K Number
K140615
Date Cleared
2014-07-15

(127 days)

Product Code
Regulation Number
886.4390
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VICTUS Femtosecond Laser Platform is indicated for use in:

  • . the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea
  • for anterior capsulotomy during cataract surgery .
  • the creation of cuts / incisions in the cornea in patients undergoing cataract surgery or . other ophthalmic treatment requiring cuts / incisions in the cornea.
  • . laser-assisted lens fragmentation of nuclear cataracts during cataract surgery, not for fragmentation of posterior subcapsular (PSC) and cortical cataracts
Device Description

The VICTUS Femtosecond Laser Platform (hereafter referred to as the VICTUS Platform) is a precision ophthalmic surgical laser indicated for use the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea and for anterior capsulotomy during cataract surgery (via K120426). In addition, the VICTUS has been cleared for patients undergoing cataract surgery or other ophthalmic treatment requiring penetrating arcuate cuts / incisions in the cornea (via K122386), and for corneal incisions (via K 132534). This 510(k) expands the list of indications for use to include laser-assisted lens fragmentation during cataract surgery for nuclear cataracts.

For all indications for use, laser pulses are delivered through a sterile disposable Patient Interface, consisting of a contact lens and suction clip to provide suction. The contact lens and suction clip assembly creates a reference surface for depth control and fix the eye relative to the delivery of the laser beam. Surgical effects are produced by scanning thousands of individual pulses, producing continuous incisions. The location of the tissue photodisruption is controlled by a fixed laser beam focused through a scanning optic system to the desired location.

Safety margins are built into the software, including an anterior capsule safety margin of 500 um, a posterior capsule safety margin of 700 um, and a pupil safety margin of 500 um. The graphical user interface is used to mark the anatomical boundaries and the software automatically calculates the safety margins. The safety margins are not modifiable by the end user.

The fundamental scientific technology remains the same as previously cleared for the VICTUS Femtosecond Laser Platform under K122386 and K132534.

AI/ML Overview

Here's an analysis of the provided text, focusing on the acceptance criteria and the study that proves the device meets those criteria:


Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device Performance
Primary Endpoint: Effective phacoemulsification time (EPT) in the laser group (pooled from all cataract grades) should be lower than or equal to that in the manual group.Met: Effective phacoemulsification time (EPT) in the laser group (pooled from all cataract grades) was lower than or equal to that in the manual group.
Secondary Endpoint: Less or equal adverse events in the laser group compared with the manual group.Met: Less or equal adverse events in the laser group compared with the manual group.
Safety Margins: Anterior capsule safety margin of 500 um, posterior capsule safety margin of 700 um, and pupil safety margin of 500 um. (Implied, as these are "built into the software" and not user-modifiable, suggesting they must be maintained.)Implied Met: "Safety margins are built into the software, including an anterior capsule safety margin of 500 um, a posterior capsule safety margin of 700 um, and a pupil safety margin of 500 um." The text indicates these are programmatically enforced, not reported as a result.
Reproducibility and Accuracy (Bench Testing): Laser-assisted lens fragmentation performed with the device must meet established acceptance criteria for lens fragmentation for various parameters (upper/lower depth, cross/open angle, spot/line spacing, decentration).Met: "The testing showed that laser-assisted lens fragmentation performed with the VICTUS Femtosecond Laser Platform met the established acceptance criteria for lens fragmentation." "resulted in highly reproducible and accurate treatments".
Intraocular Pressure (IOP): Pre- and post-operative IOPs should be similar between the VICTUS and manual groups.Met: Pre- and post-operative IOPs were similar between the VICTUS and manual groups at all time points during the study for the complete study group and when stratified by cataract grade.
Best Corrected Visual Acuity (BCVA): BCVDA should be similar between the VICTUS and manual groups.Met: Best corrected visual acuities (BCVDA) were similar between the VICTUS and manual groups at all time points during the study for the complete study group and also when stratified by cataract grade.
Slit Lamp Examinations: Slit lamp examinations in the laser group should be consistent with the manual group.Met: Slit lamp examinations in the laser group were consistent with the manual group at all time points for the complete study group and also when stratified by cataract grade.
Intra-operative Complications: (Implied: Acceptable rate, or no increase compared to manual).Met/Favorable: The only intra-operative complications were two anterior capsule tears in the manual group; no other operative complications were recorded for either the manual or the laser group.
Post-operative Observation Rates (Corneal edema, flare, trace anterior chamber): Comparable between laser and manual groups.Met: The post-operative observation rates of corneal edema, flare and trace anterior chamber in the laser group were comparable to those in the manual group.
OCT Image Correlation with Biometric Data: OCT images through the PI should correlate well with biometric data for lens thickness and anterior chamber depth.Met: OCT images through the PI correlate well with biometric data obtained from OCT images for lens thickness and anterior chamber depth for the complete study group and also when stratified by cataract grade.
VICTUS OCT Image Precision: High level of precision for lens thickness and anterior chamber depth.Met: VICTUS OCT images were found to have a high level of precision for lens thickness (both pooled and stratified by cataract grade) and anterior chamber depth (pooled only).

