(254 days)
The BioPlex® 2200 25-OH Vitamin D kit is a flow competitive immunoassay intended for the quantitative determination of 25-hydroxyvitamin D in human serum. The BioPlex 2200 25-OH Vitamin D assay is to be used as an aid in the assessment of vitamin D sufficiency.
The BioPlex® 2200 25-OH Vitamin D kit is intended for use with the Bio-Rad BioPlex 2200 System.
The BioPlex® 2200 25-OH Vitamin D Calibrator Set is intended for the calibration of the BioPlex® 2200 25-OH Vitamin D Reagent Pack.
The BioPlex® 2200 25-OH Vitamin D Control Set is intended for use as an assayed quality control to monitor the overall performance of the BioPlex® 2200 System and the corresponding BioPlex® 25-OH Vitamin D Reagent Packs in the clinical laboratory. The performance of the BioPlex® 25-OH Vitamin D Control Set has not been established with any other 25-hydroxyvitamin D assays.
BioPlex® 2200 25-OH Vitamin D kit includes the following components:
- One (1) 10 mL vial of Bead Set containing dyed beads coated with anti-25-OH D antibody (sheep), an Internal Standard bead (ISB), and a Serum Verification bead (SVB) in buffer with protein stabilizers (bovine). ProClin 950 (< 1.0%) and sodium azide (< 0.1%) as preservatives.
- One (1) 10 mL vial of Release Buffer containing 25-OH Vitamin D releasing reagents in citrate and trisodium citrate acid buffer at pH 4.1 and ProClin 950 (<1.0%) as preservative.
- One (1) 5 mL vial of Conjugate 1 containing biotinylated 25-OH Vitamin D conjugate and biotinylated anti-human FXIII antibody conjugate (murine) in buffer with protein stabilizers (bovine). ProClin 950 (< 1.0%) and 5-bromo-5nitro-1. 3-dioxane (<0.1%) as preservatives and chemical blockers.
- One (1) 5 mL vial of Conjugate 2 containing phycoerythrin conjugated streptavidin (SA -PE) in a buffer comprising protein stabilizers (bovine). ProClin 950 (< 1.0%) and sodium azide (< 0.1%) as preservatives, chemical blockers and detergent (Tween 20).
BioPlex® 2200 25-OH Vitamin D Calibrator set contains six (6) 0.5 mL 25-OH Vitamin D vials. Calibrator level 1 contains 25% horse serum without 25-OH Vitamin D. The calibrator levels 2 to 6 are provided in a Vitamin D depleted human serum matrix supplemented with known concentration of 25-hydroxyvitamin D3. All calibrators contain ProClin 950 (≤ 0.3%), sodium benzoate (≤ 0.1%) and 5-bromo-5-nitro-1, 3dioxane (<0.1%) as preservatives.
BioPlex® 2200 25-OH Control set contains two (2) 1.5 mL Level 1 and two (2) 1.5 mL Level 2 Control vials, each containing 25-OH Vitamin D in human serum matrix. All controls contain ProClin 950 (< 0.3%), sodium benzoate (< 0.1%) and 5-bromo-5-nitro-1, 3-dioxane (<0.1%) as preservatives.
Additional materials required but not supplied include BioPlex® 2200 Sheath Fluid containing Phosphate Buffered Saline (PBS), ProClin 300 (0.03%) and sodium azide (<0.1%) as preservatives; and BioPlex® 2200 Wash Solution containing Phosphate Buffered Saline (PBS) and Tween 20. ProClin 300 (0.03%) and sodium azide (<0.1%) as preservatives.
The provided text describes the performance characteristics of the BioPlex® 2200 25-OH Vitamin D Kit, Calibrator Set, and Control Set. This is a medical device, and the information is presented as part of a 510(k) submission to the FDA. The sections below extract the requested information based on the provided document.
Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" for each performance characteristic. Instead, it presents the results of various studies and implies that these results demonstrate adequate performance for the intended use and substantial equivalence to the predicate device. For precision, for example, it shows the calculated %CV (Coefficient of Variation), which indicates the device's variability. For linearity, it provides the slope, intercept, and r² value of the regression analysis, which are generally expected to be close to 1, 0, and 1, respectively, for good linearity. For interference studies, the implicit acceptance criterion is that the percent difference should be ≤ 10%.
Since explicit acceptance criteria are not provided for all metrics, the table below will list the reported performance and indicate the implicit target where inferable.
