation. The instruments are equipped with a mechanical, user-actuated blade for the division of sealed tissue. The instruments are dissimilar in that the subject jaws are capable of uninterrupted 360° rotation where the predicate jaws have 159° of rotation.

Discussion of Performance Testing Submitted:

Testing was performed on the subject Voyant® system to demonstrate electrical, mechanical and functional capabilities in accordance with IEC 60601-1-2 and IEC 60601-2-2. The subject system was also tested side-by-side against the predicate system to evaluate substantial equivalence with respect to performance in a laboratory setting as well as an animal model. Side-by-side testing focused on the seal quality evaluation and the local tissue effects of the applied RF energy in and surrounding the seal.

Non-clinical: Basic mechanical and functional capabilities were tested for the subject electrosurgical generator and subject sealer-divider. The following is a summary of components and testing performed:

  • Electrosurgical Generator
    • o Energy output

{5}------------------------------------------------

  • Static and Dynamic Load testing o
  • Sealer-Divider Device
    • Mechanical strength testing per IEC 60601-1 о
    • Durability testing о
    • Destructive testing per IEC 60601-2-2 o
  • . Integrated System
    • Electrical safety testing per IEC 60601-1 o
    • o Safety systems testing

Preclinical: Preclinical bench (ex vivo porcine vessels/tissue) and animal (in vivo canine model) studies were performed to evaluate system safety and efficacy and to demonstrate that the subject device performance is substantially equivalent to the predicate device; namely:

  • . Preclinical (ex vivo)
    • o Seal evaluation
    • o Thermal damage
    • o Burst pressures
  • . Acute animal study (in vivo)
    • o Seal evaluation
    • o Thermal damage

In addition, a chronic study (in vivo canine model) was conducted using the subject system to evaluate seal quality and chronic hemostasis.

Clinical: This premarket notification does not rely on human clinical trial data to demonstrate substantial equivalence

Conclusions Drawn from Testing:

The Applied Medical Voyant system is substantially equivalent in performance to the predicate system with respect to vessel sealing performance and local tissue effects.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.