Study Details:

  1. Sample size used for the test set and the data provenance:

    • Sample Size: 136 eyes from 110 subjects (26 patients treated bilaterally). 68 eyes in the VICTUS (laser) group and 68 eyes in the standard (manual) treatment group.
    • Data Provenance: The text does not explicitly state the country of origin. It mentions "multi-center, multi-surgeon clinical study," implying various clinical sites, but no geographical location is specified. It was a prospective study.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The study design describes a "randomized, controlled, open, prospective, multi-center, multi-surgeon clinical study" comparing the VICTUS device to manual methods. The "ground truth" in this context refers to the clinical outcomes and observations recorded during the study, rather than an independent expert review of images or data to establish a reference standard. The study data (e.g., EPT, adverse events, IOP, BCVA, slit lamp findings) were collected through standard clinical procedures by the participating surgeons and study personnel. The document does not specify a number of independent experts for ground truth establishment outside of the primary treating physicians and clinical staff who recorded the outcomes.
  3. Adjudication method for the test set:

    • The document does not describe a specific adjudication method (like 2+1 or 3+1) for the test set results. Clinical outcomes were collected by the participating centers. Adverse events were observed and reported, but the process of determining their relation to the intervention was clinical (e.g., "All adverse events in the laser group were determined to be related to the cataract surgery and not to the use of the VICTUS Femtosecond Laser."). This implies a clinical judgment by the investigators, not an independent adjudication committee.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This study compares a surgical device (VICTUS Femtosecond Laser Platform) to a manual surgical technique, not an AI or imaging-based diagnostic tool requiring human reader interpretation. Therefore, there's no mention of human reader improvement with/without AI assistance or effect sizes related to such a comparison.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • This is not applicable as the device is a surgical laser platform, not an algorithm for diagnosis or image analysis. The "standalone" performance mentioned refers to the device's ability to perform specific surgical actions (e.g., lens fragmentation) as intended, which was evaluated through the clinical study and non-clinical bench testing. The outcomes were compared to manual surgical techniques, which inherently involve human-in-the-loop performance.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" primarily consisted of clinical outcomes data (e.g., Effective Phacoemulsification Time (EPT), adverse event rates, intraocular pressure (IOP), best corrected visual acuity (BCVA), slit lamp examination findings, intra-operative complications) obtained directly from the patients in a prospective clinical study. Non-clinical bench testing also established a "ground truth" for the device's physical performance parameters against established acceptance criteria.
  7. The sample size for the training set:

    • This device is a surgical instrument and does not rely on a "training set" in the context of machine learning. Therefore, a training set sample size is not applicable. The device's design and parameters are based on engineering principles and prior iterations (as evidenced by its substantial equivalence to previous VICTUS and predicate devices).
  8. How the ground truth for the training set was established:

    • As explained above, there is no "training set" for this type of device. The functionality and safety were established through mechanical/bench testing and clinical trials, not through learning from a dataset with established ground truth labels.

