K Number
K141274
Date Cleared
2015-02-13

(273 days)

Product Code
Regulation Number
866.5660
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The QUANTA Flash® ß2GP1-Domain1 is an in vitro chemiluminescent immunoassay (CIA) for the semi-quantitative determination of IgG autoantibodies to ß2GP1-Domain1 in human serum. The presence of anti-B2GP1- Domain1 autoantibodies is used in conjunction with clinical and other laboratory findings as an aid in the diagnosis of antiphospholipid syndrome. The OUANTA Flash® B2GP1-Domain1 is not intended to replace assays for antibodies against the whole ß2GP1 molecule. Testing for antibodies to the whole is required according to the classification criteria for antiphospholipid syndrome.

The QUANTA Flash ß2GP1-Domain1 Controls are intended for quality control purposes of the QUANTA Flash ß2GP1-Domain1 chemiluminescent immunoassay (CIA) kit.

The HemosIL® AcuStar Anti-S2GPI Domain 1 is an in vitro chemiluminescent immunoassay (CIA) for the semiquantitative determination of IgG autoantibodies to B2GPI Domain 1 in human serum. The presence of anti-R2GPI Domain 1 autoantibodies is used in conjunction with clinical and other laboratory findings as an aid in the diagnosis of antiphospholipid syndrome. The HemosIL® AcuStar Anti-ß2GPI Domain 1 is not intended to replace assays for antibodies against the whole ß2GPI molecule. Testing for antibodies to the whole is required according to the classification criteria for antiphospholipid syndrome.

The HemosIL AcuStar Anti-B2GPI Domain 1 Controls are intended for quality control purposes of the HemosIL AcuStar Anti-ß2GPI Domain 1 chemiluminescent immunoassay (CIA) kit.

Device Description

The QUANTA Flash® ß2GP1-Domain1 assay is designed to run on the BIO-FLASH® instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash® ß2GP1-Domain1 assay utilizes a reagent cartridge format, which is compatible with the BIO-FLASH instrument.

The assays included in this submission, the QUANTA Flash ß2GP1-Domain1 marketed by Inova Diagnostics Inc. (9900 Old Grove Road, San Diego, CA 92131) and HemoslL AcuStar Anti-ß-GPI Domain 1 marketed by Instrumentation Laboratory (180 Hartwell Road Bedford, MA 01730), are equivalent assays. Therefore all data stated hereafter and referred to as: QUANTA Flash §2GP1-Domain1 data is equivalently also valid for HemosIL "AcuStar Anti-ß₂GPI Domain 1.

Recombinant ß2GP1-Domain1 protein is coated onto paramagnetic beads, which are stored lyophilized in the reagent cartridge. The reagent pack is prepared for use in the BIO-FLASH system by pressing down on the grey lid of the reagent pack to pierce the induction seals on the reagent tubes. Once the seals are broken, the beads are rehydrated by adding the entire contents of the vial of resuspension buffer to the bead reagent tube using the transfer pipette supplied with the kit. Only the hole above the bead reagent tube is accessible at this point. The beads are then mixed with the resuspension buffer by pipetting up and down 30 times. This amount of mixing ensures complete resuspension of the beads. The label covering the remaining three reagent holes is now removed, and the reagent cartridge is then loaded onto the BIO-FLASH instrument. Samples are also loaded onto the instrument in sample racks. A patient serum sample is prediluted 1:10 by the BIO-FLASH with system rinse in a small disposable plastic cuvette. Small amounts of the diluted patient serum, the beads, and assay buffer are all combined into a second cuvette, and mixed. This cuvette is then incubated at 37°C. The beads are magnetized and washed several times. Isoluminol conjugated anti-human IgG antibodies are then added to the cuvette, and again incubated at 37°C. The beads are magnetized and washed repeatedly. The isoluminol conjugate is oxidized when Trigger 1 (Fe(III) coproporphyrin in sodium hydroxide solution) and Trigger 2 (urea-hydrogen peroxide in sodium chloride solution) are added to the cuvette, and the flash of light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. The RLU are proportional to the amount of isoluminol conjugate that is bound to the human IgG, which is in turn proportional to the amount of anti- ß2GP1-Domain1 antibodies bound to the corresponding ß2GP1-Domain1 on the beads.

The QUANTA Flash® ß2GP1-Domain1 assay utilizes a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. Every new lot number of reagent cartridge must be calibrated before first use, with the QUANTA Flash® ß2GP1-Domain1 Calibrators. Based on the results obtained with the two Calibrators included in the reagent kit, an instrument specific Working Curve is created, which is used to calculate chemiluminescent units (CU) from the instrument signal (RLU) obtained for each sample.

AI/ML Overview

This document describes the analytical and clinical performance characteristics of the QUANTA Flash® β2GP1-Domain1 and HemosIL® AcuStar Anti-β2GPI Domain 1 assays, which are in vitro chemiluminescent immunoassays (CIA) for the semi-quantitative determination of IgG autoantibodies to β2GP1-Domain1 in human serum, used as an aid in diagnosing antiphospholipid syndrome (APS).