| Performance Characteristic | Implicit Acceptance Criteria (Inferred) | Reported Device Performance |
|---|---|---|
| Precision (CLSI EP5-A2) | Low %CVs expected | Total Precision (%CV):- Samples: 4.0% - 10.5% (across 6 samples from 15.0 to 110.8 ng/mL)- Controls: 6.1% - 8.5% (Control 1 at 22.1 ng/mL, Control 2 at 50.0 ng/mL) |
| Reproducibility (CLSI EP15-A2) | Low %CVs expected | Total Precision (%CV):- Samples: 5.0% - 14.8% (across 8 samples from 11.5 to 104.9 ng/mL)- Controls: 5.1% - 8.5% (Control 1 at 21.6 ng/mL, Control 2 at 58.8 ng/mL) |
| Linearity | Slope ≈ 1, Intercept ≈ 0, r² ≈ 1 | Example: For 168.9 ng/mL, Slope = 1.0001, Intercept = 0.0045, r² = 0.9988 (over dilution range 5.5 - 168.9 ng/mL). Reportable Measuring Range: 6.5 to 125.0 ng/mL. |
| Detection Limits | Low values for LoB, LoD, LoQ expected | LoB: 0.8 ng/mLLoD: 2.5 ng/mLLoQ: 6.5 ng/mL (based on accuracy goal of precision ≤ 20% CV) |
| Interfering Substances | % difference ≤ 10% | No interference observed for tested substances at specified concentrations (e.g., Hemoglobin ≤ 150 mg/dL, Bilirubin (unconjugated) ≤ 20 mg/dL, Triglycerides ≤ 400 mg/dL, etc.), except for Hemoglobin > 150 mg/dL, which "may interfere," and Paricalcitol (Zemplar), which "cross-react[s] and interfere[s]." |
| Cross-Reactivity | Low % cross-reactivity for non-target analytes | High cross-reactivity: 25-hydroxyvitamin D2 (103%), 1,25-dihydroxyvitamin D2 (>100%), 1,25-dihydroxyvitamin D3 (79%), 3-epi 25-hydroxyvitamin D3 (59%).Low cross-reactivity: Vitamin D2 (0.2%), Vitamin D3 (0.0%), 24,25-dihydroxyvitamin D3 (9%).Paricalcitol (>100% and interferes). |
| Method Comparison | High correlation (r) and agreement with predicate | Weighted Deming Regression (vs. EUROIMMUN 25-OH Vitamin D ELISA):- N = 196 samples- Slope = 1.0039 (95% CI: 0.9365 to 1.0712)- Intercept = -0.2256 (95% CI: -2.4121 to 1.9608)- Correlation Coefficient (r) = 0.9553 (95% CI: 0.9412 to 0.9661) |
For other requested details:
2. Sample Sizes Used for the Test Set and Data Provenance:
- Precision/Reproducibility (CLSI EP5-A2): 6 frozen human serum samples (N=80 data points for each sample/control for total precision).
- Reproducibility (CLSI EP15-A2): 8 human serum samples (n=20 data points for each sample/control for total precision).
- Linearity: 5 high patient serum samples.
- Detection Limits (LoB): 5 blank samples (100 data points per reagent lot).
- Detection Limits (LoD/LoQ): 6 human samples with low 25-OH vitamin D (60 data points per sample per reagent lot).
- Interfering Substances: Samples containing 10 to 90 ng/mL Vitamin D were used, spiked with various interfering substances.
- Cross-Reactivity: 2 human serum pools (at 20 and 35 ng/mL 25-hydroxyvitamin D), spiked with 9 cross-reactants.
- Method Comparison (CLSI EP09-A2-IR): 204 human samples (185 unaltered, 19 spiked with 25-hydroxyvitamin D3). 196 results included in analysis after excluding 8 out of range samples.
- Expected Values/Reference Range (CLSI C28-A3c): 287 samples from allegedly healthy donors.
- Provenance: Samples were collected from three regions (North, Central, and South) in the US, across spring, summer, and winter, including African Americans, Hispanics, and Caucasians. No specific mention of retrospective or prospective for these samples, but typical for such studies, they are likely collected specifically for the study (prospective) or from a biobank.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
This document describes an in vitro diagnostic (IVD) device for quantitative measurement of a biomarker. The "ground truth" for the test set is established by the reference method or known concentrations of the analyte.
- For precision, linearity, detection limits, interference, and cross-reactivity studies, the ground truth is based on the known concentrations or analytical characteristics of the spiked samples, dilutions, or control materials. This does not involve human expert interpretation of images or clinical cases. The performance is determined by comparing the device's measurements to these known or expected values.
- For method comparison, the ground truth is the measurement by the predicate device (EUROIMMUN 25-OH Vitamin D ELISA, K123660), which is itself a validated IVD device.
- For expected values/reference range, the "ground truth" is derived from the measurements of samples from a statistically diverse group of healthy individuals, yielding a statistical distribution rather than an expert-defined diagnosis.
Therefore, the concept of "experts establishing ground truth" as it applies to subjective assessments (e.g., radiologists for images) is not directly applicable here.
4. Adjudication Method for the Test Set:
Not applicable in the context of an IVD device for quantitative measurement. The ground truth, where applicable, is either an analytical standard, a predicate device's output, or statistical results from a large sample.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance:
Not applicable. This is an IVD device, not an AI-assisted diagnostic imaging device with human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
This refers to the "standalone" performance of the BioPlex® 2200 25-OH Vitamin D Kit itself. All the analytical performance studies (precision, linearity, detection limits, interference, cross-reactivity, method comparison) are studies of the device's standalone performance, as it quantifies 25-OH Vitamin D in human serum samples on the Bio-Rad BioPlex 2200 System. The device does not involve a human-in-the-loop for interpretive tasks in the way an AI diagnostic imaging tool would.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):
As indicated in point 3:
- For analytical studies (precision, linearity, detection limits, interference, cross-reactivity), the ground truth is established by analytical standards, known concentrations in spiked samples, or reference materials.
- For method comparison, the ground truth is the measurements obtained from the predicate device.
- For establishing expected values, the "ground truth" is derived from measurements in a defined healthy population.
