(231 days)
EUROIMMUN's 25-OH Vitamin D ELISA is intended for the quantitative determination of 25-OH Vitamin D and other hydroxylated vitamin D metabolites in human serum and plasma (EDTA, Li-heparin). Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of vitamin D sufficiency in adult populations.
EUROIMMUN 25-OH Vitamin D ELISA consists of a microwell ELISA plate coated with (sheep) monoclonal anti-25-OH vitamin D antibodies, 6 calibrators, 2 controls, Biotin, sample buffer, conjugate. wash buffer concentrate. enzyme TMB chromogen/substrate solution and stop solution.
This ELISA test kit is designed for the in vitro determination of 25-OH vitamin D in human serum or plasma samples. In the first analysis step, the calibrators and patient samples are diluted with biotin-labelled 25-OH vitamin D and added to microplate wells coated with (sheep) monoclonal anti-25-OH vitamin D antibodies. During the incubation an unknown amount of 25-OH vitamin D in the patient sample and a known amount of biotinlabelled 25-OH vitamin D compete for the antibody binding sites in the microplate wells plate. Unbound 25-OH vitamin D is removed by washing. For the detection of bound biotin-labelled 25-OH vitamin D, a second incubation is performed using peroxidase-labelled streptavidin. In a third incubation using the peroxidase substrate tetramethylbenzidine (TMB) the bound peroxidase promotes a colour reaction. The colour intensity is inversely proportional to the 25-OH vitamin D concentration in the sample. Results for the samples can be calculated directly using a standard curve.
Antibodies: sheep monoclonal antibodies which identify specifically 25-OH vitamin D3 and 25-OH vitamin D2.
Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Characteristic | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Reproducibility | CV's < 12% for samples > 10 ng/ml; CV's < 20% for concentrations < 10 ng/ml | Intra-Assay CVs: (Data from table is unreadable; however, the text states "All results passed the acceptance criteria.") Inter-Assay CVs: (Data from table is unreadable; however, the text states "All results passed the acceptance criteria.") Lot-to-Lot CVs: |
| Limit of Blank (LoB) | N/A (determined for the device) | 0.4 ng/ml |
| Limit of Detection (LoD) | N/A (determined for the device) | 1.8 ng/ml |
| Functional Sensitivity (FS) | Lowest concentration at which the potential regression line crosses the 20% CV line | 4.0 ng/ml (Results below 4 ng/mL are reported as "< 4 ng/mL") |
| Linearity/Reportable Range | Amount of nonlinearity above functional sensitivity of 4.0 ng/ml acceptable below 15%. Sufficiency linear from 4.0 to 120 ng/ml | Although polynomial regression p-values were significant (< 0.05) for two of three sample sets, the amount of nonlinearity above 4.0 ng/ml was found acceptable below 15%. Conclusion: Assay is sufficiently linear from 4.0 to 120 ng/ml. |
| Interference | Individual recovery within 90 – 110% when spiked with potential interfering substances. | Hemoglobin (up to 750 mg/dl): 90 - 110% recovery. Triglycerides (up to 2000 mg/dl): 98 - 110% recovery. Bilirubin (up to 40 mg/dl): 90 - 109% recovery. Cholesterol (up to 400 mg/dl): 94 – 109% recovery. Biotin (up to 1000 mg/dl): 91 - 107% recovery. Ascorbic acid (up to 10.0 mg/ml): 94 - 110% recovery. Note: Significant interference (> 10% deviation) seen with 1000 mg/dl of hemoglobin. |
| Method Comparison (vs. Predicate) | High correlation to predicate device | n: 240 samples. Concentration Range (Predicate): 4.1 - 110.4 ng/ml. Concentration Range (Candidate): 4.1 - 119.1 ng/ml. Regression Equation (y = Candidate, x = Predicate): y = 0.78 + 1.08 x. 95% C.I. of Intercept: -0.06 - 1.63. 95% C.I. of Slope: 1.06 - 1.11. Correlation Coefficient R: 0.9858. 95% C.I. of R: 0.9817 - 0.9890. |
| Matrix Comparison (Plasma vs. Serum) | Regression equation near ideal correlation (intercept 0; slope 1.0); Coefficients of determination > 0.99; %recovery compared to serum in range of 87 to 127 % | EDTA plasma (n=38): Y = 0.29 + 0.99x (95% CI Intercept: -0.37 - 1.18, 95% CI Slope: 0.93 - 1.02); R² = 0.996; Mean %recovery = 100% (Range: 87 - 116%). Li-heparin plasma (n=38): Y = 0.55 + 0.97x (95% CI Intercept: -0.65 - 1.33, 95% CI Slope: 0.93 - 1.04); R² = 0.993; Mean %recovery = 101% (Range: 87 - 127%). |
| Stability | Predicted shelf-life of at least 12 months at 2-8°C for control materials and calibrators; Open-vial stability of 3 months at 2-8°C for kit. | Predicted shelf-life of at least 12 months at 2-8°C for control materials and calibrators (based on accelerated testing at 37°C, real-time ongoing). Open-vial stability of 3 months when stored at 2-8°C. Biotin stable for at least 2 weeks when diluted to working strength. |
2. Sample Sizes Used for the Test Set and Data Provenance
- Reproducibility:
- Intra- and Inter-Assay: Minimum of 8 natural serum samples (2 spiked) for CV determination; 40 determinations for intra-assay CVs; 40 determinations performed in 10 different runs on 5 different days (with 4 replicates per run) for inter-assay CVs.
- Lot-to-Lot: Minimum of 8 natural serum samples (2 spiked); 32 determinations performed in 8 different runs on 4 different lots (with 2 runs per lot and 4 replicates per run).
- Data Provenance: "natural serum samples collected freshly in-house from obvious healthy blood donors".
- LoD: 200 determinations from 5 samples in the low range (2-10 ng/ml), measured in 5 independent runs with 8 replicates per run.