1. Table of Acceptance Criteria and Reported Device Performance

Feature/MetricAcceptance CriteriaReported Device Performance
PrecisionTotal %CV values within 15% (general statement), or implicitly, the ranges reported for within-run, between-day, and between-run variability for various sample concentrations.Total %CV: 5.6-10.6% across 8 precision samples. Range for within-run: 4.6-8.7%; between-day: 0.0-5.3%; between-run: 0.0-6.0%. All within the 15% acceptance limit.
Limit of Detection (LoD)Proportion of false positives (alpha) < 5% and false negatives (beta) < 5%.LoD: 1296 RLU, which is below the analytical measuring range. (Explicitly stated to meet the CLSI EP17-A guideline with alpha and beta < 5%). LoB: 425.7 RLU.
Analytical Measuring Range (AMR)Defined by the lowest and highest Master Curve Standards.3.6 CU – 1380.4 CU. (Reportable range up to 13804.0 CU with auto-rerun function).
Auto-rerun function (Recovery)% recovery values for auto-rerun results vs. manual dilution within ± 20%.% recovery: 91.9% to 99% (average 94%) for three high positive specimens after auto-rerun compared to manual dilution, meeting the ± 20% acceptance limit.
Hook EffectNo hook effect up to a specified high RLU value.No hook effect observed up to 435,000 RLU; high positive specimens always produced significantly higher RLU values when used "as is" compared to diluted ones.
Linearity (Percent Recovery)For each data point: 85%-115% recovery, or less than 4 CU difference.Percent recovery for 100 data points ranged from 83.9% to 113.9%, or less than 4 CU, meeting the criteria. Overall linear regression: Slope 1.01 (0.99 to 1.02), Y-intercept -3.18 (-9.20 to 2.83), R 1.00.
InterferenceRecovery: 85%-115%, or ±4 CU difference, whichever is greater.No interference detected with bilirubin up to 100 mg/dL (92.9% to 97.6% recovery), hemoglobin up to 200 mg/dL (95.6% to 97.2%), triglycerides up to 1000 mg/dL (92.5% to 100%), cholesterol up to 224.3 mg/dL (92.5% to 100%), and RF IgM up to 500 IU/mL (86.2% to 98.3%). All within acceptance criteria.
Cross-reactivityNo significant association with other autoantibodies (implicitly, minimal positive results for other autoantibodies).Out of 111 samples with high levels of various other autoantibodies, all but two were negative for QUANTA Flash β2GP1 Domain1 CIA. Fisher test showed no significant association. This suggests no significant cross-reactivity.
Shelf Life (Reagent Kit, Beads, Buffer)With regression analysis, the lower 95% CI interval of the regression line is ≥ 85% at 2 weeks, and no individual data point has ≤ 75% recovery at 2 weeks (for 4-week accelerated stability at 37°C, representing 1 year at 5±3°C).All components fulfilled the acceptance criteria. One year expiration dating assigned.
Shelf Life (Controls & Calibrators)With regression analysis, the lower 95% CI interval of the regression line is ≥ 90% at 2 weeks, and no individual data point has ≤ 80% recovery at 2 weeks (for 4-week accelerated stability at 37°C, representing 1 year at 5±3°C).All components fulfilled the acceptance criteria. One year expiration dating assigned.
In-use Stability (Calibrators)All five calibrations performed over 8.5 hours successful. Calibrator average RLU recovery values between 90% and 110% compared to first use. Characterized patient samples within expected range.5 successful calibrations over 8.5 hours. Calibrator RLU values within 90-110% range. All patient samples ran within their expected range. Supports use for up to 4 calibrations over 8 hours.
In-use Stability (Controls)All replicates run within established range. Linear regression line of %recovery vs. runs between 85% and 115% at run 15.Low and high controls ran within their acceptable range for all 20 runs. Linear regression line within 85-115% at run 15. Supports 15 uses with max 10 minutes onboard per use.
In-use Stability (Reagent Cartridge)Stability claim established when 95% CI of regression line reaches 85% or 115% recovery, OR when 2 data points or ≥2% of recovery data is ≤ 75% or ≥ 125%.The onboard study ended at 64 days, with all three lots fulfilling acceptance criteria. In-use stability set at 60 days.
Clinical Sensitivity for APSNot explicitly stated as acceptance criteria, but reported for diagnostic performance.51.1% (95% CI: 45.0-57.2%)
Clinical Specificity for APSNot explicitly stated as acceptance criteria, but reported for diagnostic performance.99.6% (95% CI: 98.9-99.9%)
ROC analysis (AUC for APS)Not explicitly stated as acceptance criteria, but reported for diagnostic efficiency.0.84 (95% CI: 0.81 to 0.86)
Comparison to Predicate Device (Positive Agreement)Not explicitly stated as acceptance criteria, but reported for comparison.91.0% (95% CI: 84.1-95.6%)
Comparison to Predicate Device (Negative Agreement)Not explicitly stated as acceptance criteria, but reported for comparison.78.0% (95% CI: 69.7-84.8%)
Comparison to Predicate Device (Total Agreement)Not explicitly stated as acceptance criteria, but reported for comparison.84.0% (95% CI: 78.7-88.4%)

Detailed Study Information:

2. Sample Size and Data Provenance

  • Test Set (Clinical Performance):

    • Clinical Validation Set: 1090 samples.
      • 270 samples from APS patients.
      • 820 samples from a "control population" (71 infectious disease samples (40 syphilis, 10 HCV, 21 HBV), 104 Crohn's Disease, 94 Ulcerative Colitis, 127 scleroderma, 168 rheumatoid arthritis, 49 osteoarthritis, 24 other diseases, 183 "without APS" conditions like pre-eclampsia/eclampsia, fetal loss, SLE, thrombosis, atopic dermatitis).
    • Expected Values (Normal Population): 400 apparently healthy blood donors (191 females, 209 males). This was used to validate the cutoff and assess the expected value in a normal population, differing from the reference range establishment set.
    • Provenance: Not explicitly stated, but the mention of "human serum from commercial sources" for calibrators and controls implies a general origin rather than specific countries. The description of diverse disease cohorts suggests clinical sites. It is implied these are retrospective samples collected from various patient populations.
  • Test Set (Analytical Performance - Precision): 8 quantitative samples were tested. Each sample was run in duplicate, twice a day, for 20 days (total 80 measurements per sample).

  • Test Set (Analytical Performance - LoD/LoB): 264 determinations (144 measurements on blank samples and 120 measurements of low level samples).

  • Test Set (Analytical Performance - Linearity): Five serum samples with various β2GP1-Domain1 antibody concentrations were diluted.

  • Test Set (Analytical Performance - Interference): Unspecified number of specimens, each spiked at three different concentrations and assessed in triplicates.