8. The Sample Size for the Training Set:
The concept of a "training set" typically applies to machine learning or AI models. This document describes a traditional immunoassay system. Therefore, there is no "training set" in the machine learning sense.
However, the closest analogous concept might be the data used for calibration and value assignment.
- Calibrator Assignment: For each calibrator level (except Level 1), three vials were tested in replicates of five on three BioPlex 2200 analyzers for a total of 45 data points to assign values to the kit calibrators using Master calibrators as reference.
- Control Assignment: For each control level, three vials were tested in replicates of five using each of the kit lots on three BioPlex 2200 analyzers, totaling 45 replicates per reagent lot (90 for two lots, 135 for three lots used).
- The "Master calibrators" themselves were derived from internal standards quantified by UV spectrophotometric analysis.
9. How the Ground Truth for the Training Set Was Established:
Again, not a "training set" in the AI sense. For the calibrator and control assignment (analogous to establishing reference values for the system):
- Calibrator Traceability: The BioPlex 25-OH Vitamin D Calibrators are traceable to internal standards that are determined by UV spectrophotometric analysis using the known extinction coefficient of 25-OH Vitamin D at 264 nm wavelength. This is a primary analytical method to establish absolute concentration.
- Master Calibrators: Were manufactured volumetrically from these internal standards into depleted human serum.
- Lot-specific Calibrator and Control Assignment: Performed by testing multiple replicates on multiple analyzers against these Master Calibrators, with mean values having to fall within specified acceptable ranges.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is a stylized image of a caduceus, which is a symbol of medicine. The caduceus is made up of three lines that form the shape of a snake wrapped around a staff.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 9, 2015
BIO-RAD LABORATORIES JUANG WANG REGULATORY AFFAIRS REPRESENTATIVE 5500 E. 2ND STREET BENICIA CA 94510
Re: K14114
Trade/Device Name: BioPlex® 2200 25-OH Vitamin D Kit BioPlex® 2200 25-OH Vitamin D Calibrator Set, BioPlex® 2200 25-OH Vitamin D Control Set Regulation Number: 21 CFR 862.1825 Regulation Name: Vitamin D test system Regulatory Class: II Product Code: MRG, JJX, JIT Dated: November 25, 2014 Received: November 25, 2014
Dear Juang Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the
{1}------------------------------------------------
electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Stayce Beck -S
For :Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K141114
Device Name BioPlex® 2200 25-OH Vitamin D Kit BioPlex® 2200 25-OH Vitamin D Calibrator Set BioPlex® 2200 25-OH Vitamin D Control Set
Indications for Use (Describe)
The BioPlex® 2200 25-OH Vitamin D kit is a flow competitive immunoassay intended for the quantitative determination of 25-hydroxyvitamin D in human serum. The BioPlex 2200 25-OH Vitamin D assay is to be used as an aid in the assessment of vitamin D sufficiency.
The BioPlex® 2200 25-OH Vitamin D kit is intended for use with the Bio-Rad BioPlex 2200 System.
The BioPlex® 2200 25-OH Vitamin D Calibrator Set is intended for the BioPlex® 2200 25-OH Vitamin D Reagent Pack.
The BioPlex® 2200 25-OH Vitamin D Control Set is intended for use as an assayed quality control to monitor the overall performance of the BioPlex® 2200 System and the corresponding BioPlex® 25-OH Vitamin D Reagent Packs in the clinical laboratory. The performance of the BioPlex® 25-OH Vitamin D Control Set has not been established with any other 25-hydroxyvitamin D assays.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | |
|---|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
BioPlex® 2200 25-OH Vitamin D 510(k) Summary
Bio-Rad Laboratories hereby submits this 510(k) in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. This summary of 510(k) safety and effectiveness information provides detail as a basis for a determination of substantial equivalence for the BioPlex® 2200 25-OH Vitamin D kit.
510(k) Number:
Summary Preparation Date:
December 30, 2014
Applicant:
Bio-Rad Laboratories
Contact:
Juang Wang Regulatory Affairs Representative 5500 E. Second Street Benicia, CA 94510 Tel:510-741-4609 FAX: 510-741-3941 Juang wang@bio-rad.com
Purpose for Submission:
New Device
Measurand:
25-hydroxyvitamin D
Type of Test:
Quantitative multiplexed flow immunoassay
Proprietary and Established Names:
BioPlex® 2200 25-OH Vitamin D kit BioPlex® 2200 25-OH Vitamin D Calibrator Set BioPlex® 2200 25-OH Vitamin D Control Set
Regulatory Information:
- Regulation section: 21 CFR §862.1825 – Vitamin D test system 21 CFR §862.1150 - Calibrator 21 CFR §862.1660 - Quality Control Material (assayed and unassayed)
{4}------------------------------------------------
- Classification:
Class II (Assays, Calibrator) Class I (Controls)
- Product code:
MRG, System, Test, Vitamin D JIT, Calibrator, Secondary JJX, Single (specified) Analyte Controls (Assayed and Unassayed)
- Panel:
Clinical Chemistry (75)
Intended Use:
-
- Intended use(s):
The BioPlex® 2200 25-OH Vitamin D kit is a flow competitive immunoassay intended for the quantitative determination of 25-hydroxyvitamin D in human serum. The BioPlex 2200 25-OH Vitamin D assay is to be used as an aid in the assessment of vitamin D sufficiency.