- Linearity/Assay Reportable Range: Sets of 11 sample preparations (natural negative sample + high positive samples), tested in double determinations.
- Cross-Reactivity: 25-OH Vitamin D free sample aliquoted and spiked with 7 potential cross-reacting Vitamin D metabolites.
- Interference: Sera at different 25-OH Vitamin D concentrations spiked with 6 potential interfering substances.
- Method Comparison w/Predicate Device:
- Sample Size: 240 samples in total.
- Data Provenance:
- 141 prospective samples for 25-OH-Vitamin D testing from a clinical laboratory.
- 28 samples from 25-OH-Vitamin D quality assessment programs (harvested from blood donated by venesection undergoing therapeutic for patients with hemochromatosis or polycythemia).
- 5 samples from a 25-OH-Vitamin D quality control panel (untreated routine patient material).
- 30 samples from normal blood donors.
- 36 samples (15%) were spiked with 25-OH Vitamin D3 stock solution to ensure concentration distribution.
- All patients gave informed consent. Samples are natural and not treated.
- Matrix Comparison: 38 sample pairs of serum and corresponding plasma (EDTA, Li-heparin) from donors. 3 sample pairs were spiked.
- Expected Values/Reference Range: 206 samples from healthy subjects (70 men, 136 women; average age 64 years; age range: 22–99 years) from a US commercial source (normal US blood donors from the US Midwest region, drawn early Oct 2010).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not explicitly state the number of experts or their qualifications for establishing ground truth for the test set in terms of clinical interpretation. However, the performance studies (e.g., reproducibility, linearity) rely on in-house personnel and laboratory procedures, with calibrator and control value assignments confirmed by "2 different technicians" and "a different person for final release." Method comparison was against an "FDA-cleared reference assay," implying the predicate device serves as a reference for comparison.
4. Adjudication Method for the Test Set
No specific adjudication method (e.g., 2+1, 3+1) is mentioned or implied for the evaluation of the test set results for clinical interpretation. The performance studies describe analytical methods and statistical comparisons.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is an in-vitro diagnostic (IVD) assay, not an imaging device typically evaluated with MRMC studies. The "Method Comparison" section compares the device's analytical results against a predicate device, not human readers with and without AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, the studies described are for the standalone performance of the EUROIMMUN 25-OH Vitamin D ELISA assay, which is an algorithm-only (test kit) device. Its performance characteristics like linearity, reproducibility, limits of detection, and interference are evaluated intrinsically, and its results are then compared to a predicate device. There is no human-in-the-loop component for the device's operation, only human involvement in performing the laboratory tests (technicians) and interpreting the output for clinical use.
7. The Type of Ground Truth Used
The ground truth used varies based on the type of study:
- Reproducibility, LoB, LoD, Functional Sensitivity, Linearity, Cross-Reactivity, Interference: The ground truth is effectively the known concentrations of the analytes in the prepared samples (e.g., spiked samples, calibrators, controls, or precisely characterized samples). These are established through gravimetric methods, UV spectrophotometric analysis, and comparison with NIST and DEQAS standards, as well as predicate device measurements and HPLC.
- Method Comparison w/Predicate Device: The ground truth for comparison is the results obtained from the FDA-cleared predicate device, which is considered the reference method for this type of comparison.
- Matrix Comparison: The ground truth is the serum sample concentrations when comparing to plasma.
- Expected Values/Reference Range: The ground truth for establishing population reference intervals is the 25-OH Vitamin D levels measured in healthy subjects using the new device itself.
8. The Sample Size for the Training Set
The document does not describe a "training set" in the context of machine learning, as this is an in-vitro diagnostic ELISA kit. The term "training set" is not applicable here. The assay is based on chemical reactions and optical density measurements, not a learned algorithm in the typical sense.
9. How the Ground Truth for the Training Set was Established
As noted in #8, there isn't a "training set" in the machine learning context. The calibration and standardization of the assay (which could be considered analogous to "establishing ground truth" for the assay's internal function) involves:
- Gravimetric calibration using UV-Vis verified stock standards.
- Comparison with NIST standards and DEQAS quality assessment data.
- In-house quality control sera.
- Initial values for quality control sera assigned using the predicate device in conjunction with liquid chromatography (HPLC).
- Final calibrator values verified and assigned by adjusting initial values to meet specified ranges when tested against predicate assays.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "EUROIMMUN" in large, bold, black letters. To the right of the word is a small, pixelated image. Next to the pixelated image are the letters "US" in the same font as the word "EUROIMMUN".
. اُ
510(K) SAFETY & EFFECTIVENESS SUMMARY
| Submitter | EUROIMMUN US INC. | |
|---|---|---|
| 1100 The American Road | ||
| Morris Plains, NJ 07950 | ||
| Tel: 973.656.1000/1.800.913.2022 | JUL 17 2013 | |
| Fax: 973.656.1098 | ||
| Michael Locke | ||
| Director of Regulatory Affairs | ||
| E-mail: M.Locke@euroimmun.us | ||
| Date Prepared | July 3, 2013 | |
| Submission Type | 510(k) | |
| Device Trade/Proprietary Names | 25-OH Vitamin D ELISA | |
| Device Common/Usual Name or Classification Name | Vitamin D Test System | |
| Classification Number/Class | 21 CFR § 862.1825 - Vitamin D Test System - Class II | |
| Product Code | MRG - Vitamin D Test system | |
| Panel | Clinical Chemistry and Toxicology |
As required by 21 CFR § 807.92, the following is in sufficient detail to provide an understanding of the basis for a determination of substantial equivalence.
510(k) Number
Standard/Guidance Document Referenced (if applicable): None referenced.
. .
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "EUROIMMUN" followed by a black rectangle and the letters "US". The text is in a bold, sans-serif font and is underlined. The black rectangle appears to be obscuring some text or a logo.