  • Test Set (Analytical Performance - Cross-reactivity): 111 clinical samples with unknown disease states, most having high levels of various other autoantibodies.

  • Test Set (Stability):

    • Accelerated Stability: 1 lot of Reagent Kit, 3 lots of β2GP1-Domain1 beads, 3 lots of Resuspension Buffer, 3 lots of Calibrators, 3 lots of Low and High controls.
    • In-use Stability (Calibrators): Not specified by number, but tested over 8.5 hours.
    • In-use Stability (Controls): Low and high controls assayed for a total of 20 runs over 9 days.
    • In-use Stability (Reagent Cartridge): Four serum specimens (with different reactivity levels) along with Low and High Controls tested periodically for 64 days using three lots of reagent cartridge.

3. Number of Experts (Ground Truth for Test Set)

The document does not specify the number or qualifications of experts used to establish the ground truth for the clinical test set. The ground truth (diagnoses of APS, infectious diseases, autoimmune diseases, etc.) appears to be based on pre-existing clinical diagnoses and classifications from the source of the samples.

4. Adjudication Method (Test Set)

The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for establishing the ground truth of the clinical test set. The clinical diagnoses were presumably accepted as the ground truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done. This device is an in vitro diagnostic (IVD) assay designed to semi-quantitatively determine autoantibodies in serum, not an imaging or AI-assisted diagnostic tool that involves human readers interpreting results. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply here.

6. Standalone Performance

Yes, a standalone performance study was done. The entire document describes the performance of the assay (algorithm and system) itself, without human interpretation as part of the primary diagnostic output. The clinical sensitivity and specificity, as well as analytical performance characteristics, are all measures of the algorithm's standalone performance.

7. Type of Ground Truth Used

  • Clinical Performance (APS diagnosis): The ground truth for APS and other disease categories was based on clinical diagnoses of the patients from whom the samples were collected. For example, "270 samples from APS patients" implies these patients had a clinical diagnosis of APS.
  • Analytical Performance:
    • Precision, LoD/LoB, Linearity, AMR, Hook Effect: Ground truth established through reference methods and known concentrations/dilutions of samples.
    • Interference/Cross-reactivity: Ground truth established by spiking known interfering substances or by using samples with known high levels of other autoantibodies/diseases.

8. Sample Size for the Training Set

  • Cut-off Establishment (Reference Range): 30 subjects were used to establish the cut-off. This can be considered a form of "training set" for the cut-off. These subjects included apparently healthy blood donors (5), viral hepatitis (4), Systemic Lupus erythematosus (without history of thrombotic events) (10), Syphilis (10), and HIV (1).
  • The document also implies other internal validation/development data may have been used to define the Master Curve and instrument parameters, but specific sample sizes and details for such internal "training" are not provided.

9. How the Ground Truth for the Training Set was Established

  • Cut-off (Reference Range) Ground Truth:
    • The "training set" for establishing the cut-off consisted of 30 subjects with defined conditions (apparently healthy, specific infectious diseases, SLE without thrombotic events).
    • The ground truth was established by the clinical diagnosis or health status of these subjects. The device was expected to differentiate between "negatives" and "positives" based on these known conditions.
    • The cut-off was derived using "Analyse-it for Excel, to ensure optimal differentiation between negatives and positives," which suggests statistical analysis of the results from these 30 subjects to find the optimal threshold (20 CU in this case).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

FEB 1 8 2015

INOVA DIAGNOSTICS, INC. C/O ROGER ALBESA SUPERVISOR, RESEARCH AND DEVELOPMENT 9900 OLD GROVE ROAD SAN DIEGO CA 92131-1638

Re: K141274

Trade/Device Name: QUANTA Flash® B2GP1-Domain1 QUANTA Flash® ß2GP1-Domain1 Controls HemosIL® Acustar Anti-B2GPI-Domain 1 HemosIL® Acustar Anti-B2GPI-Domain 1 Controls Regulation Number: 21 CFR 866.5660 Regulation Name: Multiple autoantibodies immunological test system Regulatory Class: Class II Product Code: MSV, JJX Dated: January 12, 2015 Received: January 14, 2015

Dear Mr. Albesa,

This letter corrects our substantially equivalent letter of February 13, 2015.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Roger Albesa

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Leonthena R. Carrington -A

Leonthena Carrington, MS, MBA, MT(ASCP) Director (Acting) Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K141274

Device Name

QUANTA Flash® ß2GP1-Domain1, QUANTA Flash® ß2GP1-Domain1 Controls, HemosIL® AcuStar Anti-ß2GPI Domain1 and HemosIL® AcuStar Anti-ß2GPI Domain 1 Controls.

Indications for Use (Describe)

The QUANTA Flash® ß2GP1-Domain1 is an in vitro chemiluminescent immunoassay (CIA) for the semi-quantitative determination of IgG autoantibodies to ß2GP1-Domain1 in human serum. The presence of anti-B2GP1- Domain1 autoantibodies is used in conjunction with clinical and other laboratory findings as an aid in the diagnosis of antiphospholipid syndrome. The OUANTA Flash® B2GP1-Domain1 is not intended to replace assays for antibodies against the whole ß2GP1 molecule. Testing for antibodies to the whole is required according to the classification criteria for antiphospholipid syndrome.

The QUANTA Flash ß2GP1-Domain1 Controls are intended for quality control purposes of the QUANTA Flash ß2GP1-Domain1 chemiluminescent immunoassay (CIA) kit.

The HemosIL® AcuStar Anti-S2GPI Domain 1 is an in vitro chemiluminescent immunoassay (CIA) for the semiquantitative determination of IgG autoantibodies to B2GPI Domain 1 in human serum. The presence of anti-R2GPI Domain 1 autoantibodies is used in conjunction with clinical and other laboratory findings as an aid in the diagnosis of antiphospholipid syndrome. The HemosIL® AcuStar Anti-ß2GPI Domain 1 is not intended to replace assays for antibodies against the whole ß2GPI molecule. Testing for antibodies to the whole is required according to the classification criteria for antiphospholipid syndrome.