- Intended use(s):
The BioPlex® 2200 25-OH Vitamin D kit is intended for use with the Bio-Rad BioPlex 2200 System.
The BioPlex® 2200 25-OH Vitamin D Calibrator Set is intended for the calibration of the BioPlex® 2200 25-OH Vitamin D Reagent Pack.
The BioPlex® 2200 25-OH Vitamin D Control Set is intended for use as an assayed quality control to monitor the overall performance of the BioPlex® 2200 System and the corresponding BioPlex® 25-OH Vitamin D Reagent Packs in the clinical laboratory. The performance of the BioPlex® 25-OH Vitamin D Control Set has not been established with any other 25-hydroxyvitamin D assays.
-
- Indication(s) for use: Same as Intended Use
-
- Special conditions for use statement(s): For prescription use only
- Special instrument requirements: 4. Bio-Rad BioPlex® 2200 System
Device Description:
BioPlex® 2200 25-OH Vitamin D kit includes the following components:
- One (1) 10 mL vial of Bead Set containing dyed beads coated with anti-25-OH D antibody (sheep), an Internal Standard bead (ISB), and a Serum Verification bead
{5}------------------------------------------------
(SVB) in buffer with protein stabilizers (bovine). ProClin 950 (< 1.0%) and sodium azide (< 0.1%) as preservatives.
- . One (1) 10 mL vial of Release Buffer containing 25-OH Vitamin D releasing reagents in citrate and trisodium citrate acid buffer at pH 4.1 and ProClin 950 (<1.0%) as preservative.
- One (1) 5 mL vial of Conjugate 1 containing biotinylated 25-OH Vitamin D conjugate and biotinylated anti-human FXIII antibody conjugate (murine) in buffer with protein stabilizers (bovine). ProClin 950 (< 1.0%) and 5-bromo-5nitro-1. 3-dioxane (<0.1%) as preservatives and chemical blockers.
- . One (1) 5 mL vial of Conjugate 2 containing phycoerythrin conjugated streptavidin (SA -PE) in a buffer comprising protein stabilizers (bovine). ProClin 950 (< 1.0%) and sodium azide (< 0.1%) as preservatives, chemical blockers and detergent (Tween 20).
BioPlex® 2200 25-OH Vitamin D Calibrator set contains six (6) 0.5 mL 25-OH Vitamin D vials. Calibrator level 1 contains 25% horse serum without 25-OH Vitamin D. The calibrator levels 2 to 6 are provided in a Vitamin D depleted human serum matrix supplemented with known concentration of 25-hydroxyvitamin D3. All calibrators contain ProClin 950 (≤ 0.3%), sodium benzoate (≤ 0.1%) and 5-bromo-5-nitro-1, 3dioxane (<0.1%) as preservatives.
BioPlex® 2200 25-OH Control set contains two (2) 1.5 mL Level 1 and two (2) 1.5 mL Level 2 Control vials, each containing 25-OH Vitamin D in human serum matrix. All controls contain ProClin 950 (< 0.3%), sodium benzoate (< 0.1%) and 5-bromo-5-nitro-1, 3-dioxane (<0.1%) as preservatives.
Additional materials required but not supplied include BioPlex® 2200 Sheath Fluid containing Phosphate Buffered Saline (PBS), ProClin 300 (0.03%) and sodium azide (<0.1%) as preservatives; and BioPlex® 2200 Wash Solution containing Phosphate Buffered Saline (PBS) and Tween 20. ProClin 300 (0.03%) and sodium azide (<0.1%) as preservatives.
Substantial Equivalence Information:
-
- Predicate device name(s): EUROIMMUN 25-OH Vitamin D ELISA, K123660
-
- Comparison with predicate:
{6}------------------------------------------------
| Device Similarities | ||
|---|---|---|
| Characteristics | New DeviceBioPlex 2200 25-OHVitamin D Kit | Predicate DeviceEUROIMMUN 25-OHVitamin D ELISAK123660 |
| IntendedUse/Indication forUse | Intended for thequantitativedetermination of 25-hydroxyvitamin D inserum. To be used as anaid in the assessment ofvitamin D sufficiency | Same |
| Measured Analyte | 25-hydroxyvitamin D | Same |
| Assay Type | Quantitative | Same |
| Test Principle | Competitiveimmunoassay | Same |
| Antibody | Monoclonal Sheepantibody against 25-OHVitamin D | Same |
| Signal Detection | Fluorescence | Same |
| Unit of Measure | ng/mL | Same |
| Calibrator(s) andCalibration | 6 calibrator levels (soldseparately); 4-PL(parameter logistic)curve fit algorithm | Same |
| Device Differences | ||||||
|---|---|---|---|---|---|---|
| Characteristics | New Device | Predicate Device | ||||
| BioPlex 2200 25-OH | EUROIMMUN 25-OH | |||||
| Vitamin D Kit | Vitamin D ELISA | |||||
| K123660 | ||||||
| Assay Technology | Automated flow | Manual competitive | ||||
| competitive immunoassay | immunoassay | |||||
| Conjugate | Biotin-labeled 25-OH | |||||