510(K) SAFETY & EFFECTIVENESS SUMMARY
Traceability /Value Assignment of Calibrators and Controls
As there is no international standard, the Traceability: calibrators and controls are calibrated gravimetrically using UV-Vis (264nm) verified stock standards and compared with NIST standards (National Institute of Standards and Technology, USA), DEQAS (Vitamin D External Quality Assessment Scheme, UK) quality assessment data and in-house quality control sera.
Calibrators and controls are traceable to concentrations determined by UV spectrophotometric analysis. An in-house stock solution is prepared gravimetrically; and the antigen concentration is spectrophotometrically calculated using the OD coefficient of 18.2 at 264 nm to calculate the concentration from the absorbance value. This is used to build an intermediate stock volumetrically by dilution into horse serum. The intermediate stock is used in the manufacturing of lot specific calibrators and controls volumetrically.
- · The calibrator levels for each batch are confirmed by running a min. of 8 quality control sera covering the whole concentration range of the calibration curve measured in a minimum of 4 runs performed by 2 different technicians on 2 different days in duplicates. The quality control sera must fall within established target ranges. Initial values for the quality control sera are assigned using the predicate device in conjunction with liquid chromatography. Once confirmation of the calibrator values is established. the new calibrators are retested by a different person for final release.
- · 25-OH Vitamin D controls prepared in human serum were tested with 7 different released EUROIMMUN 25-OH Vitamin D lots, obtained from 20 independent runs, performed by 4 different persons on 9 different days. The obtained mean values were assigned as control target values. The control target ranges were assigned as mean ±3 SD (ng/mL). Target values were confirmed using two predicate devices (K021163; K112725) and a HPLC method.
Calibrators and controls are supplied in liquid form and are horse serum based with active ingredients of 0.09% ProClin 950 and 0.09% sodium azide. Avoid skin contact. Calibrators are 25-OH vitamin D3 spiked. No special treatment necessary for the use of the calibrators and controls except reagents must be mixed thoroughly before use either manually or by vortexing.
Caution: The serum contained in the calibrators and controls are of animal origin (horse). Handle kit reagents as if capable of transmitting an infectious agent. Appropriate precautions and good laboratory practices must be used in the storage, handling and disposal of the kit reagents. Disposal of kit reagents should be in accordance with local regulations.
Continued on the next page
{2}------------------------------------------------
EUROIMMUN 25-OH Vitamin D ELISA 510(k) Sybmission K123660
Value Assignments: Target ranges for control materials are: C1 (low): 15 ng/mL (10-25 ng/mL) and C2 (high): 40 ng/mL (25-60 ng/mL). Testing and assignment of values involves a min. of 5 assays from a min. of 2 operators with 8 replicates, resulting in a min. of 40 determinations. The mean obtained values are assigned as control values. The controls are labeled with the assigned values as mean C1 ±50% and C2 ±30%.
Calibrator value assignments are based on an internal procedure; the initial value assignment for calibrators was performed using 2 predicate assays (K021163; K112725) and HPLC. The final values are verified & assigned by adjusting their initial values to meet the specified ranges when tested against the predicate assays. Once confirmation of the values is established, the new calibrators are tested again in bulk and final.
510(K): K123660
{3}------------------------------------------------
Device Description
EUROIMMUN 25-OH Vitamin D ELISA consists of a microwell ELISA plate coated with (sheep) monoclonal anti-25-OH vitamin D antibodies, 6 calibrators, 2 controls, Biotin, sample buffer, conjugate. wash buffer concentrate. enzyme TMB chromogen/substrate solution and stop solution.
This ELISA test kit is designed for the in vitro determination of 25-OH vitamin D in human serum or plasma samples. In the first analysis step, the calibrators and patient samples are diluted with biotin-labelled 25-OH vitamin D and added to microplate wells coated with (sheep) monoclonal anti-25-OH vitamin D antibodies. During the incubation an unknown amount of 25-OH vitamin D in the patient sample and a known amount of biotinlabelled 25-OH vitamin D compete for the antibody binding sites in the microplate wells plate. Unbound 25-OH vitamin D is removed by washing. For the detection of bound biotin-labelled 25-OH vitamin D, a second incubation is performed using peroxidase-labelled streptavidin. In a third incubation using the peroxidase substrate tetramethylbenzidine (TMB) the bound peroxidase promotes a colour reaction. The colour intensity is inversely proportional to the 25-OH vitamin D concentration in the sample. Results for the samples can be calculated directly using a standard curve.
Antibodies: sheep monoclonal antibodies which identify specifically 25-OH vitamin D3 and 25-OH vitamin D2.
EUROIMMUN 25-OH Vitamin D ELISA is intended for the quantitative determination of 25-OH Vitamin D and other hydroxylated vitamin D metabolites in human serum and plasma (EDTA, Li-heparin). Results are to be used in coniunction with other clinical and laboratory data to assist the clinician in the assessment of vitamin D sufficiency in adult populations.
Indication(s) for Use: Same as intended use.
Special Conditions for the Use Statement(s):
For prescription use only.
Special Instrument Requirements:
Microwell plate reader capable of measuring OD at 450nm and at 620nm for dual wavelength readings.
Predicate Device Name(s):
IDS Immunodiagnostic Systems Ltd. OCTEIA 25-Hydroxy Vitamin D
Predicate 510(k) Number(s): K021163
Continued on the next page
510(K): K123660
Intended Usel
Indications for Use
Substantial Equivalence:
{4}------------------------------------------------
EUROIMMUI
510(K) SAFETY & EFFECTIVENESS SUMMARY, Continued
Comparison of Assays, Similarities and Differences
The following table compares the EUROIMMUN 25-OH Vitamin D ELISA Test System with the Equivalence - predicate device.