The HemosIL AcuStar Anti-B2GPI Domain 1 Controls are intended for quality control purposes of the HemosIL AcuStar Anti-ß2GPI Domain 1 chemiluminescent immunoassay (CIA) kit.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of the 510(k) (ref. K141274) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Submitter:Inova Diagnostics, Inc.9900 Old Grove Road,San Diego, CA, 92131
Purpose of submission:New device(s)
Devices in the submission:QUANTA Flash® β2GP1-Domain1QUANTA Flash® β2GP1-Domain1 ControlsHemosIL® AcuStar Anti-β₂GPI Domain 1HemosIL® AcuStar Anti-β₂GPI Domain 1 Controls
Scientific contact:Roger Albesa, M.S.Supervisor of Research and DevelopmentInova Diagnostics, Inc.9900 Old Grove Road, San Diego, CA 92131e-mail: ralbesa@inovadx.comTel. (858) 586-9900/1394
Quality Systems contact:Ronda Elliott, Vice President, Quality and RegulatoryInova Diagnostics, Inc.9900 Old Grove Road, San Diego, CA, 92131e-mail: relliott@inovadx.comTel. (858) 586-9900Fax: 858-863-0025/1381
Preparation date:01/09/2015
Device name (assay kit):Proprietary names: QUANTA Flash® β2GP1-Domain1,HemosIL® AcuStar Anti-β₂GPI Domain 1
Common name: Anti-β2GPI Domain 1 Chemiluminescent immunoassayClassification name: Anti-β2GPI Domain 1 antibody, antigen and control
Regulation DescriptionMultiple autoantibodies immunological test system
Regulation Medical SpecialtyImmunology
Review PanelImmunology
Product CodeSystem, Test, Antibodies,B2-Glycoprotein I (B2-GPI) (MSV)
Regulation Number866.5660 (Reagent Kit)
Regulation Number862.1660 (Controls)
Device Class2
Predicate device:QUANTA Lite® β2GPI IgG ELISA, 510(k) number: K970551

{4}------------------------------------------------

Device description:

The QUANTA Flash® ß2GP1-Domain1 assay is designed to run on the BIO-FLASH® instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash® ß2GP1-Domain1 assay utilizes a reagent cartridge format, which is compatible with the BIO-FLASH instrument.

The assays included in this submission, the QUANTA Flash ß2GP1-Domain1 marketed by Inova Diagnostics Inc. (9900 Old Grove Road, San Diego, CA 92131) and HemoslL AcuStar Anti-ß-GPI Domain 1 marketed by Instrumentation Laboratory (180 Hartwell Road Bedford, MA 01730), are equivalent assays. Therefore all data stated hereafter and referred to as: QUANTA Flash §2GP1-Domain1 data is equivalently also valid for HemosIL "AcuStar Anti-ß₂GPI Domain 1.

Recombinant ß2GP1-Domain1 protein is coated onto paramagnetic beads, which are stored lyophilized in the reagent cartridge. The reagent pack is prepared for use in the BIO-FLASH system by pressing down on the grey lid of the reagent pack to pierce the induction seals on the reagent tubes. Once the seals are broken, the beads are rehydrated by adding the entire contents of the vial of resuspension buffer to the bead reagent tube using the transfer pipette supplied with the kit. Only the hole above the bead reagent tube is accessible at this point. The beads are then mixed with the resuspension buffer by pipetting up and down 30 times. This amount of mixing ensures complete resuspension of the beads. The label covering the remaining three reagent holes is now removed, and the reagent cartridge is then loaded onto the BIO-FLASH instrument. Samples are also loaded onto the instrument in sample racks. A patient serum sample is prediluted 1:10 by the BIO-FLASH with system rinse in a small disposable plastic cuvette. Small amounts of the diluted patient serum, the beads, and assay buffer are all combined into a second cuvette, and mixed. This cuvette is then incubated at 37°C. The beads are magnetized and washed several times. Isoluminol conjugated anti-human IgG antibodies are then added to the cuvette, and again incubated at 37°C. The beads are magnetized and washed repeatedly. The isoluminol conjugate is oxidized when Trigger 1 (Fe(III) coproporphyrin in sodium hydroxide solution) and Trigger 2 (urea-hydrogen peroxide in sodium chloride solution) are added to the cuvette, and the flash of light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. The RLU are proportional to the amount of isoluminol conjugate that is bound to the human IgG, which is in turn proportional to the amount of anti- ß2GP1-Domain1 antibodies bound to the corresponding ß2GP1-Domain1 on the beads.

The QUANTA Flash® ß2GP1-Domain1 assay utilizes a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. Every new lot number of reagent cartridge must be calibrated before first use, with the QUANTA Flash® ß2GP1-Domain1 Calibrators. Based on the results obtained with the two Calibrators included in the reagent kit, an instrument specific Working Curve is created, which is used to calculate chemiluminescent units (CU) from the instrument signal (RLU) obtained for each sample.

{5}------------------------------------------------

The QUANTA Flash® ß2GP1-Domain1 kit contains the following materials:

One (1) QUANTA Flash ß2GP1-Domain1 Reagent Cartridge, containing the following reagents for 50 determinations:

  • a. ß2GP1-Domain1 antigen coated paramagnetic beads in a suspension.
  • b. Assay Buffer - buffer containing protein stabilizers and preservatives.
  • C. Tracer IgG - Isoluminol labeled anti-human IgG antibodies in buffer, containing protein stabilizers and preservative.
  • d. Resuspension Buffer 3 - buffer containing protein stabilizers and preservatives.
  • QUANTA Flash ß2GP1-Domain1 Calibrator 1: One (1) barcode labeled tube containing e. 1.0 mL prediluted, ready to use reagent. Calibrators contain human antibodies to ß2GP1-Domain1 in stabilizers and preservatives.
  • f. QUANTA Flash ß2GP1-Domain1 Calibrator 2: One (1) barcode labeled tube containing 1.0 mL prediluted, ready to use reagent. Calibrators contain human antibodies to ß2GP1-Domain1 in stabilizers and preservatives ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

The QUANTA Flash 12GP1-Domain1 Controls kit contains three vials of Low Control and three vials of High Control.