| hydroxyvitamin D and | vitamin D, Peroxidase- | |||||
| phycoerythrin conjugated | labeled streptavidin and | |||||
| streptavidin | substrate TMB | |||||
| Solid Phase | Antibody-coated | Antibody coated 96 | ||||
| paramagnetic microbeads | microwell ELISA plate | |||||
| Measuring range | 6.5 – 125.0 ng/mL | 4 – 120 ng/mL | ||||
| Sample Matrix | Serum | Serum or EDTA or | ||||
| Lithium heparin plasma | ||||||
| Sample Size | 10 µL | 20μL | ||||
| Calibrator Matrix | 25% horse serum and | Liquid in horse serum | ||||
| depleted human serum | with preservatives |
{7}------------------------------------------------
| Device Differences | ||
|---|---|---|
| Characteristics | New DeviceBioPlex 2200 25-OHVitamin D Kit | Predicate DeviceEUROIMMUN 25-OHVitamin D ELISAK123660 |
| with ProClin 950, sodium benzoate and BND | ||
| Open Pack Stability | 60 days | Not applicable |
| Reagent Integral Storage | On-board or in refrigerator at 2-8°C | Not applicable |
| Sample Handling/Process | Automated | Manual |
| Calibrator Open storage at 2-8°C | 30 days | 3 months |
| Calibration Frequency | Every 30 days | Every 96 well plate |
| Instrumentation | Bio-Rad BioPlex 2200 System | ELISA plate reader |
| Measuring wavelength | 550 - 610 nm | 450/620 nm |
| Control Set Similarities and Differences | ||
|---|---|---|
| Characteristics | BioPlex 2200 25-OHVitamin D Kit | Predicate DeviceEUROIMMUN 25-OHVitamin D ELISAK123660 |
| Intended Use | Use as an assayed qualitycontrol to monitor the overallperformance of 25-OHVitamin D reagent. | Same |
| Storage | Store at 2 -8 °C until ready touse | Same |
| Matrix | Human serum with ProClin950, sodium benzoate andBND | Liquid in horseserum withpreservatives |
| Control OpenStability at 2 – 8 °C | 60 days | No Applicable |
Standard/Guidance Document Referenced (if applicable):
EP05-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline, Second Edition (Vol. 24 No.25)
EP06-A, Evaluation of Linearity of Quantitative Measurement: A Statistical Approach, Approved Guideline (Vol. 23 No.16)
{8}------------------------------------------------
EP07-A2, Interference Testing in Clinical Chemistry, Approved Guideline, Second Edition (Vol. 25 No.27)
EP09-A2IR, Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline - Second Edition (Interim Revision) (Vol. 30 No. 17)
EP15-A2. User Verification of Performance for Precision and Trueness. Approved Guideline, Second Edition (Vol. 25 No.17)
EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline, Second Edition (Vol. 32 No.8)
EP25-A, Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline (Vol. 29, No. 20)
C28-A3c, Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline, Third Edition (Vol. 28 No.30)
Test Principle:
The BioPlex® 2200 25-OH Vitamin D assay is a multiplesflow competitive immunoassay for the quantitative determination of 25-hydroxyvitamin D in human serum.
The BioPlex 2200 System combines an aliquot of patient sample with the Vitamin D Release Buffer to dissociate the 25-hydroxyvitamin D from its binding protein. After the first incubation, the bead reagent is added to the reaction vessel and incubated at 37°C. After the second incubation, the BioPlex 2200 System adds the vitamin D-biotin conjugate 1. The excess conjugate 1 is removed during a wash cycle and the streptavidin-phycoerythrin (SA-PE) conjugate 2 is added. The excess conjugate 2 is removed during a wash cycle, and the beads are re-suspended in wash buffer. The bead mixture then passes through the detector. The identity of the dyed beads is determined by the fluorescence of the dyes, and the amount of 25-hydroxyvitamin D captured is inversely proportional to the fluorescence of the attached SA-PE. Raw data is calculated in relative fluorescence intensity (RFI). Two additional dyed beads, Internal Standard Bead (ISB) and Serum Verification Bead (SVB) are present in each reaction mixture to verify detector response and the addition of serum to the reaction vessel, respectively.
The BioPlex 25-OH Vitamin D assay is calibrated using a set of 6 distinct calibrators supplied separately by Bio-Rad Laboratories. Results are calculated from a 4-PL (parameter logistic) calibration curve in which an inverse relationship exists between the amount of 25-hydroxyvitamin D in the patient sample and the amount of RFI detected by the system. The results are expressed in ng/mL.