{0}------------------------------------------------

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Date PreparedJuly 9, 2014
Applicant:Technolas Perfect Vision GmbHMesserschmittstr. 1 +380992 Munich, GermanyTel: 49 174 3494062
Contact:Betty M. JohnsonSr. Manager, Regulatory AffairsTechnolas Perfect Visionc/o PO Box 17190Anaheim, CA 92817Ph: (714) 273-8254Fax: (714) 855-2794Email : betty.johnson@bausch.com
Trade Name:Common Name:Regulation NumberClassification Name:Device Classification:Product CodeVICTUS Femtosecond Laser PlatformOphthalmic Laser21 CFR 886.4390Laser, OphthalmicClass IIOOE (Ophthalmic Femtosecond Laser)HOC (Unit, Lens fragmentation)

Substantially Equivalent To:

510(k)NumberProduct Trade NameCurrent Manufacturer
K122386,K135234VICTUS Femtosecond LaserPlatformTechnolas Perfect Vision GmbH
K094052,K101626,K120732LenSx LaserLenSx, an Alcon company
K102727LensAR LaserLensAR Inc.

{1}------------------------------------------------

Description of the Device Subject to Premarket Notification:

The VICTUS Femtosecond Laser Platform (hereafter referred to as the VICTUS Platform) is a precision ophthalmic surgical laser indicated for use the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea and for anterior capsulotomy during cataract surgery (via K120426). In addition, the VICTUS has been cleared for patients undergoing cataract surgery or other ophthalmic treatment requiring penetrating arcuate cuts / incisions in the cornea (via K122386), and for corneal incisions (via K 132534). This 510(k) expands the list of indications for use to include laser-assisted lens fragmentation during cataract surgery for nuclear cataracts.

For all indications for use, laser pulses are delivered through a sterile disposable Patient Interface, consisting of a contact lens and suction clip to provide suction. The contact lens and suction clip assembly creates a reference surface for depth control and fix the eye relative to the delivery of the laser beam. Surgical effects are produced by scanning thousands of individual pulses, producing continuous incisions. The location of the tissue photodisruption is controlled by a fixed laser beam focused through a scanning optic system to the desired location.

Safety margins are built into the software, including an anterior capsule safety margin of 500 um, a posterior capsule safety margin of 700 um, and a pupil safety margin of 500 um. The graphical user interface is used to mark the anatomical boundaries and the software automatically calculates the safety margins. The safety margins are not modifiable by the end user.

The fundamental scientific technology remains the same as previously cleared for the VICTUS Femtosecond Laser Platform under K122386 and K132534.

Indications for Use:

The VICTUS Platform is indicated for use in:

  • the creation of a corneal flap in patients undergoing LASIK surgery or other treatment . requiring initial lamellar resection of the cornea
  • . for anterior capsulotomy during cataract surgery
  • . the creation of cuts / incisions in the cornea in patients undergoing cataract surgery or other ophthalmic treatment requiring cuts / incisions in the cornea.
  • . laser-assisted lens fragmentation during cataract surgery for nuclear cataracts, not for fragmentation of posterior subcapsular (PSC) and cortical cataracts

Technical Characteristics Comparison:

The design principle and mode of operation of the VICTUS Platform is fundamentally the same as the VICTUS Platform cleared in K122386 and K132534, the LenSx Laser System cleared in K094052, K101626, and K120732, and the LensAR Laser cleared in K102727, all of which deliver femtosecond pulses to produce a pattern of photodisruption to fragment the crystalline lens. The VICTUS Platform delivers femtosecond pulses to produce a pattern of photodisruption

{2}------------------------------------------------

in the crystalline lens, as do the predicate devices.

The means of fixation of the patient contact portion of the VICTUS Platform is identical, and therefore substantially equivalent to that described in the VICTUS Femtosecond Laser Platform cleared in K122386 and K132534. The predicate system uses suction vacuum to affix a suction ring to the corneal surface prior to use. The vacuum suction and contact pressure are the same as the VICTUS Platform previously cleared in K132534 and therefore there are no changes in regards to the effect on intraocular pressure.