  • । QUANTA Flash ß2GP1-Domain1 Low Control: Three (3) barcode labeled tubes containing 1.0 mL, ready to use reagent. Controls contain human antibodies to ß2GP1-Domain1 in stabilizers and preservatives.
  • । QUANTA Flash ß2GP1-Domain1 High Control: Three (3) barcode labeled tubes containing 1.0 mL, ready to use reagent. Controls contain human antibodies to ß2GP1-Domain1 in stabilizers and preservatives.

Intended use(s):

QUANTA Flash ß2GP1-Domain1

The QUANTA Flash 82GP1-Domain1 is an in vitro chemiluminescent immunoassay (CIA) for the semiquantitative determination of IgG autoantibodies to ß2GP1-Domain1 in human serum. The presence of anti-ß2GP1-Domain1 autoantibodies is used in conjunction with clinical and other laboratory findings as an aid in the diagnosis of antiphospholipid syndrome. The QUANTA Flash ß2GP1-Domain1 is not intended to replace assays for antibodies against the whole ß2GP1 molecule. Testing for antibodies to the whole ß2GP1 molecule is required according to the classification criteria for antiphospholipid syndrome.

QUANTA Flash ß2GP1-Domain1 Controls

The QUANTA Flash ß2GP1-Domain1 Controls are intended for quality control purposes of the QUANTA Flash ß2GP1-Domain1 chemiluminescent immunoassay (CIA) kit.

HemosIL AcuStar Anti-ß,GPI Domain 1

The HemosIL " AcuStar Anti-R2GPI Domain 1 is an in vitro chemiluminescent immunoassay (CIA) for the semi-quantitative determination of IgG autoantibodies to ß3GPI Domain 1 in human serum. The presence of anti-ß-GPI Domain 1 autoantibodies is used in conjunction with clinical and other laboratory findings as an aid in the diagnosis of antiphospholipid syndrome. The Hemosl ® AcuStar Anti-ß,GPI Domain 1 is not intended to replace assays for antibodies against the whole ß₂GPI molecule. Testing for

{6}------------------------------------------------

antibodies to the whole ß₂GPI molecule is required according to the classification criteria for antiphospholipid syndrome.

HemosIL AcuStar Anti-ß₂GPI Domain 1 Controls

The HemosIL AcuStar Anti-ß2GPI Domain 1 Controls are intended for quality control purposes of the HemosIL Anti-ß2GPI Domain 1 chemiluminescent immunoassay (CIA) kit.

Substantial equivalence: The QUANTA Flash ß2GP1-Domain1 and the QUANTA Flash ß2GP1-Domain1 Controls have a similar intended use and assay principle as the predicate device.

{7}------------------------------------------------

Comparison to predicate device:

QUANTA Flash ® β2GP1-Domain1 reagent kit
--------------------------------------------
Similarities
ItemApplicantPredicate Device
QUANTA Flash ® β2GP1-Domain1QUANTA Lite® β2 GPI IgG ELISA
Intended useThe QUANTA Flash ® β2GP1-Domain1 isan in vitro chemiluminescentimmunoassay (CIA) for the semi-quantitative determination of IgGautoantibodies to β2GP1-Domain1 inhuman serum. The presence of anti-β2GP1-Domain1 autoantibodies isused in conjunction with clinical andother laboratory findings as an aid inthe diagnosis of antiphospholipidsyndrome. The QUANTA Flash ® β2GP1-Domain1 is not intended to replaceassays for antibodies against thewhole β2GP1 molecule. Testing forantibodies to the whole β2GP1molecule is required according to theclassification criteria forantiphospholipid syndrome.QUANTA Lite® β2 GPI IgG is an enzymelinked immunoassay (ELISA) for thesemi-quantitative detection of β2 GPIIgG antibodies in human serum. Thepresence of β₂ GPI IgG antibodies canbe used in conjunction with clinicalfindings and other laboratory tests toaid in the diagnosis of certainautoimmune disease thromboticdisorders such as those secondary tosystemic lupus erythematosus (SLE) orother lupus-like thrombotic diseases.
Assay methodologySolid phase (heterogeneous)immunoassaySolid phase (heterogeneous)immunoassay
Shelf lifeOne yearOne year
Sample typeSerumSerum
Differences
ItemQUANTA Flash β2GP1-Domain1Predicate Device
Detection/Operating principleChemiluminescent immunoassayEnzyme-linked immunosorbent assay
Solid phaseParamagnetic microparticles (beads)96-well plate
AntigenRecombinant domain1 of β2-Glycoprotein1purified β2-Glycoprotein1
ConjugateIsoluminol conjugated anti-human IgGHRP conjugated anti-human IgG
CalibrationLot specific Master Curve + twoCalibrators (included in kit)Five lot specific calibrators(Included in the kit)
Units< 20 CU Negative result≥ 20 CU Positive result< 20 SGU Negative result≥ 20 SGU Positive result
AnalyticalMeasuring Range3.6 – 1380.4 CUReportable range = up to 13804.0 CU9.4 – 150.0 SGU