Similarities
| Item | New Device | Predicate Device |
|---|---|---|
| Intended Use/Indication for Use | For the quantitative determination of 25-OH Vitamin D andother hydroxylated vitamin D metabolites in human serumor plasma. Results are to be used in conjunction with otherclinical and laboratory data to assist the clinician in theassessment of vitamin D sufficiency in adult populations. | Same |
| Sample Type | Serum and Plasma | Same |
| Test Format | 96-well microplate assay | Same |
| Assay Components | Microtiterplate; Calibrators; Biotin labeled 25-OH Vitamin D;Avidin-Conjugate; Assay buffer; Washbuffer; Substrate;Stop solution | Same |
| Instrument | ELISA plate reader | Same |
| MeasuringWavelength | 450/620nm | Same |
| Approximate AssayTime | 3.5 hours | Same |
| Antigen Used inCalibrators | 25-OH Vitamin D3 | Same |
| Assay Principle | Competitive immunoassay | Same |
| Reagent StorageTemperature | 2-8 °C | Same |
| Test Methodology | Samples are diluted with buffer containing a reagentdissociating Vitamin D metabolites from its binding protein.The diluted samples are incubated in microtiter wells whichare coated with sheep anti-25-OH vitamin D antibodies for 2hours at room temperature before aspiration and washing.Peroxidase labeled avidin is added and binds to thecaptured biotin, following a further wash step, color isdeveloped using a chromogenic substrate (TMB). Colorintensity is inversely proportional to the concentration of 25-OH vitamin D. | Same |
| Interpretation ofResults | Standard curve | Same |
| Traceability | Standardized using UV quantification of 25-OH vitamin | Same |
| Specificity | 25-OH Vitamin D and other hydroxylated vitamin Dmetabolites | Same |
| Item | New Device | Predicate Device |
| Antibody | Monoclonal sheep anti-25-OH-Vitamin D IgG antibody | Polyclonal sheep anti-25-OH-Vitamin D IgG antibody |
| Calibrators | 6 | 7 |
| Assay range | 4.0 - 120 ng/ml | 2.4 - 144 ng/ml |
| Sample Volume | 20 $\mu$ l | 25 $\mu$ l |
Differences
Continued on the next page
Image /page/4/Picture/11 description: The image shows a close-up of what appears to be a diagram or map. There are two rectangular shapes with text inside, possibly labels or identifiers. The overall image quality is low, making it difficult to discern fine details or the broader context of the diagram.
{5}------------------------------------------------
Performance Characteristics
Precision/Reproducibility:
The reproducibility of the test was investigated following CLSI standard EP05-A2. Intra- and Inter-Assay coefficients of variation (CV) were determined using samples with values at different points on the calibration curve. All samples were natural serum samples collected freshly in-house from obvious healthy blood donors. After initial screening, 2 of 8 samples (No. 7 and 8) were spiked. All samples were then aliquoted and stored at -20°C. The intra-Assay CVs are based on 40 determinations and the Inter-Assay CVs on 40 determinations performed in 10 different runs on 5 different days (with 4 replicates per run) according to the package insert. Acceptance criterion was that the CV's show results below 12% for samples over 10 ng/ml and results below 20% for concentrations below 10 ng/ml. The following results were obtained:
Intra-Assay Reproducibility
| Sample No-------------- | ﺍﻟﻤﺎﺿﻴﺔ ﺍﻟﻤﻘﺎﺑﻠﺔ ﺍﻟﻤﺘﻬﺎﺑﺎand and the state of the same of | T MAA-BATTING - NA TAUA MITTI CA A-BUT LAND C C LT THE CLE CONTRACT COLLECTION OF CO | managements and an collection program program production in the letter Program by of Concent by other collectionas and while country contribution and and submit the form be commended the comments and more of the programs | |||
|---|---|---|---|---|---|---|
| - Read Star1000PostPostMarchMan. | 6 H | g | ||||
| A - BACK COLORES AN ARRANALE STATE LABURER / BARLARDARA BE SERVICE AF ALThe count of experience and encomes in the researching and | ||||||
| 44-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-4-410-10-14 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 | œ | 1 |
Inter-Assay Reproducibility
| Presses of the Section of the Course of Second of Second of Second of Personal Property Carry . Comments and in personal more | 11 11 11 16 1992 11 1992 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 | ||||
|---|---|---|---|---|---|
| . SULCERCALLO DEBRAL A SEBRIC: SARA SE MERICA . SECURE SEBERTV PRIPELAR/W/RADI JEThe first of the state of the country of the county of the county of | |||||
| The final program of the comments of the first of the comments of the comments of the are andAustralian / June / a warmann and Pho / film / the Marker / provent | |||||
| And Printer Printers (Andrews Mark And Andress State Add AlaborDolot and more of Altel New courses is releasioner on N als 11 1999 1 - 1 1995 1999 1 | A |
The lot-to-lot reproducibility of the test was investigated following CLSI standard EP05-A2. Inter-lot coefficients of variation (CV) were determined using samples with values at different points on the calibration curve. All samples were natural serum samples collected freshly in-house from obvious healthy blood donors. After initial screening, 2 of 8 samples (No. 7 and 8) were spiked. The inter-lot CVs are based on 32 determinations performed in 8 different runs on 4 different lots (with 2 runs per lot and 4 replicates per run) according to the package insert. Acceptance criterion was that the CV's show results below 12% for samples over 10 ng/ml and results below 20% for concentrations below 10 ng/ml. The following results were obtained:
Lot 10 Lot Reproducibility
| Sample No. | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 |
|---|---|---|---|---|---|---|---|---|
| Mean (x): | 7.3 | 18.5 | 24.8 | 37.4 | 47.6 | 58.0 | 74.3 | 97.3 |
| SD. | 0.89 | 1.89 | 1.81 | 2.29 | 4.25 | 4.07 | 6.59 | 9.26 |
| CV,% | 12.2 | 10.2 | 7.3 | 6.1 | 8.9 | 7.0 | 8.9 | 9.5 |
Continued on the next page
{6}------------------------------------------------
Limit of Blank, Limit of Detection and Functional Sensitivity: Limit of blank (LoB), limit of detection (LoD) and functional sensitivity (FS) were investigated following CLSI standard EP17-A.