Performance Characteristics (if/when applicable):
-
- Analytical performance:
- a. Precision/Reproducibility:
Precision testing of the BioPlex® 2200 25-OH Vitamin D kit on the BioPlex® 2200 instrument was performed in accordance with CLSI EP5-A2 guideline. A human serum panel consisting of 6 frozen samples spanning the measuring range was assayed in duplicate per run on two runs daily over 20 days (N=80) on one
{9}------------------------------------------------
reagent lot. Two levels of the BioPlex 25-OH Vitamin D controls were also included. The data were analyzed for within-run, between-run, between-day, and total precision and the mean (ng/mL), standard deviation (ng/mL) and percent coefficient of variation (%CV) are summarized below:
| SerumPanel | N | Meanng/mL | Within Run | Between Run | Between Day | Total Precision | ||||
|---|---|---|---|---|---|---|---|---|---|---|
| SD | %CV | SD | %CV | SD | %CV | SD | %CV | |||
| Sample 1 | 80 | 15.0 | 1.19 | 7.9 | 0.69 | 4.60 | 0.77 | 5.2 | 1.58 | 10.5 |
| Sample 2 | 80 | 17.2 | 1.02 | 5.9 | 0.86 | 5.00 | 1.19 | 6.9 | 1.78 | 10.4 |
| Sample 3 | 80 | 36.1 | 1.61 | 4.5 | 1.30 | 3.60 | 1.56 | 4.3 | 2.59 | 7.2 |
| Sample 4 | 80 | 47.2 | 2.20 | 4.7 | 1.24 | 2.60 | 1.85 | 3.9 | 3.13 | 6.6 |
| Sample 5 | 80 | 77.9 | 1.99 | 2.6 | 1.42 | 1.80 | 1.98 | 2.5 | 3.15 | 4.0 |
| Sample 6 | 80 | 110.8 | 3.60 | 3.2 | 3.00 | 2.7 | 2.62 | 2.4 | 5.37 | 4.8 |
| Control 1 | 80 | 22.1 | 1.30 | 5.9 | 0.87 | 3.9 | 1.04 | 4.7 | 1.88 | 8.5 |
| Control 2 | 80 | 50.0 | 2.38 | 4.8 | 1.18 | 2.4 | 1.52 | 3.0 | 3.07 | 6.1 |
BioPlex® 2200 25-OH Vitamin D -- CLSI EP5-A2 Precision
CLSI EP15-A2 Reproducibility
Precision and reproducibility was also evaluated in accordance with CLSI EP15-A2 guideline "User Verification of Performance for Precision and Trueness, Vol 25, No 17".
A different serum panel consisting of 8 samples spanning the measuring range were assayed in 2 replicates per run, two runs per day over 5 days (n=20) using one lot of BioPlex 25-OH Vitamin D kit. Two levels of controls were also included. The data were analyzed for within-run, between run, between day, and total precision and the mean ng/mL, standard deviation and percent coefficient of variation (%CV) are summarized below:
| Serum | Mean | Within Run | Between Run | Total Precision | ||||
|---|---|---|---|---|---|---|---|---|
| Panel | N | ng/mL | SD | %CV | SD | %CV | SD | %CV |
| Sample 1 | 20 | 11.5 | 0.69 | 6.0% | 0.47 | 4.1% | 1.71 | 14.8% |
| Sample 2 | 20 | 13.6 | 0.80 | 5.9% | 0.26 | 1.9% | 1.18 | 8.7% |
| Sample 3 | 20 | 26.1 | 0.88 | 3.4% | 1.15 | 4.4% | 1.59 | 6.1% |
| Sample 4 | 20 | 30.2 | 1.99 | 6.6% | 0.00 | 0.0% | 2.71 | 9.0% |
| Sample 5 | 20 | 50.2 | 2.23 | 4.4% | 0.77 | 1.5% | 2.96 | 5.9% |
| Sample 6 | 20 | 56.4 | 2.09 | 3.7% | 2.79 | 4.9% | 5.26 | 9.3% |
| Sample 7 | 20 | 100.5 | 4.52 | 4.5% | 2.81 | 2.8% | 5.32 | 5.3% |
| Sample 8 | 20 | 104.9 | 3.97 | 3.8% | 1.54 | 1.5% | 5.24 | 5.0% |
| Control 1 | 20 | 21.6 | 0.98 | 4.5% | 1.00 | 4.6% | 1.84 | 8.5% |
| Control 2 | 20 | 58.8 | 2.44 | 4.2% | 1.29 | 2.2% | 2.99 | 5.1% |
BioPlex® 2200 25-OH Vitamin D - CLSI EP15-A2 Reproducibility
{10}------------------------------------------------
b. Linearity/assav reportable range:
Five high patient serum samples extending 20% higher than upper limit of the assay range were tested to demonstrate linearity. These samples were serially diluted with low levels of human sample near LoO in accordance with CLSI EP06-A guideline. Each sample and dilution was evaluated in replicates of four using one BioPlex 25-OH Vitamin D reagent lot on one instrument. Linear and polynomial regression analysis of 25-OH Vitamin D recovery vs. sample dilution was performed to determine if the dilution curves exhibit statistically significant non-linear regression based on the CLSI guideline EP06-A.
See one example below for the regression parameters (slope, intercept and r') of the observed values vs. predicted values.
| Conc(ng/mL) | Slope | Intercept | r2 | Dilution range |
|---|---|---|---|---|
| 168.9 | 1.0001 | 0.0045 | 0.9988 | 5.5 - 168.9 |
The BioPlex 2200 25-OH Vitamin D assay has demonstrated that the assay range is 6.5 to 125.0 ng/mL.
Over-Range (OR) results may be generated for values greater than the reportable measuring range and results are reported as > 125.0 ng/mL.
- Traceability, Stability, Expected values (controls, calibrators, or methods): c.
Traceability:
The BioPlex 25-OH Vitamin D Calibrators are traceable to internal standards which are determined by UV spectrophotometric analysis using the extinction coefficient of 18,000 mL/mmol/cm at 264 nm wavelength.
The six levels of Master calibrators are manufactured volumetrically from the internal standards into the depleted human serum except level 1 (25hydroxyvitamin D free) in 25% horse serum. The Master calibrators are immediately frozen at <-70°C.