{8}------------------------------------------------

QUANTA Flash ß2GP1-Domain1 Calibrators

ItemQUANTA Flash B2GP1-Domain1CalibratorsPredicate Device
Intended useNo separate intended use; calibrators are part of the kit.No separate intended use; calibrators are part of the kit.
AnalyteAnti-β2GP1-Domain 1 antibodiesAnti-β2GP1 antibodies
MethodQUANTA Flash® β2GP1-Domain1QUANTA Lite® β2 GPI IgG ELISA
UnitChemiluminescent Units (CU)QUANTA FlashArbitrary Units (U/ml)HemosIL AcuStarSGU
MatrixHuman serum, stabilizers, and preservativeHuman serum, stabilizers, and preservative
Physico-chemicalcharacteristicsLiquid, prediluted, ready to useLiquid, prediluted, ready to use
Storage2-8 °C2-8 °C
Shelf lifeOne yearOne year

QUANTA Flash ß2GP1-Domain1 Controls

ItemQUANTA FlashControlsß2GP1-Domain1Predicate Device
Intended useThe QUANTA Flash ß2GP1-Domain1 Controls are intended for quality control purposes of the QUANTA Flash ß2GP1-Domain1 chemiluminescent immunoassay (CIA) kit.No separate intended use; controls are part of the kit.
AnalyteAnti-ß2GP1-Domain1 antibodiesAnti- ß2GP1- antibodies
MethodQUANTA Flash ß2GP1-Domain1 chemiluminescent immunoassayQUANTA Lite® ß2 GPI IgG ELISA
UnitQUANTA Flash (CU)HemosIL AcuStar (U/ml)SGU
MatrixHuman serum, stabilizers, and preservativeHuman serum, stabilizers, and preservative
Physico-chemical characteristicsLiquid, ready to useLiquid, prediluted, ready to use
Levels2 (low and high)2 (negative and positive)
Storage2-8°C2-8°C
Shelf lifeOne yearOne year

{9}------------------------------------------------

Value assignment and traceability of Calibrators and Controls

The QUANTA Flash ß2GP1-Domain1 Calibrators and Controls are manufactured by diluting human serum that contains high titer of anti-ß2GP1-Domain1 antibody stabilizer buffer, containing preservative. The human serum is obtained from commercial sources and it is tested for markers of infectious substances.

The target CU is achieved through trial dilutions on small scale. Once a dilution is selected, the Calibrators and Control are bulked, tested, and adjusted. Upon completion of the manufacturing process, the Calibrators and Controls are tested on at least two instruments, on at least two lots of reagent cartridge, in replicates of 10 to determine final value assignment.

There are no international reference standards for anti- ß2GP1 IgG. Calibrators and controls are assigned values based on a 20 CU cut-off between positive (low) during assay development. Calibrators are specified in the labeling and supplied with the assay. Control materials are sold separately. The table below summarizes the control and calibrators target values.

Points on master curveAssigned Value (CU)
Point 13.6
Point 218.5
Point 343.7
Point 490.0
Point 5316.0
Point 61380.4

List of ß2GP1-Domain1 Standards, Calibrators and Controls:

MaterialManufacturingTarget Value (CU)
Calibrator 118.5
Calibrator 2316.0
Control Low10.0
Control High50.0

Analytical performance characteristics

Precision

Testing for precision was performed in accordance with CLSI EP5-A2, and analyzed using Analyze-it which is based on CLSI EP5 1 & 2. 8 Samples were selected to cover the analytical measuring range of the assay, including samples around the medical decision point. Patients were run using duplicate aliquots, twice a day, for 20 days. Controls were run as quality control during each run. A working curve was generated using calibrators prior to run 1 on day 1. Total %CV values were within the acceptance limit, 15%.

{10}------------------------------------------------

QUANTA Flash ß2GP1-Domain1 Precision Study Decision Summary
Within RunBetween-RunBetween-DayTotal
SampleNMean(CU)SD%CVSD%CVSD%CVSD%CV
Precision 18010.40.98.3%0.00.0%0.54.3%1.09.4%
Precision 28018.40.94.6%0.52.9%0.52.6%1.16.0%
Precision 38022.81.46.1%0.93.8%0.00.0%1.67.2%
Precision 48054.32.54.6%1.42.5%1.22.2%3.15.6%
Precision 58099.66.46.5%0.00.0%0.00.0%6.46.5%
Precision 680318.318.65.9%0.00.0%4.91.6%19.36.1%
Precision 780538.130.35.6%7.91.5%28.75.3%42.57.9%
Precision 880947.882.88.7%57.36.0%0.00.0%100.710.6%

Precision study QUANTA Flash ß2GP1-Domain1 according to the CLSI EP5-A2 guideline

Summary of Precision study QUANTA Flash ß2GP1-Domain1, according to the CLSI EP5-A2 guideline
---------------------------------------------------------------------------------------------------
No. of samplesWithin RunBetween-DayBetween-RunTotal
β2GP1-Domain184.6-8.7%0.0-5.3%0.0-6.0%5.6-10.6%

Limit of Blank (LoB) and Limit of Detection (LoD)

The LoD of the QUANTA Flash" ß2GP1-Domain1 assay is 1296 RLU, which is below the analytical measuring range of the assay. It was determined consistent with CLSI EP17-A guideline with proportions of false positives (alpha) less than 5% and false negatives (beta) less than 5%; based on 264 determinations, with 144 measurements on blank samples and 120 measurements of low level samples. The LoB is 425.7 RLU.

Analytical Measuring Range (AMR)

3.6 CU – 1380.4 CU

The AMR is defined by the values of the lowest and highest Master Curve Standards.

{11}------------------------------------------------

Auto-rerun function and reportable results

The BIO-FLASH software has an Auto-rerun option available. If this option is selected, the instrument will automatically rerun any sample that has a result >1380.4 CU by further diluting it by a factor specified in the assay definition file (10 fold), thereby bringing the measured value within the AMR. The final result will be calculated by the software by taking into account the additional dilution factor. As the highest value that can be measured is 1380.4 CU, the highest value that can be reported is 13804 CU.