LoB was determined as the concentration corresponding to the mean OD of the zero calibrator minus 1.645 times the standard deviation using the mean of 60 replicates for calibrator 1 (0 ng/ml) and the mean of 20 replicates for calibrator 2 (4 ng/ml). LoB was found to be 0.4 ng/ml.
| $[\mu_B]$ (OD) | 2.721 |
|---|---|
| $[\sigma_B]$ (OD) | 0.038 |
| $[\mu_{C1}]$ (OD) | 2.721 |
| $[\mu_{C2}]$ (OD) | 2.146 |
| LoB (OD) | 2.658 |
| LoB (ng/ml) | 0.43 |
LoD was determined as the LoB plus 1.645 times the standard deviation of 200 determinations from 5 samples in the low range of 2 to 10 ng/ml, measured in 5 independent runs with 8 replicates per run. The mean LoD was found to be 1.8 ng/ml.
| Sample | 1 | 2 | 3 | 4 | 5 |
|---|---|---|---|---|---|
| Concentration (ng/ml) | 2.0 | 4.0 | 6.0 | 8.0 | 10.0 |
| $\sigma$ s (ng/ml) | 0.731 | 0.810 | 0.783 | 0.919 | 0.834 |
| LoD (ng/ml) | 1.63 | 1.76 | 1.72 | 1.94 | 1.80 |
| Mean LoD (ng/ml) | 1.8 |
Functional sensitivity is defined as the lowest concentration at which the potential regression line crosses the 20% CV line and was determined from a plot of the mean concentrations (X-axis) vs. % CVs (Y-axis). Functional sensitivity was found to be 4.0 ng/ml. Results below 4 ng/mL are reported as "< 4 ng/mL".
Image /page/6/Figure/9 description: The image is a graph titled "Determination of functional sensitivity". The x-axis is labeled "Mean concentration (ng/ml)" and ranges from 0 to 12. The y-axis is labeled "CV (%)" and ranges from 0 to 45. The graph shows a curve that decreases as the mean concentration increases, and the equation of the curve is y = 69.006x^-0.899 with an R^2 value of 0.991.
Continued on the next page
510(K): K123660
{7}------------------------------------------------
EUROIMMUN
510(K) SAFETY & EFFECTIVENESS SUMMARY, Continued
Linearity/Assay Reportable Range:
The linearity of the test was investigated following CLSI standard EP6-A. Sets of 11 sample preparations were prepared by mixing a natural negative sample (0 ng/ml) and high positive samples. The sample preparations covered the concentration range of 2 to 129 ng/ml. Each set was run according to the package insert in the same run in double determinations per sample preparation, the mean of the two determinations for each sample was calculated and polynomial regression was performed of observed results vs expected results. Although visually the data demonstrated a linear relationship, the p-values of the polynomial regression coefficients were significant (< 0.05) for two of the three sample sets. However, the amount of nonlinearity above the functional sensitivity of 4.0 ng/ml was found acceptable below 15%. The detailed results are shown below. Based on the results of the linearity study and limit of detection study, it can be concluded that the assay is sufficiently linear from 4.0 to 120 ng/ml.
Image /page/7/Figure/5 description: The image is a scatter plot titled "Linearity". The x-axis is labeled "Expected conc." and ranges from 0 to 140. The y-axis is labeled "Observed conc." and ranges from 0 to 140. There are three sets of data plotted on the graph, labeled as Set 1, Set 2, and Set 3, which all show a positive correlation.
| Regressiony=ax3+bx2+cx+d | Set 1 | Set 2 | Set 3 | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Coeff. | Std. errorof slope | p-value | Std. errorof regr. | Coeff. | Std. errorof slope | p-value | Std. errorof regr. | Coeff. | Std. errorof slope | p-value | Std. errorof regr. | ||
| linear | c | 1.01 | - | - | 2.12 | 0.96 | - | - | 3.02 | 0.98 | - | - | 4.38 |
| d | -3.25 | - | - | -2.09 | - | - | -0.57 | - | - | ||||
| R2 | 0.997 | - | - | 0.990 | - | - | 0.999 | - | - | ||||
| 2nd orderpoly-nomial | b | 0.00 | 0.00 | 0.00 | 1.03 | 0.00 | 0.00 | 0.01 | 2.05 | 0.00 | 0.00 | 0.13 | 3.98 |
| c | 0.81 | - | - | 0.67 | - | - | 0.79 | - | - | ||||
| d | 0.24 | - | - | 2.46 | - | - | 3.23 | - | - | ||||
| R2 | 0.999 | - | - | 0.996 | - | - | 0.993 | - | - | ||||
| 3rd orderpoly-nomial | a | 0.00 | 0.00 | 0.10 | 1.03 | 0.00 | 0.00 | 0.06 | 1.43 | 0.00 | 0.00 | 0.20 | 3.56 |
| b | 0.00 | 0.00 | 0.36 | -0.01 | 0.00 | 0.02 | -0.01 | 0.00 | 0.12 | ||||
| c | 0.73 | - | - | 1.10 | - | - | 1.43 | 1.17 | - | - | 3.56 | ||
| d | 0.93 | - | - | 1.03 | -0.95 | - | - | -0.31 | - | - | |||
| R2 | 0.999 | - | - | 0.998 | - | - | 0.995 | - | - |
Continued on the next page
510(K): K123660
{8}------------------------------------------------
EUROIMMU
Image /page/8/Figure/2 description: The image is a linearity graph with representative data (Set 3). The x-axis is labeled 'Expected' and ranges from 0 to 140, while the y-axis is labeled 'Observed' and ranges from 0 to 140. A line of best fit is plotted on the graph, with the equation y = 0.94259545437413x + 1.167017828201 and an R-squared value of 0.9874.