Value Assignment:
The BioPlex 25-OH Vitamin D kit calibrators are manufactured for each lot kit with the same matrix as the Master calibrators and are stabilized with <0.3% ProClin 950, <0.1% sodium benzoate, and <0.1% 5-bromo-1.3-nitro-dioxane.
Calibrator assignment is established for the matched lot of BioPlex® 2200 25-OH Vitamin D kit using the Master calibrators as reference. Calibrator assignment is established for the matched lot of BioPlex® 2200 25-OH Vitamin D kit using the Master calibrators as reference. For each calibrator level except level 1, three vials are tested in replicates of five on three BioPlex 2200 analyzers for a total of 45 data points. The mean values obtained for each kit calibrator level are verified and must fall within specified acceptable range.
{11}------------------------------------------------
Two levels of the BioPlex 2200 25-OH Vitamin D control set are prepared from a pool of native human serum specimens. For each control level, three vials are tested in replicates of five using each of the kit lots on three BioPlex 2200 analyzers for a total of 45 replicates per reagent lot. The total number of replicates for each control level is 90 when two reagent lots are used and 135 when three reagent lots are used. For each control level, the mean values were derived from replicate analyses and should fall within the corresponding deviation.
The manufacturing target values of the Calibrator and Control Sets are listed below.
| Calibrator Set | Target (ng/mL) |
|---|---|
| Level 1 | 0.0 |
| Level 2 | 10.0 |
| Level 3 | 30.0 |
| Level 4 | 75.0 |
| Level 5 | 110.0 |
| Level 6 | 165.0 |
| Control Set | Target (ng/mL) | Range (ng/mL) |
|---|---|---|
| Level 1 | 19.0 | 14.5 - 23.5 |
| Level 2 | 55.0 | 45.0 - 65.0 |
Stability:
Stability studies have been performed to support the following claims:
Calibrator and Control:
BioPlex® 2200 25-OH Control and Calibrator Sets: Calibrator Open Vial Stability (2 to 8°C), 30 days from first opening; Control Open Vial Stability (2 to 8°C), 60 days from first opening; Onboard Calibration Curve Stability, 30 days; Calibrators and Controls Real Time Stability (2 to 8°C), 24 months; labeled as until expiration date; Calibrators and Controls Accelerated Stability (2 to 8°C), 2 years predicted. Calibrators freeze-thaw (-20℃ or -70℃), 5-freeze thaw cycles; Control freeze-thaw (-20℃ or -70℃), 1-freezethaw cycle at -20℃ and 5-freezethaw cycles at -70°C.
Kit Stability:
BioPlex® 2200 25-OH Vitamin D Kit: Real Time (unopened) Kit Stability, 9 months or until the date of expiration when stored unopened on the instrument or at 2 to 8°C; the open kit claim is 60 days.
Sample Stability:
Sample stability studies were also performed: Sample stability fresh (2 to 8°C), 7 days; Sample stability frozen (-20 or -70°C), 24 months; Sample Freeze-thaw (-20 or -70°C), up to 3-freeze thaw cycles at -20°C and 2-freeze thaw cycles at -
{12}------------------------------------------------
70℃ acceptable.
- d. Detection limit:
The study was conducted in accordance with CLSI EP17-A2 guideline for determining the Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantitation (LoQ).
Limit of Blank (LoB)
Five blank samples were tested with two BioPlex 25-OH Vitamin D reagent lots on one instrument in 4 replicates per day for 5 days yielding 100 data points per reagent lot.
A non-parametric statistical analysis at 95th percentile is used to calculate LoB.
Limit of Detection (LoD)
Six human samples with low level of 25-OH vitamin D in the range of 4 to 35 ng/mL were tested with two BioPlex 25-OH Vitamin D reagent lots on one instrument in 12 replicates per day for five days yielding 60 data points per sample per reagent lot. LoD is then calculated by the equation:
LoD = LoB + c "SD + c "SD is a multiplier to give the 95th percentile of a normal distribution and SD is from the linear regression of standard deviation versus 25-OH Vitamin D mean value.
Limit of Quantitation (LoQ)
The LoQ was evaluated based on the accuracy goal which was defined as precision ≤ 20% CV. The %CV was calculated using the same measurement results of the 6 low level samples used for determining the LoD.
The results of LoB, LoD, and LoO in ng/mL are summarized in the table below.
| LoB | LoD | LoQ |
|---|---|---|
| 0.8 | 2.5 | 6.5 |
e. Analytical specificity:
Interfering Substances:
An interfering substances study was conducted to evaluate the potential interference of specific endogenous and exogenous substances with the BioPlex® 2200 25-OH Vitamin D kit according to CLSI EP7-A2 guideline.
The effects of the test levels of potential interfering substances on the assay have been evaluated with samples containing 10 to 90 ng/mL Vitamin D. The percent difference between the mean value of each test substance and a corresponding control was calculated. No interference was observed with any of the substances
{13}------------------------------------------------
tested if the percent difference is ≤ 10%. The substances and the maximum levels of interfering substances tested are shown in the table below:
| Substance | Concentration |
|---|---|
| Hemoglobin* | ≤150 mg/dL |
| Bilirubin (unconjugated) | ≤20 mg/dL |
| Bilirubin (conjugated) | ≤30 mg/dL |
| Triglycerides | ≤400 mg/dL |
| Total Protein | ≤12 g/dL |
| Cholesterol | ≤500 mg/dL |
| Uric Acid | ≤20 mg/dL |
| HAMA | ≤100 ng/mL |
| Rheumatoid Factor | ≤350 IU/mL |
| Ascorbic Acid | <3 mg/dL |
*Hemoglobin > 150 mg/dL may interfere. Do not use visibly hemolyzed samples.