To validate the Auto-rerun function, three high positive specimens with results above the analytical measuring range were selected. The samples were run with the Auto-rerun function enabled on the BIO-FLASH. Then the specimens were manually diluted the same way as it happens in the Auto-rerun function (10 fold), and tested on the BIO-FLASH. The results were within the analytical measuring range after auto-rerun or manual dilution for all specimens. The % recovery values for results obtained with the auto-rerun results compared to the results obtained by manual dilution were between 91.9% and 99% (average 94%) and are within the ± 20% acceptance limit.

High concentration hook effect

To assess hook effect, the measurement signal (relative light units, RLU) was examined for the above mentioned four high positive specimens, before and after automatic or manual dilution. All sera produced significantly higher RLU values (above the AMR) when used "as is" compared to the manually or automatically diluted ones, thereby confirming that high positive specimens above the analytical measuring range do not show hook effect up to 435,000 RLU.

Linearity

The linearity of the AMR was evaluated by a study according to CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline. Five serum samples with various ß2GP1-Domain1 antibody concentrations were diluted with assay buffer to obtain values that cover the AMR. Percent recovery for all 100 data points ranged from 83.9% to 113.9%, or less than 4 CU. All specimens showed dilution linearity individually, and the combined data yielded the following results with linear regression:

AssayTest Range (CU)Slope (95% CI)Y-intercept (95% CI)RPercent Recovery
ß2GP1-Domain11.0 to 1640.71.01 (0.99 to 1.02)-3.18 (-9.20 to 2.83)1.0083.9-113.9%

{12}------------------------------------------------

Interference

Interfering substances were spiked into every specimen at three different concentrations in 10% of total specimen volume, and the resulting samples were assessed in triplicates with the ß2GP1-Domain1 assay. Recovery of the unit values was calculated compared to control samples spiked with the same volume of diluents (10% of total). Acceptance criteria for the interference studies were 85%-115% recovery, or ±4 CU difference, whichever is greater.

No interference was detected with bilirubin up to 100 mg/dL (recovery: 92.9% to 97.6%), hemoglobin up to 200 mg/dL (recovery: 95.6% to 97.2%), triglycerides up to 1000 mg/dL (recovery: 92.5% to 100%), cholesterol up to 224.3 mg/dL (recovery: 92.5% to 100%), and RF IgM up to 500 IU/mL (recovery: 86.2% to 98.3%).

Cross-reactivity

A potential cross-reactivity of the QUANTA Flash ß2GP1 Domain1 CIA with other autoantibodies was evaluated with 111 clinical samples with unknown disease states, most having high levels of various other autoantibodies. These samples were tested for various autoantibodies using other QUANTA Flash immunoassays: Sm, RNP, SS-A/Ro60, Ro52, SS-B, Jo-1, Scl-70, CENP, Ribosomal P, DFS70 and the QUANTA Flash 12GP1-Domain1 ClA. These are autoantibodies found in individuals with autoimmune diseases such as SLE, Sjögren's Syndrome, scleroderma, and polymyositis patients. All except for two samples were negative for QUANTA Flash ß2GP1 Domain1 CIA.

Fisher test revealed no significant association between the other positive reactivities and Domain1. This verified that there is no significant cross-reactivity on the QUANTA Flash 182GP1 Domain1 ClA.

# of Samples tested = 111# of Positives (%)# of Domain1 Positives (%)
Sm6 (5.4%)0 (0.0%)
RNP10 (9.0%)0 (0.0%)
Ro6011 (9.9%)0 (0.0%)
Ro5220 (18.0%)0 (0.0%)
SSB8 (7.2%)0 (0.0%)
Jo-11 (0.9%)0 (0.0%)
Scl-7025 (22.5%)2 (1.8%)
CENP0 (0.0%)0 (0.0%)
Ribosomal P5 (4.5%)0 (0.0%)
DFS705 (4.5%)0 (0.0%)

{13}------------------------------------------------

Stability

Shelf life

To establish the initial claim for shelf life, accelerated stability studies were performed for 4 weeks at 37 °C.

Accelerated stability testing was performed on each of the following sealed components of the QUANTA Flash® ß2GP1-Domain1 to establish initial stability claim:

• QUANTA Flash® β2GP1-Domain1 Reagent Kit(1 Lot)
• β2GP1-Domain1 beads(3 Lots)
• Resuspension Buffer(3 Lots)
• Calibrators 1 and 2(3 Lots)
• Low and High controls(3 Lots)

Each week a new sealed component was placed in the incubator, and all components were tested at the end of the experiment together with the one that was stored at 5 ± 3°C. The recovery of the measured values was calculated for each time point (compared to those obtained with 5 ± 3°C stored reagent). All calculations were performed by comparing results of sealed components stored at 5 ± 3℃ (control) to those stored at 37 ± 3℃ (test) for 1, 2, 3, and 4 weeks, where one week is equal to six months at 5 ± 3℃. Linear regression analysis was performed between recovery values and the number of days.

Acceptance criteria for one year preliminary expiration dating were:

-Microparticles (beads), Resuspension Buffer, and Reagent Kit:

With regression analysis, the lower 95% Cl interval of the regression line is ≥ 85% at 2 weeks, and no individual data point has ≤ 75% recovery at 2 weeks.

- Controls and Calibrators:

With regression analysis, the lower 95% Cl interval of the regression line is ≥ 90% at 2 weeks, and no individual data point has ≤ 80% recovery at 2 weeks.

All components tested fulfilled the acceptance criteria above, so one year expiration dating was assigned to each component

In-use (onboard) stability

Calibrators

During assessing on-board stability, Calibrators were placed, uncapped, onboard the instrument, and calibration was performed altogether five times, then a panel of characterized patient specimens were run on each calibration curve.

Acceptance criteria were: Calibrators are considered stable if all five calibrations performed in the 8.5 hour period are successful, and Calibrator average RLU recovery values are between 90% and 110% compared to the first use.