Regression Data
Image /page/8/Figure/4 description: The image shows a table with four columns labeled x, y, xy, and x^2. The sum of x from i=1 to n is 584.80, the sum of y from i=1 to n is 562.90, the sum of xy from i=1 to n is 45822.52, and the sum of x^2 from i=1 to n is 47889.10. The table appears to be summarizing data for statistical analysis.
Regression Calculations
m
$$m = \frac{n\sum_{i=1}^{n} x_i y_i - \sum_{l=1}^{n} x_l \sum_{l=1}^{n} y_l}{\sqrt{n\sum_{l=1}^{n} x_l^2 - (\sum_{l=1}^{n} x_l)^2}} = 0.94259545437413 \quad \left| \quad b = \frac{\sum_{i=1}^{n} y_i - m\sum_{l=1}^{n} x_l}{n} = 1.167017828201$$
$$\mathbf{y} = \mathbf{0}.94259545437413\mathbf{x} + \mathbf{1.167017828201}$$
9.1833 ; Ob
Standard Deviation
The deviation of the measurement $[$ from the mean is $[$ = $[$ - $\overline{[}$
Uncertainties in the Slope & Intercept
Standard error in the slope om:
Standard error in the intercept $σ_b$:
$$
\sigma_m^2 = \frac{n\sigma_Y^2}{n,\Sigma_{l=1}^n ,\mathrm{x}l^2 - \left(\Sigma{l=1}^n ,\mathrm{x}_l\right)^2} = 0.0019,\mathrm{,,s},\sigma_m = \pm 0.04385
$$
-.2 5n -.2
$$
\sigma_m^2 = \frac{n\sigma_\circ^4}{n\sum_{l=1}^n \mathbf{x}l^2 - \left(\sum{l=1}^n \mathbf{x}_l\right)^2} = 0.0019 ; ; ; \sigma_m = \pm 0.0438 ; \quad \left| ; \sigma_l \right| $$
Standard deviation of y;:
$$\sigma_{l}^{2} = \sigma_{\mathcal{Y}}^{2} = \frac{\sum_{i=1}^{n} (\mathbf{y}{i} - \mathbf{b} - m\mathbf{x}{i})^{2}}{n - 2} = 26.2521; \sigma_{\mathcal{Y}} = \pm 5.1237$$
$$\text{Correlation } \textbf{Coefficient:} \quad r = \frac{n \sum_{i=1}^{n} x_i \nu_i - \sum_{i=1}^{n} x_i \sum_{i=1}^{n} \nu_i}{\sqrt{\left(n \sum_{l=1}^{n} x_l^2 - (\sum_{l=1}^{n} x_l)^2 \right) \left(n \sum_{l=1}^{n} \nu_l^2 - (\sum_{l=1}^{n} \nu_l)^2\right)}} = 0.99147705552933$$
510(K): K123660
= ±3.0304
{9}------------------------------------------------
EUROIMI
510(K) SAFETY & EFFECTIVENESS SUMMARY, Continued
Analytical Specificity
Cross-Reactivity:
Cross reactivity was investigated following Abraham et al (Handbook of radioimmunoassay, 1977). A 25-OH Vitamin D free sample was aliquoted and spiked with potential cross reacting Vitamin D metabolites at the concentrations listed below. The following cross reactivities were observed. The study shows that the 25-OH Vitamin D ELISA detects both 25-OH Vitamin D3 and 25-OH Vitamin D2 as well as other hydroxylated Vitamin D metabolites as stated in the intended use. Expected concentrations of 1,25-OH Vitamin D3 and 1,25-OH Vitamin D2 in natural samples are below 100 pg/ml, so the obtained cross reactivity has no significant influence on results of 25-OH Vitamin D.
| Potential Cross-ReactingSubstance | MolecularWeight(g/mol) | Conc.Spiked(ng/ml) | Conc.Observed(ng/ml) | CrossReactivity(%) |
|---|---|---|---|---|
| 25-OH Vitamin D3 | 400.64 | 10.0 | 10.0 | 100 |
| 25-OH Vitamin D2 | 412.65 | 25.0 | 24.3 | 100 |
| 24,25-OH Vitamin D3 | 416.64 | 100 | 0.3 | 0.3 |
| Cholecalciferol (Vit. D3) | 384.64 | 10,000 | 3.4 | 0.03 |
| Ergocalciferol (Vitamin D2) | 396.65 | 10,000 | 5.1 | 0.05 |
| 1,25-OH Vitamin D3 | 416.64 | 10.0 | 4.3 | 45 |
| 1,25-OH Vitamin D2 | 428.65 | 10.0 | 19.8 | 212 |
| 3-epi-25-OH Vitamin D3 | 400.64 | 10.0 | 1.7 | 17 |
Continued on the next page
{10}------------------------------------------------
Interference:
Interferences were investigated following CLSI standard EP07-A2. To investigate the influence from hemoglobin, triglycerides and bilirubin as well as from cholesterol, biotin and ascorbic acid, sera at different 25-OH Vitamin D concentrations were spiked with potential interfering substances and incubated with the test system according to the package insert. The recovery in relation to the unspiked sample without interferent was calculated. The individual recovery was within the accepted individual recovery of 90 – 110 %. No significant interference was observed for concentrations of up to 750 mg/dl for hemoglobin, 2000 mg/dl for triglycerides, 40 mg/dl for bilirubin and up to 400 mg/dl for cholesterol, 1000 mg/dl for biotin and 10.0 mg/ml for ascorbic acid.