Cross-Reactivity:
The study was conducted in accordance with CLSI EP17-A2 using 2 human serum pools at 25-hydroxyvitamin D concentrations of 20 and 35 ng/mL. Nine cross reactants at levels listed below were then spiked into the human serum pools. The spiked and non-spiked samples were then evaluated in replicates of five to calculate the cross reactivity as shown below.
% Cross Reactivity = (spiked vitamin D - non-spiked vitamin D) = Cross reactant concentration × 100%
The results of each potential cross reactant are listed below.
| Cross Reactant | SpikedConcentration(ng/mL) | % CrossReactivity |
|---|---|---|
| 25-hydroxyvitamin D2 | 30 | 103% |
| 25-hydroxyvitamin D3 | 30 | 97% |
| Vitamin D2 | 1000 | 0.2% |
| Vitamin D3 | 1000 | 0.0% |
| 1,25-dihydroxyvitamin D2 | 30 | >100% |
| 1,25-dihydroxyvitamin D3 | 30 | 79% |
| 3-epi 25-hydroxyvitamin D3 | 30 | 59% |
| 24,25-dihydroxyvitamin D3 | 20 | 9% |
| Paricalcitol (Zemplar)* | 24 | >100% |
- Paricalcitol (Zemplar) has been found to cross-react and interfere with the BioPlex 2200 25-OH Vitamin D assay
{14}------------------------------------------------
High dose hook effect: Not Applicable
f. Assay cut-off:
Not Applicable
2. Comparison studies:
- a. Method comparison with predicate device: Method comparison studies were performed following CLSI EP09-A2-IR guideline.
A total of 204 human samples spanning the entire measuring assay range were tested in singlicate on both the BioPlex 2200 25-OH Vitamin D kit and the predicate assay. Of the 204 samples, there were 185 unaltered samples and 19 samples spiked with 25-hydroxyvitamin D3 to supplement the assay range. There are eight (8) samples with values lower or higher than the measuring range of the comparator method not including in the analysis. A total of 196 BioPlex 25-0H Vitamin D results were plotted using weighted Deming regression analysis for all samples spanning the measuring range of both assays. Results of the regression slope, intercept, and coefficient of correlation (r) are summarized in the table below:
| Number ofResultsAnalyzed | Slope(95% CI) | Intercept(95% CI) | Correlation Coefficient (r)(95% CI) | Test Range (ng/mL) |
|---|---|---|---|---|
| 196 | 1.0039(0.9365 to 1.0712) | -0.2256(-2.4121 to1.9608) | 0.9553(0.9412 to 0.9661) | BioPlex: 6.6 to 124.9Comparator: 4.3 to118.1 |
- b. Matrix comparison: Serum only
-
- Clinical studies:
- a. Clinical Sensitivity: Not Applicable
- b. Clinical Specificity: Not Applicable
- c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable
{15}------------------------------------------------
4. Clinical cut-off: Not Applicable
5. Expected values/Reference range:
The Expected Values study was conducted following CLSI C28-A3c guideline.
Two hundred and eighty-seven (287) samples from apparently healthy donors including 160 males ranging in age from 21 to 79 and 127 females ranging in age from 21 to 66 were collected from three regions (North, Central, and South) in the US in spring, summer and winter, including African Americans, Hispanics and Caucasians.
The 287 samples from apparently healthy donors met the following inclusion/exclusion criteria as follows and tested with the BioPlex 25-OH Vitamin D kit in singlicate.
- Age from 21 to 90 ●
- Roughly 50% female and 50% male
- 20% from Northern. 20% from Central and 60% from Southern region
- 40% collected in Spring, 30% in Summer and 30% in Winter
- At least 30% African Americans and 30% Caucasians
- . 90% not taking Vitamin D supplements and <30% of those taking Vitamin D supplements at or more than 1000 IU, but less than 2000 IU
- . Normal TSH, PTH, and Total Calcium
- No family history of parathyroid or calcium regulatory disease. In addition, . no personal history of kidney disease, GI disease, liver disease, and no bariatric surgery
The observed median, mean, and range between 2.5th to 97.5th percentile are summarized below
| Mean | Median | 2.5th to 97.5th Percentile | |
|---|---|---|---|
| 286* | 29.7 ng/mL | 27.7 ng/mL | 12.7 - 65.7 ng/mL |
- One sample <6.5 ng/mL was excluded from the data analysis
Each laboratory should establish its own reference range pertinent to their specific patient populations.
Instrument Name:
The BioPlex 2200 System, software version 4.1 cleared in K130053
Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
§ 862.1825 Vitamin D test system.
(a)
Identification. A vitamin D test system is a device intended for use in clinical laboratories for the quantitative determination of 25-hydroxyvitamin D (25-OH-D) and other hydroxylated metabolites of vitamin D in serum or plasma to be used in the assessment of vitamin D sufficiency.(b)
Classification. Class II (special controls). Vitamin D test systems must comply with the following special controls:(1) Labeling in conformance with 21 CFR 809.10 and
(2) Compliance with existing standards of the National Committee on Clinical Laboratory Standards.