{14}------------------------------------------------

A total of 5 successful calibrations were performed over an 8.5 hour period. Calibrator RLU values remained within the 90-110% range. Moreover, all characterized patient samples ran within their expected range. This supports the claim that calibrators can be used for up to 4 calibrations over an 8 hour period.

Controls

During assessing on-board stability, Low and high Controls were assayed for a total of 20 runs, over 9 days. The controls were left uncapped, onboard the instrument for 15 minutes per run. When not in use, the controls were capped, and stored at 5 ± 3°C for at least 2 hours between runs.

Acceptance criteria: Controls are considered stable when all replicates run within their established range, moreover, the linear regression line obtained by plotting %recovery values against the number of runs stays between 85% and 115% at run 15.

Low and high controls ran within their respective acceptable range for all 20 runs. The linear regression line obtained by plotting %recovery values against the number of runs was within 85% and 115% at run 15 for both Controls. The controls were given a maximum of 15 uses with a maximum of 10 minutes onboard per use.

Reagent Cartridge

To determine the in-use stability of the QUANTA Flash® ß2GP1-Domain1 reagent cartridge, four serum specimens (with different reactivity levels) along with the Low and high Control were tested. The specimens were tested periodically for 64 days. Recoveries were calculated compared to the day zero average values, and linear regression analysis was performed. The claim was established using the following criteria (using the one that is fulfilled first):

a) The stability claim is established at the day where the 95% confidence interval of the regression line reaches 85% or 115% recovery, or b) When 2 data points or ≥2% of the recovery data (whichever is greater) is ≤ 75% or ≥ 125%.

The onboard study was ended at 64 days, where all three lots of Reagent Cartridge tested still fulfill the acceptance criteria. The in-use (onboard) stability of the reagent cartridge was set at 60 days.

{15}------------------------------------------------

Cut-off, reference range

The reference population for establishing the reference interval for the ß2GP1-Domain1 assay consisted of 30 subjects:

Apparently healthy blood donors5
Viral hepatitis4
Systemic Lupus erythematosus (without history of thrombotic events)10
Syphilis10
HIV1

All specimens were the same matrix as specified in the Intended Use. All specimens were unaltered. The cut-off was established using Analyse-it for Excel, to ensure optimal differentiation between negatives and positives, and found at 7880 RLU. This point was defined as 20 CU. A result below 20 CU is considered negative, and equal or greater than 20 CU is considered positive.

{16}------------------------------------------------

Clinical performance characteristics

Clinical sensitivity, specificity

A separate set of samples, none of which were used in establishing the reference range, was used to validate the clinical performance of the QUANTA Flash ß2GP1-Domain1 ClA. A total of 1090 samples were included in the Validation Set for the QUANTA Flash ß2GP1-Domain1. This Validation Set includes:

  • 270 samples from APS patients
  • 71 infectious disease samples (40 syphilis, 10 HCV, 21 HBV)
  • 104 samples from Crohn's Disease (CD) patients
  • 94 samples from Ulcerative Colitis (UC) patients
  • 127 samples from autoimmune scleroderma patients
  • . 168 samples from patients with rheumatoid arthritis
  • 49 samples from patients with osteoarthritis
  • 24 samples from patients with other diseases (Polymyalgia Rheumatica, Degenerative Spine Disease)
  • . 183 samples from different conditions "without APS" (Pre-eclampsia/eclampsia, fetal loss, systemic lupus erythematosus (SLE), thrombosis and atopic dermatitis)

Clinical sensitivity and specificity for APS (n=270) using the control population (n=820) is calculated in the table below.

Cohort 1 (n=938)DiagnosisPercent Agreement
APSNot APSTotal(95% confidence)
QUANTA Flash®Domain 1 CIAPositive1383141Sensitivity = 51.1% (45.0-57.2%)
Negative132817949Specificity = 99.6% (98.9-99.9%)
Total2708201090

To assess diagnostic efficiency, ROC analysis was performed on the validation sample pool for APS. The results are below:

TestArea95% CISEZp
QUANTA Flash B2GP1-Domain1(CU)0.840.81 to 0.860.01522.24<0.0001

{17}------------------------------------------------

Expected values

The expected value in the normal population is "negative". Anti- ß2GP1-Domain1 antibody levels were analyzed in a cohort of 400 apparently healthy blood donors (191 females, ages 17 to 60 years, average age 32.2 years, and 209 males ages 17 to 60 years, average age 34.7 years) using the QUANTA Flash ß2GP1-Domain1. This patient population was different from the one that was used to establish the cutoff, and was only used to validate the cutoff. With a cut-off of 20 CU, one sample (0.2%) was positive (34.2 CU) on the QUANTA Flash ß2GP1-Domain1. The mean concentration was 3.8 CU, and the values ranged from <3.6 to 34.2 CU.

Comparison with predicate device

Of the samples tested in the validation set, 238 samples, within the analytical measuring range of the QUANTA Flash B2GP1-Domain1 ClA were tested also by the QUANTA Lite were positive by QUANTA Flash ß2GP1-Domain1 CIA, 111 samples were positive by QUANTA Lite ˚ ½GPI IgG ELISA. The positive, negative, and total agreements were 91.0%, 78.0% and 84.0% respectively. The seroreactivity of 22 sera in an area ±5CU of the cut off, the calculated positive, negative, and total agreements were 80.0%, 82.4% and 81.8% respectively.

All (n=238)QUANTA Lite ® β2GPI IgG ELISAPercent Agreement (95% confidence)
PositiveNegativeTotal
QUANTA Flash®β2GP1-Domain1 CIAPositive10128129Pos. Agree = 91.0% (84.1-95.6%)
Negative1099109Neg. Agree = 78.0% (69.7-84.8%)
Total111127238Total Agree = 84.0% (78.7-88.4%)

§ 866.5660 Multiple autoantibodies immunological test system.

(a)
Identification. A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).(b)
Classification. Class II (performance standards).