Significant interference (above 10 % deviation from unspiked sample) was seen with 1000 mg/dl of hemoglobin.
| Potential interfering substance | Range of sampleconcentration | Range of recoveries(%) |
|---|---|---|
| Hemoglobin(0, 250, 500 and 750 mg/dl) | 15.0 - 71.6 ng/ml | 90 - 110 % |
| Triglycerides(0, 500, 1000 and 2000 mg/dl) | 16.2 - 51.0 ng/ml | 98 - 110 % |
| Bilirubin(0, 10, 20 and 40 mg/dl) | 13.0 - 54.9 ng/ml | 90 - 109 % |
| Cholesterol(0, 200 and 400 mg/dl) | 9.6 - 27.3 ng/ml | 94 – 109 % |
| Biotin(0, 10, 100 and 1000 mg/dl) | 21.0 - 74.7 ng/ml | 91 - 107 % |
| Ascorbic acid(0, 2.5, 5.0, 7.5 and 10.0 mg/ml) | 13.0 - 56.1 ng/ml | 94 - 110 % |
Assay Cut-off:
Not Applicable
ontinued on the next page
{11}------------------------------------------------
Comparison Studies
Method Comparison w/Predicate Device:
The method comparison against the predicate device was performed following CLSI standard EP09-A2. Both the predicate device and the subject device were performed exactly as described by their respective instructions for use.
Serum samples were obtained from different sources (141 prospective samples sent in for 25-OH-Vitamin D testing from a clinical laboratory, 28 samples from 25-OH-Vitamin D quality assessment programs (harvested from blood donated by venesection undergoing therapeutic . for patients hemochromatosis or polycythemia), 5 samples from a 25-OH-Vitamin D quality control panel (untreated routine patient material) and 30 samples from normal blood donors). All patients have given their informed consent. These samples are native (natural) and not treated or adjusted in any way.
To ensure that the tested concentrations of 25-OH Vitamin D are distributed across the reportable measurement range, 36 samples (15%) in the set were spiked with 25-OH Vitamin D stock solution of 25-OH Vitamin D3. In total, 240 samples were collected, ranging from 4.1 to 119.1 ng/ml, and tested with the EUROIMMUN 25-OH Vitamin D ELISA and with a FDA-cleared reference assay. Results of linear regression analysis are shown in the table below.
| n | 240 |
|---|---|
| Concentration Range (Predicate) | 4.1 - 110.4 ng/ml |
| Concentration Range (Candidate) | 4.1 - 119.1 ng/ml |
| Regression Equation (y = Candidate, x = Predicate) | y = 0.78 + 1.08 x |
| 95% C.I. of Intercept | -0.06 - 1.63 |
| 95% C.I. of Slope | 1.06 - 1.11 |
| Correlation Coefficient R | 0.9858 |
| 95% C.I. of R | 0.9817 - 0.9890 |
Image /page/11/Figure/9 description: This image is a scatter plot comparing two methods of measuring 25-OH-Vitamin D: EUROIMMUN ELISA and IDS ELISA. The x-axis represents the IDS ELISA measurements in ng/ml, ranging from 0 to 120. The y-axis represents the EUROIMMUN ELISA measurements in ng/ml, ranging from 0 to 140. The plot shows a positive correlation between the two methods, with data points clustered around a straight line.
Continued on the next page
510(K): K123660
{12}------------------------------------------------
Matrix Comparison
The usability of plasma was investigated using sample pairs each of serum and corresponding plasma (EDTA, Li-heparin) from donors. To ensure that the tested concentrations of 25-OH Vitamin D are distributed across the reportable dynamic range, 3 sample pairs in the set were spiked with 25-OH Vitamin D stock solution of 25-OH Vitamin D3. Passing-Bablok regression was calculated for the comparison of serum to plasma. The regression equation is near the ideal correlation (intercept 0; slope 1.0) indicating equivalence of concentrations between serum and the corresponding plasma matrices. Coefficients of determination were found to be above 0.99 and %recovery compared to serum was in the range of 87 to 127 % (serum = 100 %).
| n | EDTA plasma | Li-heparin plasma | |
|---|---|---|---|
| Concentration Range (Serum) | 8.6 - 105.1 ng/ml | 8.6 - 105.1 ng/ml | |
| Concentration Range (Plasma) | 8.8 - 104.2 ng/ml | 8.5 - 102.6 ng/ml | |
| Regression Equation:(y = plasma, x = serum)95% C.I. of intercept95% C.I. of slope | $y = 0.29 + 0.99 x$ $-0.37 - 1.18$ $0.93 - 1.02$ | $y = 0.55 + 0.97 x$ $-0.65 - 1.33$ $0.93 - 1.04$ | |
| Coefficient of determination R2 | 0.996 | 0.993 | |
| Mean %recoveryRange of %recovery | 100 %87 - 116 % | 101 %87 - 127 % | |
| 38 | 38 |
Image /page/12/Figure/6 description: The image contains two scatter plots comparing serum concentrations with EDTA plasma and Li-heparin plasma concentrations. In the first plot, titled "EDTA plasma", the x-axis represents serum concentrations in ng/ml, ranging from 0 to 120, while the y-axis represents EDTA plasma concentrations in ng/ml, also ranging from 0 to 120. The data points show a positive correlation between serum and EDTA plasma concentrations. The second plot, titled "Li-heparin plasma", similarly compares serum concentrations (x-axis) with Li-heparin plasma concentrations (y-axis), both ranging from 0 to 120 ng/ml, and also shows a positive correlation.
Continued on the next noge
510(K): K123660
{13}------------------------------------------------
Expected Values/Reference Range
Reference intervals were established following CLSI standard C28-A3. The levels of 25-OH vitamin D were analyzed in a panel of 206 samples from healthy subjects (70 men and 136 women with an average age of 64 years: age range: 22 – 99 years) from a US commercial source. The samples are known to be from normal US blood donors from the US Midwest region and were drawn in early October 2010. The observed median, minimum and maximum values as well as the 2.5% and 97.5% percentiles were similar to those reported for other devices cleared in the US (e.g. K102432, K112725, K110619, K091849, K110586, K071480). The results are shown in the table below.
| (ng/ml) | (nmol/l) | |
|---|---|---|
| Minimum | < 4 | < 10 |
| Maximum | 64.8 | 162.0 |
| Mean | 20.8 | 51.9 |
| Median | 19.4 | 48.4 |
| 2.5% Percentile | 5.4 | 13.6 |
| 97.5% Percentile | 47.0 | 117.4 |
This data is provided for guidance only. It is important for each laboratory to establish its own reference ranges, representative of its typical population. Also published studies representing the local population can be taken into consideration.