The VICTUS Platform mode of operation and the technology used to create the cutting action are similar or identical to these previously mentioned devices, and therefore substantially equivalent to these legally marketed predicate devices.

Non-Clinical Performance Data

A variety of test procedures were conducted to assess the performance of the proposed VICTUS Platform in support of this premarket submission. The collected data were evaluated by comparing the mean values to the specified acceptance criteria. Three different materials were used for the bench performance testing: agarose gel, polyethylene terephthalate (PETG), and polymethyl methacrylate (PMMA). The testing showed that laser-assisted lens fragmentation performed with the VICTUS Femtosecond Laser Platform met the established acceptance criteria for lens fragmentation.

The non-clinical performance testing showed that laser-assisted lens fragmentation performed with the VICTUS Femtosecond Laser Platform resulted in highly reproducible and accurate treatments, including parameters such as upper and lower depth, cross angle, open angle, spot spacing, line spacing, and decentration.

Clinical Performance Data

Clinical performance was evaluated in a clinical study using the VICTUS Femtosecond Laser Platform (VICTUS) in cataract procedures. This study included evaluation of femtosecond laser-assisted lens fragmentation for nuclear cataracts as compared to manual methods for lens fragmentation for nuclear cataracts. This was a randomized, controlled, open, prospective, multi-center, multi-surgeon clinical study. Follow-up visits were conducted at one day, one week, and one month post-operatively.

Enrolled subjects were randomly assigned to one of two groups: Laser, which had lens fragmentation performed by the VICTUS device prior to phacoemulsification (treatment), and Manual, which had manual phacoemulsification.

One hundred thirty six (136) eyes of 110 subjects were included in the analysis (26 patients were treated bilaterally); 68 eyes were treated in the VICTUS study arm and 68 eyes in the standard treatment group. The mean age of the manual group over all cataract grades was 63.5 ± 10.4 years (range: 40 to 80 years), and the mean age of the laser group over all cataract grades was 62.3 ± 8.7 years (range: 45 to 83 years).

{3}------------------------------------------------

The demographics of the subjects included in the analysis of the clinical study are summarized in the following table.

DemographicsNumber%ManualLaser
NUMBER OF EYES
Gender
Female7756.64037
Male5943.42831
Race
African-American0000
Asian0000
Caucasian136100%6868
Hispanic0000
OtherN/AN/AN/AN/A
Age (Years)
Mean (SD)63.5 ± 10.462.3 ± 8.7
Minimum, Maximum40, 8045, 83
Study eye
OD3627
OS3241

Demographics-Subgroup Eyes (All Cataract Grades Combined)

The types of cataracts included in the analysis of the study included nuclear cataracts with Grades 1-5, as noted in the following table.

Cataract Grade and Type

ParameterLaser(N=68)n (%)Manual(N=68)n (%)Total (N=136)n (%)
Grade 116 (23.5)16 (23.5)32 (23.5)
Grade 217 (25.0)17 (25.0)34 (25.0)
Grade 314 (20.6)14 (20.6)28 (20.6)
Grade 413 (19.1)13 (19.1)26 (19.1)
Grade 58 (11.8)8 (11.8)16 (11.8)

{4}------------------------------------------------

Summary of Results

  • Primary endpoint was met: Effective phacoemulsification time (EPT) in the laser group (pooled from all cataract grades) was lower than or equal to that in the manual group.
  • . Secondary endpoint was met: Less or equal adverse events in the laser group compared with the manual group.
  • . Other results:
    • o Pre- and post-operative IOPs were similar between the VICTUS and manual groups at all time points during the study for the complete study group and when stratified by cataract grade.
    • o Best corrected visual acuities (BCVDA) were similar between the VICTUS and manual groups at all time points during the study for the complete study group and also when stratified by cataract grade.
    • o Slit lamp examinations in the laser group were consistent with the manual group at all time points for the complete study group and also when stratified by cataract grade.
    • o The only intra-operative complications were two anterior capsule tears in the manual group; no other operative complications were recorded for either the manual or the laser group. The post-operative observation rates of corneal edema. flare and trace anterior chamber in the laser group were comparable to those in the manual group.
    • OCT images through the PI correlate well with biometric data obtained from OCT o images for lens thickness and anterior chamber depth for the complete study group and also when stratified by cataract grade.
    • o VICTUS OCT images were found to have a high level of precision for lens thickness (both pooled and stratified by cataract grade) and anterior chamber depth (pooled only).