Interpretation criteria are provided by the US CDC/NCHS National Health and Nutrition Examination Surveys (NHANES):
| Status | (ng/ml) | (nmol/l) |
|---|---|---|
| At risk of vitamin D deficiency | < 12 | < 30 |
| At risk of vitamin D inadequacy | 12-19 | 30-49 |
| Sufficient in vitamin D | 20 - 50 | 50 - 125 |
| Possibly harmful vitamin D | > 50 | > 125 |
1Looker AC, Johnson CL, Lacher DA, Pfeiffer CM, Schleicher RL, Sempos CT. Vitamin D status: United States, 2001-2006. NCHS data brief, no 59. Hyattsville, MD: National Center for Health Statistics. 2011.
The Endocrine Society Clinical Practice Guideline (2011) recently suggested a higher target level of at least 30 na/mi:
| Status | (ng/ml) | (nmol/l) |
|---|---|---|
| Deficiency | < 20 | < 50 |
| Insufficiency | 20 - 29 | 50 - 75 |
| Sufficiency | 30 - 100 | 75-250 |
"Holick MF, Binkley NC, Bischoff-Ferrari HA, Gordon CM, Hanlev DA, Heaney RP, Murad MH, Weaver CM. Evaluation, Treatment, and Prevention of Vitamin D Deficiency: an Endocrine Society Clinical Practice Guideline. J Clin Endocrinol Metabol 96 (2011) 1911-1930.
Continued on the next page
510(K): K123660
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the word "EUROIMMUN" in all caps, followed by a black rectangle and the letters "US" in all caps. The text is in a bold, sans-serif font. A horizontal line is present above and below the text.
Stability Studies
Stability studies are conducted following the international standard EN 13640:2002: Stability testing of in vitro diagnostic reagents.
Real-time testing at 2-8°C and accelerated testing at 37°C were conducted for all kit components (Biotin, Conjugate, Microtiterplate, control materials and the calibrators). The stability study protocol and the acceptance criteria are shown below. The predicted shelf-life, based on results of accelerated testing at 37°C, is at least 12 months at 2-8°C for the control materials and the calibrators. Real-time stability testing is ongoing to support the predicted shelf-life of 12 months at 2-8°C.
The stability study protocol and the acceptance criteria to determine open-vial stability of the complete kit are shown below. Open-vial stability of the kit is 3 months when stored at 2-8°C. Biotin was found to be stable for at least 2 weeks when diluted to working strength.
Continued on the next page
{15}------------------------------------------------
Conclusion
The data presented is complete and provides a reasonable assurance to support the basis of the claim for substantial equivalence. The labeling is sufficient and satisfies the requirements of 21 CFR § 809.10.
The EUROIMMUN 25-OH Vitamin D ELISA assay (including calibrators and controls} is substantially equivalent to other products presently in commercial distribution intended for similar use; and has been demonstrated to be safe and effective. Most notably, the EUROIMMUN 25-OH Vitamin D ELISA is substantially equivalent to the currently marketed IDS Immunodiagnostic Systems Ltd. OCTEIA 25-Hydroxy Vitamin D (K021163) for in vitro diagnostic use for the quantitative determination of 25-OH Vitamin D and other hydroxylated vitamin D metabolites in human serum and plasma (EDTA, Liheparin). The EURÓIMMUN 25-OH Vitamin D ELISA assay is to be used in conjunction with other clinical and laboratory data in the assessment of vitamin D sufficiency in adult populations.
Michael Locke
Signature
Signature
Michael Locke/Director of Requlatory Affairs Printed Name/Title
7/3/2013 Date
510(K): K123660
E
{16}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/16/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three talons, representing service, science, and security. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the eagle in a circular fashion.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 17, 2013
EUROIMMUN US INC. C/O Michael Locke 1100 The American Road MORRIS PLAINS NJ 07950
Re: K123660
Trade/Device Name: 25-OH Vitamin D ELISA Regulation Number: 21 CFR 862.1825 Regulation Name: Vitamin D test system Regulatory Class: II Product Code: MRG Dated: June 06, 2013 Received: June 07, 2013
Dear Mr. Locke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for
{17}------------------------------------------------
Page 2-Mr. Locke
the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Carol C. Benson -S for
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{18}------------------------------------------------
Indications for Use
510(k) Number (if known): K123660
Device Name: 25-OH Vitamin D ELISA
Indications for Use:
EUROIMMUN's 25-OH Vitamin D ELISA is intended for the quantitative determination of 25-OH Vitamin D and other hydroxylated vitamin D metabolites in human serum and plasma (EDTA, Li-heparin). Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of vitamin D sufficiency in adult populations.
Prescription Use X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
YungW.Chan-S
Division Sign-Off Office of In Vitro Diagnostics and Radiological Health
K123660 510(k)
Page 1 of 1
§ 862.1825 Vitamin D test system.
(a)
Identification. A vitamin D test system is a device intended for use in clinical laboratories for the quantitative determination of 25-hydroxyvitamin D (25-OH-D) and other hydroxylated metabolites of vitamin D in serum or plasma to be used in the assessment of vitamin D sufficiency.(b)
Classification. Class II (special controls). Vitamin D test systems must comply with the following special controls:(1) Labeling in conformance with 21 CFR 809.10 and
(2) Compliance with existing standards of the National Committee on Clinical Laboratory Standards.