Complications

.

Two intraoperative complications were reported during the study: two mild anterior capsular tears in the manual group. These complications were determined to be adverse events and are therefore also listed in the following Adverse Events section.

Adverse Events

During the study, a total of nine adverse events (5 events from 4 eyes in manual group and 4 events from 3 eyes in laser group) and serious adverse events (one in laser group and one in manual group) were observed. Seven events were judged as adverse events (three in the laser group and four in the manual group) and two as serious adverse events (one in laser group and one in manual group) were observed.

At one day, two events were documented in a single eye of one subject in the VICTUS group, cortex in the capsular bag (SAE) and increased IOP (AE). Another subject in the VICTUS group was found to have mild iritis at the one day visit (AE). The last subject in the VICTUS group was diagnosed with moderate iritis at the one visit (AE). All adverse events in the laser group were determined to be related to the cataract surgery and not to the use of the VICTUS Femtosecond Laser.

{5}------------------------------------------------

On the day of treatment, two mild anterior capsular tears were noted in the manual group (both AEs). One subject in the manual group was found to have moderate cystoid macular edema at the one month visit. The last subject in the manual group was diagnosed with iritis (SAE) and cystoid macular edema (AE) at an unscheduled visit that took place between the one week and the one month follow-up visit. All adverse events in the manual group were determined to be related to the cataract surgery.

Basis for Determination of Substantial Equivalence:

The technological characteristics of the VICTUS Femtosecond Laser Platform are substantially equivalent to the technological characteristics of the VICTUS Femtosecond Laser Platform (K122386 and K132534), the LenSx Laser (K094052, K101626, K120732).

The proposed additional indication for use of laser-assisted lens fragmentation during cataract surgery for the VICTUS Femtosecond Laser Platform is the same or very similar to that cleared for the LenSx Laser (K094052, K101626, and K120732), as well as for the LensAR Laser (K102727).

Overall, it can be concluded that the VICTUS can be successfully used for lens fragmentation during cataract surgery for nuclear cataracts.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its wing and body. The eagle is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the circle's perimeter.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60S Silver Spring, MD 20993-0002

July 15, 2014

Technolas Perfect Vision, GmbH c/o Ms. Betty M. Johnson Sr. Manager, Regulatory Affairs PO Box 17190 Anaheim Hills, CA 92817

Re: K140615

Trade/Device Name: VICTUS Femtosecond Laser Platform Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Codes: OOE, HQC Dated: June 5, 2014 Received: June 9, 2014

Dear Ms. Johnson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

Page 2 - Ms. Betty M. Johnson

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Kesia Y. Alexander -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(k) Number (if known): K140615

VICTUS Femtosecond Laser Platform Device Name(s):

Indications for Use:

The VICTUS Femtosecond Laser Platform is indicated for use in:

  • . the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea
  • for anterior capsulotomy during cataract surgery .
  • the creation of cuts / incisions in the cornea in patients undergoing cataract surgery or . other ophthalmic treatment requiring cuts / incisions in the cornea.
  • . laser-assisted lens fragmentation of nuclear cataracts during cataract surgery, not for fragmentation of posterior subcapsular (PSC) and cortical cataracts

Prescription Use X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

Image /page/8/Picture/13 description: The image shows the text "Charles" and "2014.07". There is also an image of what appears to be a logo. The logo is difficult to make out, but it appears to be a stylized image of a building or structure. The text and logo are all in black and white.

Page __ of ___________________________________________________________________________________________________________________________________________________________________

§ 886.4390 Ophthalmic laser.

(a)
Identification. An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.(b)
Classification. Class II.